MDR Risk Class: Class I, IIa, IIb and III and ÜTS Registration

SUMMARY
The MDR risk class is the determination of a medical device as Class I, IIa, IIb or III according to the risk arising from its intended purpose and design. Classification is carried out under the 22 rules in Annex VIII of the Medical Device Regulation; the class determines the conformity assessment route and the documents to be linked to the device in ÜTS.
In this article we give the basis of the classes, the rule groups of Annex VIII, the conformity assessment route by class and the selection of the risk class in ÜTS, with verbatim quotations from the Medical Device Regulation and TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03. The details of each class are in separate articles.
For ÜTS registration of your devices: Medical Device ÜTS Registration Consultancy
Medical Device Classification Wizard
The wizard below applies the rules of Annex VIII of the Medical Device Regulation through questions about the intended purpose. It shows the estimated class, the applied rule with a verbatim quotation from the regulation text, whether a notified body is required, and the basic documents to be linked to the device in ÜTS. The result is not binding; the exact class is determined according to the intended purpose determined by the manufacturer and the technical file.
Contents
1. What Is the MDR Risk Class?
“(1) Devices are classified as class I, IIa, IIb or III, taking into account the risks they carry by virtue of their intended purpose and design. Classification is carried out in accordance with Annex VIII.”
This provision is in Article 51 of the Medical Device Regulation. Article 56 of the TİTCK ÜTS Registration Guideline also states that devices under the MDR are divided into four classes and are classified under a total of 22 rules in the categories "İnvaziv Olmayan Cihazlar" (Non-Invasive Devices), "İnvaziv Cihazlar" (Invasive Devices), "Aktif Cihazlar" (Active Devices) and "Özel Kurallar" (Special Rules).
The guideline explains in the same article that additional classification is made in ÜTS according to characteristics such as sterile packaging or measuring function, for example that "Sınıf-Ir" and "Sınıf-I Diğer" devices are Class I under the MDR. For the Class I subgroups, see our article on Class I medical devices.
2. Application Rules in Classification
Chapter II of Annex VIII determines how the rules are applied:
“3.1. The application of the classification rules shall be governed by the intended purpose of the devices.”
“3.2. If the device in question is intended to be used in combination with another device, the classification rules shall apply separately to each of the devices. Accessories for a medical device shall be classified in their own right separately from the device with which they are used.”
“3.3. Software, which drives a device or influences the use of a device, shall fall within the same class as the device.”
“3.5. If several rules, or if, within the same rule, several sub-rules, are applicable to the same device based on the intended purpose of the device, the strictest rule and sub-rule resulting in the higher classification shall apply.”
The durations of use mentioned in the rules are defined in Chapter I of Annex VIII:
“1.1. Transient: Normally intended for continuous use for less than 60 minutes.”
“1.2. Short term: Normally intended for continuous use for between 60 minutes and 30 days.”
“1.3. Long term: Normally intended for continuous use for more than 30 days.”
For examples of how the rules are applied, see our page on medical device classification rules.
3. Non-Invasive Devices: Rules 1-4
The starting rule is Rule 1:
“All non-invasive devices are classified as class I, unless one of the rules set out below applies.”
Rule 2 covers devices intended for channelling or storing blood, body liquids, tissues or gases; Rule 3 covers devices that modify their biological or chemical composition; Rule 4 covers devices that come into contact with injured skin or mucous membrane. For example, Rule 2 contains the following sentence for blood bags:
“Blood bags are classified as class IIb.”
4. Invasive Devices: Rules 5-8
Rule 5 covers invasive devices in relation to body orifices, Rule 6 surgically invasive devices intended for transient use, Rule 7 those intended for short-term use, and Rule 8 implantable devices and long-term surgically invasive devices. The first indents of Rule 5 give the class according to duration of use:
“- If intended for transient use, it is classified as class I,”
The opening sentences of Rule 6 and Rule 7:
“All surgically invasive devices intended for transient use are classified as class IIa.”
“All surgically invasive devices intended for short-term use are classified as class IIa.”
Rule 6 has a separate indent for reusable surgical instruments:
“- Reusable surgical instruments are classified as class I,”
The opening sentence of Rule 8:
“All implantable devices and long-term surgically invasive devices are classified as class IIb.”
The exception indents in Rules 6, 7 and 8 raise the class in cases such as direct contact with the heart, the central circulatory system or the central nervous system, biological effect, being absorbed, and administration of medicinal products. The full text of these indents is in our article on Class III medical devices.
5. Active Devices: Rules 9-13
Rule 9 covers active therapeutic devices, Rule 10 active devices intended for diagnosis and monitoring, Rule 11 software, and Rule 12 active devices intended to administer or remove substances. The first paragraph of Rule 11:
“Software intended to provide information which is used to take decisions with diagnosis or therapeutic purposes is classified as class IIa, except if such decisions have an impact that may cause:”
“- death or an irreversible deterioration of a person's state of health, in which case it is in class III;”
“- a serious deterioration of a person's state of health or a surgical intervention, in which case it is classified as class IIb.”
Rule 13 collects the remaining active devices:
“All other active devices are classified as class I.”
6. Special Rules: Rules 14-22
The special rules assign a class according to the content or function of the device: Rule 14 devices incorporating a medicinal substance, Rule 15 contraception and prevention of sexually transmitted diseases, Rule 16 disinfecting and sterilising devices, Rule 17 recording of diagnostic images generated by X-ray radiation, Rule 18 devices utilising tissues or cells of human or animal origin, Rule 19 nanomaterials, Rule 20 invasive devices in relation to body orifices intended to administer medicinal products by inhalation, Rule 21 devices composed of substances that are absorbed, and Rule 22 active therapeutic devices with an integrated diagnostic function. As an example, Rule 15:
“All devices used for contraception or the prevention of the transmission of sexually transmitted diseases are classified as class IIb, unless they are implantable or long term invasive devices, in which case they are classified as class III.”
For the rules resulting in Class IIa and IIb, see our article on Class IIa and IIb medical devices.
7. Conformity Assessment Route by Class
Article 52 of the Medical Device Regulation sets the route separately for each class. Class III:
“(3) Manufacturers of class III devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Annex IX. Alternatively, the manufacturer may choose to apply a conformity assessment as specified in Annex X coupled with a conformity assessment as specified in Annex XI.”
Class IIb (fourth paragraph, point a):
“a) Manufacturers of class IIb devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Chapters I and III of Annex IX, including an assessment of the technical documentation as specified in Section 4 of that Annex of at least one representative device per generic device group.”
Class IIa (fifth paragraph, point a):
“a) a conformity assessment as specified in Chapters I and III of Annex IX, including an assessment of the technical documentation as specified in Section 4 of that Annex of at least one representative device for each category of devices,”
Class I (sixth paragraph):
“(6) Manufacturers of class I devices, other than custom-made or investigational devices, shall declare the conformity of their products by issuing the EU declaration of conformity referred to in Article 19 after drawing up the technical documentation set out in Annexes II and III.”
Under the same paragraph, for class I devices that are sterile, have a measuring function or are reusable surgical instruments, the manufacturer applies the procedures in Chapters I and III of Annex IX or in Part A of Annex XI; the involvement of the notified body is limited to the relevant aspects.
8. Risk Class Selection in ÜTS
According to Article 27 of the guideline, for the registration of a device under the MDR the "Sınıf" (Class) field under "Sınıflandırma Bilgileri" (Classification Information) is mandatory, and the risk class of the device under the MDR is selected from this field. The options are:
"Sınıf-I Diğer (steril ve ölçme fonksiyonu olmayan)" (Class I Other, neither sterile nor with a measuring function)
"Sınıf-Ir (Yeniden kullanılabilir cerrahi aletler)" (Class Ir, reusable surgical instruments)
"Sınıf-Im (Ölçme fonksiyonu olan)" (Class Im, with a measuring function)
"Sınıf-Is (Steril fonksiyonu olan)" (Class Is, with a sterile function)
"Sınıf-IIa", "Sınıf-IIb" and "Sınıf-III" (Class IIa, Class IIb and Class III)
The selected class determines under which business rule the documents to be linked to the device are required. Article 66 of the guideline:
“(3) For devices to be registered in ÜTS under the MDR as "Sınıf-I Diğer", "Sınıf-Is", "Sınıf-Im" or "Sınıf-Ir", the appropriate one of the ÜTS document business rules shown in Annex-88 is applied.”
“(4) For devices to be registered in ÜTS under the MDR as "Sınıf-IIa", the appropriate one of the ÜTS document business rules shown in Annex-89 is applied.”
“(8) For devices to be registered in ÜTS under the MDR as "Sınıf-III", the appropriate one of the ÜTS document business rules shown in Annex-93 is applied.”
Annex-90 applies to Class IIb, Annex-91 to implantable IIb devices, and Annex-92 to IIb devices within the scope of Rule 12 (Article 66, fifth to seventh paragraphs). For the registration steps, see our articles on ÜTS product registration and ÜTS document registration.
9. Document Rules That Vary by Class
The obligation to upload the instructions for use varies by class (Guideline, Article 17):
“1) For Class III and Class IIb devices, the instructions for use issued by the manufacturer,”
“2) For Class IIa devices, the instructions for use if the manufacturer has issued instructions for use;”
Uploading instructions for use is optional for Sınıf-I Diğer, Sınıf-Im, Sınıf-Is and Sınıf-Ir devices; the Authority may request it when it deems necessary (Article 17, third paragraph). The notified body information in the EU declaration of conformity also depends on the class (Article 11):
“ı) For EU declarations of conformity other than those covering only Sınıf-I Diğer or Class A devices; the name and identification number of the notified body, a description of the conformity assessment procedure carried out, and the identification of the EU certificate or certificates issued (certificate number or numbers),”
10. Classification Dispute
For a disagreement between the notified body and the manufacturer arising from the application of Annex VIII, the second paragraph of Article 51 provides the following:
“a) Disputes between the notified bodies concerned and a manufacturer established in Türkiye, arising from the application of Annex VIII, shall be referred to the Authority for a decision.”
11. Frequently Asked Questions
How many risk classes are there under the MDR?
Four: Class I, IIa, IIb and III. In ÜTS, Class I is further selected as Sınıf-I Diğer, Sınıf-Is, Sınıf-Im or Sınıf-Ir; all of these are Class I under the MDR.
Which rule applies if more than one rule fits a device?
Under Section 3.5 of Annex VIII, the strictest rule and sub-rule resulting in the higher classification apply.
Is an accessory in the same class as the main device?
No. Under Section 3.2 of Annex VIII, accessories are classified in their own right, separately from the device with which they are used.
For which classes is uploading instructions for use to ÜTS mandatory?
Under Article 17 of the guideline it is mandatory for Class III and IIb; for Class IIa it is uploaded if the manufacturer has issued instructions for use. It is optional for the Class I groups.
Does a Class I device need a notified body?
For a Sınıf-I Diğer device, the manufacturer declares conformity itself with the EU declaration of conformity. For Class I devices that are sterile, have a measuring function or are reusable surgical instruments, the notified body assesses only the relevant aspect (Article 52, sixth paragraph).
For documents and ÜTS registration appropriate to the risk class: Medical Device ÜTS Registration Consultancy