ÜTS Product Registration: Device Registration Fields Under MDR, Barcode Rules and Device Application

12 hours ago
12 min read
ÜTS ürün kaydı alanları ve cihaz başvurusu

SUMMARY

ÜTS product registration means entering a medical device into the Product Tracking System (ÜTS) with its barcode, class, linked documents and images, and bringing it to “Registered” status by submitting a device application to the Authority. Sales centers cannot sell devices that are not registered in ÜTS.

In this article we explain device registration under MDR by quoting verbatim from Articles 27 and 40 of the TİTCK ÜTS Registration Guideline coded TCKKD-KLVZ-03 (Medical Device Product Tracking System Registration Processes Guideline). Devices under IVDR are registered under Article 30 of the Guideline, and devices under the older regulations under Article 25.

For registering your products in ÜTS in line with the Guideline: Medical Device ÜTS Registration Consultancy

Contents

  • 1. What Is ÜTS Product Registration and Why Is It Mandatory?

  • 2. Prerequisite: Company Registration and Document Registration

  • 3. Barcode: UDI Issuing Entity and UDI-DI Number

  • 4. Brand, Product Name, Version/Model and Reference Number

  • 5. Import/Manufacture Information and Country of Origin

  • 6. Class, GMDN and EMDN

  • 7. Product Documents and Basic UDI-DI

  • 8. Product Images

  • 9. Ready for Application Status and Device Application

  • 10. Device Application Types and Number of Devices per Application

  • 11. Grouping Rule: Same Basic UDI-DI, Same GMDN

  • 12. Objecting to a Rejection or Revision Decision: Submit a Request Before Reapplying

  • 13. Accessories, Spare Parts, Number of Products in Content and Product Description

  • 14. Frequently Asked Questions

1. What Is ÜTS Product Registration and Why Is It Mandatory?

Article 26 of the Medical Device Sales, Advertising and Promotion Regulation:

“(2) Sales centers cannot sell devices that are not registered in the Authority’s registration and information management system.”

The first paragraph of Article 27 of the Guideline lists the headings of the MDR registration screen:

“ARTICLE 27- (1) The information to be entered into the system on the ÜTS registration screen during device registration (device registration/update notification) under MDR, together with its explanations, appears under the headings ‘Identifying Information’, ‘Import/Manufacture Information’, ‘Classification Information’, ‘Product Documents’, ‘Product Images’, ‘Features’, ‘Dimension Information’, ‘Storage/Use Condition Information’, ‘CMR and/or Endocrine Disrupting Substance Information’ and ‘Application Status’.”

There is a training requirement for the person who will enter the registration into ÜTS. Article 11 of Guideline TCS-KLVZ-08:

“(3) As of 1/1/2025, persons who do not hold a medical device registration clerk qualification certificate are not permitted to register medical devices in the Product Tracking System (ÜTS).”

2. Prerequisite: Company Registration and Document Registration

The company must be registered in ÜTS before product registration. For details of company registration, see our ÜTS company registration article.

The documents to be linked to the device must also be registered in ÜTS beforehand. Paragraph five of Article 27 of the Guideline:

  • “1) This field is a mandatory selection in ÜTS.”

  • “2) The documents that must be added under the applicable one of the document business rules referred to in Article 66 of this Guideline, in line with the risk class of the relevant device, are selected from among the documents that are already registered in ÜTS as a result of the application made under Article 39 of this Guideline, and added to this field.”

For document types and document application, see our ÜTS document registration article.

3. Barcode: UDI Issuing Entity and UDI-DI Number

The first field under the “Identifying Information” heading is the UDI issuing entity:

“3) The UDI issuing entities and standards registered in ÜTS under MDR and authorized for medical devices are GS1, HIBCC, ICCBBA and IFA.”

The barcode is entered exactly as it is written in the documents and on the label:

“4) For a device registration whose barcode is not entered into ÜTS as it appears in the technical documents (label/sales packaging, EU declaration of conformity, EU certificate, etc.), the device application made under Article 40 of this Guideline results in a negative outcome.”

Registration is made with the barcode of the smallest package:

“6) As a rule, device registration in ÜTS is made with the barcode of the smallest package placed on the market (unit package or single package).”

The barcodes of sales packages containing multiple products are added, together with their content quantities, to the “Market Placement Information” field after registration is completed. The Guideline requires a new UDI-DI for the following changes:

  • “a. Change of device name or trade name,”

  • “b. Change of device version/model,”

  • “c. Change related to whether it is single-use,”

  • “ç. Change related to whether it is packaged sterile,”

  • “d. Change related to whether it needs to be sterilized before use,”

  • “e. Change in the quantity of devices in the package/box (number of products in content),”

You can check the check digit of a GS1 barcode with our GTIN check digit calculation tool.

4. Brand, Product Name, Version/Model and Reference Number

“2) Devices for which no brand is declared, for the purpose of ensuring product safety, are not registered in ÜTS.”

The brand information is entered as it appears on the label or in the linked documents. For details, see our article on brand name rules in ÜTS product registration. The product name is entered in Turkish:

“6) A device application whose product name/trade name does not state what the device is, cannot be understood, is incomplete, is not in Turkish, or is not entered into the system consistently with what is shown on the label is rejected.”

At least one of the version/model and reference/catalog number fields must be filled in:

“7) Both this field and the field mentioned in subparagraph (ğ) of this paragraph cannot be left empty, and “YOK” (NONE) cannot be written in the relevant fields. Information must be entered in at least one of the two fields.”

5. Import/Manufacture Information and Country of Origin

“2) If the manufacturer that issues the declaration of conformity to which the device is linked (legal manufacturer) is established in Türkiye, this field is selected as ‘Manufacture’. If the manufacturer that issues the declaration of conformity to which the device is linked (legal manufacturer) is not located in Türkiye (that is, a foreign manufacturer), this field is selected as ‘Import’.”

If “Import” is selected, the “Country of Import” and “GTİP Number” fields become mandatory. In the country of origin field, the country where the device is physically manufactured is selected, and more than one country can be selected. For the GTİP field, see our article on the GTİP code in ÜTS registration.

6. Class, GMDN and EMDN

The class field is mandatory. The following options exist under MDR (Article 27, fourth paragraph):

  • Class I Other (non-sterile and without measuring function)

  • Class Ir (reusable surgical instruments)

  • Class Im (with measuring function)

  • Class Is (with sterile function)

  • Class IIa

  • Class IIb

  • Class III

The GMDN field is mandatory and each device has a single GMDN code. If the document contains a GMDN code, that code is selected. For EMDN, the Guideline sets the following rule:

“5) If the EMDN code that identifies the device appears in at least one of the documents belonging to the device (EU certificate, EU Declaration of Conformity), selecting the most detailed code and terminal term of the EMDN code stated in that document or documents in this field is mandatory.”

For code selection, see our articles on GMDN code, and for class, on MDR risk class.

7. Product Documents and Basic UDI-DI

The class of the device must have been selected before documents are selected. After the documents are selected, the Basic UDI-DI is selected:

“4) Through the ‘Select Basic UDI-DI’ screen in this field, only the Basic UDI-DI that identifies the relevant device is selected from among the Basic UDI-DI numbers associated with the EU Declaration of Conformity selected in the field mentioned in subparagraph (a) of this paragraph. More than one selection cannot be made in this field.”

“7) The Basic UDI-DI is not present on the market placement label and/or sales packaging of the device.”

For the difference between Basic UDI-DI and UDI-DI, see our article What Is Basic UDI-DI.

8. Product Images

The image field is mandatory. The rules in the sixth paragraph of Article 27 of the Guideline:

  • “2) Photographs and/or technical drawings of the relevant device that clearly show the device are uploaded to this field as images.”

  • “4) The catalog itself, or an image containing more than one device that makes the device hard to understand, is not uploaded as a product image.”

  • “7) A maximum of 10 (ten) product images can be uploaded to this field.”

  • “8) The product image is uploaded to this field as a jpeg, jpg, pdf or png file, not exceeding the maximum file size specified in ÜTS. If images that are unclear and unreadable are uploaded, the application results in a negative outcome.”

Not only a package or packaging image is uploaded; a photograph or technical drawing of the device is also uploaded. If the packaging carries an image of the device itself, the packaging image can be uploaded. If opening the packaging would break sterility, or the device is in liquid form, only the packaging image can be uploaded.

9. Ready for Application Status and Device Application

The “Is it ready for application?” question on the registration screen determines the status of the registration:

  • “3) If this field is selected as ‘No’ (if the relevant medical device registration notification is not ready for application), the status of this medical device notification is saved in ÜTS as ‘Draft’.”

  • “4) If this field is selected as ‘Yes’ (if the relevant medical device registration notification is ready for application), the status of this medical device notification is saved in ÜTS as ‘Ready for Application’ and it is listed on the ‘List My Medical Devices’ screen.”

Registration is not finished when it is entered. Under Article 40 of the Guideline, a device application is made to the Authority. The application is created from the “Medical Device Application Procedures” screen in the “Medical Device” tab. Rules:

  • “c) Since medical device applications can only be created from devices linked to one declaration, only one declaration can be selected from this list.”

  • “ğ) This type of device application includes a maximum of 500 medical device notifications.”

  • “(8) In device applications under the new Medical Device Regulations (excluding ITCs), applications are only permitted for devices linked to the same Basic UDI-DI code.”

The “Confirm” action is carried out only by the company’s users of the “Medical Device Registration Clerk” type. The application is created in “Awaiting Signature” status, the authorized signatory signs electronically, it moves to “Awaiting Payment” status, and once the fee is paid it reaches “Under Review” status. Applications that do not reach “Under Review” status cannot be evaluated by the Authority’s review specialist.

For a bulk registration route for many devices, see our ÜTS bulk device notification Excel article. For having the registration prepared on your behalf, see our ÜTS product registration service page.

10. Device Application Types and Number of Devices per Application

The application type is selected from the “Create Medical Device Application” list in ÜTS. According to the third paragraph of Article 40 of the Guideline, the list contains the following types: “Optical Products Application”, “Active Implantable Medical Devices Application”, “Class 3 Products Application”, “Other Products Application”, “Custom-Made Medical Device Application”, “System Procedure Pack Application” and “Products under 2023/KK-1 & 2023/KK-2 Application”.

The number of devices that can be added to one application varies by type. Subparagraph (ğ) of the fifth paragraph and subparagraph (d) of the sixth paragraph:

  • “This type of device application includes a maximum of 500 medical device notifications.”

  • “This type of device application includes a maximum of 5 medical device notifications.”

The limit of 500 applies to optical, active implantable, Class 3, other products and 2023/KK-1 & KK-2 applications; the limit of 5 applies to custom-made device and system procedure pack applications. There is no declaration selection step in custom-made and SPP applications; the product is selected directly (sixth paragraph).

11. Grouping Rule: Same Basic UDI-DI, Same GMDN

The regulation scope determines which devices can be placed together in one application. The seventh and eighth paragraphs of Article 40 of the Guideline:

  • “In device applications under the old Medical Device Regulations, applications are only permitted for devices linked to the same GMDN code.”

  • “In device applications under the new Medical Device Regulations (excluding ITCs), applications are only permitted for devices linked to the same Basic UDI-DI code.”

In practice this combines with the single-declaration rule: under MDR, an application consists of devices linked to a single declaration of conformity and a single Basic UDI-DI. A separate application is opened for devices with different Basic UDI-DIs. To determine the Basic UDI-DI, see our guide What Is Basic UDI-DI.

12. Objecting to a Rejection or Revision Decision: Submit a Request Before Reapplying

If the Authority issues a “Rejection” or “Revision” decision for a device notification and the decision is considered contrary to the legislation, the right route is not a new application but a request. The ninth paragraph of Article 40 of the Guideline:

“before making a new device application for the relevant device notification or notifications, the company may request the re-evaluation of the decision in question by using the ‘Submit Request’ screen in ÜTS, selecting ‘Medical Device Document and Product Registration Procedures’ as the ‘Request Area’, and entering the Device Application Number assigned to that device application in the ‘Medical Device/Document Application Number’ field of that area.”

The same paragraph sets two conditions for the request: the application number must be entered, and a reference must be made to the legislation or Guideline provision objected to; requests that meet neither condition are not taken into consideration. The last sentence of the paragraph:

“The Authority is not held responsible for applications that are sent for review by reapplying without being corrected through the request.”

For common rejection reasons and their solutions, see our guide Why Is a ÜTS Application Rejected.

13. Accessories, Spare Parts, Number of Products in Content and Product Description

Three fields on the registration screen frequently produce errors. For the “Is it a component/accessory of a device?” field, subparagraph (dd) of the first paragraph of Article 27 of the Guideline:

“This field is selected as ‘Yes’ if the relevant device is an accessory of another medical device, or is a part intended to replace a part or component of a device and significantly changes the performance or safety characteristics or intended use of the device; it is selected as ‘No’ if it is not a component/accessory of another device.”

When “Yes” is selected, the “Branch Types of the Main Device” field opens and at least one branch type code is selected (subparagraph (ee)). For spare parts, sub-subparagraph 5 of the same subparagraph is definitive:

“Parts defined as spare parts are not registered in ÜTS.”

The “Number of Products in Content” field is mandatory and the smallest package is taken as the basis. Sub-subparagraph 2 of subparagraph (i):

“Since registration in ÜTS is made over the smallest package placed on the market (unit package or single package), the value entered in the relevant field equals the number of products in the smallest device package.”

According to the same subparagraph, for kits made up of more than one component 1 is entered in this field; if the smallest package contains several identical kits, the number of kits is entered. Multiple sales packages are added later to the “Market Placement Information” field.

“Product Description” is optional but becomes mandatory for configurable devices. Sub-subparagraph 3 of subparagraph (ı):

“For configurable devices, it is mandatory to write in this field that the relevant device is a configurable device with more than one configuration, and to enter the name, variant/model and, if any, reference/catalog number of each configuration of this device and other information identifying the relevant variant.”

The information entered in these fields must be consistent with the label, instructions for use and declaration of conformity; a registration that contradicts the documents is a reason for rejection. For the whole registration, see our ÜTS consultancy page.

14. Frequently Asked Questions

Which barcode is used for ÜTS product registration?

As a rule, it is made with the barcode of the smallest package placed on the market (unit or single package). The barcode is entered unchanged, exactly as allocated under the GS1, HIBCC, ICCBBA or IFA standard.

Can a product be registered without document registration?

No. The documents in a product registration are selected from among the documents that were applied for under Article 39 of the Guideline and have become registered in ÜTS.

How many devices can be added to one device application?

In optical, active implantable, Class 3 and other products applications there are at most 500 device notifications. A single declaration is selected in the application, and under MDR only devices linked to the same Basic UDI-DI are placed in the same application.

Can a catalog page be uploaded as a product image?

No. According to the Guideline, the catalog itself or an image containing more than one device that makes the device hard to understand is not uploaded.

What is the difference between Draft and Ready for Application?

If the “Is it ready for application?” field is selected as “No”, the registration becomes Draft; if “Yes”, it becomes Ready for Application. The device application screen lists only the declarations linked to devices that are Ready for Application.

For the preparation and application of your ÜTS product registrations: Medical Device ÜTS Registration Consultancy

Sources

How many devices can be added to a custom-made device or SPP application?

At most 5 medical device notifications; for other application types the limit is 500 (TCKKD-KLVZ-03 Article 40, fifth and sixth paragraphs).

Can devices with different Basic UDI-DIs be in the same application?

No. In device applications under MDR and IVDR (excluding ITCs), applications are only permitted for devices linked to the same Basic UDI-DI code; under the old regulations the criterion is the same GMDN code.

How is a rejection decision objected to?

Before reapplying, a re-evaluation of the decision is requested from the “Submit Request” screen in ÜTS by selecting the “Medical Device Document and Product Registration Procedures” request area and entering the device application number together with a reference to the legislation.

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara