Importing Medical Devices from South Korea: ÜTS Registration and Document Requirements
Importing medical devices from South Korea is a common question for product groups such as dermal fillers, aesthetic devices and diagnostic kits. A Korean manufacturer's MFDS approval is not used for ÜTS registration in Türkiye; what is required is the CE marking, an authorized representative and apostilled documents. This page explains the import and ÜTS steps for Korean-origin devices based on the text of the legislation.
Does MFDS approval suffice for ÜTS registration in Türkiye?
The competent authority for medical devices in South Korea is the MFDS (Ministry of Food and Drug Safety). According to the MFDS English approval process page, devices are divided into four classes (I-IV) based on their potential risk to human health; Class I is subject to notification, Class II to certification or approval, and Classes III and IV to approval. The regulatory framework is the Medical Device Act of 2003 (Act No. 6909) and its secondary legislation. These approvals and certificates belong to the Korean market. In Türkiye, the condition for placing a device on the market is CE marking and conformity with the Regulation.
For Korean-origin fillers and aesthetic devices, the first thing an importer should check is whether the manufacturer has completed the CE process in addition to its MFDS approval. ÜTS document registration requires the EU declaration of conformity and, depending on the class, a notified body certificate; the MFDS document does not replace these.
Under Article 20/1 of the Regulation, a device placed on the market in Türkiye bears the CE marking; MFDS approval does not meet this requirement. The importer's pre-placement verification list is also in the Regulation:
“a) In order to place a device on the market; 1) verifies that the device bears the CE marking and that the EU declaration of conformity has been drawn up, 2) verifies that the manufacturer is identified and that an authorized representative has been appointed by the manufacturer in accordance with Article 11, 3) verifies that the device is labelled in accordance with this Regulation and that the required instructions for use accompany the device, 4) where applicable, verifies that a UDI has been assigned by the manufacturer in accordance with Article 27.”
Source: Medical Device Regulation (OG 31499), Article 13/2-a.
In Korean files, the second item, the authorized representative, is the document most often missing; MFDS approval does not replace the CE marking, and a local representative in Korea does not replace the authorized representative required by the Regulation. For the difference between the declaration and the certificate, see our article on the differences between the EC certificate and the declaration of conformity.EC sertifikası ve uygunluk beyanı farkları yazısına bakın.
The Korean manufacturer's authorized representative in the EU or Türkiye
South Korea is not an EU member. A device from a Korean manufacturer can be placed on the market in Türkiye only if an authorized representative established in Türkiye or the EU has been appointed. A representative appointed by the manufacturer for the EU is also valid for Türkiye under the Customs Union; a separate Turkish representative is not sought. The authorization document is uploaded in the ÜTS document record together with the declaration of conformity.
“If the manufacturer is not established in Türkiye or in EU member states, the device may be placed on the market only if the manufacturer appoints an authorized representative.”
Source: Medical Device Regulation, Article 11, paragraph 1.
“If the manufacturer of a device is not established in a Member State, the device may be placed on the Union market only if the manufacturer appoints a single authorized representative [MDR and IVDR Article 11(1)].”
Source: MDCG 2022-16 (TİTCK translation), Section 4 Appointment and mandate.
The Commission's Customs Union notice also states that for a manufacturer established outside the EU or Türkiye, appointing a single authorized representative for the two markets is sufficient. Medex provides this service as an EU Authorised Representative in EUDAMED under number TR-AR-000057550; for Korean manufacturers, see our pages on EU authorized representation for South Korean manufacturers and authorized representation for importers.EU Authorised Representative olarak verir; Koreli üreticiler için Güney Kore üreticileri için AB yetkili temsilcilik, ithalatçı için yetkili temsilcilik hizmeti sayfalarına bakın.
Two ÜTS rules for the Korean manufacturer's authorization document:
“For ‘EU Declaration of Conformity (MDR/IVDR)’ type document registration notifications covered by this article that belong to manufacturers not established in an EU member state or in Türkiye, it is mandatory to make an entry in the ‘EU Authorized Representative’ field on the ÜTS registration screen.”
Source: ÜTS Registration Guide TCKKD-KLVZ-03 V4.0, Article 12/7-a.
“The original of the document and its Turkish translation by a sworn translator are uploaded together in the ‘Authorization Document’ field. Neither an apostille nor embassy/consulate legalization is required for the original of the authorization document.”
Source: ÜTS Registration Guide TCKKD-KLVZ-03 V4.0, Article 12/7-c.
Apostille and translation for Korean documents
According to the HCCH status table, the Republic of Korea acceded to the Hague Apostille Convention on 25.10.2006, and the Convention entered into force on 14.7.2007. For the Korean manufacturer's declaration of conformity, an apostille obtained in Korea is accepted in ÜTS; the consular route is not needed. Even if Korean documents are drawn up in Korean or English, a sworn Turkish translation is uploaded in every case.
The Korean apostille and consular legalization appear in the Guide in the following order:
“a) Apostille-certified document or, 1) If the manufacturer is in an EU member state, the apostille certification obtained from the competent authority of the country where the manufacturer is established; if the manufacturer is established in a country that is not an EU member, the apostille certification obtained from the competent authority of the country where the manufacturer or the EU Authorized Representative appointed by the manufacturer is established is accepted.”
Source: ÜTS Registration Guide, Article 12, paragraph 1, subparagraph (a).
“b) The document certified by the Turkish embassy/consulate in the country of the manufacturer or, 1) If the manufacturer is established in a country that is not an EU member, certification by the Turkish embassy/consulate located in the country where the manufacturer's EU Authorized Representative is established is also accepted.”
Source: ÜTS Registration Guide, Article 12/1-b.
A sworn Turkish translation is uploaded alongside the Korean or English original:
“A complete Turkish translation, made by a sworn translator, of the document uploaded in the ‘Original Document’ field is uploaded in the ‘Turkish Document’ field.”
Source: ÜTS Registration Guide (TCKKD-KLVZ-03, 13.05.2026), Article 12/2.
The apostilled original upload (Guide Article 12/1-a-2), e-Apostille (12/1-c) and e-signed document (12/1-ç) routes, and the translator's stamp and declaration requirement (12/2-b), are explained on our apostille guide, e-Apostille verification and sworn translation and document preparation pages.apostil rehberi, e-Apostil sorgulama ve yeminli tercüme ve belge hazırlığı sayfalarında anlatılır.
TAREKS and ÜTS import notification for Korean consignments
An A.TR certificate does not apply to consignments from Korea; TAREKS risk analysis is applied to products on the Annex-1 list of the Communiqué. Some aesthetic devices are listed separately in Annex-1 as products under Annex XVI of the Regulation. After customs, product movement begins in ÜTS with the import notification.
“All procedures related to the import control of medical devices are carried out through TAREKS and according to risk analysis.”
Source: Communiqué on Import Control of Medical Devices (Product Safety and Inspection: 2026/16), Article 4/1.
“Permission to import a product or the creation of a TAREKS reference number for the product does not mean that the product complies with the legislation and/or is safe.”
Source: Communiqué on Import Control of Medical Devices (2026/16), Article 12/2.
For a single product from Korea, an ÜTS notification can be opened only if the product is registered:
“The importer or the dealer authorized for import notification can make an import notification only for medical devices that are registered in ÜTS or that meet the conditions announced by the Agency within the scope of the transitional provisions of the Medical Device Regulation.”
Source: TİTCK Guide on Individual Movement Notifications, Article 7/4.
An import notification is made for every product whose origin is not Türkiye (Individual Movement Guide Article 7/1); for fillers and aesthetic devices from Korea, the notification comes after product registration and before the giving notification. The process is described in our article on ÜTS registration for medical device imports.tıbbi cihaz ithalatında ÜTS kaydı yazısında.
ÜTS steps for a device of Korean origin
The ÜTS sequence for a device of Korean origin: ÜTS company registration as an importer with a sales center authorization certificate, ÜTS document registration with the apostilled declaration and authorization document, ÜTS product registration linked to the notified body certificate, and an import notification for each consignment. The general explanation is in our article on ÜTS registration for medical device imports; the licence requirement for company registration is as follows:satış merkezi yetki belgesi ile ithalatçı türünde ÜTS firma kaydı, apostilli beyan ve yetki belgesiyle ÜTS belge kaydı, onaylanmış kuruluş sertifikasına bağlı ÜTS ürün kaydı, her parti için ithalat bildirimi. Genel anlatım tıbbi cihaz ithalatında ÜTS kaydı yazısında; firma kaydındaki ruhsat şartı şöyledir:
“Companies that will make a company registration in ÜTS as ‘Manufacturer/Importer/Dealer/Exporter’ must hold a medical device sales center licence, and the ÇKYS number must be entered during company registration.”
Source: ÜTS Registration Guide (TCKKD-KLVZ-03, 13.05.2026), Article 5/5.
In the product registration, “Imported” is selected in the Imported/Manufactured field, and the Country of Import and GTİP fields become mandatory (Guide Article 27); for the GTİP, see our pages on the GTİP code in ÜTS registration and on contract manufacturing and country of origin for devices manufactured outside Korea.ÜTS kaydında GTİP kodu, Kore dışında üretilen cihazda menşe için fason üretim ve menşe ülke yazısına bakın.
Summary table for South Korea
TopicSouth KoreaCountry authorityMFDS (Ministry of Food and Drug Safety) (mfds.go.kr)Is MFDS approval used in ÜTS?No; for fillers and aesthetic devices, a notified body certificate is requiredAuthorized representativeMandatory (Regulation Article 11/1); a representative in the EU is also valid for TürkiyeApostilleIn force for Korea since 14.7.2007; no need for the consular routeDocument languageKorean or English original + sworn Turkish translationCustomsNo A.TR; aesthetic devices under Annex XVI are separately listed in Annex-1
For other countries: importing medical devices from Taiwan, importing medical devices from Malaysia.Tayvan'dan tıbbi cihaz ithalatı, Malezya'dan tıbbi cihaz ithalatı.
Medical Device ÜTS Registration Consultancy for Imports from South Korea
We prepare the Korean manufacturer's declaration of conformity, notified body certificate and authorization document for ÜTS with apostille and sworn translation; for filler and aesthetic device files, we handle company, document and product registrations and import notifications. Request a quote or reach us at +90 505 580 46 22 and info@medexmedikal.com.Fiyat teklifi alın veya +90 505 580 46 22 numarasından ve info@medexmedikal.com adresinden ulaşın.
Frequently Asked Questions
Can I register an MFDS-approved Korean filler product in ÜTS without CE marking?
No. Only the EU declaration of conformity and the notified body certificate are used for ÜTS document registration. Dermal fillers are in a class that requires a notified body certificate; MFDS approval does not replace this certificate.
Is the Dutch authorized representative of a Korean manufacturer sufficient for Türkiye?
Yes. Article 11/1 of the Regulation accepts a representative established in Türkiye or in an EU member state. The Commission's Customs Union notice also states that a single representative is sufficient for both markets. In ÜTS, the original authorization document and its sworn translation are uploaded.
Kore apostili yerine konsolosluk onayı alabilir miyim?
Article 12/1-b of the Guide still lists consular legalization as an option. However, since South Korea is a party to the Convention, the apostille is both sufficient and the shorter route. The two authentications are not requested together.
The Korean manufacturer's documents are bilingual, in Korean and English. Which language is the translation made from?
Article 12/2 of the Guide requires a sworn translation of the document uploaded in the Original Document field. The whole bilingual document is uploaded as the original; the translation is made from this document and, under Article 12/2-a, only the Turkish part is placed in the Turkish Document field.
Are aesthetic devices from Korea under Annex XVI also registered in ÜTS?
Yes. The Annex-1 list of the Communiqué includes cavitation, lipolysis, skin care and hair removal devices and similar products under Annex XVI of the Regulation as subject to import control. Since these products fall under the Medical Device Regulation, ÜTS registration and an import notification are required.