System and Procedure Pack (SPP): Kit and Set Preparation, Article 22 Declaration, Sterilization, Device Status and ÜTS SPP Registration

12 hours ago
8 min read
Sistem ve işlem paketi SİP beyanı ve ÜTS kaydı

SUMMARY

A system and procedure pack (SPP) is the placing on the market, as a kit or set, of devices bearing the CE marking, assembled with other devices or products in accordance with their intended purpose. The person who assembles the pack prepares a declaration under Article 22 of the Medical Device Regulation.

In this article we give, with verbatim quotations (unofficial translation) from the Regulation and the TİTCK ÜTS Registration Guidance TCKKD-KLVZ-03, the definitions, the content of the declaration, sterilization, the cases in which the pack is considered a device in its own right, and SPP registration in ÜTS.

​For ÜTS SPP registration of your kits and sets: Medical Device ÜTS Registration Consultancy

Contents

  • 1. What Is a System and Procedure Pack?

  • 2. Which Products May Be Put in the Pack?

  • 3. What Does the SPP Declaration Contain?

  • 4. The Person Who Sterilizes the Pack

  • 5. When Is the Pack Considered a Device in Its Own Right?

  • 6. CE Marking, Labelling and Retention of the Declaration

  • 7. SPP Declaration Document Record in ÜTS

  • 8. SPP Record Under the MDR in ÜTS

  • 9. Frequently Asked Questions

1. What Is a System and Procedure Pack?

The definitions are in Article 3 of the Medical Device Regulation:

“v) Procedure pack: A combination of products packaged together and placed on the market with the purpose of being used for a specific medical purpose,”

“bbb) System: A combination of products, either packaged together or not, which are intended to be inter-connected or combined to achieve a specific medical purpose,”

The ÜTS Registration Guidance separately defines the person who forms the pack:

“ccc) SPP manufacturer: The natural or legal person who forms the SPP and issues the SPP Declaration for the SPP,”

​For the difference between the two concepts, see our page on what is a system and procedure pack.

2. Which Products May Be Put in the Pack?

“ARTICLE 22 -(1) Natural or legal persons shall draw up a declaration if they combine devices bearing the CE marking with the following other devices or products, in a manner compatible with the intended purpose of the devices or other products and within the limits of use specified by their manufacturers, in order to place them on the market as a system or procedure pack:”

  • “a) Other devices bearing the CE marking.”

  • “b) In vitro diagnostic medical devices bearing the CE marking in accordance with the In Vitro Diagnostic Medical Device Regulation.”

  • “c) Other products that comply with the legislation applicable to them, only where their use within a medical procedure or their presence in the system or procedure pack is justified.”

​Conditions of the paragraph: the starting point is a device bearing the CE marking; the combination is compatible with the intended purpose of the devices and products and stays within the limits of use specified by their manufacturers. For the full text of the Article, see our page on systems and procedure packs: Article 22.

3. What Does the SPP Declaration Contain?

“(2) In the declaration made pursuant to the first paragraph, the natural or legal person concerned declares that:”

  • “a) It has verified the mutual compatibility of the devices and, where applicable, other products in accordance with the manufacturers' instructions and has carried out its activity in accordance with these instructions,”

  • “b) It has packaged the system or procedure pack and supplied relevant information to users, including information to be supplied by the manufacturers of the devices or other products which are put together,”

  • “c) The activity of combining devices and, where applicable, other products as a system or procedure pack has been subject to appropriate methods of internal monitoring, verification and validation,”

  • “declares.”

​The preparation of the declaration is the subject of our article on the SPP declaration.

4. The Person Who Sterilizes the Pack

“(3) Any natural or legal person who sterilizes the systems or procedure packs referred to in the first paragraph for the purpose of placing them on the market applies, at its choice, one of the procedures specified in Annex IX or the procedure specified in Annex XI Part A. The application of these procedures and the involvement of the notified body are limited to the aspects of the procedure relating to ensuring sterility until the sterile package is opened or damaged. The natural or legal person draws up a declaration stating that sterilization has been carried out in accordance with the manufacturer's instructions.”

​According to the third paragraph of Article 14 of the ÜTS Registration Guidance, this provision applies where the pack is sterilized separately before being placed on the market, independently of whether the products in it are individually sterile. On sterilization, see our article on sterile devices and sterilization service.

5. When Is the Pack Considered a Device in Its Own Right?

“(4) Where the system or procedure pack incorporates devices that do not bear the CE marking, where the chosen combination of devices is not compatible in view of their original intended purpose, or where the sterilization has not been carried out in accordance with the manufacturer's instructions, the system or procedure pack shall be treated as a device in its own right and shall be subject to the relevant conformity assessment procedure under Article 52. The natural or legal person shall assume the obligations incumbent on manufacturers.”

Three cases are listed:

  • The pack contains a device that does not bear the CE marking.

  • The combination of devices is not compatible in view of their original intended purposes.

  • Sterilization is not carried out in accordance with the manufacturer's instructions.

​In these cases the person forming the pack assumes the obligations of a manufacturer. For manufacturer obligations, see our article on manufacturer obligations.

6. CE Marking, Labelling and Retention of the Declaration

“(5) The systems or procedure packs referred to in the first paragraph shall not bear an additional CE marking but shall bear the name, registered trade name or registered trade mark of the person referred to in the first and third paragraphs, as well as the address at which that person can be contacted. Systems or procedure packs shall be accompanied by the information referred to in item 23 of Annex I. The declaration referred to in the second paragraph shall be kept at the disposal of the Agency, after the system or procedure pack has been put together, for the period applicable under the eighth paragraph of Article 10 to the devices that have been put together. Where those periods differ, the longer period shall apply.”

According to the second paragraph of Article 29 of the Regulation, before placing an SPP that is not a custom-made device on the market, the responsible person assigns to the pack a Basic UDI-DI in accordance with the rules of the issuing entities.

7. SPP Declaration Document Record in ÜTS

The SPP Declaration is a separate document type in ÜTS:

“ARTICLE 14- (1) The "SPP Declaration (MDR) (Article-22)" document is the declaration of natural or legal persons defined as SPP manufacturers, together with devices within the scope of the MDR bearing the CE marking;”

For the application of a domestic SPP manufacturer, Article 16 of the Guidance:

“a) It is mandatory to apply by selecting "e-Signed Document Application for Manufacturer Companies" as the Document Application Type.”

“c) Documents whose original language is Turkish are accepted.”

​The document application for a declaration of a foreign SPP manufacturer is made according to the procedure in Article 15 of the Guidance; the sworn Turkish translation is uploaded to the "Turkish Document" field. For the forms of approval, see our medical device apostille guide and for the document type, our page on ÜTS document type: SPP declaration.

8. SPP Record Under the MDR in ÜTS

Article 28 of the ÜTS Registration Guidance:

“ARTICLE 28- (1) The registration in ÜTS of an SPP of which at least one of the devices in its content has completed its transition to the MDR is carried out under the MDR. If all devices in the SPP concerned are placed on the market under the MDD, the registration of that SPP is made as specified in Article 26 of this Guidance.”

“(3) The information to be entered in the system on the ÜTS registration screen during SPP registration (registration/update notification) under the MDR, and their explanations, are under the headings "Identifying Information", "Import/Manufacture Information", "Classification Information", "Detail Information", "System/Procedure Pack Documents", "Product Images", "Label", "Medical Devices in the Content", "Dimension Information", "Storage/Use Condition Information", "Non-Medical Device Products in the Content" and "Application Status".”

Medical devices in the content (eleventh paragraph):

  • “2) In order to register the SPP under the MDR, it is mandatory that at least one of the devices forming the SPP is registered in ÜTS as a medical device under the MDR.”

  • “3) In order for the SPP to be registered in ÜTS, it is mandatory that all devices within the scope of the Medical Device Regulations in the SPP content are registered in ÜTS.”

  • “4) If the company that will register the SPP in ÜTS is not the importer or manufacturer of the medical devices forming the SPP, it is mandatory that there is a ÜTS dealership record for the relevant medical devices.”

​Devices are selected from the "Add Manufactured Medical Device" or "Add Dealt Medical Device" screen. If a device whose status is not "Registered" is selected, ÜTS automatically blocks the SPP application; this rejection is covered on our page device in SPP content is not registered.

Sterility field is under the "Detail Information" heading:

“a) Has the formed System or Procedure Pack been sterilized? ( MDR Article 22(3) ):”

Documents (eighth paragraph): if the sterility field is "No", the declaration of the type "SPP Declaration (MDR) (Article-22)" is selected. If "Yes", the certificate of the type "EU Certificate (MDR/IVDR)", issued under the Annex IX (Chapters I and III) or Annex XI (Part A) procedure and whose document application in ÜTS has been completed, is selected together with the declaration. The Basic UDI-DI field under the same heading is also mandatory.

​Products that are not medical devices (fourteenth paragraph): if the pack contains products outside the scope of the medical device regulations, the field "Are there products outside the scope of medical devices in the SPP content?" is set to "Yes" and these products are added to the list. After the registration is completed, "System Procedure Pack Application" is selected as the device application type. For ÜTS product registration in general, see our article on ÜTS product registration.

9. Frequently Asked Questions

What happens if a device that does not bear the CE marking is put in a set?

Under the fourth paragraph of Article 22, the pack is considered a device in its own right, is subject to conformity assessment under Article 52, and the person forming the pack assumes the manufacturer's obligations.

Does the SPP bear an additional CE marking?

No. Under the fifth paragraph, an SPP does not bear an additional CE marking; it bears the name, registered trade name or registered trade mark of the person forming the pack, together with an address at which that person can be contacted.

What does a company preparing a sterile set do?

A person who sterilizes the pack for placing it on the market applies one of the procedures in Annex IX or Annex XI Part A. The involvement of the notified body is limited to ensuring sterility until the sterile package is opened or damaged.

Must the devices in the SPP content be registered in ÜTS?

Yes. Under Article 28 of the Guidance, all devices within the scope of the Medical Device Regulations in the content must be registered in ÜTS; for registration under the MDR, at least one must be registered under the MDR.

Can we register an SPP if we are not the manufacturer of the devices in the pack?

Under Article 28 of the Guidance, if the company registering the SPP is neither the manufacturer nor the importer of the devices in the content, a ÜTS dealership record for these devices is mandatory.

​For ÜTS SPP registration of your kits and sets: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara