The manufacturer's declaration of conformity is the document in which the manufacturer declares, on its own responsibility, that its device meets the requirements of the regulation. The Medical Device Regulation (MDR) calls this document the "EU declaration of conformity" and lists its content in Annex IV.

In this article we explain, with verbatim quotations from the regulation and the TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03, who issues the declaration, what Annex IV contains and how the information in the declaration is checked during ÜTS document registration. For the difference from the certificate, see our article on differences between the EC certificate, the EU certificate and the declaration of conformity.

For your MDR and IVDR declaration of conformity: Medical Device ÜTS Registration Consultancy

Contents

1. Who Issues the Manufacturer's Declaration of Conformity?

The manufacturer issues the declaration. Article 10 lists this among the general obligations of manufacturers:

“(6) Manufacturers of devices other than custom-made or investigational devices, where the applicable conformity assessment procedures demonstrate that their devices comply with the applicable requirements, shall draw up an EU declaration of conformity in accordance with Article 19 and affix the CE marking of conformity in accordance with Article 20.”

Responsibility also rests with the manufacturer:

“(3) By drawing up the EU declaration of conformity, the manufacturer assumes responsibility for compliance with the requirements of this Regulation and all other legislation applicable to the device.”

The ÜTS Registration Guideline repeats the same rule: "The EU Declaration of Conformity is issued by the legal manufacturer." For details, see our question on who prepares and signs the declaration of conformity.

2. The Manufacturer's Obligations Before the Declaration

The declaration is the result of the conformity assessment. Under Article 10, the manufacturer does the following before it:

  • Establishes and maintains a risk management system in accordance with point 3 of Annex I (second paragraph).

  • Carries out a clinical evaluation in accordance with Article 61 and Annex XIV (third paragraph).

  • Draws up and keeps technical documentation up to date in accordance with Annex II and Annex III (fourth paragraph).

  • Establishes a quality management system proportionate to the risk class and type of the device (ninth paragraph).

  • Implements the post-market surveillance system in accordance with Article 81 (tenth paragraph).

“(4) Manufacturers of devices other than custom-made devices shall draw up and keep up to date technical documentation for those devices. The technical documentation shall be such as to allow an assessment of the conformity of the device with the requirements of this Regulation. It shall include the elements set out in Annex II and Annex III.”

For the full text of the article, see our page on manufacturer obligations, Article 10.

3. Annex IV: Information the Declaration Must Contain

“The EU declaration of conformity shall state that the requirements specified in this Regulation have been fulfilled in relation to the device that is covered. The manufacturer shall continuously update the EU declaration of conformity. The EU declaration of conformity shall as a minimum contain the information set out in Annex IV”

Annex IV: "The EU declaration of conformity shall contain all of the following information:"

  • “1. Name, registered trade name or registered trade mark of the manufacturer and, where applicable, of its authorised representative, the single registration number if already issued as referred to in Article 31, and the address of their registered place of business where they can be contacted;”

  • “2. A statement that the EU declaration of conformity is issued under the sole responsibility of the manufacturer,”

  • “3. The Basic UDI-DI as referred to in Annex VI Part C,”

  • “4. Product and trade name, product code, catalogue number or other unambiguous reference allowing identification and traceability of the device covered by the EU declaration of conformity, such as a photograph where appropriate, as well as its intended purpose. Except for the product name or trade name, the information allowing identification and traceability may be provided by the Basic UDI-DI referred to in point 3 of this Annex,”

  • “5. Risk class of the device according to the rules set out in Annex VIII,”

  • “6. A statement that the device covered by the present declaration is in conformity with this Regulation and, where applicable, with any other relevant Union legislation that provides for the issuing of an EU declaration of conformity,”

  • “7. References to any common specifications used and in relation to which conformity is declared,”

  • “8. Where applicable, the name and identification number of the notified body, a description of the conformity assessment procedure performed and identification of the certificate or certificates issued,”

  • “9. Where applicable, any additional information,”

  • “10. Place and date of issue of the declaration, name and function of the person who signed it as well as an indication for, and on behalf of whom, that person signed, signature.”

For a plain-language explanation of Annex IV, see our glossary entry EU declaration of conformity: what it is and what it contains, for IVD devices our IVDR declaration of conformity article, and for the relationship with the CE marking our EU declaration of conformity and CE marking articles.

4. Turkish Translation and the Single Declaration Rule

“The EU declaration of conformity shall as a minimum contain the information set out in Annex IV, and where the declaration is drawn up in other languages for devices placed on the domestic market, a Turkish translation of the declaration shall be attached.”

“(2) Where, in respect of matters not covered by this Regulation, devices are subject to other legislation which also requires an EU declaration of conformity, a single EU declaration of conformity shall be drawn up in respect of all legislation applicable to the device. The declaration shall contain all the information required for identification of the legislation to which it relates.”

5. Retention of the Declaration

“a) Retains a copy of the technical documentation, the EU declaration of conformity and, where applicable, a copy of any relevant certificate issued in accordance with Article 56, including amendments and supplements, for at least 10 years after the last device covered by the EU declaration of conformity has been placed on the market, and submits it to the Authority upon request. For implantable devices, this period is at least 15 years after the last device has been placed on the market.”

6. Role of the Authorized Representative, Importer and Distributor

These actors do not issue the declaration; they verify that it has been issued. The minimum task in the authorised representative's mandate:

“a) To verify that the EU declaration of conformity and the technical documentation have been drawn up by the manufacturer and, where applicable, that an appropriate conformity assessment procedure has been carried out.”

Before placing a device on the market or making it available, the importer (Article 13) and the distributor (Article 14) carry out the same check:

“1) That the device bears the CE marking and that the EU declaration of conformity has been drawn up,”

The importer also keeps a copy of the declaration for the period set out in the eighth paragraph of Article 10 (Article 13, ninth paragraph). For details, see our article on importer and distributor obligations.

7. Document Type in ÜTS

“ARTICLE 11- (1) A document of the type "AB Uygunluk Beyanı (MDR/IVDR)" (EU Declaration of Conformity (MDR/IVDR)) is used only in notifications and applications relating to declarations of conformity prepared under the new Medical Device Regulations. For declarations of conformity issued under the old Medical Device Regulations, a document application of the type "Uygunluk Beyanı" (Declaration of Conformity) is made.”

For declarations issued under the old directives, Article 10 of the guideline defines the document type "Uygunluk Beyanı" (Declaration of Conformity) and sets the same rule: "The Declaration of Conformity is issued by the legal manufacturer." For document types, see our page on ÜTS document types.

8. Fields Checked During ÜTS Document Review

According to Article 11 of TCKKD-KLVZ-03, the EU declaration of conformity issued by the manufacturer must contain the following fields (selected items):

  • “a) Name or trade name and address of the manufacturer,”

  • “c) Single Registration Number (SRN) of the manufacturer, if already issued (entered in the "Münferit Kimlik No" field in ÜTS),”

  • “ç) A statement that the manufacturer issues the EU declaration of conformity under its sole responsibility,”

  • “d) Basic UDI-DI,”

  • “g) Risk class of the device,”

  • “h) References to any common specifications used and in relation to which conformity is declared,”

  • “ı) For EU declarations of conformity other than those covering only Class I Other or Class A devices: name and identification number of the notified body, a description of the conformity assessment procedure performed, and identification of the EU certificate or certificates issued (certificate number or numbers),”

  • “j) Document start (issue) date,”

  • “l) Place and date of issue of the declaration,”

  • “m) Name and surname of the authorised person together with function and signature/stamp information (an e-signature is also valid)”

“are the fields that must be present and are checked during document review in ÜTS. The information entered in the system must be identical to the information in the document.”

Missing or incorrect information in the declaration therefore directly affects the document review. The guideline also requires the following:

  • For a manufacturer that has its device produced by another manufacturer, the name and address of that manufacturer are expected to appear in the declaration or the EU certificate (fourth paragraph).

