Notified Body Certificate: Validity Period, Suspension, Withdrawal and Restriction of the Certificate (Art. 56, Annex XII) and ÜTS

SUMMARY
Suspension of a certificate is one of the decisions a notified body takes under Article 56 of the Tıbbi Cihaz Yönetmeliği (Medical Devices Regulation) when it finds that the manufacturer no longer meets the requirements of the regulation and conformity is not restored through corrective action within the set period. The other options are withdrawal and restriction of the certificate.
In this article we give the paragraphs of Article 56, the relevant provisions in Annex VII, Annex IX and Annex XII, the status of certificates when a designation is suspended, and the corresponding ÜTS document record as direct quotations.
For the ÜTS document registration of your EU certificate: Medical Device ÜTS Registration Consultancy
Contents
1. Language and Minimum Content of the Certificate
Article 56, first paragraph of the Tıbbi Cihaz Yönetmeliği:
“(1) Certificates issued by notified bodies in accordance with Annex IX, Annex X and Annex XI shall be in Turkish and/or in an official EU language determined by the Agency or, failing such a determination by the Agency, in an official EU language acceptable to the notified body and/or in Turkish. The minimum content of the certificates shall be as set out in Annex XII.”
Of the minimum content listed in Chapter II of Annex XII, the items directly related to suspension and restriction are:
“3. the unique number identifying the certificate,”
“5. the date of issue,”
“6. the expiry date,”
“14. conditions or limitations on the validity of the certificate,”
Item 8 of Chapter I of Annex XII covers a modified certificate:
“8. In the event of a certificate having been supplemented, modified or re-issued, the new certificate shall contain a reference to the previous certificate and its date of issue, together with a description of the changes.”
To find out what an EU certificate is, see our page on what is an EU certificate.
2. How Long Does a Certificate Remain Valid?
“(2) Certificates shall be valid for the period indicated in them, which shall not exceed five years. On application by the manufacturer, the validity of a certificate may be extended for further periods, each not exceeding five years, based on a re-assessment in accordance with the applicable conformity assessment procedures. Any supplement to a certificate shall remain valid as long as the certificate which it supplements is valid.”
Point 4.11 of Annex VII covers recertification:
“The re-certification of approved quality management systems or of EU technical documentation assessment certificates or EU type-examination certificates shall be carried out at least every five years.”
Five years is the upper limit; the validity is the period stated in the certificate. A supplement to a certificate depends on the validity of the main certificate.
3. Restriction of the Intended Purpose: Art. 56(3)
“(3) Notified bodies may restrict the intended purpose of a device to certain groups of patients, or require manufacturers to undertake specific post-market clinical follow-up studies pursuant to Part B of Annex XIV.”
Under point 14 of Chapter II of Annex XII, a restriction appears in the certificate as "conditions or limitations of the validity". For PMCF studies, see our article on clinical evaluation and PMCF.
4. Suspension, Withdrawal and Restriction of a Certificate: Art. 56(4)
“(4) Where a notified body finds that the requirements of this Regulation are no longer met by the manufacturer, it shall, taking account of the principle of proportionality, suspend or withdraw the certificate issued or impose any restrictions on it unless compliance with such requirements is ensured by appropriate corrective action taken by the manufacturer within an appropriate deadline set by the notified body. The notified body shall give reasons for its decision.”
Elements of the paragraph:
The notified body finds that the manufacturer no longer meets the requirements.
The manufacturer is given an appropriate deadline, set by the notified body, for corrective action.
If compliance is not ensured within that deadline, the certificate is suspended, withdrawn or restricted.
The choice is made taking account of the principle of proportionality.
The notified body gives reasons for its decision.
5. Relevant Provisions in Annex VII and Annex IX
Point 4.3 of Annex VII covers the contract between the notified body and the manufacturer:
“The contract shall contain clear terms and conditions and include obligations enabling the notified body to act as required under this Regulation, including the obligation of the manufacturer to inform the notified body of vigilance reports, the right of the notified body to suspend, restrict or withdraw certificates issued and the duty of the notified body to fulfil its information obligations.”
Point 4.10 of Annex VII lists one of the notified body's procedures for surveillance audits:
“- where necessary, to impose specific restrictions on the relevant certificate, or to suspend or withdraw it.”
Point 3.7 of Chapter I of Annex IX covers a difference in samples during surveillance:
“3.7. Where the notified body finds a divergence between the sample taken from the devices manufactured or from the market and the specifications laid down in the technical documentation or the approved design, it shall suspend or withdraw the relevant certificate, or impose restrictions on it.”
6. A Suspended Certificate Is Entered in the Electronic System
“(5) The notified body shall enter into the electronic system referred to in Article 57 any information regarding certificates issued, including amendments and supplements thereto, and regarding suspended, reinstated, withdrawn or refused certificates and restrictions imposed on certificates. Such information shall be accessible to the public.”
One of the items of information contained in the electronic system under Article 57:
“(g) information relating to certificates referred to in Article 56(5).”
The EUDAMED Q&A text published in Turkish by the TİTCK (Turkish Medicines and Medical Devices Agency):
“Where the NB/CRF module is used to record certificate information, all subsequent updates and decisions relating to the initially recorded certificate (for example withdrawal, suspension, lifting of suspension) must be recorded in Eudamed (this also applies during the voluntary period).”
For the EUDAMED modules, see our page on what is EUDAMED.
7. If the Designation of a Notified Body Is Suspended
Art. 46, fourth and fifth paragraphs:
“The suspension shall not exceed one year and may be renewed once for the same period.”
“(5) Where its designation is suspended, restricted or fully or partially withdrawn, the notified body shall inform the manufacturers concerned at the latest within 10 days.”
Under the seventh paragraph, where the designation is restricted, suspended or withdrawn, the Agency:
“(a) shall assess the impact on the certificates issued by the notified body.”
“(c) shall request the notified body to suspend or withdraw, within a reasonable period of time determined by the Agency, any certificates which were unduly issued, in order to ensure the safety of devices on the market.”
The eighth paragraph lists the conditions under which certificates remain valid when a designation is suspended or restricted:
“(8) With the exception of certificates unduly issued, and where a designation has been suspended or restricted, the certificates shall remain valid in the following circumstances:”
“(a) the Agency has confirmed, within one month of the suspension or restriction, that there is no safety issue in relation to certificates affected by the suspension or restriction, and the Agency has outlined a timeline and actions anticipated to remedy the suspension or restriction”
Where, under point (b), the Agency determines that the notified body is not capable of supporting existing certificates, the manufacturers submit, within three months of the suspension or restriction, written confirmation that another qualified notified body is temporarily assuming the functions. The ninth paragraph separately regulates the case where the designation is withdrawn.
8. EU Certificate in the ÜTS Document Record
TİTCK ÜTS Kayıt Kılavuzu (ÜTS Registration Guide, TCKKD-KLVZ-03), Art. 7, on the validity of a document of the "AB Sertifikası (MDR/IVDR)" (EU Certificate (MDR/IVDR)) type:
“(3) EU certificates issued under the new medical device regulations remain valid for the period stated in the certificate, which shall not exceed five years, and as long as no change or update is made to the certificate during that period.”
The second paragraph of the same article lists the fields checked during document review. One of them:
“(m) conditions or limitations of the validity of the certificate,”
“are fields that must be present and are checked during document review in ÜTS. The information entered in ÜTS must be the same as the information in the certificate.”
“(4) If the document has been revised, the revision date is entered as the document start date.”
The Guide does not define a separate ÜTS procedure for a suspended or withdrawn certificate. The steps of document registration are on our pages ÜTS document registration and ÜTS document registration service.
For EC certificates issued under the former regulations, Provisional Article 1 refers to certificates that are "not subsequently withdrawn":
“certificates that are not subsequently withdrawn and are still valid on 26/5/2021 remain valid after the end of the period indicated in the certificate until the dates specified in point (b) of the fifth paragraph for the relevant risk class of the devices.”
The transition timetable is in our article on the Regulation (EU) 2023/607 MDR transition timetable. The change of notified body by the manufacturer is regulated in Article 58; for details see change of notified body.
9. Frequently Asked Questions
When is a certificate suspended?
Under the fourth paragraph of Article 56, the notified body finds that the requirements are no longer met and, if compliance is not ensured through corrective action within the deadline it sets, it suspends, withdraws or restricts the certificate, taking account of the principle of proportionality.
Must a suspension decision state reasons?
Yes. Under the same paragraph, the notified body gives reasons for its decision.
Is information on a suspension public?
Under the fifth paragraph of Article 56, information on suspended, reinstated, withdrawn or refused certificates and on restrictions is entered into the electronic system and is accessible to the public.
Is a supplement to a certificate subject to a separate period?
No. Under the second paragraph of Article 56, a supplement remains valid as long as the certificate it supplements is valid.
What happens to certificates if the designation of a notified body is suspended?
Under the eighth paragraph of Article 46, certificates other than those unduly issued remain valid under the conditions set out in that paragraph. The notified body informs the manufacturers concerned at the latest within 10 days.
For the ÜTS document registration of your EU certificate: Medical Device ÜTS Registration Consultancy
Sources