Importer Obligations and Distributor Obligations: Articles 13 and 14 of the Tıbbi Cihaz Yönetmeliği and ÜTS

SUMMARY
The importer’s obligations are regulated in Article 13 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation), and the distributor’s obligations in Article 14. In both, certain conditions are verified before the device is placed on the market, the relevant parties are informed in case of non-conformity, and records are kept.
This article gives Articles 13 and 14, the traceability provision and their counterparts in ÜTS as direct quotations from the official texts. For the cases in which an importer or distributor is considered the manufacturer, see our article on considered a manufacturer and Article 16.
For ÜTS registration and notifications of your imported products: Medical Device ÜTS Registration Consultancy
Contents
1. Who Is an Importer and a Distributor?
“(y) Importer: the natural or legal person established in Türkiye who places a device from a third country on the market in Türkiye,”
“(j) Distributor: any natural or legal person in the supply chain, other than the manufacturer or the importer, that makes a device available on the market up to the point of putting into service,”
The importer is the person who makes the device available on the Turkish market for the first time (places it on the market), and the distributor is the person who makes it available on the market at later stages. Under Article 3 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation), placing on the market is the first making available of a device on the Turkish market.
2. The Importer’s Verification Before Placing on the Market
“ARTICLE 13 - (1) Importers shall place on the market only devices that are in conformity with this Regulation.”
Under subparagraph (a) of the second paragraph, to place a device on the market, the importer verifies the following:
“(1) that the device has been CE marked and that the EU declaration of conformity of the device has been drawn up,”
“(2) that a manufacturer is identified and that an authorized representative has been designated by the manufacturer in accordance with Article 11,”
“(3) that the device is labelled in accordance with this Regulation and accompanied by the required instructions for use,”
“(4) where applicable, that a UDI has been assigned by the manufacturer in accordance with Article 27,”
“(b) Where an importer considers or has reason to believe that a device is not in conformity with the requirements of this Regulation, it shall not place the device on the market until it has been brought into conformity, and shall inform the manufacturer and the manufacturer’s authorized representative. Where the importer considers or has reason to believe that the device presents a serious risk or is a falsified device, the importer shall also inform the Agency.”
For the authorized representative requirement, see our article on authorized representative and mandate.
3. The Importer’s Details on the Device
“(3) Importers shall indicate on the device or on its packaging or in a document accompanying the device their name, registered trade name or registered trade mark, their registered place of business and the address at which they can be contacted. They shall ensure that any additional label does not obscure any information on the label provided by the manufacturer.”
The information is given on the device, the packaging or the accompanying document. An additional label cannot cover any information on the manufacturer’s label.
4. Verification of the Record in the Electronic System
“(4) Importers shall verify that the device is registered in the electronic system in accordance with Article 29, and shall add their details to the registration in accordance with Article 31.”
The second paragraph of Article 30 ties a time limit to this:
“(2) Within 2 weeks of placing a device other than a custom-made device on the market, importers shall verify that the manufacturer or authorized representative has entered into the electronic system the information referred to in the first paragraph. Where the information referred to in the first paragraph is missing or incorrect, importers shall inform the relevant authorized representative or the manufacturer. Importers shall add their details to the relevant entry or entries.”
The electronic system mentioned here is EUDAMED. For actor registration, see our article on EUDAMED registration.
5. Storage, Record Keeping and Informing
“(5) Importers shall ensure that, while a device is under their responsibility, storage or transport conditions do not jeopardize its compliance with the general safety and performance requirements set out in Annex I, and shall comply with the conditions set by the manufacturer, if any.”
“(6) Importers shall keep a register of complaints, of non-conforming devices and of recalls and withdrawals, and shall provide the manufacturer, authorized representative and distributors with the information requested by them in order to enable them to investigate complaints.”
“(8) Importers that receive complaints or reports from healthcare professionals, patients or users about suspected adverse events related to a device which they have placed on the market shall immediately forward this information to the manufacturer and the authorized representative.”
Under the seventh paragraph, an importer that considers that a device it has placed on the market is not in conformity immediately informs the manufacturer and the authorized representative and cooperates in corrective action; if the device presents a significant risk, it also immediately informs the Agency and, where applicable, the notified body that issued the certificate. For storage, see our article on medical device storage and transport conditions.
6. Keeping Documents and Cooperation with the Agency
“(9) For the period referred to in the eighth paragraph of Article 10, importers shall keep a copy of the EU declaration of conformity and, if applicable, a copy of any relevant certificate issued in accordance with Article 56, including any amendments and supplements.”
The period in the eighth paragraph of Article 10:
“at least 10 years after the last device covered by the EU declaration of conformity has been placed on the market, and present them to the Agency upon request. For implantable devices, this period is at least 15 years after the last device has been placed on the market.”
Under the tenth paragraph, at the request of the Agency, the importer provides free of charge samples of the device or, where this is not practicable, provides access to the device.
7. The Distributor’s Verification and Sampling
“ARTICLE 14 - (1) When making a device available on the market, distributors shall act with due care in relation to the applicable requirements, within the context of their activities.”
Under subparagraph (a) of the second paragraph, before making a device available on the market, the distributor verifies the following:
“(1) that the device has been CE marked and that the EU declaration of conformity of the device has been drawn up,”
“(2) that the information to be supplied by the manufacturer in accordance with the eleventh paragraph of Article 10 accompanies the device,”
“(3) in the case of imported devices, that the importer has complied with the requirements set out in the third paragraph of Article 13,”
“(4) where applicable, that a UDI has been assigned by the manufacturer.”
“(b) In order to meet the requirements referred to in subparagraph (a), points (1), (2) and (4), the distributor may apply a sampling method that is representative of the devices supplied by that distributor.”
Sampling does not cover point (3), that is, the check of the importer’s details.
8. The Distributor’s Storage, Notification and Record Obligations
“(3) Distributors shall ensure that, while a device is under their responsibility, the storage or transport conditions comply with the conditions set by the manufacturer.”
“(5) Distributors that receive complaints or reports from healthcare professionals, patients or users about suspected incidents related to a device they have made available shall forward this information without delay to the manufacturer and, where applicable, the authorized representative and the importer. Distributors shall keep a register of complaints, of non-conforming devices and of recalls and withdrawals, shall keep the manufacturer and, where available, the authorized representative and the importer informed of such monitoring, and shall provide them with any information they request.”
Where a device is non-conforming, the distributor informs the manufacturer and, where applicable, the authorized representative and the importer; where there is a serious risk or a suspected falsified device, it also informs the Agency (Article 14, second paragraph, subparagraph (c) and fourth paragraph). Under the sixth paragraph, once the manufacturer or authorized representative has provided the necessary information, the distributor is deemed to have met its obligation to provide information to the Agency.
9. Traceability
“ARTICLE 25 - (1) Distributors and importers shall cooperate with manufacturers or authorized representatives in order to achieve an appropriate level of traceability of devices.”
Under the second paragraph, economic operators must be able to provide the Agency, for the period set out in the eighth paragraph of Article 10, with the following:
“(a) any economic operator to which they have directly supplied a device,”
“(b) any economic operator which has directly supplied them with a device,”
“(c) any health institution or healthcare professional to which they have directly supplied a device.”
10. Importer and Dealer in ÜTS
One of the company types in ÜTS is “Üretici/İthalatçı/Bayi/İhracatçı” (Manufacturer/Importer/Dealer/Exporter). TCKKD-KLVZ-03, Article 5, fifth paragraph:
“In other words, companies that will register in ÜTS under the company type “Üretici/İthalatçı/Bayi/İhracatçı” must hold a medical device sales center license, and the ÇKYS number must be entered when registering the company.”
When “İthal” (Imported) is selected in the product registration of an imported device, the “İthal Edilen Ülke” (Country of Import) and “GTİP Numarası” (Customs Tariff Number) fields must be filled in (Article 27, third paragraph). The import notification is made on the individual product side. Individual Movement Notifications Guide, Article 7:
“(1) For products whose country of origin in the product registration is not “Türkiye”, an “import notification” is made.”
“(3) The import notification may also be made by a dealer authorized by the importer, together with the importer.”
The consequence of granting import authorization to another sales center through ÜTS is set out in the implementation guide of the Medical Device Sales, Advertising and Promotion Regulation (TCS-KLVZ-08, Article 17):
“(9) Sales centers that grant another sales center import authorization through ÜTS for a device they have registered in ÜTS are jointly and severally liable, together with the sales center to which they granted the import authorization, for the warranty certificate and technical service obligations under the Regulation for the devices imported with this authorization, for the entire service life of those devices. Where no import authorization is granted through ÜTS, no liability under this paragraph is sought.”
The document issued by the manufacturer to the importer can be registered in ÜTS as “Yetkili Distribütörlük Belgesi” (Authorized Distributorship Document); under Article 23 of TCKKD-KLVZ-03, this document is not mandatory (it is an optional document). For company registration, see our post on ÜTS company registration, for notifications our post on ÜTS individual movement notifications and our page on ÜTS individual notification service.
11. Frequently Asked Questions
What does the importer verify before placing a device on the market?
Under the second paragraph of Article 13, the importer verifies the CE marking and the EU declaration of conformity, that the manufacturer is identified and an authorized representative has been designated, the label and instructions for use, and, where applicable, the assignment of a UDI.
Can the importer’s label be placed over the manufacturer’s label?
The importer’s details are given on the device, on its packaging or in an accompanying document. Under the third paragraph of Article 13, the additional label must not cover any information on the manufacturer’s label.
Must the distributor check every device individually?
Under subparagraph (b) of the second paragraph of Article 14, the distributor may apply a sampling method that is representative of the devices it supplies for the CE marking and declaration of conformity, manufacturer details and UDI checks.
How long does the importer keep the documents?
Copies of the EU declaration of conformity and the certificate are kept for at least 10 years after the last device covered by the declaration has been placed on the market, and for at least 15 years for implantable devices.
Can a dealer make an import notification in ÜTS?
Yes. Under Article 7 of the Individual Movement Notifications Guide, a dealer authorized by the importer can also make the import notification.
For your ÜTS registration and notifications as an importer and dealer: Medical Device ÜTS Registration Consultancy