SUMMARY

The implant card is the card that carries the information, provided by the manufacturer of an implantable device together with the device, that allows the device to be identified. Article 18 of the Medical Device Regulation (Tıbbi Cihaz Yönetmeliği) regulates the card, the other information to be given to the patient and the exempt devices.

In this article we give Article 18 word for word, and the fields relating to implantable devices in ÜTS from the TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03 and the Individual Movement Notifications Guideline (Tekil Hareket Bildirimleri Kılavuzu).

İmplante edilebilir cihazlarınızın ÜTS kaydı için: Tıbbi Cihaz ÜTS Kayıt Danışmanlığı

Contents

1. What Is an Implantable Device?

Definition in Article 3 of the Medical Device Regulation:

  • “1) A device, including those that are partially or wholly absorbed, that is intended to be totally introduced into the human body or to replace an epithelial surface or the ocular surface by clinical intervention, and which is intended to remain in place after the procedure,”

  • “2) A device intended to be partially introduced into the human body by clinical intervention and intended to remain in place after the procedure for at least 30 days,”

2. Information the Manufacturer Must Provide with the Device

  • “ARTICLE 18 - (1) The manufacturer of an implantable device, together with the device;”

  • “a) Information allowing the identification of the device, including the device name, serial number, lot number, UDI and the device model, as well as the name, address and website of the manufacturer,”

  • “b) Any warnings, precautions or measures to be taken by the patient or a healthcare professional with regard to reasonably foreseeable external influences, medical examinations or environmental conditions,”

  • “c) Any information about the expected lifetime of the device and any necessary follow-up,”

  • “ç) Any other information to ensure safe use of the device by the patient, including the information in item (r) of point 23.4 of Annex I,”

  • “shall provide.”

Annex I, point 23.4 (r), referred to by item (ç):

“r) In the case of implantable devices, the overall qualitative and quantitative information on the materials and substances to which patients can be exposed,”

3. Language, Presentation and Updating of the Information

“(2) The information referred to in the first paragraph shall be provided to the patient in whom the device has been implanted, in a way that allows quick access, in Turkish and, where necessary, together with its English version. This information shall be written in a way that is easily understood by a lay person and shall be updated where appropriate. Updates to the information shall be made available to the patient via the website referred to in item (a) of the first paragraph.”

The information is in Turkish, with an English version added where necessary. The text is written so that a lay person can understand it. Updates are made available from the manufacturer's website in item (a). For the rules on Turkish information, see our article on Turkish labelling and instructions for use.

4. What Does the Implant Card Contain?

“In addition, manufacturers shall provide the information referred to in item (a) of the first paragraph on an implant card delivered with the device.”

The card contains the information in item (a):

  • Name, address and website of the manufacturer

  • Device name and device model

  • Serial number and lot number

  • UDI

For the structure of the UDI, see our article on the UDI system and Basic UDI-DI.

5. Duty of the Healthcare Institution

“(3) Healthcare institutions shall provide the patients in whom the device has been implanted, together with the implant card bearing identification information, with the information referred to in the first and second paragraphs by means that allow quick access.”

The manufacturer provides the card and the information with the device; it is the healthcare institution that presents them to the patient.

6. Exempt Devices

“(4) Sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors are exempt from the obligations set out in this Article. The Authority updates this list of exempt devices in line with the relevant measures adopted by the Commission, by adding other types of implants to it or removing implants from it.”

The exemption is for the devices listed in this list. The list is updated by the Authority in accordance with the Commission's measures.

7. UDI Recording and Retention for Implantable Devices

UDI recording is regulated separately for class III implantable devices in Article 27 (eighth paragraph and item (a) of the ninth paragraph):

  • “(8) Economic operators shall record and keep, preferably by electronic means, the UDIs of the devices which they have supplied or which have been supplied to them, if those devices are class III implantable devices or devices, device categories or groups determined by the Commission.”

  • “a) Healthcare institutions shall record and keep, preferably by electronic means, the UDIs of the devices which they have supplied or which have been supplied to them, if those devices are class III implantable devices.”

The manufacturer's document retention period for implantable devices is at least 15 years (Article 10, eighth paragraph). A summary of safety and clinical performance is also prepared for implantable devices (Article 32); for details, see our article on the summary of safety and clinical performance (SSCP).

8. The "Vücuda İmplante Edilebilir mi?" Field in ÜTS

This field is mandatory in the registration of a device under the MDR in ÜTS. TCKKD-KLVZ-03, Article 27, seventh paragraph, item (h):

  • “1) This field is mandatory to select in ÜTS.”

  • “2) This field is selected as "Evet" (Yes) if the relevant device is a device implantable in the body, and as "Hayır" (No) if it is not.”

According to item (j) of the same paragraph, when this field is selected as "Evet", ÜTS automatically enters "1" in the "Sınırlı Kullanım Sayısı" (Limited Number of Uses) field:

“(The limited number of uses of devices implantable in the body and/or single-use devices is 1.)”

For product registration, see our articles on ÜTS medical device product registration and our ÜTS product registration service page.

9. Usage Notification in ÜTS

An implanted device leaves the system when a usage notification (Kullanım Bildirimi) is made with the patient information. Individual Movement Notifications Guideline, Article 12:

“Article 12- (1) When a medical device implanted in the body by a healthcare service provider and/or covered by reimbursement and required to be directly associated with the patient is used, a "Kullanım Bildirimi" (Usage Notification) is made for that product with the person-identifying patient information.”

According to the second paragraph of the same article, one of the conditions enabling a usage notification is that the answer "evet" (yes) was given to the question "Vücuda implante edilebilir mi?" in the product registration. For products removed from the patient's body, the reason "Hastanın vücudundan çıkarılmış" (Removed from the patient's body) can be selected in the destruction/disposal notification (Article 21). For the steps, see our article on ÜTS usage notification.

10. Frequently Asked Questions

What information does the implant card contain?

The information in item (a) of the first paragraph of Article 18: the name, address and website of the manufacturer, and the device name, serial number, lot number, UDI and device model.

In which language is the information provided?

According to the second paragraph, it is provided in Turkish and, where necessary, together with its English version; it is written so that a lay person can easily understand it.

Who gives the card to the patient?

According to the third paragraph, healthcare institutions provide the implant card bearing identification information, and the relevant information, to the patient by means that allow quick access. It is the manufacturer that provides the card and the information with the device.

Which implants are exempt from the implant card?

According to the fourth paragraph, sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors are exempt.

How is an implantable device marked in ÜTS?

The "Vücuda implante edilebilir mi?" field, which is mandatory in the registration of a device under the MDR, is selected as "Evet"; in this case ÜTS automatically sets the limited number of uses to "1".

İmplante edilebilir cihazlarınızın ÜTS kayıt ve bildirimleri için: Tıbbi Cihaz ÜTS Kayıt Danışmanlığı

Sources

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara