Manufacturer Obligations: Article 10 of the Tıbbi Cihaz Yönetmeliği, Quality System, Retention Periods and ÜTS

12 hours ago
8 min read
imalatçının genel yükümlülükleri madde 10 ve ÜTS

SUMMARY

Manufacturer obligations are regulated in Article 10 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation). The Article consists of 16 paragraphs, covering everything from design and manufacture to risk management, technical documentation, document retention, the quality management system and financial coverage.

In this post we give the paragraphs of Article 10 as verbatim quotations, together with the ÜTS equivalent from the TİTCK ÜTS Registration Guide TCKKD-KLVZ-03 and the Individual Movement Notifications Guide. For the registration of a domestic manufacturer, see also our post on medical device manufacturer ÜTS registration.

For your declaration of conformity and ÜTS registrations: Medical Device ÜTS Registration Consultancy

Contents

1. Who Is the Manufacturer?

“(t) Manufacturer: the natural or legal person who manufactures or fully refurbishes a device, or has a device designed, manufactured or fully refurbished, and markets that device under its name or trademark,”

The cases in which a distributor or importer also assumes manufacturer obligations are in Article 16; the details are in our post on being treated as a manufacturer and Article 16.

2. Design, Risk Management, Clinical Evaluation

  • “ARTICLE 10 - (1) When placing their devices on the market or putting them into service, manufacturers shall ensure that they have been designed and manufactured in accordance with the requirements of this Regulation.”

  • “(2) Manufacturers shall establish, document, implement and maintain a risk management system as described in Section 3 of Annex I.”

  • “(3) Manufacturers shall conduct a clinical evaluation in accordance with the requirements set out in Article 61 and Annex XIV, including a post-market clinical follow-up.”

For clinical evaluation, see our post on clinical evaluation and PMCF.

3. Technical Documentation

  • “(4) Manufacturers of devices, other than custom-made devices, shall draw up and keep up to date technical documentation for those devices. The technical documentation shall be such as to allow the assessment of the conformity of the device with the requirements of this Regulation. The technical documentation shall include the elements set out in Annexes II and III.”

  • “(5) Manufacturers of custom-made devices shall draw up documentation in accordance with Section 2 of Annex XIII, keep it up to date and keep it available for the Agency on request.”

4. EU Declaration of Conformity, CE, UDI and Registration

  • “(6) Manufacturers of devices, other than custom-made or investigational devices, that have demonstrated conformity of their devices with the applicable requirements following the applicable conformity assessment procedure, shall draw up an EU declaration of conformity in accordance with Article 19 and affix the CE marking of conformity in accordance with Article 20.”

  • “(7) Manufacturers shall comply with the obligations relating to the Unique Device Identification system referred to in Article 27 and with the registration obligations referred to in Articles 29 and 31.”

Articles 29 and 31 concern the registration of the device and of the manufacturer in EUDAMED. For the declaration, see our post on EU declaration of conformity and CE marking; for UDI, our post on UDI system and Basic UDI-DI.

5. Document Retention Periods

“(a) keep the technical documentation, the EU declaration of conformity and, if applicable, a copy of any relevant certificate issued in accordance with Article 56, including any amendments and supplements, available for the Agency for a period of at least 10 years after the last device covered by the EU declaration of conformity has been placed on the market. In the case of implantable devices, the period shall be at least 15 years after the last device has been placed on the market.”

The period runs from the placing on the market of the last device covered by the declaration. Under subparagraph (b) of the eighth paragraph, the manufacturer provides the whole or a summary of the technical documentation at the Agency’s request; under subparagraph (c), if it is established outside Türkiye or the EU, it ensures that its authorized representative has the necessary documentation permanently available.

6. Quality Management System

“Manufacturers of devices, other than investigational devices, shall establish, document, implement, maintain, keep up to date and continually improve a quality management system that shall ensure compliance with this Regulation in the most effective manner and in a manner that is proportionate to the risk class and the type of device.”

Under the ninth paragraph, the quality management system covers at least the following aspects:

  • “(a) a strategy for regulatory compliance, including compliance with conformity assessment procedures and procedures for management of modifications to the devices covered by the system;”

  • “(b) identification of applicable general safety and performance requirements and exploration of options to address those requirements;”

  • “(c) responsibility of the management;”

  • “(ç) resource management, including selection and control of suppliers and sub-contractors;”

  • “(d) risk management as set out in Section 3 of Annex I;”

  • “(e) clinical evaluation in accordance with Article 61 and Annex XIV, including PMCF;”

  • “(f) product realisation, including planning, design, development, production and service provision;”

  • “(g) verification of the UDI assignments made in accordance with the third paragraph of Article 27 to all relevant devices and ensuring the consistency and validity of information provided in accordance with Article 29;”

  • “(ğ) setting-up, implementation and maintenance of a post-market surveillance system in accordance with Article 81;”

  • “(h) handling communication with competent authorities, notified bodies, other economic operators, customers and/or other stakeholders;”

  • “(ı) processes for reporting of serious incidents and field safety corrective actions in the context of vigilance;”

  • “(i) management of corrective and preventive actions and verification of their effectiveness;”

  • “(j) monitoring and measurement of outputs, data analysis and product improvement.”

7. Surveillance, Language, Corrective Action and Cooperation

  • “(10) Manufacturers shall implement and keep up to date the post-market surveillance system in accordance with Article 81.”

  • “The particulars on the label shall be indelible, easily legible and clearly comprehensible to the intended user or patient.”

  • “(13) Manufacturers shall have a system for recording and reporting of incidents and field safety corrective actions as described in Articles 85 and 86.”

Under the eleventh paragraph, the information in Annex I, section 23 (label and instructions for use) accompanies the device; for the language requirement, see our article on Turkish label and instructions for use. Under the twelfth paragraph, where a device is not in conformity, the manufacturer promptly takes the necessary corrective action and informs the distributors, the authorized representative and the importers:

“Where the device presents a serious risk, manufacturers shall immediately inform the Agency and/or the competent authorities of the EU member states in which they made the device available, and, where applicable, the notified body that issued a certificate for the device in accordance with Article 56, in particular of the non-compliance and of any corrective action taken.”

For the notification process, see our article on vigilance and serious incident reporting.

“(b) If the manufacturer does not cooperate or the information and documentation provided are incomplete or incorrect, the Agency may take all appropriate measures to prohibit or restrict the making available of the device on the market, withdraw it from the market or recall it, in order to ensure the protection of public health and patient safety, until the manufacturer cooperates or provides complete and correct information.”

Under point (a) of the fourteenth paragraph, the manufacturer provides, on request, the information and documentation demonstrating conformity in Turkish and/or the relevant official EU languages; the Agency may request free samples of, or access to, the device.

8. Contract Manufacturing and Financial Coverage

  • “(15) Where manufacturers have their devices designed or manufactured by another legal or natural person, the information on the identity of that person shall be part of the information to be submitted in accordance with paragraph 4 of Article 29.”

  • “Manufacturers shall ensure that measures are in place to provide sufficient financial coverage in respect of their potential liability under the applicable legislation, proportionate to the risk class, type of device and the size of the enterprise.”

For the country of origin in contract manufacturing, see our article on contract manufacturing and country of origin. The manufacturer's notification obligation in case of a supply interruption is set out in Article 10/A; the details are in our article on notification of supply interruption.

9. Person Responsible for Regulatory Compliance

“ARTICLE 15 - (1) Manufacturers shall have within their organization at least one person responsible for regulatory compliance who possesses the requisite expertise in the field of medical devices.”

For the qualifications and the micro/small enterprise exception, see our article on the person responsible for regulatory compliance (PRRC), and for the service, see our page on the regulatory compliance person service.

10. The Domestic Manufacturer in ÜTS

A domestic manufacturer's declaration of conformity is registered in ÜTS through an e-signed application. TCKKD-KLVZ-03 (ÜTS Registration Guide), Article 13, first paragraph:

  • “(a) The application must be made by selecting “İmalatçı Firmalar için e-imzalı Belge Başvurusu” (e-signed Document Application for Manufacturer Firms) as the Document Application Type.”

  • “(b) The option “Belge kapsamındaki ürünlerin üreticisi firması siz misiniz?” (Are you the manufacturer of the products covered by the document?) is selected as “Evet” (Yes).”

  • “(c) Documents whose original language is Turkish are accepted.”

“ÜTS allows a document application only if the document in question is electronically signed by the “Mesul Müdür” (Responsible Manager) or the “İmza Yetkilisi” (Authorized Signatory).”

In the product registration, if the legal manufacturer that issued the declaration of conformity is established in Türkiye, “İmal” (Manufactured) is selected in the "İthal/İmal Bilgisi" (Import/Manufacture Information) field (Article 27, third paragraph). Contract manufacturing is also asked separately in the product registration (Article 27, seventh paragraph):

“(gg) Was the product manufactured for its legal manufacturer by another manufacturer?:”

“2) This field is selected as “Evet” (Yes) if the legal manufacturer has the device designed and/or manufactured by another natural or legal person, and as “Hayır” (No) if it does not have the device designed and/or manufactured by another natural or legal person (that is, if the legal manufacturer produces in its own facilities).”

The individual units of a registered device enter the system through a production notification. Individual Movement Notifications Guide, Article 6:

  • “(1) A “production notification” is made for products whose country of origin in the product registration is “Türkiye”.”

  • “(2) The manufacturer makes a production notification only for medical devices that are registered in ÜTS or that meet the conditions announced by the Agency under the transitional provisions of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation).”

For declaration preparation, see our page on MDR/IVDR declaration of conformity preparation service, and for document registration, see our article on ÜTS document registration.

11. Frequently Asked Questions

How long do manufacturers keep their documents?

Under the eighth paragraph of Article 10, the technical documentation, the EU declaration of conformity and copies of certificates are kept for at least 10 years after the last device covered by the declaration has been placed on the market, and for at least 15 years for implantable devices.

Is a quality management system mandatory for every manufacturer?

Under the ninth paragraph, with the exception of investigational devices, manufacturers establish a quality management system proportionate to the risk class and type of the device.

If we use contract manufacturing, do we have to notify it?

Under the fifteenth paragraph of Article 10, the identity details of the person carrying out the manufacture are part of the information submitted under the fourth paragraph of Article 29. In the ÜTS product registration, the field “Ürün, yasal imalatçısı tarafından başka bir imalatçıya mı ürettirildi?” (Was the product manufactured for its legal manufacturer by another manufacturer?) is also filled in.

Who e-signs the domestic manufacturer's declaration of conformity?

Under Article 13 of TCKKD-KLVZ-03 (ÜTS Registration Guide), the person registered in ÜTS as the “Mesul Müdür” (Responsible Manager) or “İmza Yetkilisi” (Authorized Signatory) of the company signs electronically, and the application is made with the type “İmalatçı Firmalar için e-imzalı Belge Başvurusu” (e-signed Document Application for Manufacturer Firms).

How many items does the quality management system cover?

The ninth paragraph lists at least 13 items, from point (a) to point (j); these include verification of UDI assignment, post-market surveillance and vigilance reporting.

For your declaration of conformity and ÜTS registrations as a domestic manufacturer: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara