ÜTS IVD Product Registration: IVDR Risk Class Selection, Registration Fields and Document Business Rules

12 hours ago
8 min read
ÜTS IVD ürün kaydı IVDR risk sınıfları

SUMMARY

ÜTS IVD product registration means registering an in vitro diagnostic device in ÜTS through the registration screen under the In Vitro Diagnostic Medical Device Regulation (IVDR). The fields to be entered in the registration are regulated in Article 30 of the TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03, and the documents to be linked to the device, by class, in Article 67.

In this article we give the headings of IVD product registration, the selection of the risk class, the transition field from IVDD and the IVD-specific fields, quoting verbatim from the Guideline. For details of document registration, see our article on IVDR ÜTS document registration.

For the ÜTS registration of your IVD products: Medical Device ÜTS Registration Consultancy

Contents

1. IVD Registration Screen Headings

“ARTICLE 30- (1) The information to be entered into ÜTS on the ÜTS registration screen during device registration (device registration/update notification) under IVDR, together with its explanations, appears under the headings ‘Identifying Information’, ‘Import/Manufacture Information’, ‘Classification Information’, ‘Product Documents’, ‘Product Images’, ‘Features’, ‘Dimension Information’, ‘Storage/Use Condition Information’, ‘CMR and/or Endocrine Disrupting Substance Information’ and ‘Application Status’.”

The headings are the same as for MDR registration. For the general rules of identifying fields such as UDI-DI, brand name, product name, version/model and reference/catalog number, see our article on ÜTS product registration. Below we explain the fields specific to IVD registration.

2. Device Transitioning from IVDD: Different Product Number Field

There is a field specific to IVDR registration under the “Identifying Information” heading:

“2) This field is selected as ‘Yes’ if the relevant device to be registered in ÜTS under IVDR was previously registered in ÜTS under IVDD with a different barcode, and as ‘No’ if it was not previously registered in ÜTS under IVDD with a different barcode.”

If the “Did it transition to IVDR with a different product number?” field is selected as “Yes”, the “Product Number under 98/79/EC” field opens. In this field, the primary product number (barcode) of the device previously registered under IVDD is selected from the “Select Medical Device” screen. Only one barcode can be selected in this field.

3. IVDR Risk Classes and Class Selection in ÜTS

“(1) Devices are classified as Class A, B, C or D, taking into account the risks they carry by virtue of their intended purpose and structure. Classification is carried out in accordance with Annex VIII.”

This provision is in the In Vitro Diagnostic Medical Device Regulation. In ÜTS, these four classes are shown with additional options according to the device’s characteristics. Article 57 of the Guideline:

“(In ÜTS, additional classifications have been made according to characteristic product features and intended purposes, such as whether the device is packaged sterile or whether it is, for example, a Companion Diagnostic Device. For example, a device that is a ‘Near-Patient Test Device (Class C)’ in ÜTS is also classified as Class C under IVDR.)”

The options in the mandatory “Class” field under the “Classification Information” heading are as follows:

“3) According to the risk classes that can be selected under IVDR, the appropriate one is selected from the options ‘Class A Sterile’, ‘Class A’, ‘Class B’, ‘Class C’, ‘Class D’, ‘Self-Test Device (Class B)’, ‘Self-Test Device (Class C)’, ‘Self-Test Device (Class D)’, ‘Near-Patient Test Device (Class B)’, ‘Near-Patient Test Device (Class C)’, ‘Near-Patient Test Device (Class D)’, ‘Companion Diagnostic Device (Class C)’ and ‘Companion Diagnostic Device (Class D)’ in the relevant field.”

For the class of self-test devices, Rule 4 as reproduced in Article 57 of the Guideline:

“1) Devices intended for self-testing are classified as Class C. However, devices intended for the detection of pregnancy, fertility testing, determining cholesterol level, and the detection of glucose, erythrocytes, leukocytes and bacteria in urine are classified as Class B.”

GMDN, Branch Type Code and EMDN are also selected in the same field. For GMDN, see our article What Is a GMDN Code.

4. Product Documents and Basic UDI-DI

Under the “Product Documents” heading, documents are selected from among the documents already registered in ÜTS. The order matters:

“3) Before document selection in this field, it is mandatory that the class of the relevant device has been selected on the ÜTS Registration screen. In this field, selection is made from the documents that can be selected according to the class of the device.”

The Basic UDI-DI is also a mandatory field under this heading, and a single selection is made only from among the Basic UDI-DI numbers associated with the selected EU declaration of conformity. The Guideline also states that the Basic UDI-DI does not appear on the label:

“7) The Basic UDI-DI is not present on the market placement label and/or sales packaging of the device.”

For details, see our article What Is Basic UDI-DI.

5. Document Business Rules by Class

Article 67 of the Guideline links the documents to be associated with an IVD device to separate annexes according to class:

  • “(3) For devices to be notified for registration in ÜTS under IVDR as ‘Class A’ or ‘Class A Sterile’, the appropriate one of the ÜTS document business rules shown in Annex-94 is applied.”

  • “(4) For devices to be notified for registration in ÜTS under IVDR as ‘Class B’ or ‘Class C’, the appropriate one of the ÜTS document business rules shown in Annex-95 is applied.”

  • “(5) For devices to be notified for registration in ÜTS under IVDR as ‘Class D’, the appropriate one of the ÜTS document business rules shown in Annex-96 is applied.”

  • “(6) For devices to be notified for registration in ÜTS under IVDR as ‘Self-Test Device (Class B)’ or ‘Near-Patient Test Device (Class B)’, the appropriate one of the ÜTS document business rules shown in Annex-97 is applied.”

  • “(7) For devices to be notified for registration in ÜTS under IVDR as ‘Self-Test Device (Class C)’ or ‘Self-Test Device (Class D)’, the appropriate one of the ÜTS document business rules shown in Annex-98 is applied.”

  • “(8) For devices to be notified for registration in ÜTS under IVDR as ‘Near-Patient Test Device (Class C)’ or ‘Near-Patient Test Device (Class D)’, the appropriate one of the ÜTS document business rules shown in Annex-99 is applied.”

  • “(9) For devices to be notified for registration in ÜTS under IVDR as ‘Companion Diagnostic Device (Class C)’ or ‘Companion Diagnostic Device (Class D)’, the appropriate one of the ÜTS document business rules shown in Annex-100 is applied.”

For this reason, if the class is selected wrongly, the correct document cannot be selected. Which documents are required in which class is covered in our article on the documents required for ÜTS registration of products under IVDR. For the content of the declaration of conformity, see our article on the IVDR declaration of conformity.

6. IVD-Specific Feature Fields

Under the “Features” heading there are fields that do not exist in MDR registration or are defined separately for IVD:

  • Is it a new type of device?: mandatory field.

  • Is it a device intended for self-testing or near-patient testing?: mandatory field.

  • Does it contain cells or substances of microbial origin?: mandatory field.

  • Does it contain tissues or cells of human origin or their derivatives?: mandatory field.

The Guideline gives the definition of a “new device” as follows:

  • “a. For the relevant analyte or other parameters, a device of this type has not been continuously available in Türkiye / EU member states during the last 3 years,”

  • “b. The procedure involves an analytical technology that has not been continuously used in connection with a specific analyte or other parameters in Türkiye / EU member states during the last three years”

For the self-test or near-patient test field:

“3) If the device is a Self-Test Device or a Near-Patient Test Device, it is mandatory that this information (the indication of whether the device is intended for self-testing or near-patient testing) appears on the label and/or sales packaging and in the instructions for use, and that this field is selected as ‘Yes’.”

7. IVD Information Checked on the Label

Among the common information checked under IVDR Annex I on the label uploaded to the “Original Label and Sales Packaging” field, the following IVD-specific items are included:

  • “m. An indication that the device is an in vitro diagnostic device or, if the device is a ‘device for performance study’, an indication to that effect,”

  • “n. If the device is intended for self-testing or near-patient testing, an indication to that effect,”

  • “o. In case rapid tests are not intended for self-testing or near-patient testing, a clear statement regarding the exclusion,”

The same list also includes the CE mark and, if a notified body is involved, the four-digit notified body number. For label and image rules, see our page on ÜTS product image and label rules.

8. Rule for Uploading Instructions for Use

According to Article 17 of the Guideline, uploading the manufacturer’s instructions for use to ÜTS is mandatory for the following IVD devices:

“b) For Class B, Class C, Class D devices certified under IVDR, as well as Self-Test Devices (Class B/C/D), Near-Patient Test Devices (Class B/C/D) and Companion Diagnostic Devices (Class C/D), instructions for use drawn up by the manufacturer,”

“(3) For other risk groups of devices (for Class I Other, Class Im, Class Is, Class Ir, Class A, Class A Sterile and IVD Other devices), uploading instructions for use is optional.”

Where it deems necessary, the Authority can also request the upload of instructions for use of these devices (Article 17, third paragraph).

9. Application Status and Application to the Authority

  • “3) If this field is selected as ‘No’ (if the relevant medical device registration notification is not ready for application), the status of this medical device notification is saved in ÜTS as ‘Draft’.”

  • “4) If this field is selected as ‘Yes’ (if the relevant medical device registration notification is ready for application), the status of this medical device notification is saved in ÜTS as ‘Ready for Application’ and it is listed on the ‘List My Medical Devices’ screen.”

The registration notification alone is not enough; for the conformity assessment, a device application is made to the Authority under Article 40 of the Guideline (Article 30, thirteenth paragraph).

For devices under IVDD to benefit from the transitional provisions, an application is first made in EBS and then in ÜTS (Guideline Article 48, fifth paragraph). For the dates, see our article on the (EU) 2024/1860 IVDR transition timeline.

10. Frequently Asked Questions

Which screen is used to register an IVD device in ÜTS?

It is registered through the device registration screen under IVDR. The fields are regulated in Article 30 of the ÜTS Registration Guideline, and the document business rules in Article 67.

Is a near-patient test device a separate class in ÜTS?

It is shown as a separate option in ÜTS, but the IVDR class does not change. For example, a device selected as “Near-Patient Test Device (Class C)” is Class C under IVDR.

What is done if a device registered under IVDD transitions to IVDR with a different barcode?

The “Did it transition to IVDR with a different product number?” field is selected as “Yes” and the primary product number of the old registration is selected from the “Product Number under 98/79/EC” field that opens. Only one barcode can be selected.

Is it mandatory to upload instructions for use for a Class A IVD device?

No, it is optional for Class A and Class A Sterile devices. It is mandatory for Class B, C, D, self-test, near-patient test and companion diagnostic devices.

Why is document selection made after the class?

According to the Guideline, it is mandatory that the class has been selected before document selection, and the documents that can be selected are determined by class.

For the ÜTS product registration of your IVD products: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application