How to Obtain an EU Conformity Certificate for a Medical Device: Notified Body, Validity, Annex XII Content and ÜTS Document Registration

SUMMARY
A medical device EU conformity certificate is a certificate issued by notified bodies in accordance with Annex IX, Annex X and Annex XI under Article 56 of the Medical Device Regulation, titled "Certificates of Conformity". To obtain the certificate, the manufacturer applies to the notified body of its choice under Article 53.
In this article we cover, with verbatim quotations from the regulation and TİTCK guideline TCKKD-KLVZ-03, who issues the certificate, the conformity assessment routes by device class, the validity period, the minimum content in Annex XII, the transition of certificates issued under the old regulations, and the registration of the certificate as a document in ÜTS.
For the ÜTS document registration of your certificate: Medical Device ÜTS Registration Consultancy
Contents
1. Who Issues the Certificate?
Article 56, first paragraph of the Regulation:
“(1) Certificates issued by notified bodies in accordance with Annex IX, Annex X and Annex XI shall be in Turkish and/or an official EU language determined by the Authority, or, if not determined by the Authority, in an official EU language and/or Turkish acceptable to the notified body. The minimum content of the certificates shall be as set out in Annex XII.”
The notified body is defined in Article 3:
“oo) Notified body: a conformity assessment body designated in accordance with this Regulation,”
Only a notified body issues the certificate. Medex is not a notified body and does not issue certificates. For the concept, see our page on what is a notified body.
2. How to Obtain a Certificate: Application to the Notified Body
Article 53 regulates the application:
“(1) Where the involvement of a notified body in the conformity assessment procedure is required, the manufacturer may apply to a notified body of its choice, provided that it has been designated for conformity assessment activities relating to the relevant types of devices. The manufacturer may not lodge an application in parallel with another notified body for the same conformity assessment procedure.”
“(3) When applying to a notified body under the first paragraph, the manufacturer declares whether it has withdrawn a previous application to another notified body before the decision of that notified body, and submits information about any previous application for the same conformity assessment that was refused by another notified body.”
“(4) The notified body may require from the manufacturer any information or data necessary to properly conduct the chosen conformity assessment procedure.”
The application is made by the manufacturer.
The notified body must be designated for the relevant device types.
A parallel application to another notified body cannot be made for the same procedure.
Previously withdrawn and refused applications are declared.
3. Conformity Assessment Route by Device Class
Article 52 requires the manufacturer to carry out a conformity assessment before placing the device on the market:
“(1) Before placing a device on the market, manufacturers shall carry out an assessment of the conformity of that device in accordance with the applicable conformity assessment procedures set out in Annex IX to Annex XI.”
For class III, the third paragraph:
“(3) Manufacturers of class III devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Annex IX. Alternatively, the manufacturer may choose to apply a conformity assessment as specified in Annex X coupled with a conformity assessment as specified in Annex XI.”
For class IIa, point (a) of the fifth paragraph:
“a) Be subject to a conformity assessment as specified in Chapters I and III of Annex IX, including an assessment of the technical documentation as specified in point 4 of Annex IX for at least one representative device per category of devices,”
For class I, the sixth paragraph; the involvement of the notified body is limited to sterile devices, devices with a measuring function and reusable surgical instruments, and the listed aspects:
“(6) Manufacturers of class I devices, other than custom-made or investigational devices, shall declare the conformity of their products by issuing the EU declaration of conformity referred to in Article 19 after drawing up the technical documentation set out in Annex II and Annex III. If those devices are placed on the market in a sterile condition, have a measuring function or are reusable surgical instruments, the manufacturer shall apply the procedures set out in Chapters I and III of Annex IX or in Part A of Annex XI.”
For class IIb the fourth paragraph provides separately. For choosing the procedure, see our article on choosing the conformity assessment procedure.
4. Validity Period, Suspension and Withdrawal
“(2) Certificates shall be valid for the period indicated on the certificate, which shall not exceed 5 years. On application by the manufacturer, the validity of a certificate may be extended, on the basis of a re-assessment in accordance with the applicable conformity assessment procedures, for further periods, each not exceeding 5 years. Any supplement to a certificate shall remain valid as long as the certificate which it supplements is valid.”
“(4) Where a notified body finds that the requirements of this Regulation are no longer met by the manufacturer, it shall, taking account of the principle of proportionality, suspend or withdraw the certificate issued or place restrictions on it, unless compliance with such requirements is ensured by appropriate corrective action taken by the manufacturer within an appropriate deadline set by the notified body. The notified body shall give reasons for its decision.”
For details on suspension, see our article on suspension of a notified body certificate.
5. Annex XII: Minimum Content of the Certificate
From the general requirements in Annex XII Chapter I:
“2. Each certificate shall refer to only one conformity assessment procedure.”
“3. Certificates shall be issued to only one manufacturer. The name and address of the manufacturer included in the certificate shall be the same as that registered in the electronic system referred to in Article 30.”
