Medical Device Storage Conditions: Importer and Distributor Obligations, Sales Center Preservation and ÜTS Record Entry

12 hours ago
7 min read
tıbbi cihaz depolama ve nakil koşulları, ithalatçı ve dağıtıcı sorumluluğu

SUMMARY

The storage conditions of a medical device are determined by the manufacturer. Under Articles 13 and 14 of the Medical Device Regulation, the importer and the distributor keep the storage and transport conditions in line with those conditions while the device is under their responsibility.

In this article we give, with verbatim quotations from the regulations and TİTCK guidance (unofficial translation), the importer and distributor provisions, the transport and storage requirements in Annex I, the sales center's obligation to preserve devices, how the conditions are entered in the ÜTS product record, and the ÜTS notification for a product whose conditions have been compromised.

​For ÜTS product registration and individual notifications of your devices: Medical Device ÜTS Registration Consultancy

Contents

  • 1. The Manufacturer Determines the Conditions: Annex I 11.3 and 11.4

  • 2. Conditions on the Label and in the Instructions for Use

  • 3. The Importer's Obligation: Article 13/5

  • 4. The Distributor's Obligation: Article 14/3

  • 5. The Same Provision in In Vitro Diagnostic Devices

  • 6. The Sales Center's Preservation and Storage Requirement

  • 7. Storage/Use Condition Information in the ÜTS Product Record

  • 8. ÜTS Notification for a Product Whose Conditions Have Been Compromised

  • 9. Frequently Asked Questions

1. The Manufacturer Determines the Conditions: Annex I 11.3 and 11.4

General safety and performance requirements in Annex I of the Medical Device Regulation (unofficial translation):

  • “11.3. Devices labelled as having a specific microbial state shall be designed, manufactured and packaged to ensure that they remain in that state when placed on the market and under the transport and storage conditions specified by the manufacturer.”

  • “11.4. Devices delivered in a sterile state shall be designed, manufactured and packaged in accordance with appropriate procedures, to ensure that they are sterile when placed on the market and that, unless the packaging intended to maintain their sterile condition is damaged, they remain sterile, under the transport and storage conditions specified by the manufacturer, until that packaging is opened at the point of use. It shall be ensured that the integrity of this packaging is clearly evident to the final user.”

In both items, the criterion is "the transport and storage conditions specified by the manufacturer". Item 10.2 of Annex I also covers persons involved in transport and storage:

“10.2. Devices shall be designed, manufactured and packaged in such a way as to minimize the risk posed by contaminants and residues to patients and to persons involved in the transport, storage and use of the devices, taking account of the intended purpose of the devices.”

2. Conditions on the Label and in the Instructions for Use

Under item 23.2 of Annex I, the information on the label includes the following indication:

“ı) An indication of any special storage and/or handling conditions,”

Under Part B of Annex VI, the storage conditions are also among the key data elements the manufacturer enters in the UDI database together with the UDI-DI:

“14. Where applicable, storage and/or handling conditions (as indicated on the label or in the instructions for use),”

​For the label and instructions for use, see our article on Turkish label and instructions for use.

3. The Importer's Obligation: Article 13/5

“(5) Importers shall ensure that, while a device is under their responsibility, storage or transport conditions do not jeopardize its compliance with the general safety and performance requirements set out in Annex I and, where applicable, shall comply with the conditions set by the manufacturer.”

  • The conditions must not jeopardize the device's compliance with the Annex I requirements.

  • Where the manufacturer has set conditions, they must be complied with.

  • The obligation applies for as long as the device is under the importer's responsibility.

The importer under Article 4 of the Regulation:

“y) Importer: A natural or legal person established in Türkiye who places a device from a third country on the market in Türkiye,”

4. The Distributor's Obligation: Article 14/3

“(3) Distributors shall ensure that, while a device is under their responsibility, storage or transport conditions comply with the conditions set by the manufacturer.”

Definition of distributor:

“j) Distributor: A natural or legal person in the supply chain, other than the manufacturer or the importer, who makes a device available on the market up to the point of putting into service,”

​Under Article 14/2-a, before making a device available on the market, the distributor verifies that the information provided by the manufacturer accompanies the device. For the other obligations of importers and distributors, see our article on importer and distributor obligations.

5. The Same Provision in In Vitro Diagnostic Devices

In the In Vitro Diagnostic Medical Device Regulation, Article 14/5 for importers and Article 15/3 for distributors are identical to the text in the Medical Device Regulation. Item 11.2 of Annex I to the IVD Regulation:

“11.2. Devices labelled as sterile or as having a specific microbial state shall be designed, manufactured and packaged to ensure that, unless the packaging intended to maintain their sterile or microbial state is damaged, they remain sterile, under the transport and storage conditions specified by the manufacturer, until that packaging is opened at the point of use.”

6. The Sales Center's Preservation and Storage Requirement

The Regulation on Medical Device Sales, Advertising and Promotion regulates the preservation requirement in three articles:

  • “(1) Devices shall be preserved under the conditions set by their manufacturer and required by international standards.”

  • “(4) Sales centers shall establish the infrastructure necessary to preserve devices under the conditions required by the manufacturer and international standards, and carry out their activities accordingly.”

  • “(1) The sales center shall provide the conditions set by the manufacturer and required by international standards regarding the area, vehicles, equipment, transport, lighting and air conditioning for the sale, preservation, storage and distribution of the devices it sells.”

These provisions are in Articles 8/1, 9/4 and 14/1 respectively. Row 2 of the Annex-2 inspection form is the question "Are the devices preserved under the conditions set by their manufacturer and required by international standards?" The sanction applied is a warning; if product safety is considered to be affected, the Agency is notified. If the requirement is not met at a second consecutive inspection, the sales activity is temporarily suspended for 15 days.

​For physical conditions such as the work area and fitting room, see our articles on sales center physical conditions, and for inspection, sales center inspection.

7. Storage/Use Condition Information in the ÜTS Product Record

Under Article 27 of guidance TCKKD-KLVZ-03, the record of a device within the scope of the MDR contains a "Storage/Use Condition Information" heading. The field "Is there a storage and/or use condition?" is mandatory; if "Yes" is selected, the conditions are entered on the "Add Storage/Use Condition" screen. The condition types include:

“'Storage/Warehouse Environment Temperature', 'Storage/Warehouse Environment Humidity', 'Storage/Warehouse Environment Altitude', 'Storage/Warehouse Environment Pressure'”

Relationship of the entered condition to the label and instructions for use:

“3) The storage/warehouse condition entered in this field, whether or not it has specific values, must be in the instructions for use of the device. If the storage/warehouse condition has specific values, it must be on the label of the device and/or the sales packaging. Conversely, if the storage/warehouse condition has no specific values, there is no requirement for the condition to be on the label and/or sales packaging.”

Information on transport is entered in the "Special Storage/Use Condition" field:

“Additional information related to the relevant storage or use condition is entered in this field. (For example, if the condition relates to the transport of the device or is a long-term storage condition, this information is entered in this field).”

​For product registration, see our page on the ÜTS product registration service.

8. ÜTS Notification for a Product Whose Conditions Have Been Compromised

Article 21 of the ÜTS Individual Movement Notifications Guidance:

“(1) Where products that would endanger patient and user health when used are identified, a "Destruction/Disposal Notification" is made to remove these individual products from the system.”

One of the Destruction/Disposal reason options in the third paragraph of the same article is:

“Transport and storage conditions have been compromised”

​The fields Quantity Destroyed/Disposed, Destruction/Disposal Reason and Destruction/Disposal Document Number are mandatory in the notification. If the importer returns an imported product, the Return to Origin Notification in Article 24 of the guidance is used. For notifications, see our article on ÜTS individual movement notifications.

9. Frequently Asked Questions

Who determines the storage conditions of a medical device?

Items 11.3 and 11.4 of Annex I are based on "the transport and storage conditions specified by the manufacturer". The importer and the distributor comply with these conditions under Articles 13/5 and 14/3.

What is the importer's obligation if the manufacturer has not specified conditions?

Under Article 13/5, the importer ensures that storage or transport conditions do not jeopardize the device's compliance with the Annex I requirements; it complies with the manufacturer's conditions "where applicable".

Are preservation conditions inspected at the sales center?

Yes. Row 2 of the Annex-2 inspection form asks about preservation conditions; a warning is applied at the first finding, and a 15-day temporary suspension of sales activity at a second consecutive inspection.

Is a storage condition entered in the ÜTS record?

Under TCKKD-KLVZ-03 Article 27, the field "Is there a storage and/or use condition?" is mandatory and, if "Yes" is selected, the conditions are added to the system. The entered condition must be in the instructions for use.

How is a product whose storage condition has been compromised notified in ÜTS?

Under Article 21 of the Individual Movement Notifications Guidance, a Destruction/Disposal Notification is made, and "Transport and storage conditions have been compromised" can be selected as the reason.

​For ÜTS product registration and individual notifications of your devices: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara