top of page

GMDN Code: What It Is, GMDN Selection in ÜTS Product Registration, Addition Request and the Difference from EMDN

12 hours ago
8 min read
GMDN kodu ÜTS ürün kaydı

SUMMARY

The GMDN code is the device code in the Global Medical Devices Nomenclature (GMDN) system. The GMDN field is mandatory in the ÜTS product registration, and each device has only one GMDN code. The code is taken first from the documents of the device; if it is not in the documents, it is obtained from the manufacturer; if the manufacturer does not have it either, the most suitable code describing the device is selected.

In this article we give the rules from the TİTCK Guideline on the Registration Processes of Medical Devices to the Product Tracking System (TCKKD-KLVZ-03) with article numbers and quotations translated from the Guideline.

To have the GMDN and EMDN fields in your product registrations entered according to the documents: Medical Device ÜTS Registration Consultancy

Contents

1. What Is the GMDN Code?

Article 4 of the ÜTS Registration Guideline defines GMDN among the definitions as follows:

“ö) GMDN (Global Medical Devices Nomenclature): means the Global Medical Devices Nomenclature,”

The Guideline on Individual Movement Notifications (Tekil Hareket Bildirimleri Kılavuzu) gives the same abbreviation as follows:

“GMDN: Global Medical Devices Nomenclature - Global Medical Device Classification,”

For the accuracy of the code, ÜTS takes the data of the GMDN Agency as the basis (Article 27, fourth paragraph, item (b)):

“8) In GMDN code queries, the data of the GMDN Agency are accepted as correct and valid.”

2. GMDN Field in the ÜTS Product Registration

In the registration of a device under the MDR, GMDN is under the "Sınıflandırma Bilgileri" (Classification Information) heading (Article 27, fourth paragraph):

  • “1) This field is a mandatory field to be selected in ÜTS.”

  • “2) Each device has only one GMDN code.”

The same rule also exists in device registration under the old regulations (Article 25) and device registration under the IVDR (Article 30). The single-code rule also applies to systems/procedure packs (SİP):

“2) Each SİP has only one GMDN code.”

For the other fields of the product registration, see our ÜTS product registration article.

3. In What Order Is the GMDN Code Selected?

The order in a registration under the MDR is as follows:

  • “3) If the GMDN code describing the device appears in at least one of the documents of the device (EU certificate, EU Declaration of Conformity), the GMDN code stated in that document or those documents is selected in this field.”

  • “4) If the GMDN code of the device does not appear in the documents of the device (EU certificate, EU Declaration of Conformity), the GMDN code obtained from the manufacturer is selected or, if the manufacturer does not have a GMDN code, the most suitable GMDN code describing the device is selected.”

In a registration under the old regulations (Article 25), the EC certificate and the declaration of conformity count as documents:

“3) If the GMDN code describing the device appears in at least one of the documents of the device (EC certificate, Declaration of Conformity), the GMDN code stated in that document or those documents is selected in this field.”

For code determination, see our EMDN and GMDN code determination service page.

4. Consistency with Single-Use and Sterile Information

  • “5) There must be consistency between the information in the description of the GMDN code to which the device is matched and the product characteristics such as whether the device is single-use and/or sterile.”

  • “6) The description of the GMDN code of a single-use device contains the expression single-use. On the other hand, if the description of the GMDN code of a reusable device contains an expression related to single-use and this device is a device used multiple times on a single patient, it is mandatory to select the field mentioned in item (k) of the seventh paragraph of this article as "Evet" (Yes).”

The item (k) referred to here is the "Tek bir hastaya mı kullanılabilir?" (Can it be used on a single patient?) field.

5. If the GMDN Code Is Not in ÜTS: GMDN Addition Request

“7) If the GMDN code describing the device is not defined in ÜTS, a request is opened through the "GMDN Ekleme" (GMDN Addition) field, using the "Talep Bildir" (Report Request) screen in ÜTS, for the addition of that GMDN code. If the request is found appropriate and it is confirmed in the GMDN query that the GMDN code in question is correct and valid, and at the same time, if the GMDN code in question covers a device or devices under the Medical Device Regulations, the relevant GMDN code is defined in this field. Only correct and valid GMDN codes are defined in ÜTS.”

Three conditions are required together for the acceptance of the request: the request being found appropriate, the code being correct and valid in the GMDN query, and the code covering a device under the Medical Device Regulations.

6. The GMDN Rule in an Application

In applications for devices under the old regulations, only devices linked to the same GMDN code are put in the same application (Article 40):

“(7) In applications for devices under the Old Medical Device Regulations, applications are allowed to be made only for devices linked to the same GMDN code.”

Under the new regulations, the rule works through the Basic UDI-DI. For details, see our article on Basic UDI-DI, and for an example of a rejection arising from this rule, see our page on devices linked to different GMDN codes in the same application.

7. Changing the GMDN of a Registered Product

If more than one company has a registration with the same barcode, the GMDN difference is resolved through an imported product data change request. GMDN is among the fields requiring Authority approval in this request (Article 43):

“(7) Data change requests concerning fields requiring evaluation by the Authority, such as "Sınıf" (Class), "GMDN" and "EMDN", are submitted to the evaluation and approval of the Authority in every case mentioned in the fourth, fifth and sixth paragraphs of this article.”

For the process, see our imported product data change article. According to Article 10 of the Guideline, a declaration of conformity under the MDD or AIMDD may also be updated for a GMDN code change, provided that the devices benefit from the MDR transition provisions:

“can be updated only for changes such as a change of manufacturer address, change of production site, change of EU authorised representative and change of GMDN code, other than "önemli değişiklikler" (significant changes) in the product scope of the declaration and in the design and intended purpose of the relevant devices.”

8. Difference Between GMDN and EMDN

To search for an EMDN code in the official EMDN v2026 list: EMDN code query tool.

EMDN is a separate field from GMDN in ÜTS ("EMDN Terim", EMDN Term). Article 26 of the Medical Device Regulation regulates the use of terminology:

“(1) Manufacturers and other natural or legal persons who are required by this Regulation to use the European database on medical devices (EUDAMED) shall use the terminology provided free of charge by the Commission.”

Article 27 of the ÜTS Registration Guideline regulates the EMDN field as follows:

  • “1) Until a new decision is taken by the Authority on the subject, and unless the EMDN code describing the device appears in at least one of the documents of the device (EU certificate, EU Declaration of Conformity), this is an optional field to be selected in ÜTS.”

  • “4) Each device has only one EMDN code.”

  • “5) If the EMDN code describing the device appears in at least one of the documents of the device (EU certificate, EU Declaration of Conformity), it is mandatory to select in this field the most detailed code and terminal term of the EMDN code stated in that document or those documents.”

In short: GMDN is mandatory in every registration. EMDN is mandatory if it appears in the document and optional if it does not. For details, see our EMDN code page.

For the seven-level structure of the EMDN code, its relation to EUDAMED and the EMDN Terim (EMDN Term) field in ÜTS, see our glossary entry What Is EMDN and Which Classification Does It Give.

How Is the GMDN Code Obtained?

The GMDN code is not a code obtained from ÜTS; what is done in the ÜTS product registration is to select the code describing the device according to the order in the Guideline. The Guideline lists where the code is to be obtained from as follows (Article 27, fourth paragraph, item (b)):

“3) If the GMDN code describing the device appears in at least one of the documents of the device (EC certificate, Declaration of Conformity), the GMDN code stated in that document or those documents is selected in this field.”

“4) If the GMDN code of the device does not appear in the documents of the device (EC certificate, Declaration of Conformity), the GMDN code obtained from the manufacturer is selected or, if the manufacturer does not have a GMDN code, the most suitable GMDN code describing the device is selected.”

The order is: first the code in the documents of the device, if not in the documents the code obtained from the manufacturer, and if not that either, the most suitable code describing the device. If the selected code is not in the ÜTS list, the GMDN Addition request in Section 5 of this article is opened; for the accuracy of the code, ÜTS takes the data of the GMDN Agency as the basis (rule number 8 of the same item).


For the product registration steps in which the code is entered, see our ÜTS registration guide, and for having the registrations carried out on your behalf, see our ÜTS consultancy page.


9. Frequently Asked Questions

Can a device have more than one GMDN code?

No. According to the ÜTS Registration Guideline, each device has only one GMDN code.

What is done if there is no GMDN code in the declaration of conformity?

The GMDN code obtained from the manufacturer is selected. If the manufacturer does not have a GMDN code, the most suitable GMDN code describing the device is selected.

What is done if the GMDN code does not appear in the ÜTS list?

A "GMDN Ekleme" request is opened from the "Talep Bildir" screen. If the code is correct and valid in the GMDN query and covers a device under the Medical Device Regulations, it is defined in ÜTS.

What happens if the GMDN description of a reusable device contains the expression single-use?

If the device is used multiple times on a single patient, the "Tek bir hastaya mı kullanılabilir?" field must be selected as "Evet" in the product registration.

Is the EMDN code mandatory too?

The EMDN code is mandatory if it appears in the EU certificate or EU declaration of conformity of the device. If it is not in the documents, it is optional until the Authority takes a new decision.

Where is the GMDN code query made?

The code is selected in the GMDN field in the ÜTS product registration. Item (b) of the fourth paragraph of Article 25 of the ÜTS Registration Guideline (TCKKD-KLVZ-03) determines the data to be taken as the basis in the query as follows:

“8) In GMDN code queries, the data of the GMDN Agency are accepted as correct and valid.”

If the code you are looking for does not appear in the ÜTS list, the GMDN Addition request route in Section 5 is followed.

For an ÜTS product registration with GMDN and EMDN fields consistent with the documents: Medical Device ÜTS Registration Consultancy

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara