EUDAMED Actor and Device Registration Consultancy: Where SRN, Basic UDI-DI and ÜTS Intersect
If you think your ÜTS registration is ready because you are registered in EUDAMED, or assume the opposite, you are confusing the two systems

EUDAMED Registration Consultancy and the SRN Process
If you think your ÜTS registration is ready because you are registered in EUDAMED, or assume the opposite, you are confusing the two systems. EUDAMED is the database on the European side, and ÜTS is Türkiye's national product tracking system; one does not replace the other, and the legislation regulates them as separate obligations.
This page explains EUDAMED from the places where Turkish legislation refers to it: in which ÜTS field the SRN is entered, which burden the EUDAMED output lightens in which field, and what EMDN and Basic UDI-DI correspond to in the two systems. The limit is clear from the start: the module rules of EUDAMED itself are set by the European Commission, and we manage the intersection on the Turkish side.
EUDAMED and ÜTS Are Not the Same System
Subparagraph (n) of Article 4 of the Guide on Registration Processes of Medical Devices to the Product Tracking System defines EUDAMED as "the European database system on medical devices". The first paragraph of Article 33 of the Medical Device Regulation published in the Official Gazette dated 2/6/2021 and numbered 31499 (repeated) is more detailed: EUDAMED is the system defined as the European database on medical devices, established and managed by the Commission, into which the relevant parties enter data.
The second paragraph of the same article lists seven electronic systems that make up EUDAMED: device registration, the UDI database, the economic operator registration, notified bodies and certificates, clinical investigations, vigilance and post-market surveillance, and market surveillance and inspection.
The provision that establishes the relationship between the two systems is Article 34 of the Medical Device Regulation. The first paragraph states that, without prejudice to the obligations relating to EUDAMED specified in the Regulation, economic operators, notified bodies, health institutions and health professionals established in the country will fulfil their obligations relating to ÜTS. The second paragraph states that ÜTS obligations will be determined by the regulations published by the Agency.
The result is clear: being registered in one does not remove the obligation in the other.
What Is the SRN and How Does It Differ from the ÜTS Company Number
Subparagraph (rr) of Article 4 of the Guide defines the Single Registration Number (SRN) as "the single registration number generated by EUDAMED to be allocated to the manufacturer, authorized representative or importer". In other words, the SRN is not a product number but an actor number.
How the number comes into being is written in the Medical Device Regulation. The first paragraph of Article 30 describes the electronic system established and managed by the Commission to generate the SRN and to identify the manufacturer and, if any, the authorized representative and the importer; the second paragraph of Article 31 states that the Agency, after verifying the data entered, will forward the number provided from this system to the relevant party.
On the ÜTS side this number has a concrete place. Subparagraph (ç) of the second paragraph of Article 7 of the Guide states that the manufacturer's Single Registration Number (SRN) must appear in the EU certificate if it has already been issued, and that this information is entered in the "Single Identification No" field in ÜTS. At the end of the same paragraph the document review rule is repeated: the information entered in ÜTS must be the same as the information in the certificate.
The situation we encounter most often in practice is this: the SRN is written on the certificate, but it is either never entered in ÜTS or the manufacturer's internal reference number is entered instead. Before a document application, we compare this field with the certificate text.
Where the EUDAMED Registration Output Is Used in ÜTS
There are two points where EUDAMED is useful in your ÜTS file, and both are clearly written in the Guide.
The first is the authorized representative document. Subparagraph (a) of the seventh paragraph of Article 12 of the Guide makes it mandatory to make an entry in the "EU Authorized Representative" field on the ÜTS registration screen for document registration notifications of the type "EU Declaration of Conformity (MDR/IVDR)" belonging to manufacturers not established in an EU member country or in Türkiye. Subparagraph (b) states that the agreement and/or authorization document made between the manufacturer and the EU authorized representative is uploaded to the "Authorization Document" field that appears under this field; alternatively, the output of the actor screen on the EUDAMED website that belongs to the relevant manufacturer and clearly shows the information of the EU Authorized Representative in question can be uploaded.
This alternative has a concrete benefit. Subparagraph (c) normally requires the original of the document and its sworn Turkish translation to be uploaded together to the "Authorization Document" field, while its first sub-item states, as an exception, that if a EUDAMED output has been uploaded to this field, the translation does not need to be uploaded. The same subparagraph also states that no apostille or embassy/consulate approval is required on the original of the authorization document.
The second is the certificate. Subparagraph (e) of the first paragraph of Article 8 of the Guide regulates that this route can be used provided that the certificate has been uploaded to EUDAMED by the notified body that issued it, with all its current and valid pages, and that the certificate can be accessed and verified in EUDAMED with all its pages.
If the certificate is e-signed: According to the first sub-item of subparagraph (e), the original of the certificate is uploaded, with the e-signature verification provisions in subparagraph (ç) of the same paragraph being valid.
If the certificate is not e-signed: According to the second sub-item of subparagraph (e), a copy of the relevant certificate certified by a Turkish notary is uploaded; this route is expressly recognized in the Guide. In both cases the document is uploaded in full, together with its annexes.
It is necessary to know the asymmetry here: the EUDAMED route exists for certificates, not for declarations of conformity. The first paragraph of Article 12 of the Guide lists, in four subparagraphs, the routes accepted for the "Original Document" field in "Declaration of Conformity" or "EU Declaration of Conformity (MDR/IVDR)" notifications of importing companies belonging to the foreign manufacturer: the apostilled copy, the copy approved by the embassy/consulate of the Republic of Türkiye in the manufacturer's country, the original of the e-Apostille certified document, and the original of the document signed electronically by the manufacturer. Neither the notified body query website nor the EUDAMED route is on this list.
2026/KKB-2 update: With the TİTCK Announcement No. 2026/KKB-2 dated 02.11.2026, a new route replaced the option of uploading the manufacturer's original e-signed document for the declaration of conformity and the SİP declaration of a foreign manufacturer. The announcement states the decision to apply the option that the document in question be uploaded to the relevant field in ÜTS together with an undertaking, signed electronically by an official of the importing company authorized by the foreign manufacturer, with an e-signature (secure e-signature) created on the basis of a qualified electronic certificate obtained from an electronic certificate service provider recognized by Türkiye.
Source: TİTCK Announcement No. 2026/KKB-2. Required documents and steps: ÜTS document registration guide.
Company Information Required for EUDAMED Actor Registration
Actor registration is not a document file but a matter of role declaration and verifiable company identity. What we ask from you when opening the file is as follows:
Role declaration: Subparagraph (v) of Article 4 of the Guide defines the economic operator as a manufacturer, authorized representative, importer, distributor or SİP manufacturer. The role in which you appear in the registration determines the framework of all subsequent obligations.
Company name and address: They must be exactly the same as the name and address in the certificate, the declaration of conformity and the trade registry record; these fields are compared in the document review.
Existing SRN and the manufacturer's SRN, if any: Under the definition in subparagraph (rr) of Article 4, the SRN arises separately for the manufacturer, authorized representative and importer; your own number and the manufacturer's number are recorded separately.
EU authorized representative agreement: The second paragraph of Article 11 of the Medical Device Regulation states that the appointment of an authorized representative becomes valid when an authorization document is issued and accepted in writing by the representative, and that it covers, at a minimum, all devices of the same generic device group.
Subparagraph (d) of the seventh paragraph of Article 12 of the Guide draws a limit: manufacturers established in an EU member country or in Türkiye cannot appoint an EU authorized representative.
Registration Periods: The Calendar Given by the Legislation
The periods below show not how long a job will take to finish, but the notification and confirmation calendar that the Medical Device Regulation places on economic operators.
Two weeks: Under the second paragraph of Article 30, importers verify, within 2 weeks of placing a device other than a custom-made device on the market, that the manufacturer or authorized representative has entered the information in the electronic system. If the information is missing or incorrect, they inform the relevant authorized representative or manufacturer and add their own information to the relevant entry.
One week: Under the fourth paragraph of Article 31, economic operators update the data in the system within 1 week of any change in the information.
One year, then every two years: Under the fifth paragraph of Article 31, economic operators confirm the accuracy of the data within 1 year of the initial submission of the information and every 2 years thereafter. The same paragraph states that if confirmation is not made within 6 months after the end of these periods, the Agency may take appropriate corrective measures.
This calendar turns actor registration from a "job done once and finished" into an ongoing duty. A change of name, address or representative must be reported within a one-week window; if we keep the registration, the only condition is that you inform us of the change in time.
Registration Requirement Before Applying to a Notified Body
An error in sequence is the most expensive mistake in this file. The first paragraph of Article 31 of the Medical Device Regulation states that, before placing a device other than a custom-made device on the market, manufacturers, authorized representatives and importers that are not already registered will enter the information into the electronic system for the purpose of registration. The second sentence of the same paragraph is critical: where the conformity assessment procedure requires the involvement of a notified body pursuant to Article 52, this information is entered into the system before applying to the notified body.
The same sequence applies to the Basic UDI-DI. The third paragraph of Article 29 states that, for devices subject to certain conformity assessments, the allocation of the Basic UDI-DI will be carried out before applying to the notified body, that the notified body will refer to the Basic UDI-DI on the certificate, and that it will confirm the accuracy of the information in EUDAMED.
There is also a provision on the fate of the data: under the seventh paragraph of Article 31, the data entered into this system is made publicly available.
EMDN: The ÜTS Counterpart of EUDAMED Terminology
EMDN is defined in subparagraph (k) of Article 4 of the Guide as the European Medical Device Nomenclature (Terminology). The rules of use are collected in subparagraph (ç) of the fourth paragraph of Article 27.
What it is for: According to the second sub-item, EMDN is the terminology to be used when registering devices in the EUDAMED database and in ÜTS. The same sub-item states that the structure of EMDN is a Seven-Level Tree-Like Hierarchy Model and refers to Annex-22 of the Guide for basic explanations.
Is it mandatory: According to the first sub-item, until a new decision is taken by the Agency, and if the EMDN code that defines the device is not included in at least one of the documents of the device (EU certificate, EU Declaration of Conformity), this field is optional to select in ÜTS.
Mandatory if in the document: According to the fifth sub-item, if the EMDN code appears in at least one of the documents, selecting the most detailed code and terminal term of that code is mandatory. Even if the document states a higher-level code, the most detailed related code suitable for the device is selected.
How many codes: According to the fourth sub-item, each device has only one EMDN code.
If not in the document: According to the sixth sub-item, the code obtained from the manufacturer is selected; if the manufacturer has no code, the most appropriate code defining the device may be selected.
Access: According to the seventh sub-item, all EMDN codes can be accessed free of charge from the EMDN link of the European Commission; the same codes are also available on the "Select EMDN Term" and "List EMDN" screens in ÜTS.
If the code is not defined in ÜTS, according to the eighth sub-item, a request is opened from the "Report Request" screen through the "Adding an EMDN" area. The ninth sub-item sets a limit: only the most detailed codes and terminal terms defined in the official link are added to ÜTS; less detailed codes are not added.
GMDN Is Still a Mandatory Field
The arrival of EMDN has not removed GMDN; in the Guide the two are separately defined, separate fields. Subparagraph (ö) of Article 4 defines GMDN as the Global Medical Device Nomenclature and subparagraph (k) defines EMDN as the European Medical Device Nomenclature.
For devices under the old regulations: The first sub-item of subparagraph (b) of the fourth paragraph of Article 25 of the Guide states that selecting the GMDN field in ÜTS is a mandatory field; the second sub-item states that each device has only one GMDN code.
For devices under MDR: The first and second sub-items of subparagraph (b) of the fourth paragraph of Article 27 repeat the same two rules.
For devices under IVDR: The first sub-item of subparagraph (b) of the fourth paragraph of Article 30 also makes this field mandatory.
The way code selection is tied to documents is also the same under all three regimes: if the GMDN code that defines the device appears in at least one of the documents, that code is selected; if not, the code obtained from the manufacturer is selected, and if the manufacturer has no code, the most appropriate code that defines the device. The fifth sub-item of the same subparagraph of Article 25 adds a check: there must be consistency between the information in the description of the GMDN code and characteristics of the device such as whether it is single-use or sterile.
Basic UDI-DI: The Common Key of the Two Systems
Subparagraph (ggg) of Article 4 of the Guide defines the Basic UDI-DI as the identifier code assigned at the level of the device's unit of use, serving as the main key in the database and in the relevant documentation for linking devices with the same intended purpose, risk class and essential design and manufacturing characteristics.
Subparagraph (c) of the fifth paragraph of Article 27 opens this field on the ÜTS side, and according to its first sub-item the field is mandatory. The second sub-item states that the Basic UDI-DI is the main key in EUDAMED and in ÜTS and in documents such as the EU certificate, EU declaration of conformity, free sale certificate and technical documentation.
Mandatory in the declaration: According to the sixth sub-item, the Basic UDI-DI information must be included in the EU Declaration of Conformity of the device.
In the certificate, depending on the case: According to the fifth sub-item, the Basic UDI-DI information is mandatory in the types EU Technical Documentation Certificate, EU Type-Examination Certificate and EU Product Verification Certificate; there is no such requirement for the types EU Quality Management System Certificate and EU Production Quality Assurance Certificate.
Not on the label: According to the seventh sub-item, the Basic UDI-DI does not appear on the device's market placement label and/or sales packaging. This is the rule that prevents confusion with the barcode.
Code length: According to the tenth sub-item, the Basic UDI-DI code value that issuing entities may allocate consists of at most 25 characters.
Check character: According to the eleventh sub-item, a check digit/character based on the algorithm determined by the issuing entities must be part of the Basic UDI-DI. According to the eighth sub-item, the issuing entities and standards authorized under MDR are GS1, HIBCC, ICCBBA and IFA.
The fourth sub-item also ties the selection flow: through the "Select Basic UDI-DI" screen, a selection is made only among the numbers associated with the EU Declaration of Conformity selected from the documents field, and more than one selection cannot be made.
Device Registration in EUDAMED
Actor registration introduces the company to EUDAMED, while device registration introduces the product. The two are separate steps and the order does not change: first the SRN, then the device. The basis of device registration is Article 29 of the Medical Device Regulation:
"Before placing a device other than a custom-made device on the market, the manufacturer shall, in accordance with the rules of the issuing entities referred to in the second paragraph of Article 27, assign to the device a Basic UDI-DI as defined in Annex VI Part C and enter it in the UDI database together with the other core data elements referred to in Annex VI Part B related to that device." (Medical Device Regulation, Art. 29/1)
Each variant comes under this registration with its own UDI-DI. We explained the step-by-step flow for a domestic manufacturer in our EUDAMED device registration guide.
Deadline for Device Registration
The obligations linked to the EUDAMED modules depend on the date on which the Commission publishes in the Official Journal of the European Union the notice that the relevant electronic system is functional. The seventeenth paragraph of provisional Article 1, added to the Medical Device Regulation by the amendment dated 17/8/2024, states the following for device data:
"Manufacturers shall ensure that the information required to be entered in Eudamed pursuant to Article 29 is entered in the electronic system referred to in subparagraphs (a) and (b) of the second paragraph of Article 33, including for the following devices, within 12 months at the latest from the date of publication of the notice referred to in subparagraph (a) of the first paragraph of Article 111, provided that they are placed on the market within 6 months from the date of publication of this notice" (Medical Device Regulation, Provisional Art. 1/17-a)
The notice was published in November 2025. Accordingly the calendar is as follows:
28 May 2026: Obligations linked to the first modules started to apply (Art. 111/1-a, 6 months after the notice).
28 November 2026: Device information must have been entered in EUDAMED; this is the manufacturer's obligation (Provisional Art. 1/17-a, 12 months).
28 May 2027: Certificate information must have been entered in EUDAMED; this is the notified body's obligation (Provisional Art. 1/17-b, 18 months).
The manufacturer's date for device registration is 28 November 2026.
Devices with Certificates Under the Old Regulations
Devices on the market with an old directive (MDD) certificate are also within the scope of registration. Sub-item (2) of subparagraph (a) of the same paragraph lists these devices:
"Devices, excluding custom-made devices, placed on the market pursuant to subparagraph (a), (b) or (c) of the fifth paragraph of provisional Article 1, for which the manufacturer has carried out a conformity assessment pursuant to Article 52 and which are not yet registered in Eudamed." (Medical Device Regulation, Provisional Art. 1/17-a-2)
We also carry out the EUDAMED registration of these devices, with the same Basic UDI-DI and the same product information as in the ÜTS registration.
Three Ways of Entering Data
Device data is entered into EUDAMED in three ways, and the choice is made according to the number of devices. The EUDAMED help page states:
"If you must register up to 100 devices, then use the EUDAMED user interface. If you must register 100-1000 devices, then use the bulk upload/download. If you must register more than 1000 devices, then use the M2M data exchange." (EUDAMED, Modes of data input)
That is, entry one by one from the screen up to 100 devices, bulk upload between 100 and 1000, and system-to-system data transfer above 1000. You also choose how the data will be entered into EUDAMED: we prepare the upload file and you upload it from your own account, we do it together via remote connection, or we take authorization in your account and do the entry ourselves.
Bulk XML Upload
For product groups with many variants, we collect the data in an Excel template. The template is compared with your ÜTS records and declaration, and then an XML file is produced with our own tool. The file is uploaded from the bulk upload section of EUDAMED and the result is checked record by record. Records that are not accepted are corrected in the template and uploaded again; this way the single source for the whole product group is the Excel template.
How Is the SRN Number Obtained?
The SRN (single registration number) is the unique number given to a manufacturer, authorized representative or importer whose EUDAMED actor registration has been completed. There is no separate application to obtain an SRN; the number is generated when the actor registration is verified.
EUDAMED Actor Registration for the SRN
The first paragraph of Article 31 of the Medical Device Regulation determines when actor registration is to be made:
"Before placing a device other than a custom-made device on the market, manufacturers, authorized representatives and importers that are not already registered in accordance with this article shall, for the purpose of registration, enter the information referred to in item 1 of Annex VI Part A into the electronic system referred to in Article 30."
Who Issues the SRN?
According to the second paragraph of the same article, the SRN is forwarded by the competent authority that verifies the data entered. For economic operators established in Türkiye this authority is TİTCK:
"After verifying the data entered pursuant to the first paragraph, the Agency shall forward the single registration number provided by the electronic system referred to in Article 30 to the manufacturer, authorized representative or importer."
Where Is the SRN Used?
According to the third paragraph of the article, the manufacturer uses the SRN when applying to the notified body for conformity assessment and when accessing EUDAMED. On the ÜTS side, the SRN is entered in the "Single Identification No" field in EU certificate and EU declaration of conformity records (TCKKD-KLVZ-03, Articles 7 and 11).
Who Needs EUDAMED Registration Consultancy?
Domestic Manufacturers
A domestic manufacturer that places its product on the EU market or under MDR/IVDR carries out the actor registration before placing the device on the market and, if a notified body is required, before applying to it. EUDAMED actor registration, the SRN and then device registration with the Basic UDI-DI are the basic steps for these companies.
EU Authorized Representatives
The authorized representatives of manufacturers established outside the EU also register in EUDAMED as actors and receive their own SRNs. Medex also provides this service as an EU authorized representative in Türkiye through medexeurep.com.
Importers
The importer's obligation in EUDAMED is not limited to its own registration. The second paragraph of Article 30 of the Medical Device Regulation states:
"Importers shall, within 2 weeks of a device other than a custom-made device being placed on the market, verify that the manufacturer or authorized representative has entered the information referred to in the first paragraph into the electronic system."
Carrying out EUDAMED actor and device registration in the same file as the ÜTS registration records prevents the same information from being written differently in the two systems. How this is done is explained in the earlier sections of this page; for the general operation of EUDAMED, see our EUDAMED registration and SRN guide.
Data We Ask From You
The risk class and EMDN code of each device.
Under which Basic UDI-DI your devices will be grouped; the grouping is clarified before registration.
For devices with a notified body certificate, the certificate.
The countries in which the device is placed on the market.
Why a Registration Goes Back for Correction
One reason we have encountered: the contact person, person responsible for regulatory compliance or local actor administrator information in the actor registration was not exactly the same as the information in the ÜTS company registration, and the registration came back with a correction request. For this reason, before entering anything into EUDAMED, we compare all name, title and address information with the ÜTS record.
The Second Leg of the EUDAMED Transition: Supply Interruption Notification
The first paragraph of Article 48 of the Guide states that Regulation (EU) 2024/1860 entered into force upon being published in the Official Journal of the EU on 09.07.2024, and lists its three purposes: reducing the risk of high-risk IVD devices becoming unavailable on the market; making EUDAMED modules mandatory gradually, from the last quarter of 2025, immediately after they are declared functional; and an obligation of prior notification for devices whose supply interruption is assessed to pose a risk of serious harm.
This third leg has also been reflected in national legislation. Article 10/A of the Medical Device Regulation was added by the Official Gazette dated 17/8/2024 and numbered 32635. According to subparagraph (a) of the first paragraph, where the manufacturer foresees an interruption or discontinuation in the supply of a device other than a custom-made device, and it is reasonably foreseeable that this could result in serious harm or a risk of serious harm to patients or public health, the manufacturer informs the Agency, and also the economic operators, health institutions and health professionals to which it directly supplies the device.
Subparagraph (b) sets the timing: except in exceptional circumstances, this information is provided at least six months before the foreseen interruption, and the manufacturer clearly states the reasons for the interruption in the information given to the Agency. The third paragraph states that economic operators who receive the information will inform, without delay, other operators, health institutions and health professionals that they directly supply. If you are in the middle of the chain as an importer, this notification is expected from you as well.
Limit of the Service Scope on the EUDAMED Side
On this page we have explained EUDAMED to the extent that Turkish legislation refers to it. The reason is simple: as the first paragraph of Article 33 of the Medical Device Regulation also states, EUDAMED is a system established and managed by the Commission. The screen flows of the modules, when which module becomes mandatory and the behavior of fields are determined by the European Commission; a consultant cannot change these rules and cannot give a promise that goes beyond the Commission's calendar.
There are two jobs that we can do: carrying out actor and device registration in EUDAMED, and correctly managing the intersection of the two systems: checking the consistency of the company information in the actor registration with the certificate and declaration, entering the SRN correctly in the "Single Identification No" field, uploading the EUDAMED actor screen output to the "Authorization Document" field in accordance with the seventh paragraph of Article 12, carrying a certificate published in EUDAMED to ÜTS through the route in Article 8, comparing EMDN and GMDN codes with the documents, and matching the Basic UDI-DI chain with the declaration and certificate.
The division of responsibility does not change. The company opens user authorization for the consultant from its own ÜTS company page, the consultant creates document and product records with the "Medical Device Registration Officer" authorization; the e-signature remains with the company's authorized signatory. The sixth paragraph of Article 31 of the Medical Device Regulation also reserves the responsibility of economic operators for the data. The party that takes the data from the manufacturer and declares it is you; we undertake the preparation and the registration process.
If you would like to see the scope and fee for EUDAMED actor registration, Basic UDI-DI and device registrations item by item, you can fill in the EUDAMED registration price quote form.
Home page for all service headings: Medical Device ÜTS Registration Consultancy