The UDI System: UDI-DI, UDI-PI, UDI Carrier and Barcode Registration in ÜTS

12 hours ago
7 min read
UDI sistemi UDI-DI UDI-PI ve ÜTS barkod

SUMMARY

The UDI system is the unique device identification system that provides identification and traceability of medical devices other than custom-made and investigational devices. The UDI consists of two parts: the UDI-DI, which identifies the device model, and the UDI-PI, which identifies the production unit.

This article gives, as direct quotations from Article 27 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation), Annex VI Part C and the TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03, the components of the system, the difference between UDI-DI and UDI-PI, the location of the UDI carrier and how the UDI-DI number is entered in ÜTS.

For ÜTS registration with your devices’ UDI-DI data: Medical Device ÜTS Registration Consultancy

Contents

1. What Is the UDI System?

Tıbbi Cihaz Yönetmeliği, Article 27, paragraph 1:

“ARTICLE 27 -(1) The Unique Device Identification System (UDI System), established and controlled by the Commission and defined in Annex VI Part C, allows the identification of devices other than custom-made and investigational devices and facilitates their traceability, and consists of the following:”

  • “a) The creation of a UDI containing the following;”

  • “b) The placing of the UDI on the label or packaging of the device.”

  • “c) The keeping of records of the UDI by economic operators, health institutions and health professionals, under the conditions specified in the eighth and ninth paragraphs.”

  • “ç) The establishment of an electronic system on unique device identification referred to in Article 28 (UDI database).”

For the full text of the article, see our page on the unique device identification system: Article 27.

2. Definitions of UDI, UDI-DI and UDI-PI

Definitions in Annex VI Part C:

  • “UDI is a series of numeric or alphanumeric characters created through a globally accepted device identification and coding standard. It allows the unambiguous identification of a specific device on the market. The UDI consists of the UDI-DI and the UDI-PI.”

  • “UDI-DI is a unique numeric or alphanumeric code specific to a device model, and is also used as the ‘access key’ to the information stored in the UDI database.”

  • “UDI-PI is a numeric or alphanumeric code that identifies the production unit.”

The same part also explains the meaning of the word ‘unique’:

“The word ‘unique’ does not imply serialization of individual production units.”

The UDI-DI varies by packaging level:

“3.4. The UDI-DI is unique at each level of device packaging.”

3. What Information Goes into the UDI-PI?

“The different types of UDI-PIs include the serial number, lot number, software identification and manufacturing date or expiry date, or both date types.”

Its relation to the information on the label is in item 3.5:

“3.5. If a lot number, serial number, software identification or expiry date appears on the label, it forms part of the UDI-PI. If a manufacturing date also appears on the label, it does not need to be included in the UDI-PI. If only a manufacturing date appears on the label, it is used as the UDI-PI.”

4. When Is a New UDI-DI Required?

“3.9. A new UDI-DI is required whenever there is a change that could lead to misidentification of the device and/or ambiguity in its traceability, in particular any change in any of the following UDI database data elements:”

  • “a) Name or trade name of the device,”

  • “b) Device version or model,”

  • “c) Labelled as single use,”

  • “ç) Packaged sterile,”

  • “d) Need for sterilization before use,”

  • “e) Quantity of devices provided in a package,”

  • “f) Critical warnings or contra-indications: for example, containing latex or DEHP.”

The ÜTS Registration Guideline gives the same list in Article 27 and adds, as an example of critical warnings, containing CMR/endocrine-disrupting substances.

5. The Manufacturer’s UDI Obligations

Article 27, paragraph 3:

  • “a) Before placing a device other than a custom-made device on the market, applies the UDI that it has created in accordance with the rules of an issuing entity designated under the second paragraph to the device and, where applicable, to all higher levels of packaging.”

  • “b) Before placing a device other than a custom-made or investigational device on the market, ensures that the information referred to in Annex VI Part B concerning that device is correctly submitted and transferred to the UDI database referred to in Article 28.”

General requirements of Annex VI Part C:

  • “2.3. Only the manufacturer can place the UDI on the device or its packaging.”

  • “2.4. Only coding standards provided by issuing entities designated by the Commission may be used.”

“(7) As part of the technical documentation referred to in Annex II, the manufacturer keeps up to date a list of all UDIs it has assigned.”

For the issuing entities, see our article on UDI issuing entities: GS1, HIBCC, ICCBBA and IFA.

6. Where Is the UDI Carrier Placed?

“The UDI carrier is the means of conveying the UDI by using AIDC and, if applicable, its HRI.”

“(4) UDI carriers are placed on the label of the device and on all higher levels of packaging. Higher levels of packaging do not include shipping containers.”

Special cases in Annex VI Part C:

  • “4.2. Where there are significant space constraints on the unit-of-use packaging, the UDI carrier may be placed on the next higher packaging level.”

  • “4.10. Reusable devices carry a UDI carrier on the device itself.”

  • “4.11. The UDI carrier is readable during normal use and throughout the intended lifetime of the device.”

UDI carriers can be in the form of 1D/linear barcode, 2D/matrix barcode and RFID (Part C, 1.13). For the barcode format, see our article on the UDI carrier and barcode format.

7. Basic UDI-DI and the UDI Database

“The Basic UDI-DI is the primary identifier of a device model. It is the DI assigned at the level of the unit of use of the device. It is the main key for records in the UDI database and is referenced in relevant certificates and EU declarations of conformity.”

The allocation of the Basic UDI-DI and its relationship with the EU declaration of conformity and the certificate is the subject of a separate article: What is the Basic UDI-DI. Article 28 for the UDI database:

“The core data elements to be provided to the UDI database referred to in Annex VI Part B of this Regulation are publicly accessible free of charge.”

“5.5. The presence of a device’s UDI-DI in the UDI database does not mean that the device conforms to this Regulation.”

8. Who Keeps the UDI Record?

“(8) Economic operators record and keep, preferably by electronic means, the UDIs of the devices they have supplied or that have been supplied to them, if these devices are class III implantable devices or belong to devices, categories or groups of devices determined by the Commission.”

“a) Health institutions record and keep, preferably by electronic means, the UDIs of the devices they have supplied or that have been supplied to them, if these devices are class III implantable devices.”

For other devices, health institutions and health professionals are encouraged to keep UDI records, and this can be requested where deemed necessary (Article 27, paragraph 9, subparagraphs b and c).

9. The UDI-DI Number in ÜTS

Under Article 27 of the ÜTS Registration Guideline, when registering a device under the MDR, the “UDI Tahsis Kuruluşu” (UDI Issuing Entity) is first selected under the heading “Tanımlayıcı Bilgiler” (Identifying Information):

“3) The UDI issuing entities and standards registered in ÜTS under the MDR and authorized for medical devices are GS1, HIBCC, ICCBBA and IFA.”

Then the barcode given to the device by the issuing entity is entered unchanged in the “UDI-DI Numarası” (UDI-DI Number) field. The Guideline’s rules for this field:

  • “4) A device application made under Article 40 of this Guideline for a device registration whose barcode is not entered in ÜTS as it appears in its technical documents (label/sales packaging, EU declaration of conformity, EU certificate etc.) results in a negative outcome.”

  • “6) In principle, device registration in ÜTS is made with the barcode of the smallest packaging placed on the market (unit packaging or single packaging).”

The barcodes of sales packages containing multiple products are added, together with their content quantities, to the “Piyasaya Arz Bilgisi” (Placing on the Market Information) field after registration with the smallest package barcode is completed. The UDI-PI side is used in individual movement notifications; Article 5 of the Individual Movement Notifications Guideline:

“Article 5- (1) Medical devices are tracked by serial and/or lot number according to the production process determined by the manufacturer.”

For the other fields of product registration, see our article on ÜTS product registration, and for notifications, see our article on ÜTS individual movement notifications.

10. Frequently Asked Questions

What is the difference between UDI-DI and UDI-PI?

The UDI-DI is a code specific to a device model and is the access key to the information in the UDI database. The UDI-PI identifies the production unit; it may include the serial number, lot number, software identification and manufacturing or expiry date.

Is a UDI assigned to a custom-made device?

Under Article 27, paragraph 1, the UDI system covers devices other than custom-made and investigational devices.

Is the UDI carrier placed on the shipping container?

No. Under Article 27, paragraph 4, UDI carriers are placed on the label and on all higher levels of packaging; higher levels do not include shipping containers.

Is a new UDI-DI required if the number of devices in the package changes?

Yes. Item 3.9 of Annex VI Part C lists a change in the quantity of devices in a package among the cases requiring a new UDI-DI.

With which package’s barcode is registration made in ÜTS?

According to the ÜTS Registration Guideline, in principle with the barcode of the smallest packaging placed on the market (unit or single packaging). If the barcode is not entered as it appears in the technical documents, the device application results in a negative outcome.

For ÜTS product registration with UDI-DI and barcode data: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara