ÜTS transactions consist of company registration, user and authorization definition, document registration, product (device) registration, updating of records and individual movement notifications. The screen, the performing user and the rule of each transaction are regulated in separate articles of the TİTCK guidance.
This article is a list of transaction types. It gives each transaction briefly with its legal basis and links to the relevant article for details. For the general map of ÜTS registration, see our ÜTS registration article.
To have your ÜTS transactions carried out from a single source: Medical Device ÜTS Registration Consultancy
Contents
1. The Basis of ÜTS Transactions
2. Company Registration
3. User and Authorization Transactions
4. Document Registration
5. Product (Device) Registration and Device Application
6. Bulk Transactions
7. Updating, Deletion and Data Change
8. Individual Movement Notifications
9. Free Sale Certificate and Report a Request
10. Frequently Asked Questions
1. The Basis of ÜTS Transactions
Article 8 of the Medical Device Sales, Advertising and Promotion Regulation places both a registration and a notification obligation on the sales center (unofficial translation):
“Sales centers are obliged to register the medical devices they place on the market or make available on the market in the registration and information management system established by the Authority, to keep the registered information up to date, and to make individual movement notification transactions in order to ensure the traceability of these devices.”
Registration transactions are carried out under TİTCK's ÜTS Registration Guidance coded TCKKD-KLVZ-03, and notifications under the Individual Movement Notifications Guidance.
2. Company Registration
“ARTICLE 5- (1) The registration of companies that sell and/or apply medical devices to ÜTS is carried out if they are authorized under the relevant legislation.”
Company records are MERSİS- or VEDOP-based; those of pharmacies are EBS-based. According to the ÜTS User and Company Transactions Help Guide, a medical device company record has a single field of activity:
“A medical device company in ÜTS can have only 1 field of activity.”
For details, see our article on ÜTS company registration; for the e-signature and e-Government flow of the application, see How to Apply for ÜTS.
3. User and Authorization Transactions
“There are 3 types of users in ÜTS: Normal User, High-Ranking User, Authorized User”
Authorized users are the Signatory, the Authority Official and the Hospital Officer. A new user is added from the "User Account Transactions" menu and assigned to user groups. Users are not deleted, they are deactivated. For the person who will perform registration transactions, Article 11 of TCS-KLVZ-08 provides:
“(3) As of 1/1/2025, persons who do not hold a medical device registration staff qualification certificate are not allowed to register medical devices in the Product Tracking System (ÜTS).”
For login routes and switching accounts, see our ÜTS login article.
4. Document Registration
Certificates, declarations of conformity, instructions for use and other documents are entered into ÜTS and a document application is made to the Authority (TCKKD-KLVZ-03, Article 39). The application is created from the "Document Application Transactions" screen in the "Document" tab. An application contains at most 5 document notifications. Document types and application types are in our ÜTS document registration article.
5. Product (Device) Registration and Device Application
A device is registered with its barcode, class, linked documents and images; then a device application is made to the Authority (Articles 27 and 40). The guidance regulates devices under the former regulations in Article 25, those under MDR in Article 27 and those under IVDR in Article 30. System/procedure pack (SİP) and custom-made device (ITC) records are in separate articles (Articles 26, 28 and 29).
For registration fields and application steps, see our ÜTS product registration article.
6. Bulk Transactions
The following transactions can be made in bulk from the "Bulk Transactions" area on the "List My Medical Devices" screen (Articles 31-38):
Bulk adding/updating of MDD/AIMDD/IVDD, MDR and IVDR medical devices
Bulk device deletion
Bulk update of products without application
Bulk adding of placing-on-the-market information
Bulk adding of storage/use conditions
Bulk adding of measurement information
To use the template, see our ÜTS bulk device notification Excel article.
7. Updating, Deletion and Data Change
When a device record is deleted, the result depends on the status of the record. Article 34 of the guidance:
“(7) Device records in "Draft" and "Ready for Application" status are deleted from ÜTS.”
“(9) Device records can be deleted only by the relevant company. Device record deletion in ÜTS is not performed by the Authority.”
Records in "Registered" status and, except those with a Reject-3 decision, "Rejected" status are not deleted but are set to "Deleted" status. Some information, such as associating a new catalog or uploading a new product image, can be updated without an application (Article 35). For data differences between records with the same barcode, an imported product data change request is opened.
Some changes in company information come from ÇKYS:
“(7) Where the responsible manager, title and address of the company change, if the relevant information is updated in ÇKYS, these changes are updated automatically in ÜTS.”
For transfer and title change, see our article on company transfer, title and address change.
8. Individual Movement Notifications
Movements of registered devices are notified under the Individual Movement Notifications Guidance. The main types are:
Entry into the system: production (Article 6), import (Article 7) and inventory (Article 8) notification.
Movement within the system: giving (Article 9) and receiving (Article 10) notification.
Exit from the system: use (Article 12), consumption (Article 13), giving to the consumer (Article 14), HEK/loss (Article 15), giving to an undefined place (Article 16) and export (Article 17) notification.
Recall: recall giving (Article 18) and recall receiving (Article 19) notification.
“No training from an authorized organization is required for the notification transactions (receiving-giving notifications, etc.) of a medical device registered in ÜTS.”
For a map of all types, see our article ÜTS individual movement notifications. One by one: inventory notification, giving and receiving notification, use notification.
9. Free Sale Certificate and Report a Request
“ARTICLE 41- (1) The Authority issues, for export purposes and upon the request of the domestic manufacturer or of the EU authorized representative of a foreign manufacturer established in Türkiye, a free sale certificate declaring that the manufacturer or, as applicable, the authorized representative has a registered place of business in the country and that the device or devices in question bearing the CE mark under the Medical Device Regulations may be marketed in Türkiye and in EU member countries.”
The application is made through ÜTS; for details, see our article on the free sale certificate. Requests such as adding GMDN, EMDN, GTİP or a notified body, re-evaluation of a rejection decision and system errors are submitted from the "Report a Request" screen.
10. Frequently Asked Questions
In which order are ÜTS transactions made?
First company registration, then document registration and document application, then product registration linked to the registered documents and the device application. Individual movement notifications are made after the device is registered.
Can a company register in ÜTS with more than one field of activity?
No. According to the Help Guide, a medical device company has only one field of activity in ÜTS; a separate company application is made for each field of activity.
Is a registration staff certificate required for individual notifications?
No. According to TCS-KLVZ-08, no training requirement applies to notification transactions of a registered device. The qualification certificate requirement is for medical device registration.
Can a registered device record be deleted?
A record in "Registered" status is not deleted from ÜTS, it is set to "Deleted" status. Only the relevant company performs the deletion.
Is ÜTS updated separately when the title or address changes?
If the responsible manager, title and address are updated in ÇKYS, the change is reflected in ÜTS automatically.
For your company, document, product registration and notification transactions: Medical Device ÜTS Registration Consultancy
Sources