Is an FDA-Approved Product Sold in Türkiye? Conformity Assessment, CE Marking, Authorized Representative and ÜTS Document Registration

12 hours ago
7 min read
FDA onaylı ürünün Türkiye'de CE ve ÜTS kaydı

SUMMARY

When an FDA-approved product is to be placed on the market in Türkiye, the benchmark is the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation). The Regulation ties placing on the market to the device meeting the requirements of the Regulation, to the conformity assessment procedures in Annex IX to Annex XI, to the EU declaration of conformity and to the CE marking. FDA approval is not one of these procedures.

This article gives the relevant articles, the authorized representative requirement for a manufacturer established outside Türkiye and the EU, the document types in ÜTS and import control, as direct quotations from the official texts.

For ÜTS registration of devices of a manufacturer established abroad: Medical Device ÜTS Registration Consultancy

Contents

1. The Condition for Placing on the Market in Türkiye: Compliance with the Regulation

Article 5 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation):

“(1) Devices that satisfy the requirements of this Regulation, when duly supplied and properly installed, maintained and used in accordance with their intended purpose, may be placed on the market or put into service.”

“(2) A device shall meet the general safety and performance requirements set out in Annex I which apply to it, taking into account its intended purpose.”

“(3) Demonstration of conformity with the general safety and performance requirements shall include a clinical evaluation in accordance with Article 61.”

The criterion of the article is not the legislation of another country but the requirements of this Regulation and the general safety and performance requirements in Annex I.

2. Conformity Assessment: Annex IX to Annex XI

Article 52, paragraph 1:

“(1) Before placing a device on the market, manufacturers shall carry out an assessment of the conformity of that device, in accordance with the applicable conformity assessment procedures set out in Annexes IX to XI.”

The procedure varies with the class of the device. For class III, for example:

“(3) Manufacturers of class III devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Annex IX. Alternatively, the manufacturer may choose to apply a conformity assessment as specified in Annex X coupled with a conformity assessment as specified in Annex XI.”

For class I, paragraph 6 applies:

“(6) Manufacturers of class I devices, other than custom-made or investigational devices, shall declare the conformity of their products by issuing the EU declaration of conformity referred to in Article 19 after drawing up the technical documentation set out in Annexes II and III.”

Classification is made according to Article 51 and the rules in Annex VIII. For the choice of procedure, see our article on conformity assessment procedure selection, and for classes, our article on MDR risk class.

3. EU Declaration of Conformity

Article 19, paragraphs 1 and 3:

“ARTICLE 19 - (1) The EU declaration of conformity shall state that the requirements specified in this Regulation have been fulfilled in relation to the device covered. The manufacturer shall continuously update the EU declaration of conformity. The EU declaration of conformity shall contain, as a minimum, the information set out in Annex IV, and where the declaration is drawn up in another language for devices placed on the market in Türkiye, a Turkish translation of the declaration shall be attached.”

“(3) By drawing up the EU declaration of conformity, the manufacturer shall assume responsibility for compliance with the requirements of this Regulation and with all other legislation applicable to the device.”

For the content of the declaration, see our article on EU declaration of conformity and CE marking

4. CE Marking

Article 20:

“ARTICLE 20 - (1) Devices, other than custom-made or investigational devices, that are considered to be in conformity with the requirements of this Regulation shall bear the CE marking of conformity, as set out in Annex V.”

“(3) The CE marking shall be affixed visibly, legibly and indelibly to the device or its sterile packaging. Where such affixing is not possible or not warranted on account of the nature of the device, the CE marking shall be affixed to the packaging. The CE marking shall also appear in any instructions for use and on any sales packaging.”

“(4) The CE marking shall be affixed before the device is placed on the market.”

“(5) Where there is a notified body involved in the conformity assessment procedures referred to in Article 52, the CE marking shall be followed by the identification number of that notified body.”

Article 21 sets separate conditions for custom-made and investigational devices that do not bear the CE marking.

5. Authorized Representative for a US Manufacturer

Article 11, paragraphs 1 and 2:

“ARTICLE 11 - (1) Where the manufacturer of a device is not established in Türkiye or in an EU Member State, the device may be placed on the market only if the manufacturer designates an authorized representative.”

“(2) The designation shall constitute the authorized representative’s mandate only when it is accepted in writing by the authorized representative and shall be effective at least for all devices of the same generic device group.”

The first of the minimum tasks in the mandate:

“(a) verify that the EU declaration of conformity and the technical documentation have been drawn up and, where applicable, that an appropriate conformity assessment procedure has been carried out by the manufacturer.”

For all of the tasks, see our article on authorized representative, mandate and responsibility and for the scope of the service, see our page on the TR/EU authorized representation service.

6. Document Types Used in ÜTS

The third section of the TİTCK ÜTS Kayıt Kılavuzu (TCKKD-KLVZ-03) (the Agency’s ÜTS Registration Guide) regulates the document types in separate articles: EC Certificate (Art. 6), EU Certificate (MDR/IVDR) (Art. 7), Declaration of Conformity (Art. 10), EU Declaration of Conformity (MDR/IVDR) (Art. 11), SİP Declaration (MDR) (Article 22) (Art. 14), Instructions for Use (Art. 17), Intended Use Form/Declaration (Art. 19), Quality Management Certificate - ISO 13485 (Art. 20), Domestic Goods Certificate (Art. 22), Authorized Distributorship Certificate (Art. 23) and Catalogue (Art. 24). Among these document types there is no approval or license document issued by a foreign authority.

Scope of the EU Declaration of Conformity type:

“ARTICLE 11 - (1) The document type “EU Declaration of Conformity (MDR/IVDR)” is the document type used only in notifications and applications relating to declarations of conformity prepared under the new medical device regulations.”

“ARTICLE 7 - (1) The document type “EU Certificate (MDR/IVDR)” is the document type used only in notifications and applications relating to certificates issued by a notified body under the new medical device regulations.”

Under which approval declarations and certificates of a foreign manufacturer are uploaded to ÜTS is regulated in Articles 8 and 12 of the Guide; for details, see our article on the medical device apostille guide. For the steps of document registration, see our article on ÜTS document registration.

7. Import Control at Customs

The Tıbbi Cihazların İthalat Denetimi Tebliği (Communiqué on the Import Control of Medical Devices) (Product Safety and Inspection: 2026/16) sets the criterion of the control:

“ARTICLE 1- (1) The purpose of this Communiqué is to regulate the procedures and principles for the control, at import, of the conformity of the products listed in Annex-1 with the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation) published in the Official Gazette of 2/6/2021, repeated issue 31499, and the İn Vitro Tanı Amaçlı Tıbbi Cihaz Yönetmeliği (In Vitro Diagnostic Medical Device Regulation) published in the Official Gazette of 2/6/2021, repeated issue 31499.”

“(5) If it is determined that the EU Declaration of Conformity, test report or other documents requested by the inspection unit and uploaded to TAREKS have not been issued by the relevant party, the physical inspection shall be concluded negatively even if the other conditions are met.”

“ARTICLE 12-(1) Whether or not it is inspected under this Communiqué, the importer is in every case responsible, under Law No. 7223, for the conformity and safety of the products it imports with all relevant legislation, including the regulations referred to in Article 1, and for the accuracy of the documents accompanying the products.”

8. Frequently Asked Questions

Is FDA approval sufficient for placing on the market in Türkiye?

No. Under Article 5 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation), the device meets the requirements of this Regulation; under Article 52, the manufacturer carries out the conformity assessment with the procedures in Annex IX to Annex XI before placing on the market; and under Article 20, the device bears the CE marking.

Can an FDA document be registered in ÜTS as a document?

Among the document types listed in TCKKD-KLVZ-03, there are no approval or license documents of foreign authorities. Under the new regulation, the EU Certificate (MDR/IVDR) and EU Declaration of Conformity (MDR/IVDR) types are used.

Must a US manufacturer designate an authorized representative in Türkiye?

Yes. Under Article 11, paragraph 1, a device whose manufacturer is not established in Türkiye or in an EU Member State may be placed on the market only if the manufacturer designates an authorized representative.

What if the EU declaration of conformity is in English?

Under Article 19, paragraph 1, if the declaration for devices placed on the market in Türkiye is drawn up in another language, a Turkish translation is attached.

Which legislation is checked at customs?

Under Article 1 of the Import Control Communiqué, conformity with the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation) and the İn Vitro Tanı Amaçlı Tıbbi Cihaz Yönetmeliği (In Vitro Diagnostic Medical Device Regulation) is checked.

For ÜTS registration of devices of a manufacturer established abroad: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara