IVDR ÜTS Document Registration: Certificate or Declaration for Class A, B, C and D?

12 hours ago
10 min read
IVDR ÜTS belge kaydı sınıfa göre belgeler

SUMMARY

IVDR ÜTS document registration means registering the EU certificate and EU declaration of conformity issued under the IVDR in ÜTS in the correct type and linking them to the device. The IVDR class of the device determines which document is required.

In this article we give the document types, the classification rules and the document business rules by class, with verbatim quotations from the TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03 and the Regulation on In Vitro Diagnostic Medical Devices. For the general list of required documents, see our article on IVDR document list.

For ÜTS registration of your IVDR documents: Medical Device ÜTS Registration Consultancy

Contents

1. IVDR ÜTS Document Registration: In Which Type Are Documents Registered?

“ARTICLE 7- (1) A document of the type "AB Sertifikası (MDR/IVDR)" (EU Certificate (MDR/IVDR)) is used only in notifications and applications relating to certificates prepared by a notified body under the New Medical Device Regulations. For certificates issued under the Old Medical Device Regulations, a document registration/update notification and application of the type "EC Sertifikası" (EC Certificate) is made.”

“ARTICLE 11- (1) A document of the type "AB Uygunluk Beyanı (MDR/IVDR)" (EU Declaration of Conformity (MDR/IVDR)) is used only in notifications and applications relating to declarations of conformity prepared under the New Medical Device Regulations. For declarations of conformity issued under the Old Medical Device Regulations, a document application of the type "Uygunluk Beyanı" (Declaration of Conformity) is made.”

That is, an IVDR certificate is registered as "AB Sertifikası (MDR/IVDR)" and an IVDR declaration of conformity as "AB Uygunluk Beyanı (MDR/IVDR)". IVDD (98/79/EC) documents remain in the old types.

2. IVDR Classes and Classification Rules

“(1) Devices are classified as Class A, B, C or D, taking into account the risks they carry by virtue of their intended purpose and design. Classification is carried out in accordance with Annex VIII.”

Annex VIII contains seven rules. Rule 1, which leads to Class D:

  • “Devices intended to be used for the following purposes are classified as class D:”

  • “Detection of the presence of, or exposure to, a transmissible agent in blood, blood components, cells, tissues or organs, or in any of their derivatives, in order to assess their suitability for transfusion, transplantation or cell administration.”

  • “Detection of the presence of, or exposure to, a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation.”

  • “Determining the infectious load of a life-threatening disease where monitoring is critical in the patient management process.”

Rule 2 places blood grouping and tissue typing devices in Class C; those determining certain markers in the ABO, Rhesus, Kell, Kidd and Duffy systems are Class D. Rule 3 lists the Class C purposes (such as sexually transmitted agents, cancer screening, human genetic tests, companion diagnostics). For self-testing devices, Rule 4:

“a) Devices for self-testing are classified as class C. However, devices intended for the detection of pregnancy, fertility testing, determining cholesterol level, and the detection of glucose, erythrocytes, leucocytes and bacteria in urine are classified as class B.”

Class A and Class B rules:

  • “The following devices are classified as class A:”

  • “a) General culture media and histological stains, intended by the manufacturer to make them suitable for in vitro diagnostic procedures relating to a specific examination; products for general laboratory use; accessories without critical characteristics; buffer solutions; washing solutions,”

  • “b) Instruments intended by the manufacturer specifically to be used for in vitro diagnostic procedures,”

  • “c) Specimen receptacles.”

  • “All other in vitro diagnostic devices are classified as class B.”

  • “Controls without a quantitative or qualitative assigned value are classified as class B.”

If more than one rule applies, the rule giving the highest class is applied (Annex VIII, 1.9). For classification, see also our page on IVD device classification.

3. By Class: Certificate or Declaration?

For a Class A device, the manufacturer declares conformity by its own declaration:

“(10) Manufacturers of Class A devices, other than devices for performance studies, declare the conformity of their products by issuing the EU declaration of conformity referred to in Article 18 after drawing up the technical documentation set out in Annex II and Annex III.”

For a Class A sterile device the notified body becomes involved, but only on sterility:

“If these devices are placed on the market in a sterile condition (Class A Sterile), the manufacturer also applies one of the procedures set out in IVDR Annex IX (Chapters I and III) or in IVDR Annex XI. However, the involvement of the notified body in these procedures is limited to the aspects relating to establishing, securing and maintaining sterile conditions.”

For Class B, C and D, a conformity assessment carried out by a notified body is mandatory:

  • “(9) Manufacturers of Class B devices, other than devices for performance studies, are subject to a conformity assessment as specified in Chapters I and III of Annex IX and to an assessment of the technical documentation of at least one representative device per category of devices, as specified in point 4 of the same Annex.”