  • If deemed necessary, the technical documentation of the device is uploaded to the "Belgedeki Ürün(ler)e Ait Teknik Doküman" (Technical Documentation of the Product(s) in the Document) field (fifth paragraph).

  • "The EU Declaration of Conformity must be uploaded completely, with all its annexes and additional pages." (sixth paragraph)

For declarations under the old directives, the guideline requires the scope of the products to be descriptive:

“(5) The scope of the medical devices in the declaration of conformity must be descriptive enough to identify the product. General expressions are not sufficient for identifying the product. (e.g. instead of surgical hand instruments, it should be detailed as forceps, retractor, dilator, etc.)”

9. Who Registers the Declaration in ÜTS?

The manufacturer always issues the declaration. The guideline regulates its registration in ÜTS in two separate articles:

  • Domestic manufacturer: registers the declaration of its own products itself (Article 13).

  • Declaration of a foreign manufacturer: registered by the importer (Article 12).

For a domestic manufacturer:

  • “a) It is mandatory to apply by selecting "İmalatçı Firmalar için e-imzalı Belge Başvurusu" (e-signed Document Application for Manufacturer Firms) as the Document Application Type.”

  • “c) Documents whose original language is Turkish are accepted.”

“The original of the document, electronically signed by the person registered as "Mesul Müdür" (Responsible Manager) or "İmza Yetkilisi" (Signatory) on behalf of the relevant domestic manufacturer, is uploaded completely to the "Orijinal Doküman" (Original Document) field.”

The importer uploads to the "Orijinal Doküman" field the apostille-certified copy of the document, the copy certified by a Republic of Türkiye embassy/consulate, the e-Apostille-certified original, or the manufacturer's e-signed original (Article 12, first paragraph). A Turkish translation is also uploaded separately:

“(2) Where the condition mentioned in the first paragraph of this article exists, the Turkish translation of the document uploaded to the "Orijinal Doküman" field, made by a sworn translator, is uploaded completely to the "Türkçe Doküman" (Turkish Document) field.”

2026/KKB-2 update: With TİTCK's Announcement No. 2026/KKB-2 dated 02.11.2026, a new route was introduced for the declaration of conformity and SİP declaration of a foreign manufacturer, replacing the option of uploading the manufacturer's e-signed original:

“it has been decided to implement the option that the document be uploaded to the relevant field in ÜTS by the authorised official of the importer firm authorised by the foreign manufacturer, together with an undertaking letter (taahhütname) electronically signed with an e-signature (secure e-signature) created on the basis of a qualified electronic certificate obtained from an electronic certificate service provider recognised by Türkiye.”

Source: TİTCK Announcement No. 2026/KKB-2. Required documents and steps: ÜTS document registration guide.

If the manufacturer is not established in an EU member state or in Türkiye, the "AB Yetkili Temsilcisi" (EU Authorised Representative) field must be filled in on the registration screen and the authorisation document uploaded (Article 12, seventh paragraph). For the document registration process, see our page on ÜTS document registration.

10. Frequently Asked Questions

Can a consultancy firm issue the declaration of conformity?

No. The manufacturer issues the declaration and, by preparing it, assumes responsibility for compliance (Article 19, third paragraph). A consultancy firm can support the preparation of the declaration in line with Annex IV. Signature and responsibility remain with the manufacturer.

Is the declaration of conformity the same as the EU certificate?

No. The manufacturer issues the declaration; the notified body issues the certificate. According to the guideline, notified body and certificate information is not required only in declarations covering solely Class I Other or Class A devices.

Who registers a foreign manufacturer's declaration in ÜTS?

The importer firm registers it. The certified or e-signed original of the document and the Turkish translation made by a sworn translator are uploaded together.

Is the Basic UDI-DI mandatory in the declaration?

Yes. Point 3 of Annex IV lists the Basic UDI-DI. The guideline also counts the Basic UDI-DI among the fields checked during ÜTS document review.

How long is the declaration retained?

At least 10 years after the last device covered by the declaration has been placed on the market; at least 15 years for implantable devices.

For an MDR and IVDR declaration of conformity prepared in line with Annex IV: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara