Medical Gas System Classification: Annex VIII Rules 2, 3, 5, 12 and 20, Active Device Definition, Annex I Gas Requirements and ÜTS Class Structure

SUMMARY
The classification rules for medical gas systems are in Annex VIII of the Medical Device Regulation. Annex VIII mentions gases directly in the texts of Rule 2 (channelling or storage for administering gases to the body) and Rule 3 (gas exchange); active devices that administer substances to the body are subject to Rule 12, and devices that administer medicinal products by inhalation are subject to Rule 20.
In this article we give, with verbatim quotations (unofficial translation) from the Regulation and the TİTCK guidance, the scope provisions, the classification implementing rules, the definition of active device, the relevant classification rules, the Annex I requirements relating to gases, and the class structure in ÜTS. The class of a given product is determined by applying these rules according to the intended purpose set by the manufacturer.
For the declaration of conformity and ÜTS registration of your classified devices: Medical Device ÜTS Registration Consultancy
Contents
1. Scope: Medicinal Product and the Device That Administers It
2. Implementing Rules: Intended Purpose, Separate Classification, Strictest Rule
3. Definition of Active Device
4. Rule 1 and Rule 2: Non-Invasive Devices Channelling Gases
5. Rule 3: Gas Exchange
6. Rule 5: Body Orifice Devices Connected to an Active Device
7. Rule 12 and Rule 20: Active Devices Administering Substances and Inhalation
8. Annex I Requirements Relating to Gases
9. Class Structure in ÜTS and the Sales Center Provision
10. Frequently Asked Questions
1. Scope: Medicinal Product and the Device That Administers It
According to the fourth paragraph of Article 1 of the Medical Device Regulation, the Regulation does not apply to:
“b) (Amended: OG-29/7/2022-31907) Medicinal products as defined in subparagraph (b) of the first paragraph of Article 4 of the Human Medicinal Products Licensing Regulation published in the Official Gazette dated 11/12/2021 and numbered 31686,”
A device intended to administer a medicinal product, however, is subject to this Regulation. The seventh paragraph of Article 1:
“(7) Without prejudice to the provisions of the Human Medicinal Products Licensing Regulation relating to medicinal products, a device intended to administer a medicinal product as defined in subparagraph (b) of the first paragraph of Article 4 of the same Regulation (Amended expression: OG-29/7/2022-31907) is subject to this Regulation.”
In deciding which regulation a product is subject to, the principal mode of action is considered under the sixth paragraph:
“(6) In deciding whether a product falls within the scope of the Human Medicinal Products Licensing Regulation or this Regulation, the principal mode of action of the product is taken into account in particular.”
For borderline products, see our article on borderline products: device or medicine.
2. Implementing Rules: Intended Purpose, Separate Classification, Strictest Rule
“3.1. The application of the classification rules is based on the intended purpose of the devices.”
“3.2. If the device is intended to be used in combination with another device, the classification rules apply separately to each of the devices. Accessories of a medical device are classified in their own right, separately from the device with which they are used.”
“3.5. If several rules, or several sub-rules within the same rule, apply to the same device based on the intended purpose of the device, the strictest rule and sub-rule resulting in the higher classification applies.”
The class is determined by the intended purpose.
Each of the devices used in combination is classified separately; accessories are also classified in their own right.
If more than one rule applies, the rule giving the highest class is decisive.
For the classification of accessories, see our article on medical device accessories.
3. Definition of Active Device
Rules 9-13 apply to active devices. The definition is in Article 3:
“1) A device whose operation depends on a source of energy other than that generated by the human body or by gravity, and which acts by changing the density of this energy or by converting it (Devices intended to transmit energy, substances or other elements between an active device and the patient without any significant change are not considered active devices.),”
According to the sentence in parentheses, a device that transmits a substance between an active device and the patient without significant change is not considered an active device.
4. Rule 1 and Rule 2: Non-Invasive Devices Channelling Gases
“All non-invasive devices are classified as class I unless one of the rules set out below applies.”
“All non-invasive devices intended for channelling or storing blood, body fluids, cells or tissues, liquids or gases for the purpose of eventual infusion, administration or introduction into the body:”
“- if they may be connected to an active device in class IIa, class IIb or class III, they are classified as class IIa,”
“In all other cases, such devices are classified as class I.”
Rule 1: a non-invasive device that falls under no other rule is Class I.
Rule 2: a non-invasive device intended for channelling or storing gases for administration to the body is Class IIa if it may be connected to an active device in Class IIa, IIb or III.
Rule 2, other cases: Class I.
5. Rule 3: Gas Exchange
“All non-invasive devices intended for modifying the biological or chemical composition of human tissues or cells, blood, other body fluids or other liquids intended for implantation or administration into the body are classified as class IIb. However, if the treatment consists of filtration, centrifugation or exchanges of gas or heat, they are classified as class IIa.”
6. Rule 5: Body Orifice Devices Connected to an Active Device
“All invasive devices with respect to body orifices, other than surgically invasive devices, intended to be connected to an active device in class IIa, class IIb or class III, are classified as class IIa.”
Body orifice under the special definitions of Annex VIII:
“2.1. Body orifice: Any natural opening in the body, as well as the external surface of the eyeball, or any permanent artificial opening, such as a stoma.”
7. Rule 12 and Rule 20: Active Devices Administering Substances and Inhalation
“All active devices intended to administer and/or remove medicinal products, body liquids or other substances to or from the body are classified as class IIa. However, if this is done in a manner that is potentially hazardous, taking account of the nature of the substances involved, the part of the body concerned and the mode of application, they are classified as class IIb.”
“All invasive devices with respect to body orifices, other than surgically invasive devices, intended to administer medicinal products by inhalation are classified as class IIa. However, if their mode of action has an essential impact on the efficacy and safety of the administered medicinal product or if they are intended to treat life-threatening conditions, they are classified as class IIb.”
Rule 12: an active device administering substances is Class IIa; Class IIb if potentially hazardous.
Rule 20: an invasive device with respect to a body orifice administering a medicinal product by inhalation is Class IIa; Class IIb under the conditions in the Rule.
Other active devices are Class I under Rule 13.
“All other active devices are classified as class I.”
For the general structure of the classes, see our article on MDR risk classes and our page on medical device classification rules.
8. Annex I Requirements Relating to Gases
Irrespective of class, the general safety and performance requirements of Annex I also mention gases:
“10.3. Devices shall be designed and manufactured in such a way that they can be used safely with the materials and substances, including gases, with which they come into contact during their intended use.”
“c) the risks associated with the use of the device, to which exposure occurs under normal conditions of use, when it comes into contact with materials, liquids and substances, including gases,”
“20.4. Terminals and connectors to electricity, gas or hydraulic and pneumatic energy sources which the user or other persons have to handle shall be designed and constructed in such a way as to minimize all possible risks.”
Subparagraph (c) is in the list of risks in Annex I 14.2.
9. Class Structure in ÜTS and the Sales Center Provision
According to Article 56 of the ÜTS registration guidance TCKKD-KLVZ-03, devices within the scope of the MDR are divided into four classes, Class-I, Class-IIa, Class-IIb and Class-III, according to the risks they carry by virtue of their intended purpose and structure. The same paragraph explains the additional distinction in ÜTS:
“(In ÜTS, additional classifications have been made according to characteristic product properties, such as whether the device is packaged sterile or has a measuring function. For example, a device that is "Class-Ir" in ÜTS and a device that is "Class-I Other" are both classified as Class I under the MDR.)”
“(2) Devices within the scope of the MDR are classified according to a total of 22 rules under the categories of "Non-invasive Devices", "Invasive Devices", "Active Devices" and "Special Rules".”
Article 24 of the Medical Device Sales, Advertising and Promotion Regulation lists oxygen cylinders among the auxiliary materials that may be provided free of charge:
“(9) Auxiliary and complementary materials necessary for the use of devices, such as carriers, batteries and oxygen cylinders, may be provided free of charge by sales centers.”
For placing different devices on the market together, see our article on kit, set and system and procedure pack (SPP) registration; for registration, see our page on ÜTS product registration service.
10. Frequently Asked Questions
Which class is a non-invasive device that channels gas?
Under Annex VIII Rule 2, it is Class IIa if it may be connected to an active device in Class IIa, IIb or III, and Class I in other cases.
Which rule covers devices that perform gas exchange?
Under Rule 3, non-invasive devices that modify the composition of blood or body fluids are Class IIb; if the process consists only of gas or heat exchange, they are Class IIa.
Which class is an active device that administers a substance to the body?
Under Rule 12 it is Class IIa; it is Class IIb if it is potentially hazardous, taking account of the nature of the substance, the part of the body concerned and the mode of application.
Are devices used in combination assessed as a single class?
No. Under Annex VIII 3.2, the classification rules apply separately to each device; accessories are also classified in their own right.
Are ÜTS "Class-Ir" and MDR Class I the same?
According to Article 56/1 of TCKKD-KLVZ-03, a device that is "Class-Ir" in ÜTS is also classified as Class-I under the MDR; ÜTS makes additional distinctions.
For the declaration of conformity and ÜTS registration of your classified devices: Medical Device ÜTS Registration Consultancy