Person Responsible for Regulatory Compliance (PRRC): Qualifications, Tasks, Exception and Registration

12 hours ago
6 min read
Mevzuata uyumdan sorumlu kişi PRRC nitelikleri ve görevleri

SUMMARY

The person responsible for regulatory compliance (PRRC) is the person with the necessary expertise in the field of medical devices whom the manufacturer must have within its organisation. The basis is Article 15 of the Medical Device Regulation (Tıbbi Cihaz Yönetmeliği); the same provision is in Article 16 of the IVD Regulation.

In this article we give the expertise requirements of the person, the minimum tasks, the exception for micro and small enterprises, the rule for the authorised representative and where the person appears in the registration, quoting word for word the Regulation text and the TİTCK Turkish EUDAMED user guide.

For your manufacturer and authorised representative registrations: Medical Device ÜTS Registration Consultancy

Contents

1. Who Is the Person Responsible for Regulatory Compliance?

First paragraph of Article 15 of the Medical Device Regulation:

“ARTICLE 15 -(1) Manufacturers shall have within their organisation at least one person responsible for regulatory compliance who possesses the requisite expertise in the field of medical devices. The requisite expertise shall be demonstrated by either of the following qualifications:”

The obligation lies with the manufacturer and at least one person is required. For the full text of the article, see our page PRRC: Article 15 and, for a short definition, our page who is the PRRC.

2. How Is Expertise Demonstrated?

The first paragraph lists three qualifications. According to the expression "either" in the paragraph, one of them is sufficient:

  • “a) A diploma, certificate or other evidence of formal qualification, awarded on completion of a university degree programme or of a course recognised as equivalent by the Council of Higher Education, in law, medicine, dental medicine, pharmacy, engineering or another relevant scientific discipline, together with at least one year of professional experience in regulatory affairs or in quality management systems relating to medical devices.”

  • “b) Four years of professional experience in regulatory affairs or in quality management systems relating to medical devices.”

  • “c) Without prejudice to the legislation on professional qualifications, in the case of manufacturers of custom-made devices, at least two years of professional experience in the relevant field of manufacturing.”

Item (b) has no diploma requirement; the experience must be in regulatory affairs or in quality management systems relating to medical devices. Item (c) applies only to manufacturers of custom-made devices. For custom-made devices, see our article on the custom-made device.

3. Micro and Small Enterprise Exception

“(2) Micro and small enterprises within the scope of the Regulation on the Definition, Characteristics and Classification of Small and Medium-Sized Enterprises, put into force by Council of Ministers Decision dated 19/10/2005 and numbered 2005/9617, are not required to have the person responsible for regulatory compliance within their organisation; however, such enterprises shall have permanent and continuous recourse to such a person in accordance with the procedures and principles determined by the Authority.”

The exception removes the requirement of having the person within the organisation. The obligation to obtain the services of the person continues, and this service is permanent and continuous. The exception is written only for micro and small enterprises.

4. Minimum Tasks

Third paragraph:

“(3) The person responsible for regulatory compliance shall, as a minimum, be responsible for ensuring;”

  • “a) that the conformity of the devices is appropriately checked, in accordance with the quality management system under which the devices are manufactured, before a device is released,”

  • “b) that the technical documentation and the EU declaration of conformity are drawn up and kept up-to-date,”

  • “c) that the post-market surveillance obligations are complied with in accordance with the tenth paragraph of Article 10,”

  • “ç) that the reporting obligations referred to in Articles 85 to 88 and in the measures adopted by the Commission are fulfilled,”

  • “d) that, in the case of investigational devices, the statement referred to in point 4.1 of Section II of Annex XV is issued.”

  • (End of the third paragraph.)

The paragraph makes the person "responsible for ensuring" these tasks. The tasks cover the quality management system, the technical documentation and EU declaration of conformity, post-market surveillance, and the reporting headings in Articles 85-88. For the general obligations of the manufacturer, see our articles on manufacturer obligations, and for reporting, our articles on vigilance and serious incident reporting.

5. More Than One Person and Protection of the Person

“(4) If several persons are jointly responsible for regulatory compliance in accordance with the first, second and third paragraphs, their respective areas of responsibility shall be stipulated in writing.”

“(5) The person responsible for regulatory compliance shall not suffer any disadvantage within the manufacturer's organisation in relation to the proper fulfilment of his/her tasks, regardless of whether or not he/she is an employee of the organisation.”

If there is more than one person, their areas of responsibility are in writing. The protection provision applies even if the person is not an employee of the organisation.

6. Person Responsible for Regulatory Compliance for the Authorised Representative

The sixth paragraph regulates authorised representatives separately:

“(6) Authorised representatives shall have permanent and continuous recourse to at least one person responsible for regulatory compliance who possesses the requisite expertise regarding the regulatory requirements for medical devices. The requisite expertise shall be demonstrated by either of the following qualifications:”

  • “a) A diploma, certificate or other evidence of formal qualification, awarded on completion of a university degree programme or of a course recognised as equivalent by the Council of Higher Education, in law, medicine, dental medicine, pharmacy, engineering or another relevant scientific discipline, together with at least one year of professional experience in regulatory affairs or in quality management systems relating to medical devices,”

  • “b) Four years of professional experience in regulatory affairs or in quality management systems relating to medical devices.”

Two qualifications are listed for the authorised representative; item (c), which concerns manufacturers of custom-made devices, is not in this paragraph. For the tasks and the mandate of the authorised representative, see our article on the authorised representative.

7. The Counterpart in the IVD Regulation

Article 16 of the In Vitro Diagnostic Medical Device Regulation (In Vitro Tanı Amaçlı Tıbbi Cihaz Yönetmeliği) has the same structure:

“ARTICLE 16 -(1) Manufacturers shall have within their organisation at least one person responsible for regulatory compliance who possesses the requisite expertise in the field of in vitro diagnostic devices.”

This article lists two qualifications for the manufacturer (items a and b); it refers to the ninth paragraph of Article 11 for post-market surveillance and to Articles 82-85 for reporting.

8. The Person Responsible for Regulatory Compliance in the Registration

Annex VI Part A of the Medical Device Regulation lists the person among the information to be submitted in the registration of the economic operator:

“1.4. Name, address and contact details of the person or persons responsible for regulatory compliance referred to in Article 15.”

In the TİTCK Turkish EUDAMED user guide, step 4 of the actor registration is the "Düzenleyici kişiler" (Regulatory persons) step:

“The person within your organisation who is responsible for ensuring your organisation's compliance with the Regulations. See Article 15 of Regulation 2017/745.”

“8. Enter the details of this person and specify the area(s) for which he/she is responsible. To add another person with this role, select Add a new Regulatory person (Yeni bir Düzenleyici kişi ekle).”

The TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03 has no separate registration step for the person responsible for regulatory compliance. For the EUDAMED actor registration, see our article on EUDAMED registration, and for the scope of the service, see our page on the PRRC service.

9. Frequently Asked Questions

Is a diploma required to be the person responsible for regulatory compliance?

No. Item (b) of the first paragraph of Article 15 considers four years of professional experience in regulatory affairs or in quality management systems relating to medical devices sufficient on its own.

Are micro and small enterprises exempt from this person?

No. According to the second paragraph, they are not required to have the person within the organisation, but they obtain the services of such a person on a permanent and continuous basis.

Can there be more than one person responsible for regulatory compliance?

Yes. The article says "at least one person". If several persons are jointly responsible, their areas of responsibility are stipulated in writing.

Does the authorised representative also have a person responsible for regulatory compliance?

Yes. According to the sixth paragraph, authorised representatives have permanent and continuous recourse to at least one person with the requisite expertise.

Where is the information of the person entered?

Annex VI Part A 1.4 lists the name, address and contact details of the person among the information to be submitted in the economic operator registration. In the EUDAMED actor registration, this information is entered in the "Düzenleyici kişiler" (Regulatory persons) step.

For your manufacturer and authorised representative registrations: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara