Exporting Medical Devices to Estonia: ÜTS, EUDAMED, CE and Label Language
The competent authority for medical devices in Estonia is the State Agency of Medicines, Ravimiamet, and its website answers a Turkish manufacturer's questions directly: Estonian on a device for lay users, Estonian or English only on a device for professionals; for Class IIa, IIb, III and IVD Class B-D devices, the distributor's EMDDB notification is not affected by EUDAMED. This page combines those answers with the MDR and ÜTS side.
Because Ravimiamet leaves manufacturer registration to EUDAMED, the authorised representative question is answered by the MDR text:
Source: Regulation (EU) 2017/745 (MDR), Article 11(1).ktığından yetkili temsilci sorusu MDR metninde cevaplanır:
"Where the manufacturer of a device is not established in a Member State, the device may only be placed on the Union market if the manufacturer designates a sole authorised representative."
Article 11/1 of the Turkish Medical Device Regulation also treats Türkiye and the EU Member States as a single area; the TİTCK translation of MDCG 2022-16 explains this in a footnote:
"'Union market' refers to the territory of the Member States of the European Union and, by virtue of the European Economic Area (EEA), is extended to Norway, Liechtenstein and Iceland, and, through the Customs Union Agreement, to Türkiye."
Source: MDCG 2022-16, TİTCK translation, footnote 1.nin topraklarını ifade eder ve Avrupa Ekonomik Alanı (AEA) nedeniyle Norveç, Lihtenştayn ve İzlanda'ya ve Gümrük Birliği Anlaşması yoluyla Türkiye'ye genişletilmiştir."
The Commission's March 2022 notice reaches the same conclusion: a manufacturer established in Türkiye does not appoint a representative for Estonia, and the EMDDB notification is the distributor's task. The representative obligation for imports into Türkiye from outside the EU is covered on the authorised representative service page.
Competent Authority: Ravimiametusu aynı sonucu yazar: Türkiye'de yerleşik üretici Estonya için temsilci atamaz, EMDDB bildirimi dağıtıcının işidir. AB dışından Türkiye'ye ithalatta temsilci yükümlülüğü yetkili temsilcilik hizmeti sayfasında.
The competent authority in Estonia is the Medical Devices Department of Ravimiamet (State Agency of Medicines); in the Commission's May 2026 list of national authorities the address appears as Nooruse 1, Tartu, with the contact MSO@ravimiamet.ee. The Commission's language table still refers to terviseamet.ee (Health Board) for Estonia; the current point of contact is Ravimiamet, and its medical device pages are under ravimiamet.ee/en/medical-devices.
Ravimiamet's page 'Regulation in the field of medical devices in Estonia' states that the manufacturer must register itself and its devices in EUDAMED, and that manufacturer information previously given to the national database is now provided only through EUDAMED. It then separates the national obligation: atıf yapar; güncel irtibat Ravimiamet'tir ve tıbbi cihaz sayfaları ravimiamet.ee/en/medical-devices altındadır.
Ravimiamet'in 'Regulation in the field of medical devices in Estonia' sayfası, üreticinin EUDAMED'e kendisini ve cihazlarını kaydetmesi gerektiğini ve daha önce ulusal veri tabanına verilen üretici bilgilerinin artık yalnız EUDAMED üzerinden verildiğini yazar. Ardından ulusal yükümlülüğü ayırır:
According to the page, under § 26 of the Medical Devices Act in Estonia, every operator that distributes an MDR Article 22 system/procedure pack, a Class IIa, IIb or III device, an active implant, or an IVD of Class B, C or D under IVDR Annex VIII notifies Ravimiamet within 10 days of the first distribution of the device concerned. The page states clearly who is obliged:
Kaynak: Ravimiamet, Regulation in the field of medical devices in Estonia (ravimiamet.ee).
Source: Ravimiamet, same page. § 26 uyarınca Estonya'da MDR Madde 22 sistem/işlem paketi, Sınıf IIa, IIb veya III cihaz, aktif implant ya da IVDR Ek VIII'e göre Sınıf B, C veya D IVD dağıtan her aktör, ilgili cihazın ilk dağıtımından itibaren 10 gün içinde Ravimiamet'e bildirir. Yükümlünün kim olduğu açıkça yazılıdır:
"This obligation lies with the economic operator acting in the role of a distributor and notification will continue to be submitted via the national Medical Devices and Appliances Database (EMDDB)."
Source: Ravimiamet, page above.
"The notification of Class I, Class A devices is recommended, but not mandatory."
There is no fee for registration in the EMDDB (Estonian Medical Devices and Appliances Database). Result for a Turkish manufacturer: in Estonia the distributor makes the national notification, and you provide the EUDAMED registration and the documents. For Class I products such as orthoses, bandages or examination gloves, notification is only a recommendation.
Which Language Must the Label and IFU Be in for Estoniaces Database) kaydı için ücret alınmaz. Türk üretici için sonuç: Estonya'da ulusal bildirimi distribütör yapar, siz EUDAMED kaydını ve belgeleri sağlarsınız. Sınıf I ortez, bandaj veya muayene eldiveni gibi ürünlerde bildirim yalnız tavsiyedir.
The EU-side basis for the Estonian-or-English distinction is MDR Article 10(11):
Source: Regulation (EU) 2017/745 on medical devices (MDR), Article 10, paragraph eleven.
"Manufacturers shall ensure that the device is accompanied by the information set out in Section 23 of Annex I in an official Union language(s) determined by the Member State in which the device is made available to the user or patient. The particulars on the label shall be indelible, easily legible and clearly comprehensible to the intended user or patient."
Estonia's choice can be seen in the Commission's 'Language requirements for manufacturers' table (Rev. 3, August 2025) and on the Ravimiamet page.
Relying on § 16 of the Medical Devices Act, the Ravimiamet page 'Labelling and language requirements' states that the information accompanying a device that is placed on the market, made available, distributed and put into service is to be provided as follows:
Source: Ravimiamet, Labelling and language requirements (ravimiamet.ee).Kanunu § 16'ya dayanarak piyasaya arz edilen, bulundurulan, dağıtılan ve hizmete sunulan cihaza eşlik eden bilginin şu şekilde sunulacağını yazar:
"in the Estonian language and in an appropriate manner if the medical device is intended for the use of lay users;"
"in the Estonian or English language and in an appropriate manner if the medical device is intended only for the use of professional users;"
That is: Estonian on a device for lay users; Estonian or English on a device intended only for professional users; and on a custom-made device, the language understood by the specific user. The page adds two further warnings: if the information is to be given in English only, the professional user and the supplier must agree on this in advance, and
Source: Ravimiamet (ravimiamet.ee), same page.a; yalnız profesyonel kullanıcıya yönelik cihazda Estonca veya İngilizce; ısmarlama cihazda belirli kullanıcının anladığı dil. Sayfa iki uyarı daha ekler: bilgi yalnız İngilizce verilecekse profesyonel kullanıcı ile tedarikçi bunu önceden anlaşmalıdır ve
"The professional user still has the right to demand that the relevant information is made available in Estonian."
The Commission table ties the same rule to § 16 paragraph 3 items 1 and 2 and to § 17 of the Medical Devices Act; the implant card is in Estonian or translated into Estonian, the declaration of conformity in Estonian or English (§ 16 paragraph 5), and a field safety notice may be given in English (§ 27). Practical rule for a Turkish manufacturer: an English IFU is possible on a hospital device, but a written agreement is made with the Estonian distributor, and a translation is provided if a hospital asks for Estonian. On a patient device, Estonian is unavoidable.
EUDAMED Actor and Device Registration for the Estonian Market§ 16 fıkra 3 bent 1 ve 2 ile § 17'ye bağlar; implant kartı Estonca veya Estonca'ya çevrilmiş, uygunluk beyanı Estonca veya İngilizce (§ 16 fıkra 5), acil saha güvenlik bildirimi İngilizce verilebilir (§ 27). Türk üretici için pratik kural: hastane cihazında İngilizce IFU mümkündür ama Estonyalı distribütörle yazılı anlaşma yapılır ve bir hastane Estonca isterse çeviri sağlanır. Hasta cihazında Estonca kaçınılmazdır.
The basis for Ravimiamet's sentence that manufacturer registration is only in EUDAMED is the Commission's EUDAMED page:
Source: EUDAMED Overview page, European Commission, health.ec.europa.eu.n'un EUDAMED sayfasıdır:
"As of 28 May 2026, the following 4 modules of EUDAMED became mandatory to use:"
The four modules are actor registration, UDI/device registration, notified bodies and certificates, and market surveillance; for device data, MDR Article 123(3)(e) as amended by Regulation (EU) 2024/1860 allows a 12-month period from publication. According to Ravimiamet, manufacturer registration and device information are now provided only here; the first step is actor registration, and the competent authority issues the SRN after verification (Article 31(2)):
Source: Regulation (EU) 2017/745 on medical devices (MDR), Article 31(1).ifikalar ile piyasa gözetimidir; cihaz verisi için (AB) 2024/1860 ile değişik MDR Madde 123(3)(e) yayımdan itibaren 12 aylık süre tanır. Ravimiamet'e göre üretici kaydı ve cihaz bilgisi artık yalnız buradan verilir; ilk adım aktör kaydıdır, SRN'i yetkili otorite doğrulama sonrası verir (Madde 31(2)):
"Before placing a device, other than a custom-made device, on the market, manufacturers, authorised representatives and importers shall, in order to register, submit to the electronic system referred to in Article 30 the information referred to in Section 1 of Part A of Annex VI, provided that they have not already registered in accordance with this Article."
The SRN and the device are registered under the Basic UDI-DI (MDR Article 29(1)); the screens are covered in the EUDAMED device registration article, the actor step in the EUDAMED registration article, the concept in the Basic UDI-DI article, and MDD-certified products in the legacy device registration article. Because the Estonian distributor's EMDDB notification is tied to the 10-day period, device registration must be finished before shipment; EUDAMED registration consultancy sets the schedule accordingly.
ÜTS Registration of a Product Going to Estoniaa kaydedilir (MDR Madde 29(1)); ekranlar EUDAMED cihaz kaydı, aktör adımı EUDAMED kaydı, kavram Temel UDI-DI, MDD sertifikalı ürünler legacy cihaz kaydı yazılarında. Eston dağıtıcının EMDDB bildirimi 10 günlük süreye bağlı olduğundan cihaz kaydı sevkiyattan önce bitmiş olmalıdır; EUDAMED kayıt danışmanlığı takvimi buna göre kurar.
For products under individual tracking going to Estonia, the following provision applies in ÜTS once the customs declaration is closed:
"An “Export Notification” is made for individually tracked products exported by the company."a ÜTS'de şu hüküm işler:
"When an Export Notification is made in ÜTS, after a query by product number, lot and/or serial number,"
The declaration number and quantity are entered; an incorrect Estonia notification is withdrawn with an 'Export Cancellation Notification' (Article 17/3).
Kaynak: TİTCK Tekil Hareket Bildirimlerine İlişkin Kılavuz, Madde 17.
For a shipment going to the Baltics through a bonded warehouse or free zone, the chain is as in the free zone and bonded warehouse article; all types are covered in the individual movement notifications article, and the service in the individual notification services page.İptal Bildirimi' ile geri alınır (Madde 17/3).
Free sale certificate for Estoniaaltık'a giden sevkiyatta zincir serbest bölge ve antrepo yazısındaki gibidir; türlerin tamamı tekil hareket bildirimleri, hizmet tekil bildirim işlemleri.
If an Estonian buyer or contracting authority asks for a free sale certificate, TİTCK issues the document under Article 60/1 of the Medical Device Regulation.
The application is made in ÜTS under Free Sale Certificate Application Procedures; only Estonia is selected in the country field (Registration Guide Article 42), each English copy is subject to a separate fee, and the Basic UDI-DI is written the same as in EUDAMED. The steps are in the FSC article.
Estonia Summary Tablest Satış Sertifikası Başvurusu İşlemleri bölümünden yapılır; ülke alanından yalnız Estonya seçilir (Kayıt Kılavuzu Madde 42), her İngilizce nüsha ayrı ücrete tabidir ve Temel UDI-DI EUDAMED ile aynı yazılır. Adımlar FSC yazısında.
