IVDR Declaration of Conformity: Annex IV Content, Conformity Route by Class and ÜTS Checks

SUMMARY
The IVDR declaration of conformity is the EU declaration of conformity by which the manufacturer declares, under its own sole responsibility, that its in vitro diagnostic device meets the requirements of the In Vitro Diagnostic Medical Device Regulation. The minimum content of the declaration is listed in Annex IV of the Regulation. Whether notified body information appears in the declaration depends on the class of the device.
In this article we give the basis of the declaration, the information in Annex IV, the conformity route by class and the fields checked in ÜTS, quoting verbatim from the Regulation and from the TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03. For the declaration under MDR, see our article on the manufacturer EU declaration of conformity.
For the preparation of your IVDR declaration of conformity: Medical Device ÜTS Registration Consultancy
Contents
1. What Is an IVDR Declaration of Conformity?
Article 18 of the In Vitro Diagnostic Medical Device Regulation:
“The EU declaration of conformity states that the requirements specified in this Regulation have been fulfilled in relation to the device it covers. The manufacturer keeps the EU declaration of conformity up to date. The EU declaration of conformity contains, as a minimum, the information specified in Annex IV, and for devices placed on the market domestically, if the declaration is drawn up in other languages, its Turkish translation is attached. At the same time, for devices placed on the EU market, the EU declaration of conformity is provided in the language required by the EU member state(s) in which the device is made available on the market.”
“(2) With regard to matters not covered by this Regulation, where devices are also subject to other legislation requiring an EU declaration of conformity, a single EU declaration of conformity is drawn up in respect of compliance with all legislation applicable to the device. The declaration contains all the information required to identify all the legislation to which it relates.”
“(3) By drawing up the EU declaration of conformity, the manufacturer assumes responsibility for compliance with the requirements of this Regulation and all other legislation applicable to the device.”
Three consequences follow: the manufacturer draws up the declaration and keeps it up to date; for a device placed on the market domestically, a Turkish translation is attached to a declaration in another language; and where the device is also subject to other legislation requiring an EU declaration of conformity, a single declaration is drawn up.
2. Annex IV: Information the Declaration Must Contain
“The EU declaration of conformity contains all of the following information:”
“1. Name, registered trade name or registered trademark of the manufacturer and, if applicable, its authorized representative, and, if already issued, the single identification number referred to in Article 29, and the address of their registered places of business at which they can be contacted.”
“2. A statement that the EU declaration of conformity is issued under the sole responsibility of the manufacturer.”
“3. The Basic UDI-DI as referred to in Annex VI Part C.”
“4. Product and trade name, product code, catalog number or other unambiguous reference allowing identification and traceability of the device covered by the EU declaration of conformity, such as, where applicable, a photograph, as well as its intended purpose. Except for the product name or trade name, the information allowing identification and traceability may be provided by the Basic UDI-DI referred to in item 3 of this Annex.”
“5. The risk class of the device in accordance with the rules set out in Annex VIII.”
“6. A statement that the device covered by the present declaration is in conformity with this Regulation and, if applicable, with any other relevant EU legislation that provides for the issuing of an EU declaration of conformity.”
“7. References to any common specifications used and in relation to which conformity is declared.”
“8. Where applicable, the name and identification number of the notified body, a description of the conformity assessment procedure performed, and identification of the certificate or certificates issued.”
“9. Where applicable, additional information.”
“10. Place and date of issue of the declaration, name and function of the person who signed it, as well as an indication of on whose behalf that person signed, and the signature.”
For details of the Basic UDI-DI, see our pages What Is Basic UDI-DI, and for the single identification number, What Is SRN.
3. How Is the Risk Class Determined in the Declaration?
“(1) Devices are classified as Class A, B, C or D, taking into account the risks they carry by virtue of their intended purpose and structure. Classification is carried out in accordance with Annex VIII.”
This provision is in Article 49 of the Regulation. Under item 5 of Annex IV, the class determined according to the Annex VIII rules is written in the declaration. For the class options in ÜTS, see our article on ÜTS IVD product registration.
4. Conformity Route by Class
The class for which the declaration alone suffices is Class A. Paragraph ten of Article 50 of the Regulation:
“(10) Excluding devices for performance study, manufacturers of Class A devices declare the conformity of their products by drawing up the EU declaration of conformity referred to in Article 18, after preparing the technical documentation specified in Annex II and Annex III. However, if these devices are placed on the market in a sterile condition, the manufacturer applies the procedures specified in Annex IX or Annex XI. The involvement of the notified body is limited to aspects relating to establishing, securing and maintaining sterile conditions.”
For Class B:
“(9) Excluding devices for performance study, manufacturers of Class B devices are subject to a conformity assessment as specified in Chapters I and III of Annex IX, and to an assessment of the technical documentation of at least one representative device for each category of devices, as specified in item 4 of the same Annex. In the case of self-test and near-patient test devices, the manufacturer follows, in addition to these procedures, the technical documentation assessment procedure specified in item 5.1 of Annex IX.”
The seventh and eighth paragraphs of Article 50 apply to Class C, and the third and fourth paragraphs to Class D; in these classes too a notified body takes part in the conformity assessment. Which documents are uploaded to ÜTS in which class is covered in our article on IVDR ÜTS document registration.
5. In Which Declaration Is Notified Body Information Sought?
Item 8 of Annex IV requires notified body information “where applicable”. Article 11 of the ÜTS Registration Guideline clarifies this by class:
“ı) For EU declarations of conformity other than those covering only Class I Other or Class A devices; the name and identification number of the notified body, a description of the conformity assessment procedure performed, and identification of the EU certificate or EU certificates issued (certificate number or numbers),”
Accordingly, notified body information is not sought in a declaration covering only Class A devices. A declaration covering Class A Sterile, B, C and D devices contains the name of the notified body, its identification number, the procedure description and the certificate number.
6. Document Type in ÜTS and Fields Checked
“ARTICLE 11- (1) A document of the ‘EU Declaration of Conformity (MDR/IVDR)’ type is the document type used only in notifications and applications relating to declarations of conformity prepared under the New Medical Device Regulations. For declarations of conformity issued under the Old Medical Device Regulations, a document application of the ‘Declaration of Conformity’ type is made.”
The second paragraph of Article 11 of the Guideline also lists the following fields entered in ÜTS in addition to the information in Annex IV:
“c) If already issued, the manufacturer’s Single Registration Number (SRN) (entered in the ‘Unique Identification No’ field in ÜTS),”
“j) Document start (issue) date,”
“k) Document end (validity) date,”
“1) If the document is of unlimited duration, ‘Unlimited Duration’ is selected on the ÜTS screens.”
“2) If there is a document end date, it is entered in the relevant field in ÜTS.”
“l) Place and date of issue of the declaration,”
The paragraph ends with the following sentence:
“The information entered into the system must be the same as the information in the document.”
The Guideline also specifies by whom the declaration is drawn up and how it is uploaded:
“(3) The EU Declaration of Conformity is drawn up by the legal manufacturer company.”
“(6) The EU Declaration of Conformity must be uploaded completely, with all its annexes and additional pages.”
The Basic UDI-DI information is mandatory in the declaration; in ÜTS product registration, the Basic UDI-DI is selected only from among the numbers associated with the selected EU declaration of conformity (Guideline Article 30, fifth paragraph).
7. IVDR Declaration of Conformity of a Domestic Manufacturer
For the declaration of a manufacturer established in Türkiye, Article 13 of the Guideline:
“a) It is mandatory to apply by selecting ‘E-Signed Document Application for Manufacturing Companies’ as the Document Application Type.”
“c) Documents whose original language is Turkish are accepted.”
The document is electronically signed by the person registered on behalf of the company as “Responsible Manager” or “Authorized Signatory”. Annex-7 and Annex-8 of the Guideline contain EU declaration of conformity templates especially for domestic manufacturers (Article 11, seventh paragraph). For the preparation of your declaration, see our IVDR declaration of conformity preparation service page.
8. Frequently Asked Questions
Who draws up the IVDR declaration of conformity?
The legal manufacturer draws it up. By preparing the declaration, the manufacturer assumes responsibility for compliance with the requirements of the Regulation and keeps the declaration up to date.
Is a notified body needed for a Class A device?
Not for a non-sterile Class A device; the manufacturer prepares the technical documentation and draws up the declaration. For Class A Sterile, the notified body is involved only in matters relating to sterile conditions, and the declaration contains notified body information.
Is an English IVDR declaration of conformity sufficient?
For a device placed on the market domestically, a Turkish translation is attached to a declaration drawn up in another language (Regulation Article 18).
In which type is a declaration from the IVDD period registered in ÜTS?
Declarations under the old regulations are registered under the “Declaration of Conformity” type. The “EU Declaration of Conformity (MDR/IVDR)” type is only for declarations under the new regulations.
Is the Basic UDI-DI mandatory in the declaration?
Yes. Item 3 of Annex IV lists the Basic UDI-DI, and the ÜTS Registration Guideline makes it mandatory for the EU declaration of conformity to contain the Basic UDI-DI information.
For the preparation of your IVDR declaration of conformity and its ÜTS registration: Medical Device ÜTS Registration Consultancy
Sources