top of page

Reprocessing of a Single-Use Device: Article 17 of the Tıbbi Cihaz Yönetmeliği, Label and ÜTS

12 hours ago
7 min read
Tek kullanımlık cihazın yeniden işlenmesi madde 17 ve ÜTS

SUMMARY

The reprocessing of a single-use device is regulated in Article 17 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation). A natural or legal person who reprocesses a single-use device to make it suitable for further use is considered the manufacturer of the reprocessed device.

In this article we give the definitions, all the paragraphs of Article 17, the label and instructions-for-use rules and the single-use field in the ÜTS registration, with word-for-word quotations from the regulation and the TİTCK ÜTS Registration Guide TCKKD-KLVZ-03.

For the ÜTS registration of your reprocessed devices: Medical Device ÜTS Registration Consultancy

Contents

1. Definitions of Single-Use Device and Reprocessing

Article 3 of the Tıbbi Cihaz Yönetmeliği:

“(ççç) Single-use device: a device that is intended to be used on one individual during a single procedure,”

“(jjj) Reprocessing: a process carried out on a used device in order to allow its safe reuse, including cleaning, disinfection, sterilization and related procedures, as well as testing and restoration of the technical and functional safety of the used device,”

Besides cleaning, disinfection and sterilization, the definition also lists the testing and restoration of the technical and functional safety of the used device. For the definition, see our page on what is a single-use device.

The complete remaking of a used device is a separate concept (Art. 3, point ccc: fully refurbishing). This subject is covered in our article on used and refurbished medical devices.

2. Article 17: The Reprocessor Is Treated as the Manufacturer

“(1) (Amended: Official Gazette 29/7/2022-31907) The reprocessing and reuse of single-use devices shall be carried out in accordance with the guidelines published by the Agency and with this Article.”

The paragraph gives two bases: the guidelines published by the Agency, and the Article itself.

“(2) Any natural or legal person who reprocesses a single-use device to make it suitable for further use shall be considered the manufacturer of the reprocessed device and shall assume the obligations incumbent on manufacturers under this Regulation, including the obligations relating to the traceability of the reprocessed device in accordance with Part Three of this Regulation. The reprocessor of the device shall be considered a manufacturer for the purposes of Article 6 of the Law on Technical Regulations on Product Safety.”

The paragraph uses the phrase "any natural or legal person". The reprocessor assumes the manufacturer obligations of the regulation, including traceability. For these obligations, see our article on manufacturer obligations.

3. Reprocessing within a Healthcare Institution

Point (a) of the third paragraph gives the Agency a power of exception:

“a) By way of derogation from the second paragraph, in respect of single-use devices that are reprocessed and used within a healthcare institution, it may decide that some of the rules relating to the manufacturer's obligations laid down in this Regulation shall not apply, provided that the following are ensured:”

  • “1) The safety and performance of the reprocessed device are equivalent to those of the original device, and the requirements in points (a), (b), (ç), (d), (e), (f) and (g) of the fifth paragraph of Article 5 are complied with.”

  • “2) The reprocessing is carried out in accordance with common specifications;”

  • “i) risk management, including the analysis of the structure and material, the relevant properties of the device (reverse engineering) and procedures for detecting changes in the design of the original device, as well as its planned application after reprocessing,”

  • “ii) the validation of procedures for the entire process, including cleaning steps,”

  • “iii) product release and performance testing,”

  • “iv) the quality management system,”

  • “v) the reporting of adverse events involving reprocessed devices,”

  • “vi) the traceability of reprocessed devices.”

  • “detailing the requirements relating to the above.”

The text uses the phrases "may decide" and "some of the rules": the exception depends on a decision of the Agency and covers only part of the manufacturer rules. The fifth paragraph of Article 5 sets out the conditions for devices manufactured within a healthcare institution; for details, see our article on a device manufactured within a healthcare institution.

Points (b) and (c) of the same paragraph:

“b) It encourages, and may request, healthcare institutions to provide patients with information on the use of reprocessed devices within their own premises and, where appropriate, other relevant information on the reprocessed device used on patients.”

“c) It notifies the Commission and the EU Member States of the provisions of the legislation applied under this paragraph and the grounds for applying them, to be made publicly available.”

4. External Reprocessor

“(4) At the request of a healthcare institution, the Agency may choose to apply the provisions referred to in the third paragraph to single-use devices reprocessed by an external reprocessor, provided that all of the reprocessed devices are delivered back to that healthcare institution and that the external reprocessor complies with the requirements referred to in point (a) of the third paragraph.”

There are two conditions: all reprocessed devices are delivered to the requesting healthcare institution, and the external reprocessor meets the requirements in point (a) of the third paragraph.

5. Common Specifications and the Notified Body Certificate

“(5) The common specifications adopted by the Commission in relation to reprocessing shall apply. Compliance with the common specifications or, in the absence of common specifications, with the relevant harmonised standards and the relevant legislation, shall be certified by a notified body.”

For the concept of common specification, see our article on harmonised standards and common specifications.

6. Which Devices Can Be Reprocessed?

“(6) (Amended: Official Gazette 29/7/2022-31907) Only single-use devices that have been placed on the market in accordance with this Regulation, or before 26/5/2021 in accordance with the former Tıbbi Cihaz Yönetmeliği (Medical Devices Regulation) repealed by Article 110, may be reprocessed.”

“(7) Single-use devices may only be reprocessed in a way that is considered safe in the light of the most recent scientific evidence.”

7. Label and Instructions for Use

“(8) The name and address of the natural or legal person referred to in the second paragraph and the other relevant information referred to in item 23 of Annex I shall be indicated on the label of the reprocessed device and, where applicable, in the instructions for use. The name and address of the manufacturer of the original single-use device shall no longer appear on the label, but shall be indicated in the instructions for use of the reprocessed device.”

Item 23.2 of Annex I (information on the label) requires a separate indication for a reprocessed device:

“l) if the device is a reprocessed single-use device, an indication of this fact, the number of reprocessing cycles already performed, and any limitation as regards the number of reprocessing cycles,”

For the original manufacturer, item 23.4 of Annex I (instructions for use):

“m) if the device bears an indication that it is a single-use device, information on the characteristics and technical factors known to the manufacturer that could pose a risk if the device were to be re-used.”

For the labelling rules, see our article on the Turkish label and instructions for use.

8. The Agency's Power to Restrict

“(9) The Agency may introduce provisions of legislation that are stricter than those in this Regulation and that restrict or prohibit the following, and shall notify them to the Commission and the EU Member States so that they are made publicly available:”

  • “a) The reprocessing of single-use devices and the transfer of such devices to another country for the purpose of reprocessing.”

  • “b) The making available or reuse of reprocessed single-use devices.”

9. Registration Data and ÜTS

Among the information to be submitted in the device registration under Annex VI Part A is the following item:

“2.6. Reprocessed single-use device (yes/no),”

For the "Tek kullanımlık mı?" (Is it single-use?) field in the seventh paragraph of Article 27 of the ÜTS Kayıt Kılavuzu (ÜTS Registration Guide):

“3) If the device concerned is a single-use device, this information must appear both on the label and/or sales packaging and in the instructions for use. In addition, this field must also be selected as "Evet" (Yes).”

Under the second paragraph of the same Article, a change in whether the device is single-use requires a new UDI-DI number. The "Yeniden İşleme Bildirimi" (Reprocessing Notification) in Article 22 of the ÜTS Tekil Hareket Bildirimleri Kılavuzu (Individual Movement Notifications Guide) is a different process: it is made when an individual product is used as a raw material in the manufacture of another individual product. For product registration in general, see our article on ÜTS product registration.

10. Frequently Asked Questions

Who is considered the reprocessor of a single-use device?

Under the second paragraph of Article 17, the natural or legal person who reprocesses is considered the manufacturer of the reprocessed device and assumes the manufacturer obligations.

What happens if a hospital reprocesses devices in-house?

Under the third paragraph, if the conditions listed in the Article are met, the Agency may decide that some of the manufacturer rules shall not apply to devices reprocessed and used within a healthcare institution.

Whose name appears on the label of a reprocessed device?

Under the eighth paragraph, the label shows the name and address of the reprocessor. The name and address of the original manufacturer do not appear on the label; they are stated in the instructions for use.

Which single-use devices can be reprocessed?

Under the sixth paragraph, only single-use devices placed on the market in accordance with this Regulation, or before 26/5/2021 in accordance with the repealed Tıbbi Cihaz Yönetmeliği (Medical Devices Regulation), can be reprocessed.

Can an external reprocessor supply the devices to another institution?

The arrangement in the fourth paragraph depends on all of the reprocessed devices being delivered to the requesting healthcare institution.

For the ÜTS registration of your reprocessed devices: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara