What Format Can a UDI Carrier Have? Annex VI Part C: AIDC, HRI, Linear and 2D Barcode, Placement Rules and UDI-DI Entry in ÜTS

SUMMARY
Under Part C of Annex VI to the Tıbbi Cihaz Yönetmeliği (Medical Devices Regulation), a UDI carrier is the means of conveying the UDI using AIDC and, if applicable, HRI. The Regulation lists 1D linear barcode, 2D matrix barcode and RFID as examples of carriers and regulates how the carrier is placed on the label and packaging levels.
In this article we give, with verbatim quotations, the definition, general requirement and carrier provisions of Part C of Annex VI, the dates in Art. 27 and Art. 111, and the UDI-DI field in the TİTCK ÜTS Kayıt Kılavuzu (ÜTS Registration Guide) TCKKD-KLVZ-03.
For the ÜTS registration of your medical device: Medical Device ÜTS Registration Consultancy
Contents
1. Definitions: UDI Carrier, AIDC and HRI
Annex VI Part C, point 1. Definitions:
“AIDC is a technology used to automatically capture data. AIDC technologies include bar codes, smart cards, biometrics and RFID.”
“HRI is a legible interpretation of the data characters encoded on the UDI carrier.”
“UDI carrier is the means of conveying the UDI by using AIDC and, if applicable, its HRI.”
“UDI carriers include, among others, 1D / linear bar code, 2D / Matrix bar code and RFID.”
AIDC is automatic data capture technology; HRI is the human-readable interpretation of the data on the carrier. The Regulation lists the carrier types with the words "among others"; it does not make a single barcode type mandatory.
The UDI within the carrier consists of two parts:
“3.3. The UDI shall consist of two parts: a UDI-DI and a UDI-PI.”
The concepts of UDI-DI, UDI-PI and Basic UDI-DI are covered in our article on the UDI system, UDI-DI and Basic UDI-DI.
2. General Requirements and Coding Standard
“2.1. The affixing of the UDI is an additional requirement, it does not replace any other marking or labelling requirements laid down in Annex I to this Regulation.”
“2.2. The manufacturer shall assign and maintain unique UDIs for its devices.”
“2.3. Only the manufacturer may place the UDI on the device or its packaging.”
“2.4. Only coding standards provided by issuing entities designated by the Commission may be used.”
Under point 2.1, the UDI carrier does not replace other information on the label such as the batch number and expiry date. The HRI format also depends on the issuing entity:
“4.8. The HRI format shall follow the rules of the UDI code-issuing entity.”
The issuing entities are covered in our article on UDI issuing entities: GS1, HIBCC, ICCBBA, IFA.
3. Where the Carrier Is Placed: 4.1-4.4
“4.1. The UDI carrier (AIDC and HRI representation of the UDI) shall be placed on the label or on the device itself and on all higher levels of device packaging. Higher levels do not include shipping containers.”
“4.2. In the event of significant space constraints on the unit of use packaging, the UDI carrier may be placed on the next higher packaging level.”
“4.3. For single-use class I and IIa devices packaged and labelled individually, the UDI carrier shall not be required to appear on the packaging but it shall appear on a higher level of packaging, e.g. a carton containing several individually packaged devices. However, when the healthcare provider is not expected to have access to that higher level of device packaging, such as in cases of home healthcare settings, the UDI shall be placed on the packaging of the individual device.”
“4.4. For devices exclusively intended for retail point of sale the UDI-PIs in AIDC shall not be required to appear on the point of sale packaging.”
Art. 27(4) gives the same rule in the body of the Regulation:
“(4) UDI carriers shall be placed on the label of the device and on all higher levels of packaging. Higher levels of packaging shall not include shipping containers.”
4. Barcode Format: Linear, 2D and RFID
“4.5. When AIDC carriers other than the UDI carrier are part of the product labelling, the UDI carrier shall be easily identifiable.”
“4.6. If linear bar codes are used, the UDI-DI and UDI-PI may be concatenated or non-concatenated in two or more bar codes. All parts and elements of the linear bar code shall be distinguishable and identifiable.”
“4.7. If there are significant constraints limiting the use of both AIDC and HRI on the label, only the AIDC format shall be required to appear on the label. However, for devices intended to be used outside of healthcare facilities, such as devices for home care, the HRI shall appear on the label even if this results in no space for the AIDC.”
“4.9 If the manufacturer uses RFID technology, a linear or 2D bar code in accordance with the standard provided by the issuing entities shall also be provided on the label.”
“4.15. Bar code carriers which contain both a UDI-DI and a UDI-PI may also include essential data for the device to operate or other data.”
According to these provisions:
The label may carry a barcode other than the UDI carrier; the UDI carrier must be easily identifiable (4.5).
In a linear barcode, the UDI-DI and UDI-PI may be combined in one barcode or placed in separate barcodes (4.6).
If RFID is used, the label also carries a linear or 2D barcode (4.9).
For devices used outside healthcare facilities, the HRI must appear on the label (4.7).
5. Readability and Special Cases: 4.10-4.14
“4.11. The UDI carrier shall be readable during normal use and throughout the intended life of the device.”
“4.12. If the UDI carrier is readily readable or, in the case of AIDC, scannable through the device's packaging, the placement of the UDI carrier on the packaging shall not be required.”
“4.13. In the case of single finished devices made up of multiple parts that must be assembled before their first use, it shall be sufficient to place the UDI carrier on only one part of each device.”
“4.14. The UDI carrier shall be placed in a manner such that the AIDC can be accessed during normal operation or storage.”
Point 4.10, for reusable devices:
“4.10. Reusable devices shall bear a UDI carrier on the device itself. The UDI carrier for reusable devices that require cleaning, disinfection, sterilisation or refurbishing between patient uses shall be permanent and readable after each process performed to make the device ready for the subsequent use, throughout the intended life of the device. This requirement shall not apply to devices in the following circumstances:”
“(a) Any type of direct marking would interfere with the safety or performance of the device;”
“(b) The device cannot be directly marked because it is not technologically feasible.”
6. Carrier on Procedure Packs (SİP) and Software
Point 6.3.3, for systems and procedure packs:
“(a) The UDI carrier of the system or procedure pack shall, as a general rule, be affixed on the outside of the packaging.”
“(b) The UDI carrier shall be readable or, in the case of AIDC, scannable, whether it is affixed to the outside of the system or procedure pack packaging or is inside a transparent packaging.”
Point 6.5.4, for software:
“(a) Where the software is delivered on a physical medium, such as a CD or DVD, each packaging level shall bear the human readable and AIDC representation of the complete UDI. The UDI that is applied to the physical medium containing the software and its packaging shall be identical to the UDI assigned to the software at system level,”
“(ç) On the electronic displays of the software, only the human readable part of the UDI is required. UDI marking using AIDC is not required on electronic displays such as the ‘about’ menu, pop-up window and the like,”
For the classification of software, see our article on medical device software and Rule 11.
7. Dates of Application: Art. 111
The entry into force of the fourth paragraph of Art. 27 (placing the carrier on the label and packaging) is in point (c) of Art. 111(1):
“(1) for implantable devices and class III devices, on the date of publication, effective from 26/5/2021,”
“(2) for class IIa and class IIb devices, on 26/5/2023,”
“(3) for class I devices, on 26/5/2025,”
“(4) for reusable devices that bear the UDI carrier on the device itself: for implantable devices and class III devices on 26/5/2023, for class IIa and class IIb devices on 26/5/2025, and for class I devices on 26/5/2027,”
Art. 27 of the ÜTS Kayıt Kılavuzu (ÜTS Registration Guide) lists the carrier, with reference to these dates, among the label information assessed in the ÜTS device registration review:
“h. UDI carrier (the dates of application set out in point (c) of the first paragraph of Article 111 of the MDR shall be taken into account),”
8. Barcode and UDI-DI Entry in ÜTS
Under Art. 27 of the Guide, the field "UDI Tahsis Kuruluşu" (UDI Issuing Entity) is selected first under the heading "Tanımlayıcı Bilgiler" (Identifying Information):
“(3) The UDI issuing entities and standards registered in ÜTS under the MDR and authorised for medical devices are GS1, HIBCC, ICCBBA and IFA.”
The Guide's provisions for the "UDI-DI Numarası" (UDI-DI Number) field:
“(4) A device application made under Article 40 of this Guide for a device registration whose barcode is not entered in ÜTS as it appears in its technical documents (label/sales packaging, EU declaration of conformity, EU certificate, etc.) shall result in a negative outcome.”
“(5) The format of the UDI-DI Number must comply with the barcode assignment rules set by the UDI Issuing Entity to which it is assigned.”
“(6) In principle, device registration in ÜTS is made with the barcode of the smallest packaging placed on the market (unit packaging or single packaging).”
According to the Guide, upper packaging barcodes are added with their content quantities to the "Piyasaya Arz Bilgisi" (Placing on the Market Information) field after registration with the smallest packaging barcode is completed. A summary of the barcode rules is on our page on UDI-DI and barcode rules, and the related rejection reason is on our page on the barcode entered differently from the documents.
9. Frequently Asked Questions
Is a specific barcode type mandatory for the UDI carrier?
Part C of Annex VI lists the carriers as "among others" 1D linear barcode, 2D matrix barcode and RFID. The binding condition is the use of the coding standard of the issuing entities designated by the Commission (2.4).
Can the label carry a barcode other than the UDI carrier?
Under point 4.5 of Part C of Annex VI, when AIDC carriers other than the UDI carrier are part of the labelling, the UDI carrier shall be easily identifiable.
Are the batch number and expiry date written separately when there is a UDI carrier?
Yes. Under point 2.1 of Part C of Annex VI, the affixing of the UDI does not replace the other marking and labelling requirements in Annex I.
Does a manufacturer that uses RFID also add a barcode?
Yes. Under 4.9, if RFID is used, a linear or 2D barcode in accordance with the standard of the issuing entity is also provided on the label.
Which packaging's barcode is entered in ÜTS?
Under Art. 27 of the Guide, registration is made in principle with the barcode of the smallest packaging placed on the market (unit or single packaging), and the barcode is entered as it appears in the technical documents.