Minimum content in Annex XII Chapter II:
“1. Name, address and identification number of the notified body,”
“2. Name and address of the manufacturer and, where applicable, of the authorised representative,”
“3. Unique number identifying the certificate,”
“4. If already issued, the single registration number of the manufacturer referred to in the second paragraph of Article 31,”
“5. Date of issue,”
“6. Date of expiry,”
“9. Reference to this Regulation and the relevant Annex in accordance with which the conformity assessment has been carried out,”
“14. Conditions or limitations on the validity of the certificate,”
“15. Legally valid signature of the notified body in accordance with the applicable legislation.”
6. Certificates Issued Under the Old Regulations
Article 109 of the Medical Device Regulation lists Regulation (EU) 2023/607 among the EU legislation it harmonises:
“e) (Added: RG-2/4/2023-32151) Regulation (EU) 2023/607 of the European Parliament and of the Council of 15/3/2023 amending Regulations (EU) 2017/745 and (EU) 2017/746 as regards the transitional provisions for certain medical devices and in vitro diagnostic medical devices,”
Point (b) of the fourth paragraph of provisional Article 1:
“Certificates issued by notified bodies from 25/5/2017 in accordance with the Medical Device Regulation and the Active Implantable Medical Devices Regulation repealed by Article 110, that have not been withdrawn and are still valid on 26/5/2021, remain valid after the end of the period indicated on the certificate until the dates specified in point (b) of the fifth paragraph for the relevant risk class of the devices.”
The dates in point (b) of the fifth paragraph:
“class IIb implantable devices, except screws, wedges, plates, wires, pins, clips and connectors, until 31/12/2027.”
“2) Class IIb devices other than those covered by sub-point (1), class IIa devices and class I devices placed on the market in a sterile condition or with a measuring function, until 31/12/2028.”
The first sub-point begins with the words "All class III devices and sutures, staples, dental fillings, dental braces, tooth crowns,". To place devices on the market until these dates, the conditions in point (ç) of the fifth paragraph must be met; the conditions and the calendar are in our article on the (EU) 2023/607 MDR transition calendar.
7. Registration of the Certificate as a Document in ÜTS
Guideline TCKKD-KLVZ-03 defines two document types. Article 6:
“ARTICLE 6- (1) A document of the type "EC Sertifikası" (EC Certificate) is used only in notifications and applications relating to certificates prepared by notified bodies under the Old Medical Device Regulations. For certificates issued under the New Medical Device Regulations, a document registration/update notification and application of the type "AB Sertifikası (MDR/IVDR)" (EU Certificate (MDR/IVDR)) is made.”
Article 7:
“ARTICLE 7- (1) A document of the type "AB Sertifikası (MDR/IVDR)" is used only in notifications and applications relating to certificates prepared by a notified body under the New Medical Device Regulations.”
The rules to be followed in registration are in the fourth, fifth and sixth paragraphs of Article 7:
“(4) If the document has been revised, the revision date is entered as the document start date. If the document has not been revised and the document start date is not clearly stated on the certificate, the most recent issue (signature) date on the certificate is entered as the document start date.”
“(5) Where the revision number is stated on the EU certificate and shown together with the document (certificate) number, the certificate is registered with its revision number. (For example XXXXX.Rev.1)”
“(6) The EU certificate must be uploaded and submitted to the Authority completely, with all its annexes and additional pages.”
If the notified body is not defined in ÜTS, the seventh paragraph applies:
“(7) If the notified body that issued the EU certificate is not defined in ÜTS as a notified body, before the ÜTS document registration of that EU certificate, a request is opened via the "Talep Bildir" (Report Request) screen in ÜTS, under the "Onaylanmış/Uygunluk Değerlendirme Kuruluş İşlemleri" (Notified/Conformity Assessment Body Procedures) field, for adding the notified body to ÜTS or defining its authorisation under the relevant legislation.”
For document types, see our pages on ÜTS document type: EC certificate and ÜTS document type: EU certificate, for the registration steps our article on ÜTS document registration, and our page on ÜTS document registration service.
8. Frequently Asked Questions
Who issues the medical device EU conformity certificate?
Under Article 56, notified bodies issue it in accordance with Annex IX, Annex X and Annex XI. Medex is not a notified body and does not issue certificates.
Who makes the certificate application?
Under Article 53, the manufacturer applies to a notified body of its choice that has been designated for the relevant device types. A parallel application cannot be made for the same procedure.
How long is the certificate valid?
Under Article 56/2, for the period indicated on the certificate, not exceeding 5 years; it can be extended after re-assessment for periods each not exceeding 5 years.
Is a certificate required for a class I device?
Under Article 52/6, the manufacturer of a class I device issues an EU declaration of conformity. If the device is sterile, has a measuring function or is a reusable surgical instrument, the notified body is involved, limited to the listed aspects.
Which document type is selected in ÜTS?
According to TCKKD-KLVZ-03, certificates under the old regulations are registered as "EC Sertifikası" and those under the new regulations as "AB Sertifikası (MDR/IVDR)".
For the ÜTS document registration of your certificate: Medical Device ÜTS Registration Consultancy
Sources