Topic. Status. Competent authority: Ravimiamet - State Agency of Medicines, Medical Devices Department, Tartu; ravimiamet.ee. Label / IFU language: Estonian (lay user); Estonian or English (professional user only), and the professional user keeps the right to ask for Estonian (Medical Devices Act § 16 and § 17). Implant card in Estonian or translated into Estonian. Additional national registration / notification: Yes, on the distributor's side: in Estonia, an operator distributing Class IIa, IIb, III devices, Article 22 systems/packs, active implants and IVD Class B, C, D notifies Ravimiamet through the EMDDB within 10 days of first distribution (§ 26); recommended for Class I and A. There is no national registration for a manufacturer established in Türkiye. Is an AR needed: No; MDR Article 11(1) does not cover Türkiye (MDCG 2022-16 footnote 1), and Ravimiamet leaves manufacturer registration to EUDAMED. EUDAMED / ÜTS: Actor and device registration is completed before the 10-day period of the EMDDB notification; export notification in ÜTS, and a free sale certificate on request.
Baltic and Northern European marketset - State Agency of Medicines, Tıbbi Cihaz Dairesi, Tartu; ravimiamet.eeEtiket / IFU diliEstonca (sıradan kullanıcı); Estonca veya İngilizce (yalnız profesyonel kullanıcı), profesyonel kullanıcı Estonca talep etme hakkını korur (Tıbbi Cihaz Kanunu § 16 ve § 17). İmplant kartı Estonca veya Estonca'ya çevrilmiş.Ulusal ek kayıt / bildirimEvet, distribütör tarafında: Estonya'da Sınıf IIa, IIb, III cihaz, Madde 22 sistem/paket, aktif implant ve IVD Sınıf B, C, D dağıtan operatör ilk dağıtımdan itibaren 10 gün içinde EMDDB üzerinden Ravimiamet'e bildirir (§ 26); Sınıf I ve A için tavsiye. Türkiye'de yerleşik üretici için ulusal kayıt yok.AR gerekir miHayır; MDR Madde 11(1) Türkiye'yi kapsamaz (MDCG 2022-16 dipnot 1), Ravimiamet üretici kaydını EUDAMED'e bırakırEUDAMED / ÜTSAktör ve cihaz kaydı EMDDB bildiriminin 10 günlük süresinden önce tamamlanır; ÜTS'de ihracat bildirimi, talep halinde serbest satış sertifikası
Medical Device ÜTS Registration Consultancy for Exports to Estonia
- Letonya'ya tıbbi cihaz ihracatı
- Litvanya'ya tıbbi cihaz ihracatı
- Lüksemburg'a tıbbi cihaz ihracatı
- Malta'ya tıbbi cihaz ihracatı
Estonya'ya İhracat için Tıbbi Cihaz ÜTS Kayıt Danışmanlığı
EUDAMED SRN, Temel UDI-DI ve cihaz kaydı, ÜTS ihracat bildirimleri ve Estonya için serbest satış sertifikası tek dosyada toplanır. Teklif formu, +90 505 580 46 22, info@medexmedikal.com.
Frequently Asked Questions
Is an English IFU enough in Estonia for a professional user?
According to Ravimiamet, information on a device intended only for professional users may be in Estonian or English; the supplier and the professional user agree on this in advance, and the professional user keeps the right to ask for Estonian. Estonian is mandatory on a device for lay users.
Who submits the EMDDB notification, and for which classes?
In Estonia, the economic operator acting as distributor, within 10 days of first distribution, for Class IIa, IIb and III devices, Article 22 systems/procedure packs, active implants and IVD Class B, C and D. For Class I and A, notification is recommended but not mandatory.
Does a Turkish manufacturer have to register with Ravimiamet?
No. Ravimiamet states that manufacturer registration and device information are now provided only through EUDAMED. National notification is the distributor's task.
Is an authorised representative needed for Estonia?
Not required. MDR Article 11(1) requires an authorised representative only for manufacturers established outside the Union; footnote 1 of MDCG 2022-16 treats Türkiye as part of the Union market through the Customs Union. Since Ravimiamet leaves manufacturer registration to EUDAMED, the Turkish manufacturer appears with its own SRN, and the EMDDB notification is the distributor's task.
Which ÜTS step applies to exports to Estonia?
For products under individual tracking, the export notification is made with the customs declaration number and quantity; if the buyer requests it, a free sale certificate specific to Estonia is obtained from ÜTS.