Conformity Assessment Procedure: Class I, IIa, IIb and III Routes (Annex IX, X, XI) under Article 52 of the Medical Device Regulation

12 hours ago
7 min read
Tıbbi cihaz uygunluk değerlendirme prosedürü sınıflara göre Ek IX Ek X Ek XI

SUMMARY

The conformity assessment procedure is the route a manufacturer follows to demonstrate compliance with the requirements of the regulation before placing a device on the market. Article 52 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation) determines which procedure applies to each risk class, with reference to Annex IX, Annex X and Annex XI.

This article gives the paragraphs of Article 52 class by class, together with Article 61 of the TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03, as direct quotations.

For the ÜTS registration of your medical device: Medical Device ÜTS Registration Consultancy

Contents

1. General Rule: Before Placing on the Market and Putting into Service

Article 52 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation):

“(1) Before placing a device on the market, manufacturers carry out an assessment of the conformity of that device in accordance with the relevant conformity assessment procedures set out in Annex IX to Annex XI.”

“(2) Before putting into service a device that is not placed on the market, manufacturers carry out an assessment of the conformity of that device in accordance with the relevant conformity assessment procedures set out in Annex IX to Annex XI.”

The procedure is chosen according to the class of the device. For classification, see our article on MDR risk class.

Titles of the annexes of the Regulation referred to in Article 52:

  • Annex IX: CONFORMITY ASSESSMENT BASED ON A QUALITY MANAGEMENT SYSTEM AND ON ASSESSMENT OF TECHNICAL DOCUMENTATION. Chapter I covers the quality management system, Chapter II the assessment of technical documentation, and Chapter III administrative provisions.

  • Annex X: CONFORMITY ASSESSMENT BASED ON TYPE-EXAMINATION.

  • Annex XI: CONFORMITY ASSESSMENT BASED ON PRODUCT CONFORMITY VERIFICATION. Part A covers production quality assurance and Part B product verification.

Section 1 of Annex X defines EU type-examination:

“1. EU type-examination is the procedure by which a notified body ascertains and certifies that a device, including its technical documentation and relevant life-cycle processes and a corresponding representative sample of the device production envisaged, fulfils the relevant provisions of this Regulation.”

2. Class III Devices

“(3) Manufacturers of class III devices, other than custom-made or investigational devices, are subject to a conformity assessment as specified in Annex IX. Alternatively, the manufacturer may choose to apply a conformity assessment as specified in Annex X coupled with a conformity assessment as specified in Annex XI.”

There are two routes for class III: Annex IX, or Annex X together with Annex XI. For the ÜTS registration of class III devices, see our article on class III medical devices.

3. Class IIb Devices

“a) Manufacturers of class IIb devices, other than custom-made or investigational devices, are subject to a conformity assessment as specified in Chapters I and III of Annex IX, including an assessment of the technical documentation, as specified in Section 4 of Annex IX, of at least one representative device per generic device group.”

“b) However, for class IIb implantable devices other than sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors, the assessment of the technical documentation as specified in Section 4 of Annex IX applies to every device.”

“c) Alternatively, the manufacturer may choose to apply a conformity assessment based on type-examination as specified in Annex X coupled with a conformity assessment based on product conformity verification as specified in Annex XI.”

For implantable class IIb devices other than the twelve types of product listed in subparagraph (b), the technical documentation assessment is carried out for every device.

4. Class IIa Devices

“a) are subject to a conformity assessment as specified in Chapters I and III of Annex IX, including an assessment of the technical documentation, as specified in Section 4 of Annex IX, of at least one representative device per device category,”

“b) Alternatively, the manufacturer may choose to draw up the technical documentation set out in Annex II and Annex III coupled with a conformity assessment as specified in Section 10 or Section 18 of Annex XI. The assessment of the technical documentation applies to at least one representative device for each category of devices.”

For class IIb the representative device is chosen for each generic device group, and for class IIa for each device category. The ÜTS registration of both classes is covered in our article on class IIa and IIb medical devices.

5. Class I Devices: Is, Im and Ir

“(6) Manufacturers of class I devices, other than custom-made or investigational devices, declare the conformity of their products by issuing the EU declaration of conformity referred to in Article 19, after drawing up the technical documentation set out in Annex II and Annex III.”