  • “a) Manufacturers of Class C devices, other than devices for performance studies, are subject to a conformity assessment as specified in Chapters I and III of Annex IX, including an assessment of the technical documentation of at least one representative device per generic device group as specified in point 4 of the same Annex.”

  • “(8) Manufacturers of Class D devices, other than devices for performance studies, are subject to a conformity assessment as specified in IVDR Annex IX (Chapters I and III) and IVDR Annex IX (Chapter II (excluding Article 5)).”

These provisions are in Article 50 of the Regulation on In Vitro Diagnostic Medical Devices and Article 62 of the guideline (Article 48 of the IVDR). For self-testing and near-patient testing devices, the Annex IX 5.1 procedure, and for companion diagnostics the Annex IX 5.2 procedure, is applied in addition.

4. ÜTS Document Business Rules: Annex-94, Annex-95, Annex-96

  • “(3) For devices to be notified to ÜTS under the IVDR as "Sınıf A" (Class A) or "Sınıf A Steril" (Class A Sterile), the appropriate one of the ÜTS document business rules shown in Annex-94 is applied.”

  • “(4) For devices to be notified to ÜTS under the IVDR as "Sınıf B" (Class B) or "Sınıf C" (Class C), the appropriate one of the ÜTS document business rules shown in Annex-95 is applied.”

  • “(5) For devices to be notified to ÜTS under the IVDR as "Sınıf D" (Class D), the appropriate one of the ÜTS document business rules shown in Annex-96 is applied.”

There are separate tables for self-testing, near-patient testing and companion diagnostic devices (Annex-97 to Annex-100). In the Annex-95 and Annex-96 tables, the mandatory documents are the EU Declaration of Conformity appropriate to the class, the EU Certificate and the Instructions for Use (1 piece). The EU certificate may be attached with "EK IX (BÖLÜM I&III) KALİTE YÖNETİM SİSTEMİ DEĞERLENDİRMESİ" (Annex IX (Chapters I&III) Quality Management System Assessment) and "EK IX BÖLÜM II (MADDE 5 HARİÇ) TEKNİK DOKÜMANTASYON DEĞERLENDİRMESİ" (Annex IX Chapter II (excluding Article 5) Technical Documentation Assessment); for Class C and D there is also the option of a certificate attached with "EK-X TİP İNCELEMESİ" (Annex X Type Examination) and "EK XI ÜRETİM KALİTE GÜVENCESİ DEĞERLENDİRMESİ" (Annex XI Production Quality Assurance Assessment). ISO 13485, Authorised Distributorship Certificate, Domestic Goods Certificate, Technical File and Catalogue are optional documents.

The obligation to upload instructions for use also depends on the class:

“b) For Class B, Class C and Class D devices certified under the IVDR, and for Self-Testing Devices (Class B/C/D), Near-Patient Testing Devices (Class B/C/D) and Companion Diagnostic Devices (Class C/D), the instructions for use issued by the manufacturer,”

“(3) For device groups in other risk groups (Class-I Other, Class-Im, Class-Is, Class-Ir, Class A, Class A Sterile and IVD Other devices), uploading instructions for use is optional.”

Documents are first brought to "Kayıtlı" (Registered) status by a document application, then selected in the "Belgeler" (Documents) field of the IVDR device registration (Article 30). For device registration, see our article on ÜTS IVD product registration.

5. Information Required in the EU Declaration of Conformity and EU Certificate

In IVDR declarations other than Class A, notified body information is required:

“ı) For EU declarations of conformity other than those covering only Class-I Other or Class A devices: name and identification number of the notified body, a description of the conformity assessment procedure performed, and identification of the EU certificate or certificates issued (certificate number or numbers),”

The Basic UDI-DI must be in the declaration; the Basic UDI-DI is also required in EU technical documentation, type examination and product verification certificates, but not in quality management certificates (Articles 27 and 30). For the declaration of a manufacturer outside the EU and Türkiye:

“a) In document registration notifications of the "AB Uygunluk Beyanı (MDR/IVDR)" type mentioned in this article belonging to manufacturers not established in an EU member state or in Türkiye, the "AB Yetkili Temsilcisi" (EU Authorised Representative) field must be filled in on the ÜTS registration screen.”

For the preparation of the declaration, see our article on IVDR declaration of conformity.

6. Document Application: E-Signature, Apostille and Translation

For a declaration of a foreign manufacturer, the apostille-certified copy, the copy certified by a Republic of Türkiye embassy/consulate, the e-Apostille-certified original or the e-signed original is uploaded to the "Orijinal Doküman" (Original Document) field (Article 12). The condition for an e-signed document:

“ç) For documents electronically signed (e-signed) by the manufacturer, the original of the relevant e-signed document, provided that it can be verified through e-signature verification tools/data that the document has been electronically signed by its manufacturer, that the e-signature is valid and that no changes have been made to the document afterwards,”

2026/KKB-2 update: With TİTCK's Announcement No. 2026/KKB-2 dated 02.11.2026, a new route was introduced for the declaration of conformity and SİP declaration of a foreign manufacturer, replacing the option of uploading the manufacturer's e-signed original:

“it has been decided to implement the option that the document be uploaded to the relevant field in ÜTS by the authorised official of the importer firm authorised by the foreign manufacturer, together with an undertaking letter (taahhütname) electronically signed with an e-signature (secure e-signature) created on the basis of a qualified electronic certificate obtained from an electronic certificate service provider recognised by Türkiye.”

Source: TİTCK Announcement No. 2026/KKB-2. Required documents and steps: ÜTS document registration guide.

“(2) Where the condition mentioned in the first paragraph of this article exists, the Turkish translation of the document uploaded to the "Orijinal Doküman" field, made by a sworn translator, is uploaded completely to the "Türkçe Doküman" (Turkish Document) field.”

“(5) In a document application whose procedures and principles are referred to in Article 39 of this Guideline, a maximum of 5 document registration notifications is allowed.”

The options for the EU certificate are in Article 8; these include the original of the notified body's e-signed certificate and a certificate that can be verified on the body's query site or in EUDAMED. For details, see our article on medical device apostille guide.

7. Class D and the EU Reference Laboratory

“(5) Without prejudice to any of the obligations under the other procedures referred to in the third and fourth paragraphs, the notified body carrying out the conformity assessment shall, in particular for devices for which one or more EU reference laboratories have been designated, request one of the EU reference laboratories to verify by laboratory testing, as specified in point 4.9 of Annex IX and point (h) of point 3 of Annex X, the performance claimed by the manufacturer and the compliance of the device with the applicable common specifications or other solutions chosen by the manufacturer to ensure a level of safety and performance that is at least equivalent.”

The obligation to test each batch produced is in Annex IX:

“4.12. In order to verify the conformity of Class D devices, the manufacturer carries out tests on every manufactured batch of the device. Following the completion of the controls and tests, the manufacturer sends the relevant reports on these tests to the notified body without delay.”

Batch test reports go from the manufacturer to the notified body. The Annex-96 table has no separate document type for the EU reference laboratory report.

8. IVDD Documents and Transition Dates

Regulation (EU) 2024/1860 has been incorporated into Article 48 of the guideline. For IVDD devices that did not require a notified body but do under the IVDR:

  • “(4) Provided that the conditions set out in the Regulation referred to in the first paragraph of this Article are met, for devices for which the conformity assessment procedure under the IVDD did not require the involvement of a notified body, for which a declaration of conformity was drawn up before 26 May 2022, and for which the conformity assessment procedure under the IVDR requires the involvement of a notified body, and”

  • “a) Devices to be classified as Class D, Self-Testing Device (Class D), Near-Patient Testing Device (Class D) and Companion Diagnostic Device (Class D) under the IVDR, until 31.12.2027,”

  • “b) Devices to be classified as Class C, Self-Testing Device (Class C), Near-Patient Testing Device (Class C) and Companion Diagnostic Device (Class C) under the IVDR, until 31.12.2028,”

  • “c) Devices to be classified as Class B, Near-Patient Testing Device (Class B) and Class A Sterile under the IVDR, until 31.12.2029,”

For devices placed on the market under the IVDD as Annex-2 List A, Annex-2 List B and Self-Testing Devices, the date is 31.12.2027 (Article 48, third paragraph). To benefit from the transition, an EBS application is made first, then an ÜTS application (Article 48, fifth paragraph). For details, see our articles on IVDR transition calendar under (EU) 2024/1860 and IVDR transition period extension.

9. Frequently Asked Questions

In which document type is an IVDR certificate registered in ÜTS?

In the type "AB Sertifikası (MDR/IVDR)". An IVDR declaration of conformity is registered in the type "AB Uygunluk Beyanı (MDR/IVDR)".

Is an EU certificate required for a Class A IVD device?

For a non-sterile Class A device, the manufacturer declares conformity with an EU declaration of conformity. For a Class A sterile device, the notified body becomes involved in the process only on matters relating to sterile conditions.

Which documents are mandatory for Class B, C and D?

According to the Annex-95 and Annex-96 tables of the guideline, the EU Declaration of Conformity, the EU Certificate and the Instructions for Use are mandatory documents.

Is the Class D batch test report uploaded to ÜTS?

Under Annex IX 4.12, the manufacturer sends the test reports of each batch to the notified body. There is no separate document type for this report in the Annex-96 table of the guideline.

How many documents can there be in one document application?

According to the guideline, a maximum of 5 document registration notifications is allowed in one document application.

For ÜTS registration of your IVDR documents and IVD devices: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara