Refurbished Medical Devices: Full Refurbishing, Being Considered a Manufacturer, Second-Hand Sales and ÜTS

SUMMARY
The term used in the legislation for refurbished medical devices is “tamamen yenileştirme” (fully refurbishing): completely rebuilding a device already placed on the market or put into service by determining a new lifetime for it, or making a new device from used devices. Under the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation), the person who fully refurbishes a device is the manufacturer.
This article gives, as direct quotations from the Tıbbi Cihaz Yönetmeliği, the TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03 and the Individual Movement Notifications Guideline, the definitions, the cases in which a person is considered a manufacturer, the place of second-hand sales in the Regulation and what the ÜTS guidelines say on this subject.
For your ÜTS company, document and product registrations as a manufacturer: Medical Device ÜTS Registration Consultancy
Contents
1. What Is Full Refurbishing?
Article 3 of the Tıbbi Cihaz Yönetmeliği:
“ccc) Fully refurbishing: For the purposes of the definition of manufacturer, completely rebuilding a device that has already been placed on the market or put into service, or making a new device from used devices, to ensure its compliance with this Regulation, by determining a new lifetime for the refurbished device,”
The definition covers two things: completely rebuilding a device that has been placed on the market or put into service to ensure its compliance with the Regulation, and making a new device from used devices. In both cases a new lifetime is determined for the refurbished device. For a short definition, see our page on what is full refurbishing.
2. The Person Who Fully Refurbishes Is the Manufacturer
“t) Manufacturer: The natural or legal person who manufactures or fully refurbishes a device, or who has a device designed, manufactured or fully refurbished, and markets that device under its own name or trademark,”
A person who fully refurbishes a device, or who markets a fully refurbished device under their own name or trademark, falls within the definition of manufacturer, and the obligations written for manufacturers in the Regulation apply to them. The definition of manufacturer in subparagraph (y) of Article 4 of the ÜTS Registration Guideline is the same text. For the obligations of the manufacturer, see our article on manufacturer obligations.
3. Are Second-Hand Sales Within the Scope of the Regulation?
Article 1, paragraph 12 of the Tıbbi Cihaz Yönetmeliği:
“(12) This Regulation does not cover the rules relating to making medical devices available on the market by change of hands after they have already been put into service, such as second-hand sales.”
The rules relating to making a device that has been put into service available through a change of hands are outside the scope of this Regulation. If a device is fully refurbished, however, the definitions of full refurbishing and manufacturer apply.
4. Modification and Change of Intended Purpose
Article 16 of the Regulation lists the cases in which a distributor, importer or other person assumes the obligations of the manufacturer. Two of them concern intervention on a device on the market:
“b) If it changes the intended purpose of a device already placed on the market or put into service,”
“c) If it modifies a device already placed on the market or put into service in such a way that compliance with the applicable requirements may be affected,”
“assumes the obligations incumbent on manufacturers.”
The second paragraph of the article separately lists the work that is not considered a modification (providing information and translation, outer packaging changes under certain conditions). For details, see our article on distributors and importers being considered manufacturers.
5. Reprocessing of a Single-Use Device
“(2) Any natural or legal person who reprocesses a single-use device to make it suitable for further use is considered the manufacturer of the reprocessed device and assumes the obligations incumbent on manufacturers set out in this Regulation, including the obligations relating to the traceability of the reprocessed device under Part Three of this Regulation.”
This subject is regulated separately in Article 17; for details, see our article on reprocessing of a single-use device. Annex VI Part A lists the following among the information to be submitted in the device registration:
“2.6. Reprocessed single-use device (yes/no),”
6. What Do the ÜTS Guidelines Say?
The TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03 contains no separate registration type or registration field for fully refurbished or second-hand devices. The provision in the Guideline directly related to this subject is the definition of manufacturer in Article 4:
“y) Manufacturer: The natural or legal person who manufactures or fully refurbishes a device, or who has a device designed, manufactured or fully refurbished, and markets that device under its own name or trademark,”
The registration order in ÜTS as a manufacturer is company, document and product registration. For details, see our articles on ÜTS company registration, ÜTS document registration and ÜTS product registration articles. For domestic manufacturer registration, our article on medical device manufacturer ÜTS registration is also relevant.
7. A Device That Has Completed Its Economic Life: “HEK/Zayiat Bildirimi” (End of Economic Life/Loss Notification)
Article 15 of the Individual Movement Notifications Guideline:
“Article 15- (1) A ‘HEK/Zayiat Bildirimi’ (End of Economic Life/Loss Notification) is made by the institution/company/hospital for individual products that have completed their economic life or been lost.”
In the notification, the HEK/Zayiat type is selected from “HEK”, “Doğal Afet” (Natural Disaster), “Yangın” (Fire), “Çalınma” (Theft), “Stok düzeltme” (Stock correction) or “Diğer” (Other); an incorrect notification is cancelled with a “HEK/Zayiat İptal Bildirimi” (HEK/Zayiat Cancellation Notification). For the notification types, see our article on ÜTS individual movement notifications.
8. Frequently Asked Questions
Is a company that fully refurbishes and sells a device considered a manufacturer?
Yes. Under subparagraph (t) of Article 3 of the Tıbbi Cihaz Yönetmeliği, the natural or legal person who fully refurbishes a device is the manufacturer.
Is the sale of second-hand medical devices within the scope of the Tıbbi Cihaz Yönetmeliği?
Under Article 1, paragraph 12, the Regulation does not cover the rules relating to making devices that have been put into service available on the market through a change of hands, such as second-hand sales.
Is making a new device from used devices full refurbishing?
Yes. Subparagraph (ccc) of Article 3 includes making a new device from used devices within the definition of full refurbishing, provided that a new lifetime is determined.
Who is considered the manufacturer when a single-use device is reprocessed?
Under Article 17, paragraph 2, the person who reprocesses a single-use device to make it suitable for further use is considered the manufacturer of the reprocessed device.
Is there a separate registration type for refurbished devices in ÜTS?
The TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03 contains no separate registration type for refurbished or second-hand devices. The definition of manufacturer in the Guideline covers the person who fully refurbishes.
For your ÜTS company, document and product registrations as a manufacturer: Medical Device ÜTS Registration Consultancy
Sources