“If these devices are placed on the market in a sterile condition, have a measuring function, or are reusable surgical instruments, the manufacturer applies the procedures set out in Chapters I and III of Annex IX or in Part A of Annex XI. However, the involvement of the notified body in these procedures is limited to:”

  • “a) In the case of devices placed on the market in a sterile condition, the aspects relating to establishing, securing and maintaining sterile conditions,”

  • “b) In the case of devices with a measuring function, the aspects relating to the conformity of the devices with metrological requirements,”

  • “c) In the case of reusable surgical instruments, the aspects relating to the reuse of the device, in particular cleaning, disinfection, sterilization, maintenance, functional testing and the related instructions for use,”

Article 61 of the ÜTS Registration Guideline names these three groups Class-Is, Class-Im and Class-Ir:

“If these devices are placed on the market in a sterile condition (Class-Is), or have a measuring function (Class-Im), or are reusable surgical instruments (Class-Ir), the manufacturer also carries out one of the procedures set out in MDR Annex IX (Chapters I and III) or MDR Annex XI (Part A).”

For class I devices, see our article on class I medical devices and our page on the MDR/IVDR declaration of conformity service.

6. Custom-Made Devices

“(7) Manufacturers of custom-made devices follow the procedure set out in Annex XIII and draw up the statement set out in Section 1 of Annex XIII before placing such devices on the market. Manufacturers of class III custom-made implantable devices are subject, in addition to the procedure applicable under the first sentence, to a conformity assessment as specified in Chapter I of Annex IX. Alternatively, the manufacturer may choose to apply a conformity assessment as specified in Part A of Annex XI.”

Paragraph 8 of Article 61 of the Guideline:

“ITCs are placed on the market without bearing the CE marking.”

The details are in our article on custom-made devices and Annex XIII.

7. Devices Requiring an Additional Procedure

Paragraphs 8, 9 and 10 of Article 52 add procedures to the basic procedure:

“(8) In addition to the procedures applicable under the third, fourth, fifth or sixth paragraphs, for the devices referred to in the second sentence of the sixth paragraph of Article 1, the procedure laid down in Section 5.2 of Annex IX or in Section 6 of Annex X, as applicable, also applies.”

The eighth paragraph covers devices incorporating a medicinal product, the ninth paragraph devices manufactured utilizing derivatives of non-viable animal or human tissues or cells, and the tenth paragraph devices composed of substances that are absorbed by or locally dispersed in the human body. For devices incorporating a medicinal product, see our article on borderline products.

For class III implantable devices and class IIb active devices under Rule 12, the clinical evaluation consultation in Art. 54 also applies. The details are in our article on clinical evaluation and PMCF.

8. Choice of Notified Body: Art. 53

Article 53 of the Tıbbi Cihaz Yönetmeliği:

“(1) Where the involvement of a notified body is required in the conformity assessment procedure, the manufacturer may apply to a notified body of its choice, provided that the body is designated for conformity assessment activities related to the relevant types of devices. The manufacturer cannot lodge an application in parallel with another notified body for the same conformity assessment procedure.”

“(3) When applying to a notified body under the first paragraph, the manufacturer declares whether it has withdrawn an earlier application to another notified body before that notified body’s decision, and provides information about any previous application for the same conformity assessment that was refused by another notified body.”

The notified body carries out the conformity assessment; the manufacturer makes the application. For a change of notified body, see our article on change of notified body.

9. Outcome of the Procedure and ÜTS Document Types

At the end of the procedure, the EU certificate issued by the notified body and the EU declaration of conformity issued by the manufacturer are registered in ÜTS under separate document types. ÜTS Registration Guideline, Article 11:

“ARTICLE 11- (1) A document of the type “AB Uygunluk Beyanı (MDR/IVDR)” (EU Declaration of Conformity (MDR/IVDR)) is the document type used only in notifications and applications relating to declarations of conformity prepared under the new Medical Device Regulations.”

The steps of document registration are in our article on ÜTS document registration.

10. Frequently Asked Questions

Is a notified body required for a class I device?

Under the sixth paragraph of Art. 52, for class I devices the manufacturer draws up the technical documentation and issues the EU declaration of conformity. If the device is placed on the market sterile, has a measuring function or is a reusable surgical instrument, Chapters I and III of Annex IX or Part A of Annex XI also apply.

What does the notified body assess for a class Is device?

Under subparagraph (a) of the sixth paragraph of Art. 52, the involvement of the notified body is limited to the aspects relating to establishing, securing and maintaining sterile conditions.

Which routes are there for a class III device?

Under the third paragraph of Art. 52, Annex IX applies, or alternatively Annex X coupled with Annex XI.

Is each class IIb implantable device assessed separately?

Under subparagraph (b) of the fourth paragraph of Art. 52, yes; except for sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors, the technical documentation assessment applies to every device.

Can two notified bodies be applied to for the same device?

No. Under the first paragraph of Art. 53, the manufacturer cannot apply in parallel to another notified body for the same conformity assessment procedure.

For the ÜTS registration of your medical device: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara