Importing Medical Devices from Pakistan: ÜTS Registration and Document Requirements
A Turkish company that wants to import medical devices from Pakistan faces two separate systems: DRAP registration on the Pakistani side, and CE marking and ÜTS registration on the Turkish side. A registration or licence issued by DRAP is not sufficient for placing a device on the market in Türkiye. This page explains which documents the importer must prepare, with which authentication and in what order, to register a Pakistani manufacturer's product in ÜTS and clear it through customs.
Why DRAP registration is not enough in Türkiye
The competent authority for medical devices in Pakistan is the Drug Regulatory Authority of Pakistan (DRAP), within the Ministry of National Health Services, Regulations and Coordination (dra.gov.pk). According to DRAP's official page, devices are evaluated by the Medical Device Board under the Medical Devices Rules, 2017; Class A devices are published in the “enlisted” lists, Class B, C and D devices in the “registered” lists, and manufacturers and importers in the lists of licensed establishments.
This registration is for the Pakistani domestic market. Placing a device on the market in Türkiye depends on conformity with the Medical Device Regulation, and this conformity is shown by the CE marking. A surgical instrument or syringe that has DRAP registration in Pakistan but no CE marking cannot be registered in ÜTS in Türkiye.
“Except for custom-made or investigational devices, devices that are considered to conform to the requirements of this Regulation bear the CE marking of conformity as shown in Annex V.”
Source: Medical Device Regulation (OG 31499, 2.6.2021), Article 20, paragraph 1.
From the Pakistani manufacturer, the importer asks not for a DRAP certificate but for the EU declaration of conformity drawn up under the MDR and, if the class requires it, the notified body certificate. The Regulation places the duty to verify this on the importer.
“1) that the device bears the CE marking and that the EU declaration of conformity has been drawn up, 2) that the manufacturer is identified and that an authorized representative has been appointed by the manufacturer in accordance with Article 11, 3) that the device is labelled in accordance with this Regulation and that the required instructions for use accompany the device, 4) where applicable, that a UDI has been assigned by the manufacturer in accordance with Article 27.”
Source: Medical Device Regulation, Article 13, paragraph 2, subparagraph (a).
A DRAP document appears in the Turkish file only as supporting information; it has no equivalent as a document type in ÜTS. The logic explained in our article on CE and ÜTS for FDA-approved products applies to DRAP as well.FDA onaylı ürünlerde CE ve ÜTS yazısında anlatılan mantık DRAP için de aynen geçerlidir.
Authorized representative requirement for the Pakistani manufacturer
Pakistan is neither an EU member nor Türkiye; therefore the manufacturer cannot supply the device to Türkiye without appointing an authorized representative established in Türkiye or an EU member state.
“If the manufacturer is not established in Türkiye or in EU member states, the device may be placed on the market only if the manufacturer appoints an authorized representative.”
Source: Medical Device Regulation (OG 31499, 2.6.2021), Article 11(1).
“If the manufacturer of a device is not established in a Member State, the device may be placed on the Union market only if the manufacturer appoints a single authorized representative [MDR and IVDR Article 11(1)].”
Source: MDCG 2022-16, TİTCK Turkish translation, Section 4 Appointment and mandate.
The footnote in the TİTCK translation of MDCG 2022-16 explicitly states that this representative's field of duty also covers Türkiye through the Customs Union.
“‘Union market’ refers to the territories of the EU Member States and, owing to the European Economic Area (EEA), is extended to Norway, Liechtenstein and Iceland and, through the Customs Union Agreement, to Türkiye.”
Source: MDCG 2022-16 Guidance on Authorised Representatives, TİTCK Turkish translation, footnote 1.
In practice, if the Pakistani manufacturer already has an authorized representative in the EU, the same authorization document is used for Türkiye; otherwise the manufacturer may appoint a representative established in Türkiye. Medex is the authorized representative registered in EUDAMED under SRN TR-AR-000057550; the scope of this service for Pakistani manufacturers is explained on the Pakistan page of medexeurep.com. When talking to the manufacturer in Pakistan, the importer should ask for the authorization document (mandate) and the representative's SRN at the very start, because document registration in ÜTS does not proceed without this information.medexeurep.com Pakistan sayfasında anlatılır. İthalatçı Pakistan'daki üreticiyle görüşürken yetki belgesini (mandate) ve temsilcinin SRN'sini en başta ister, çünkü ÜTS'de belge kaydı bu bilgi olmadan ilerlemez.
“For ‘EU Declaration of Conformity (MDR/IVDR)’ type document registration notifications covered by this article that belong to manufacturers not established in an EU member state or in Türkiye, it is mandatory to make an entry in the ‘EU Authorized Representative’ field on the ÜTS registration screen.”
Source: ÜTS Registration Guide TCKKD-KLVZ-03 V4.0, Article 12, paragraph 7, subparagraph (a).
“The original of the document and its Turkish translation by a sworn translator are uploaded together in the ‘Authorization Document’ field. Neither an apostille nor embassy/consulate legalization is required for the original of the authorization document.”
Source: ÜTS Registration Guide (TCKKD-KLVZ-03, 13.05.2026), Article 12, paragraph 7, subparagraph (c).
Apostille and sworn translation for Pakistani documents
According to the official status table on hcch.net, Pakistan acceded to the Hague Apostille Convention on 8 July 2022, and the Convention entered into force for Pakistan on 9 March 2023. The same list also records the states that objected to Pakistan's accession (Germany, Czechia, Poland, Austria, Finland, the Netherlands, Denmark, Greece); Türkiye is not on this objection list. A declaration of conformity issued in Pakistan can therefore be authenticated for ÜTS by obtaining an apostille from the Pakistani authorities.
“a) Apostille-certified document or,”
Source: ÜTS Registration Guide, Article 12(1)(a).
“If the manufacturer is in an EU member state, the apostille certification obtained from the competent authority of the country where the manufacturer is established; if the manufacturer is established in a country that is not an EU member, the apostille certification obtained from the competent authority of the country where the manufacturer or the EU Authorized Representative appointed by the manufacturer is established is accepted.”
Source: ÜTS Registration Guide, Article 8/1-a.
“The original of the apostille-certified document must be uploaded to ÜTS; there is no need to upload a copy of this document issued by a Turkish notary or a notary of another country.”
Source: ÜTS Registration Guide TCKKD-KLVZ-03 V4.0, Article 12, paragraph 1, subparagraph (a), sub-item 2.
The list of objecting states matters to the importer at one point: if the Pakistani manufacturer's authorized representative is in an objecting country such as Germany or the Netherlands, the apostille route should be planned in Pakistan, not in the representative's country, or through the other routes Türkiye accepts. The Guide accepts consular legalization and the e-signature as alternatives to the apostille.
“b) The document certified by the Turkish embassy/consulate in the country of the manufacturer or, 1) If the manufacturer is established in a country that is not an EU member, certification by the Turkish embassy/consulate located in the country where the manufacturer's EU Authorized Representative is established is also accepted.”
Source: ÜTS Registration Guide (TCKKD-KLVZ-03, 13.05.2026), Article 12/1-b.
“ç) For documents signed electronically (e-signed) by the manufacturer, the original of the relevant e-signed document, provided that it can be verified through e-signature verification tools/data that the document was electronically signed by its manufacturer, that the e-signature is valid and that no change has been made to the document afterwards,”
Source: ÜTS Registration Guide, Article 12, paragraph 1, subparagraph (ç).
2026/KKB-2 update: With TİTCK's Announcement No. 2026/KKB-2 dated 02.11.2026, a new route was introduced in place of the option of uploading the manufacturer's e-signed original for a foreign manufacturer's declaration of conformity and SİP declaration: TİTCK'nın 02.11.2026 tarihli 2026/KKB-2 Sayılı Duyurusu ile yabancı imalatçıya ait uygunluk beyanı ve SİP beyanında, imalatçının e-imzalı belge aslını yükleme seçeneğinin yerine yeni yol getirildi:
“It was decided to apply the option of uploading the relevant document to the relevant field in ÜTS together with an undertaking, electronically signed by the importer company's authorized person authorized by the foreign manufacturer, with an e-signature (secure e-signature) based on a qualified electronic certificate obtained from an electronic certificate service provider recognized by Türkiye.”
Source: TİTCK Announcement No. 2026/KKB-2. Required documents and steps: our ÜTS document registration guide.ÜTS belge kaydı rehberi.
“For the e-signature verification of the relevant e-signed (electronically signed) document, it is mandatory to upload the original of the relevant e-signed document to ÜTS.”
Source: ÜTS Registration Guide, Article 12/1-ç-1.
Whichever route is chosen, if the document is drawn up in English or Urdu, a Turkish translation certified by a sworn translator is also uploaded to ÜTS. The translation covers the entire document; translating only the first page is a ground for rejection.
“A complete Turkish translation, made by a sworn translator, of the document uploaded in the ‘Original Document’ field is uploaded in the ‘Turkish Document’ field.”
Source: ÜTS Registration Guide TCKKD-KLVZ-03 V4.0, Article 12, paragraph 2.
“The relevant document bears the name and surname, stamp/signature/seal of the sworn translator who made the translation, and the translator's declaration that the translation was made in conformity with the original document.”
Source: ÜTS Registration Guide TCKKD-KLVZ-03 V4.0, Article 12, paragraph 2, subparagraph (b).
Medex prepares these translations within its sworn translation service for ÜTS, with the stamp and translator's declaration, in the form ÜTS requires. For the difference between apostille and e-apostille, see our medical device apostille guide.ÜTS için yeminli tercüme hizmeti kapsamında, kaşe ve tercüman beyanı ile ÜTS'nin istediği biçimde hazırlar. Apostil ve e-apostil ayrımı için tıbbi cihaz apostil rehberi yazısına bakılabilir.
TAREKS at customs: the stage that follows ÜTS
Most products arriving from Pakistan, such as surgical instruments, syringes, gloves and suture materials, fall under the GTİP list in Annex-1 of the Communiqué on Import Control of Medical Devices. Imports of products on this list are controlled through TAREKS, and the control is carried out before the customs declaration is registered.
“All procedures related to the import control of medical devices are carried out through TAREKS and according to risk analysis.”
Source: Communiqué on Import Control of Medical Devices (2026/16), Article 4, paragraph 1.
“The controls under this Communiqué are carried out before the registration of the customs declaration, within the framework of Article 181, paragraph 4 of the Customs Regulation.”
Source: Communiqué on Import Control of Medical Devices (2026/16), Article 5/1.
For a TAREKS application, the company must first be defined in TAREKS and a company user must be authorized; this is a Ministry of Trade procedure independent of ÜTS registration.
“In order for transactions to be carried out on behalf of the company in TAREKS, the company must first be defined in TAREKS.”
Source: Communiqué on Risk-Based Control System in Foreign Trade (2025/28), Article 4/1.
“The company user must have a qualified electronic certificate, mobile e-signature or e-Devlet password in order to carry out transactions through TAREKS.”
Source: Communiqué on Risk-Based Control System in Foreign Trade (2025/28), Article 5/3.
For a consignment directed to physical inspection, the documents in Annex-2 are uploaded. Here, the EU declaration of conformity with sworn translation prepared for ÜTS serves a second time: the same document is also uploaded to TAREKS.
“3. EU Declaration of Conformity (together with a certified copy of its Turkish translation where drawn up in other languages) (*)”
Source: Communiqué on Import Control of Medical Devices (ÜGD: 2026/16), Annex-2, item 3.
“4. Turkish label and instructions for use”
Source: Communiqué on Import Control of Medical Devices (ÜGD: 2026/16), Annex-2, item 4.
The Turkish label and instructions for use is an item often missing for products from Pakistan. Our article on Turkish labels and instructions for use gives guidance on the label rules. The Communiqué also states explicitly that a TAREKS reference number does not mean that the product is compliant; responsibility remains with the importer.Türkçe etiket ve kullanım talimatı yazısı yol gösterir. Tebliğ, TAREKS referans numarasının ürünün uygun olduğu anlamına gelmediğini de açıkça yazar; sorumluluk ithalatçıda kalır.
“Whether or not it is inspected under this Communiqué, the importer is in every case responsible, under Law No. 7223, for ensuring that the products it imports comply with and are safe under all relevant legislation, including the regulations specified in Article 1, and for the accuracy of the documents accompanying the products.”
Source: Communiqué on Import Control of Medical Devices (2026/16), Article 12/1.
Origin, customs duty and preferential regime topics are outside the scope of this page; these should be clarified with your customs broker according to the Ministry of Trade's current legislation.
The importer's ÜTS steps for imports from Pakistan
- Sales center authorization certificate. For company registration as an importer in ÜTS, the medical device sales center authorization certificate obtained from the Provincial Health Directorate and the ÇKYS number are required. The process is explained on our sales center authorization certificate page. ÜTS'de ithalatçı türünde firma kaydı için İl Sağlık Müdürlüğü'nden alınan tıbbi cihaz satış merkezi yetki belgesi ve ÇKYS numarası gerekir. Süreç satış merkezi yetki belgesi sayfasında anlatılır.
- ÜTS company registration. The MERSİS-based registration is opened with the e-Devlet login of the signatory. Details: ÜTS company registration. MERSİS kaynaklı kayıt, imza yetkilisinin e-Devlet girişi ile açılır. Ayrıntı: ÜTS firma kaydı.
- Document registration. The Pakistani manufacturer's EU declaration of conformity (MDR/IVDR), the notified body certificate if any, and the authorized representative's authorization document are uploaded as apostilled or e-signed originals together with their sworn translations. See our ÜTS document registration consultancy. Pakistanlı üreticinin AB uygunluk beyanı (MDR/IVDR), varsa onaylanmış kuruluş sertifikası ve yetkili temsilci yetki belgesi, apostilli veya e-imzalı asılları ve yeminli tercümeleriyle yüklenir. Bkz. ÜTS belge kayıt danışmanlığı.
- Product registration. Each product is registered according to the document it is linked to, with UDI-DI, brand, model, GMDN and label images. See ÜTS product registration. Her ürün, bağlı olduğu belgeye göre UDI-DI, marka, model, GMDN ve etiket görselleriyle kaydedilir. Bkz. ÜTS ürün kaydı.
- Import notification. Once the product becomes registered in ÜTS, an import notification is made for each consignment. Ürün ÜTS'de kayıtlı hale geldikten sonra her parti için ithalat bildirimi yapılır.
“Companies that will make a company registration in ÜTS as ‘Manufacturer/Importer/Dealer/Exporter’ must hold a medical device sales center licence, and the ÇKYS number must be entered during company registration.”
Source: ÜTS Registration Guide, Article 5, paragraph 5.
“An ‘import notification’ is made for products whose country of origin in the product registration is not ‘Türkiye’.”
Source: TİTCK Guide on Individual Movement Notifications, Article 7, paragraph 1.
“The importer or the dealer authorized for import notification can make an import notification only for medical devices that are registered in ÜTS or that meet the conditions announced by the Agency within the scope of the transitional provisions of the Medical Device Regulation.”
Source: TİTCK Guide on Individual Movement Notifications of the Product Tracking System for Medical Devices, Article 7, paragraph 4.
For the lot, serial and expiry date fields of the notifications, use our ÜTS individual notification procedures page and our individual movement notifications article. For companies bringing the same Pakistani manufacturer's products via India, the differences in authentication on our page on importing medical devices from India are important.ÜTS tekil bildirim işlemleri sayfası ve tekil hareket bildirimleri yazısı kullanılır. Aynı Pakistanlı üreticinin ürünlerini Hindistan üzerinden getiren firmalar için Hindistan'dan tıbbi cihaz ithalatı sayfasındaki tasdik farkları önemlidir.
Summary table for Pakistan
TopicSituation for PakistanCountry authorityDrug Regulatory Authority of Pakistan (DRAP), Medical Devices Rules 2017, Medical Device Board (dra.gov.pk)Is an EU/TR authorized representative required?Yes. The manufacturer is not established in the EU or Türkiye; under Article 11 of the Regulation and Article 12/7 of the ÜTS Guide, the authorization document is uploaded to ÜTSIs it a party to the Apostille Convention?Yes (hcch.net: accession 8-VII-2022, entry into force 9-III-2023). Türkiye is not among the objecting statesLanguage of documents to be uploaded to ÜTSOriginal document (apostilled, consular-certified or e-signed) + Turkish translation certified by a sworn translator; authorization document original + Turkish translation, no apostille required
Medical Device ÜTS Registration Consultancy for Imports from Pakistan
Medex handles ÜTS company, document and product registrations, import notifications and sworn translation for ÜTS for the devices you will bring from Pakistan. Send us your documents and we will tell you in writing which authentication and translation is required. Request a quote, call +90 505 580 46 22 or write to info@medexmedikal.com.Fiyat teklifi alın, +90 505 580 46 22 numarasını arayın veya info@medexmedikal.com adresine yazın.
Frequently Asked Questions
Can I register a DRAP-registered device in ÜTS without the CE marking?
No. Under Article 20 of the Medical Device Regulation, devices placed on the market in Türkiye bear the CE conformity marking, and Article 13 gives the importer the duty to verify that the EU declaration of conformity has been drawn up. DRAP registration is not accepted as a document type in ÜTS.
Must a Pakistani manufacturer appoint a separate authorized representative for Türkiye?
Since the manufacturer is not established in an EU member state or in Türkiye, an authorized representative is mandatory. The representative may be established in the EU or in Türkiye; the footnote of MDCG 2022-16 states that the Union market covers Türkiye through the Customs Union. The authorized representative field is mandatory when registering the EU declaration of conformity in ÜTS.
Is an apostille obtained in Pakistan valid in ÜTS?
According to the HCCH status table, the Convention entered into force for Pakistan on 9 March 2023, and Türkiye is not among the objecting states. Article 12/1-a of the ÜTS Registration Guide accepts an apostille obtained from the competent authority of the country where the manufacturer is established.
What routes exist instead of an apostille?
Under Article 12/1-b of the Guide, certification by the Turkish embassy/consulate in the manufacturer's country is accepted, and under Article 12/1-ç, the original of a document signed with the manufacturer's valid e-signature is accepted. If an e-signed document cannot be verified, it is rejected.
Does a TAREKS application replace ÜTS registration?
No. TAREKS is the Ministry of Trade's pre-customs import control, while ÜTS is TİTCK's registration system. Article 12/3 of the Communiqué states that the TAREKS reference number cannot be used as proof of the product's conformity. An import notification can only be made for products registered in ÜTS.