EUDAMED Legacy Device Registration: What Is a Legacy Device, When Is It Registered and With Which Identifier?

12 hours ago
9 min read
EUDAMED legacy cihaz kaydı ve UDI/Cihaz modülü

SUMMARY

A EUDAMED legacy device is a device that continues to be placed on the market under the transitional provisions with a valid certificate or declaration of conformity issued under the old medical device regulations (MDD, AIMDD, IVDD). The EUDAMED Q&A text published by TİTCK in Turkish calls these devices "kalıt cihaz" (legacy device).

In this article we give, as verbatim quotations from the regulation and the official Q&A text, the provisions that apply to a legacy device, the period for registration in the EUDAMED UDI/Device module, the device identifier used, the cases where registration is not required, and the ÜTS side.

For your EUDAMED and ÜTS registrations: Medical Device ÜTS Registration Consultancy

Table of Contents

1. What Is a Legacy ("Kalıt") Device?

Paragraph 5 of Provisional Article 1 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation) lists the dates until which devices certified under the old regulations may be placed on the market:

  • “b) Devices that were issued under the Medical Device Regulation or the Active Implantable Medical Devices Regulation repealed by Article 110 and that hold a certificate valid under the fourth paragraph may be placed on the market or put into service until the following dates:”

  • “1) (Amended wording: Official Gazette 17 Aug 2024-32635)(4) All class III devices and class IIb implantable devices, except sutures, staples, dental fillings, dental braces, dental crowns, screws, wedges, plates, wires, pins, clips and connectors, until 31 December 2027.”

  • “2) Class IIb devices other than those covered by sub-paragraph (1), class IIa devices, and class I devices placed on the market in a sterile condition or with a measuring function, until 31 December 2028.”

  • “c) Devices for which, under the Medical Device Regulation repealed by Article 110, the conformity assessment procedure did not require the involvement of a notified body, for which the declaration of conformity was drawn up before 26 May 2021, and for which the conformity assessment procedure under this Regulation requires the involvement of a notified body, may be placed on the market or put into service until 31 December 2028.”

The conditions for benefiting from these dates are in paragraph (ç) of the same paragraph. Two of the conditions listed there:

  • “1) These devices continue to comply, as applicable, with the Medical Device Regulation or the Active Implantable Medical Devices Regulation repealed by Article 110.”

  • “2) There is no significant change in the design and intended purpose.”

The Q&A text refers to the MDCG 2021-25 Rev. 1 and MDCG 2022-8 guidance documents for the definition of the term "kalıt cihaz" (legacy device) and explains the expression "Tüzük cihazı" (Regulation device) as follows:

“Devices that comply with the MDR/IVDR provisions, other than legacy devices.”

For the transition dates on the IVD side, see our article on the IVDR transition timetable, and for the MDR side, our article on the MDR transition timetable.

2. Which MDR Provisions Apply to a Legacy Device?

A legacy device remains in conformity with the old regulation, but on some matters the provisions of the new regulation apply. Paragraph 5 of Provisional Article 1:

“d) By way of derogation from paragraph (a), the requirements of this Regulation relating to post-market surveillance, market surveillance and inspection, vigilance, and the registration of economic operators and of devices apply to the devices referred to in paragraphs (b) and (c), instead of the corresponding requirements in the Regulations repealed by Article 110.”

For IVD devices, the same rule is in paragraph 3d added to Article 110 of the IVDR by (EU) 2024/1860:

“By way of derogation from paragraph 3 of this Article, the requirements of this Regulation relating to post-market surveillance, market surveillance, vigilance, and to the registration of economic operators and of devices shall apply to devices referred to in paragraphs 3a and 3b of this Article, instead of the corresponding requirements in Directive 98/79/EC.”

So the EUDAMED registration of a legacy device is made under the device registration provisions of the new regulation.

3. Mandatory Use of EUDAMED and Application of the Old Provisions

With (EU) 2024/1860, EUDAMED becomes mandatory module by module. The Q&A text (Q2):

“According to the new wording of MDR Article 123(3)(d) and IVDR Article 113(3)(e), the obligations and requirements relating to a specific module of Eudamed will become applicable 6 months after the publication in the OJEU of the notice confirming the functionality of that module.”

Until that date, the registration provisions of the old regulations apply. Provisional Article 1:

“(12) (Amended: Official Gazette 17 Aug 2024-32635)(4) The relevant provisions of the Medical Device Regulation and the Active Implantable Medical Devices Regulation repealed by Article 110 concerning vigilance reporting, clinical investigations, registration of devices and economic operators, and certificate notifications continue to apply until the date of application of the provisions referred to in paragraph 1(a) of Article 111.”

For the modules that have become mandatory, see our page on mandatory use of EUDAMED modules.

4. Time Limit for Registering a Legacy Device: The 12-Month Rule

The registration period is in paragraph 17 of Provisional Article 1:

  • “a) Manufacturers ensure that the information that must be entered into Eudamed under Article 29, including that relating to the following devices, is entered into the electronic system referred to in points (a) and (b) of paragraph 2 of Article 33 within 12 months at the latest of the date of publication of the notice referred to in point (a) of paragraph 1 of Article 111, provided that the devices are placed on the market within 6 months of the date of publication of that notice:”

  • “2) Devices, other than custom-made devices, placed on the market under paragraph 5(a), (b) or (c) of Provisional Article 1, for which the manufacturer has carried out a conformity assessment under Article 52 and which are not yet registered in Eudamed.”

The same sub-point of MDR Article 123(3)(e), as amended by (EU) 2024/1860:

“(ii) devices, other than custom-made devices, placed on the market pursuant to Article 120(3), (3a) or (3b), unless the device, for which the manufacturer has undertaken a conformity assessment in accordance with Article 52, is already registered in Eudamed;”

The Q&A text (Q7) sets the following rule for a legacy device whose first unit was placed on the EU market before the date of mandatory use and whose later units will be placed on the market on or after that date:

“The relevant device registration in the UDI/DEV module must be made within 12 months of the publication in the OJEU of the notice confirming the functionality of the UDI/DEV module”

The example in the text:

“Example: If the notice confirming the functionality of the UDI/DEV module is published in the OJEU on 1 July 2025, the requirements relating to the UDI/DEV module in MDR Article 29 and IVDR Article 26 will start to apply on 1 January 2026. If individual units/items of a device have been placed on the market before 1 January 2026 and other individual units/items will be placed on the market after that date, the device will have to be registered by 1 July 2026.”

The dates in the example are the text’s own assumptions. According to the same text, for a unit placed on the market for the first time on or after the date of mandatory use, registration is made before that unit is placed on the EU market. For the calendar, see our page on the EUDAMED UDI and device registration deadline.

5. Device Identifier: EUDAMED ID or UDI-DI

The Q&A text states that device registration in EUDAMED is made at the device identifier level:

“The device/SPP identifier is the UDI-DI for devices/SPPs placed on the market under the Regulations, and the EUDAMED ID or UDI-DI for legacy devices”

“This means that in the UDI/DEV module there will be a single registration per device identifier (UDI-DI or Eudamed ID), covering all individual devices/SPPs, but with different production identifiers, UDI-PIs, representing the lot number, serial number, manufacturing date and so on.”

The relation of a legacy device to the UDI is explained in a footnote of the text:

“Legacy devices are not subject to the UDI requirements under the MDR/IVDR. However, assigning a UDI-DI to these devices is not prevented, or these devices may be placed on other markets where, for example, UDI rules are in force.”

For the Basic UDI-DI used for MDR devices, see our article What Is the Basic UDI-DI.

6. Which Legacy Devices Are Not Registered?

The Q&A text (Q8) lists three cases:

“Legacy devices and Regulation devices whose individual (sales) unit is no longer placed on the market when the UDI/DEV module becomes mandatory do not need to be registered, unless a PMSV activity takes place.”

“In addition, legacy devices do not need to be registered if the "same device" is already registered as a Regulation device.”

“Note: Devices on which changes leading to the assignment of a new UDI-DI are made in order to become a Regulation device will not be considered the "same device".”

A "legacy" device in the sense of old devices and a "kalıt" (legacy) device are separate concepts:

“"Old" devices, as defined in MDCG 2021-25 Rev. 1 and MDCG 2022-8, cannot be registered in the UDI/DEV module. Where the device is the subject of a serious incident report (MIR) or a field safety corrective action (FSCA), the manufacturer will need to provide a limited device data set in order to submit the relevant report in the VGL module.”

