top of page

Medical Devices Subject to More Than One Technical Regulation: Art. 1 References (EMC, Machinery), Art. 19 Single Declaration, Art. 20 CE Marking and ÜTS

12 hours ago
6 min read
Tıbbi cihazda birden fazla teknik düzenleme tek uygunluk beyanı ve CE işareti

SUMMARY

For a medical device that falls under more than one technical regulation, the Medical Device Regulation sets provisions in three places: Article 1 establishes the relationship with some regulations, Article 19 requires a single EU declaration of conformity, and Article 20 states that the CE marking also indicates the other legislation.

In this article we give only the Medical Device Regulation's own references and the relevant provision in the ÜTS Registration Guidance TCKKD-KLVZ-03, with verbatim quotations (unofficial translation). The scope provisions of the other regulations are not the subject of this article.

​To register your device in ÜTS (documents and product): Medical Device ÜTS Registration Consultancy

Contents

  • 1. Art. 1(9): Electromagnetic Compatibility Regulation

  • 2. Art. 1(10): Machinery Safety Regulation

  • 3. Art. 1(5), 1(11): IVD Parts and Radiation Legislation

  • 4. Medical and Non-Medical Intended Purpose Together

  • 5. Art. 19: Single EU Declaration of Conformity

  • 6. Art. 20: The CE Marking Also Indicates Other Legislation

  • 7. More Than One CE Marking in the ÜTS Guidance

  • 8. Frequently Asked Questions

1. Art. 1(9): Electromagnetic Compatibility Regulation

“(9) The provisions of the Electromagnetic Compatibility Regulation (2014/30/EU), published in the Official Gazette dated 2/10/2016 and numbered 29845, are not sought for devices subject to conformity assessment within the scope of this Regulation.”

The provisions of the Electromagnetic Compatibility Regulation are not sought for a device subject to conformity assessment within the scope of the Medical Device Regulation. The electromagnetic requirements are in Annex I of the Medical Device Regulation:

  • “18.5. Devices shall be designed and manufactured in such a way as to reduce as far as possible the risks of creating electromagnetic interference which could impair the operation of the device in question or other devices or equipment in the intended environment.”

  • “18.6. Devices shall be designed and manufactured in such a way as to provide a level of intrinsic immunity to electromagnetic interference that is adequate to enable them to operate as intended.”

Under item 6.1 of Annex II, the summaries of data and test results in the technical documentation also include the following heading:

“- Electrical safety and electromagnetic compatibility,”

2. Art. 1(10): Machinery Safety Regulation

“(10) Devices that are subject to this Regulation and that are also defined as machinery within the framework of subparagraph (ğ) of the first paragraph of Article 4 of the Machinery Safety Regulation (2006/42/EC), published in the Official Gazette dated 3/3/2009 and numbered 27158, shall, where a relevant hazard exists under that Regulation, also meet the essential health and safety requirements set out in Annex I of the Machinery Safety Regulation (2006/42/EC) to the extent to which those requirements are more specific than the general safety and performance requirements set out in Chapter II of Annex I to this Regulation.”

  • Condition 1: the device is defined as machinery under item 4/1-(ğ) of the Machinery Safety Regulation.

  • Condition 2: the relevant hazards exist under that regulation.

  • Result: the essential health and safety requirements of Annex I of the Machinery Safety Regulation are met to the extent they are more specific than the requirements in Chapter II of Annex I of the Medical Device Regulation.

3. Art. 1(5), 1(11): IVD Parts and Radiation Legislation

“(5) A device that incorporates, as an integral part, an in vitro diagnostic device defined in subparagraph (ü) of Article 3 of the In Vitro Diagnostic Medical Device Regulation, when placed on the market or put into service, is subject to the provisions of this Regulation. The provisions of the In Vitro Diagnostic Medical Device Regulation apply to the in vitro diagnostic device part of the device.”

“(11) This Regulation does not affect the application of the legislation on radiation safety and medical irradiation.”

​For the sixth and seventh paragraphs of Article 1, which regulate drug-device combinations, see our article on borderline products.

4. Medical and Non-Medical Intended Purpose Together

Sixth paragraph of Article 9:

“(6) Where a device has both a medical and a non-medical intended purpose, these devices shall cumulatively meet the requirements applicable to devices with a medical intended purpose and the requirements applicable to devices without a medical intended purpose.”

​For groups of products without a medical intended purpose, see our article on Annex XVI products.

5. Art. 19: Single EU Declaration of Conformity

“(2) Where, with regard to matters not covered by this Regulation, devices are also subject to other legislation requiring an EU declaration of conformity, a single EU declaration of conformity shall be drawn up in respect of all legislation applicable to the device. The declaration shall contain all the information required to identify the legislation to which the declaration relates.”

“(3) By drawing up the EU declaration of conformity, the manufacturer assumes responsibility for compliance with the requirements of this Regulation and all other legislation applicable to the device.”

Item 6 of Annex IV lists a statement to this effect among the contents of the declaration:

“6. A statement that the device that is covered by the present declaration is in conformity with this Regulation and, if applicable, with any other relevant Union legislation that provides for the issuing of an EU declaration of conformity,”

  • If other legislation also requires an EU declaration of conformity, a single declaration is drawn up (19/2).

  • The declaration identifies all the legislation to which it relates (19/2).

  • By the declaration, the manufacturer also assumes responsibility for compliance with all other legislation (19/3).

​For the full content of the declaration, see our article on the EU declaration of conformity and CE marking.

6. Art. 20: The CE Marking Also Indicates Other Legislation

“(6) Where devices are subject to other legislation which provides for the affixing of the CE marking, the CE marking shall indicate that the devices also fulfil the requirements of that other legislation.”

“(5) Where there is a notified body responsible for the conformity assessment procedures referred to in Article 52, the CE marking shall be followed by the identification number of the notified body.”

7. More Than One CE Marking in the ÜTS Guidance

Article 17 of guidance TCKKD-KLVZ-03 lists the CE marking among the information sought in a "Instructions for Use" type document:

“d) The CE marking together with the four-digit notified body number of the product,”

  • “1) For Class-I Other, Class A and IVD Other products, only the CE marking is sufficient.”

  • “3) In addition to the medical device regulations, the instructions for use and label of products that are also subject to the conformity assessment procedure of other main regulations (for example, the Personal Protective Equipment Regulation, the Machinery Safety Regulation) may contain more than one CE marking and notified body information.”

​According to the guidance, the instructions for use and label of a product that is also subject to the conformity assessment of another main regulation may contain more than one CE marking and notified body information. For uploading the instructions for use document, see ÜTS document registration; for product registration, see ÜTS medical device product registration.

8. Frequently Asked Questions

Is the Electromagnetic Compatibility Regulation sought separately for a medical device?

Under the ninth paragraph of Article 1 of the Medical Device Regulation, the provisions of the Electromagnetic Compatibility Regulation are not sought for devices subject to conformity assessment within the scope of this regulation. The electromagnetic requirements are in Annex I, 18.5 and 18.6.

Which requirements does a medical device that falls within the definition of machinery meet?

Under the tenth paragraph of Article 1, if the relevant hazards exist, it meets the essential health and safety requirements of Annex I of the Machinery Safety Regulation to the extent they are more specific than those in Chapter II of Annex I of the Medical Device Regulation.

How many declarations of conformity are drawn up for a device subject to more than one piece of legislation?

Under the second paragraph of Article 19, a single EU declaration of conformity is drawn up for all legislation applicable to the device.

Can there be more than one CE marking on the label?

Under TCKKD-KLVZ-03 Article 17, the instructions for use and label of products that are also subject to the conformity assessment procedure of other main regulations may contain more than one CE marking and notified body information.

​To register your device in ÜTS (documents and product): Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara