EU Declaration of Conformity and CE Marking: Medical Device Regulation Articles 19 and 20, Annexes IV and V and ÜTS Document Registration

SUMMARY
The EU declaration of conformity and the CE marking are regulated in Articles 19 and 20 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation). In the declaration, the manufacturer states, under its own responsibility, that the device meets the requirements of the Regulation; the CE marking shows this conformity on the device, its packaging and the instructions for use.
In this article we give the full text of the two articles, Annexes IV and V, the declaration route for class I devices, and the document type "AB Uygunluk Beyanı (MDR/IVDR)" (EU Declaration of Conformity (MDR/IVDR)) from the ÜTS Registration Guide TCKKD-KLVZ-03, in verbatim quotations.
For ÜTS document registration of your EU declaration of conformity: Medical Device ÜTS Registration Consultancy
Table of Contents
1. What Does the EU Declaration of Conformity State?
Article 19, paragraph 1 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation):
“ARTICLE 19 - (1) The EU declaration of conformity states that the requirements specified in this Regulation have been fulfilled in relation to the device it covers. The manufacturer keeps the EU declaration of conformity up to date. The EU declaration of conformity contains, as a minimum, the information set out in Annex IV, and where the declaration is drawn up in another language for devices placed on the domestic market, a Turkish translation of the declaration is attached. For devices that are also placed on the EU market, the EU declaration of conformity is provided in the language required by the EU Member State(s) in which the device is made available.”
The manufacturer keeps the declaration up to date.
The minimum content is set out in Annex IV.
For a device placed on the domestic market, if the declaration is drawn up in another language, its Turkish translation is attached.
On the EU market, the declaration is provided in the language required by the Member State in which the device is made available.
For the difference between the EC certificate and the EU certificate and the declaration of conformity, see our article on Differences between the EC and EU certificate and the declaration of conformity.
2. More Than One Piece of Legislation: A Single Declaration and Responsibility
Paragraphs 2 and 3 of Article 19:
“(2) Where, for aspects not covered by this Regulation, devices are also subject to other legislation which requires an EU declaration of conformity, a single EU declaration of conformity is drawn up in respect of all the legislation applicable to the device. The declaration contains all the information required to identify all the legislation to which it relates.”
“(3) By drawing up the EU declaration of conformity, the manufacturer assumes responsibility for compliance with the requirements of this Regulation and all other legislation applicable to the device.”
If a device is also subject to another technical regulation, a single declaration identifying all the legislation is drawn up instead of separate ones. The details of this are in our article on multiple technical regulations.
3. Annex IV: Information to Be Included in the Declaration
“The EU declaration of conformity contains all of the following information:”
“1. Name, registered trade name or registered trade mark of the manufacturer and, where applicable, of its authorised representative, the single registration number referred to in Article 31 if already issued, and the address of their registered places of business at which they can be contacted;”
“2. A statement that the EU declaration of conformity is issued under the sole responsibility of the manufacturer;”
“3. The Basic UDI-DI as referred to in Annex VI Part C;”
“4. Product and trade name, product code, catalogue number or other unambiguous reference allowing identification and traceability of the device covered by the EU declaration of conformity, such as a photograph where applicable, as well as its intended purpose. Except for the product or trade name, the information allowing identification and traceability may be provided by the Basic UDI-DI referred to in point 3 of this Annex;”
“5. Risk class of the device in accordance with the rules set out in Annex VIII;”
“6. A statement that the device that is covered by the present declaration is in conformity with this Regulation and, if applicable, with any other relevant legislation that provides for the issuing of an EU declaration of conformity;”
“7. References to any common specifications used and in relation to which conformity is declared;”
“8. Where applicable, the name and identification number of the notified body, a description of the conformity assessment procedure performed and identification of the certificate or certificates issued;”
“9. Where applicable, additional information;”
“10. Place and date of issue of the declaration, name and function of the person who signed it as well as an indication for, and on behalf of whom, that person signed, signature.”
The inclusion of the Basic UDI-DI in the declaration is also stated in Article 27, paragraph 6:
“(6) The Basic UDI-DI, as defined in Annex VI Part C, is included in the EU declaration of conformity referred to in Article 19.”
For the Basic UDI-DI, see our article What Is the Basic UDI-DI; for the common specifications in point 7, see our article harmonised standards and common specifications.
4. When Is the Declaration Drawn Up? Class I Devices
Article 10, paragraph 6 ties the declaration and the CE marking to the outcome of the conformity assessment:
“(6) Manufacturers of devices, other than custom-made or investigational devices, which demonstrate conformity with the applicable requirements as a result of the applicable conformity assessment procedures, draw up an EU declaration of conformity in accordance with Article 19 and affix the CE marking of conformity in accordance with Article 20.”
Article 52, paragraph 6 for class I devices:
“(6) Manufacturers of class I devices, other than custom-made or investigational devices, declare the conformity of their products by drawing up the EU declaration of conformity referred to in Article 19 after having drawn up the technical documentation set out in Annex II and Annex III. If those devices are placed on the market in a sterile condition, have a measuring function or are reusable surgical instruments, the manufacturer applies the procedures set out in Chapters I and III of Annex IX or in Part A of Annex XI.”
For class I devices, the declaration is drawn up after the Annex II and Annex III technical documentation has been prepared. For class I devices that are sterile, have a measuring function or are reusable surgical instruments, a notified body procedure also applies. For details, see our article on class I medical devices.
5. On Which Devices and Where Is the CE Marking Affixed?
Paragraphs 1, 3 and 4 of Article 20:
“ARTICLE 20 - (1) Devices, other than custom-made or investigational devices, that are considered to be in conformity with the requirements of this Regulation bear the CE marking of conformity, as presented in Annex V.”
“(3) The CE marking is affixed visibly, legibly and indelibly to the device or its sterile packaging. Where such affixing is not possible or not warranted on account of the nature of the device, the CE marking is affixed to the packaging. The CE marking also appears in any instructions for use and on any sales packaging.”
“(4) The CE marking is affixed before the device is placed on the market. It may be followed by a pictogram or any other marking indicating a special risk or use.”
Custom-made and investigational devices do not bear the CE marking (20/1).
Place: the device or its sterile packaging; if that is not possible, the packaging (20/3).
Also in every set of instructions for use and on every sales packaging (20/3).
Time: before the device is placed on the market (20/4).
For custom-made devices, see our article on custom-made devices.
6. Notified Body Identification Number
Paragraph 5 of Article 20:
“(5) Where applicable, the CE marking is followed by the identification number of the notified body responsible for the conformity assessment procedures set out in Article 52. The identification number is also indicated in any promotional material which mentions that a device fulfils the requirements for CE marking.”
The identification number follows the CE marking only where a notified body is responsible for the conformity assessment procedure. The number is also shown in promotional material stating that the CE requirements are fulfilled.
Paragraphs 2 and 6 bind the general principles and other legislation:
“(2) The CE marking is subject to the general principles set out in Article 10 of the “CE” İşareti Yönetmeliği (CE Marking Regulation), put into force by Presidential Decision No. 4021 of 26 May 2021.”
“(6) Where devices are subject to other legislation which also provides for the affixing of the CE marking, the CE marking indicates that the devices also fulfil the requirements of that other legislation.”
7. Annex V: Form and Size of the CE Marking
“1. The CE marking consists of the initials “CE” taking the following form:”
“2. If the CE marking is reduced or enlarged, the proportions given in the above graduated drawing are respected.”
“3. The various components of the CE marking have substantially the same vertical dimension, which is not less than 5 mm. This minimum dimension may be waived for small-scale devices.”
For other information to appear on the label and in the instructions for use, see our article on Turkish labels and instructions for use.
8. Document Type in ÜTS: "AB Uygunluk Beyanı (MDR/IVDR)" (EU Declaration of Conformity (MDR/IVDR))
Article 11 of the TCKKD-KLVZ-03 guide:
“ARTICLE 11- (1) A document of the “AB Uygunluk Beyanı (MDR/IVDR)” (EU Declaration of Conformity (MDR/IVDR)) type is the document type used only in notifications and applications concerning declarations of conformity prepared under the New Medical Device Regulations. For declarations of conformity issued under the Old Medical Device Regulations, a document application is made with the “Uygunluk Beyanı” (Declaration of Conformity) document type.”
Some of the fields listed in the same article:
“c) The manufacturer’s Single Registration Number (SRN), if already issued (entered in the “Münferit Kimlik No” (Single Identification No.) field in ÜTS),”
“d) Basic UDI-DI,”
“g) Risk class of the device,”
“ı) For EU declarations of conformity other than those covering only Class I-Other or Class A devices: the name and identification number of the notified body, a description of the conformity assessment procedure performed and identification of the EU certificate or EU certificates issued (certificate number or numbers),”
“j) Document start (issue) date,”
“k) Document end (validity) date,”
“are the fields that must be present and that are checked during document review in ÜTS. The information entered in the system must be identical to the information in the document.”
“(3) The EU Declaration of Conformity is issued by the legal manufacturer.”
“(6) The EU Declaration of Conformity must be uploaded in full, together with all its annexes and additional pages.”
For an importer’s registration of a foreign manufacturer’s declaration, Article 12 governs the "Türkçe Doküman" (Turkish Document) field:
“(2) Where the condition referred to in the first paragraph of this Article exists, the complete Turkish translation of the document uploaded in the “Orijinal Doküman” (Original Document) field, made by a sworn translator, is uploaded to the “Türkçe Doküman” (Turkish Document) field.”
For the approval conditions for the "Orijinal Doküman" (Original Document) field, see our medical device apostille guide, and for the registration steps, our ÜTS document registration articles. To have the declaration prepared on behalf of the manufacturer, see our MDR/IVDR declaration of conformity preparation page.
9. Frequently Asked Questions
Does the EU declaration of conformity have to be in Turkish?
Under Article 19, paragraph 1, if the declaration of a device placed on the domestic market is drawn up in another language, a Turkish translation is attached. In ÜTS, a sworn translator’s translation is uploaded to the "Türkçe Doküman" (Turkish Document) field in the importer’s registration.
How many declarations are drawn up for a device subject to more than one piece of legislation?
Under Article 19, paragraph 2, a single EU declaration of conformity is drawn up, and it identifies all the legislation to which it relates.
Is the notified body number always shown next to the CE marking?
No. Under Article 20, paragraph 5, the number follows the CE marking where a notified body is responsible for the conformity assessment procedure.
Where does the CE marking appear?
Under Article 20, paragraph 3, it is affixed to the device or its sterile packaging; it also appears in every set of instructions for use and on every sales packaging.
Under which type is a declaration from the MDD period registered in ÜTS?
Under Article 11 of TCKKD-KLVZ-03, the "Uygunluk Beyanı" (Declaration of Conformity) type is used for declarations under the old regulations, and the "AB Uygunluk Beyanı (MDR/IVDR)" (EU Declaration of Conformity (MDR/IVDR)) type for those under the new regulations.
For ÜTS document registration of your EU declaration of conformity: Medical Device ÜTS Registration Consultancy
Sources