7. Registration When Vigilance Is Required

Registration is also required if vigilance activity arises for a legacy device that is no longer placed on the market. The Q&A text (Q14):

“Regulation devices and legacy devices whose individual (sales) units are not placed on the market from the date on which the UDI/DEV module becomes mandatory need to be registered in the UDI/DEV module only when the manufacturer has to carry out any PMSV activity in the VGL module”

“To be able to do this, the manufacturer and, where applicable, the authorised representative must first register as actors in the ACT module.”

According to a footnote of the same text:

“In accordance with MDCG 2021-25 and MDCG 2022-21, legacy devices, custom-made devices and old devices do not require a PSUR in Eudamed.”

For vigilance reports, see our articles on vigilance and serious incident reporting, and for the PSUR, post-market surveillance and PSUR.

8. Who Makes the Registration? Actor Registration

Actor registration comes before device registration. The Q&A text (Q3):

“Economic operators under MDR Article 31 and IVDR Article 28 (manufacturers, importers and authorised representatives) must be registered as actors before a device is placed on the market and, where applicable, obtain a Single Registration Number (SRN).”

Under Article 29 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation), it is the manufacturer that enters device information into EUDAMED. According to the answer to Q5 of the Q&A text, distributors do not need to register in EUDAMED. For actor registration and the SRN, see our articles on EUDAMED actor registration, and for the registration of MDR devices, EUDAMED device registration.

9. The ÜTS Side: Device Registration Under the Old Regulations

EUDAMED registration does not remove the ÜTS obligation. Article 34 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation):

“(1) Without prejudice to the EUDAMED-related obligations set out in this Regulation, economic operators established in the country, notified bodies, health institutions, health professionals and other relevant persons fulfil their obligations relating to the national product tracking system for medical devices (ÜTS).”

The counterpart of legacy devices in ÜTS is registration under the "Eski Tıbbi Cihaz Yönetmelikleri" (Old Medical Device Regulations). Article 25 of the TİTCK ÜTS Registration Guide (TCKKD-KLVZ-03):

“ARTICLE 25- (1) The information that appears on the ÜTS registration screen and must be entered into the system during device registration under the Old Medical Device Regulations in ÜTS, together with its explanations, is placed under the headings “Tanımlayıcı Bilgiler” (Identifying Information), “İthal/İmal Bilgileri” (Import/Manufacture Information), “Sınıflandırma Bilgileri” (Classification Information), “Ürün Belgeleri” (Product Documents), “Ürün Görselleri” (Product Images), “Özellikler” (Features), “Ölçü Bilgileri” (Dimension Information), “Saklama/kullanım Koşulu Bilgileri” (Storage/Use Condition Information) and “Başvuru Durumu” (Application Status).”

For the ÜTS document types of old-regulation documents, see our articles on ÜTS document registration, and for product registration, ÜTS medical device product registration.

10. Frequently Asked Questions

With which identifier is a legacy device registered in EUDAMED?

According to the TİTCK Turkish EUDAMED Q&A text, the device identifier for legacy devices is the EUDAMED ID or UDI-DI. Legacy devices are not subject to UDI requirements under the MDR/IVDR, but assigning a UDI-DI is not prevented.

Until when is a legacy device registered?

For devices placed on the market before the date of mandatory use and still placed on the market afterwards, registration is made within 12 months of the publication of the notice confirming the functionality of the UDI/Device module (Provisional Article 1, paragraph 17, Q&A Q7).

If the MDR version of the same device is registered, is the legacy version also registered?

No. According to Q&A Q8, if the "same device" is registered as a Regulation device, registration of the legacy device is not required. A device on which a change leading to the assignment of a new UDI-DI has been made is not considered the "same device".

Is a legacy device that is no longer placed on the market registered?

As a rule, no. However, if the manufacturer has to carry out a PMSV activity in the VGL module, for example submit a serious incident report, the device is registered in the UDI/Device module (Q&A Q8 and Q14).

Does EUDAMED registration replace ÜTS registration?

No. Under Article 34 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation), economic operators established in the country fulfil their ÜTS obligations without prejudice to their EUDAMED obligations.

For your EUDAMED and ÜTS registrations: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application