Medical Device Classification Consultancy

Class is not just one of the fields you fill in ÜTS; it determines which document you can select, which transition date you are subject to, and whether you must upload instructions for use.

Medical Device Classification Consultancy

Medical Device Classification Consultancy: Class Decision, Its Rationale and Its Effect on the ÜTS Registration

Class is not just one of the fields you fill in ÜTS; it is the field that determines which document you can select, which transition date you are subject to, and whether you must upload instructions for use. A wrong class means a wrong document list and a rejected registration.

This page explains how we reach the class decision, what information we ask from you and in what form we deliver the rationale. We do not repeat the full text of the classification rules here; for the text of the 22 rules of the MDR and the 7 rules of the IVDR, see our page on medical device classification rules. The references are to the TİTCK “Guide on the Registration Processes of Medical Devices in the Product Tracking System” (TCKKD-KLVZ-03) and the Medical Device Regulation.

The class decision rests on intended purpose, not on product name

The answer to the classification question is not in the product name. Article 52, paragraph one of the Guide sets the framework in a single sentence: the application of the classification rules for all Medical Device Regulations is based on the intended purpose of the devices.

The source of the intended purpose is also defined. Article 4, subparagraph (jj) of the Guide defines intended purpose as the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials, or as specified by the manufacturer in the performance evaluation. In other words, the input of the class decision is the manufacturer’s declaration, not how you sell the product.

Article 52 of the Guide sets two more application rules, and these two end most of the disputes.

  • The most critical use is taken as the basis: Paragraph four of the same article states that, if a device is not intended to be used solely or specifically in a particular part of the body, it will be evaluated and classified on the basis of the most critical specified use.
  • The strictest rule applies: Paragraph five states that, if several rules, or several sub-rules within the same rule, apply to the same device based on its intended purpose, the strictest rule and sub-rule resulting in the higher classification will apply.

Read together, these two paragraphs show that the argument “our product is usually used for this purpose” does not pull the class down.

The four pieces of information needed for a class opinion

We do not need to see the product to prepare a class opinion; it is enough to know the intended purpose and four technical data points. You can request them from your manufacturer in a single e-mail.

  • The manufacturer’s written intended purpose statement: Because Article 52, paragraph one of the Guide ties classification to the intended purpose, the rules cannot be entered without this statement. The intended purpose section taken from the label or the instructions for use is also acceptable.
  • Manner and duration of body contact: Article 51, paragraph two, subparagraphs (a), (b) and (c) of the Guide define transient, short-term and long-term use. We determine which one the device falls under according to the manufacturer’s declaration.
  • Whether it is active, invasive or surgically invasive: Subparagraph (ç) of the same paragraph defines body orifice, subparagraph (d) invasive device, subparagraph (e) surgically invasive device, and subparagraph (g) active medical devices. These three definitions determine which branch of the rule set we enter.
  • Existing document and the class and annex information on the label: Article 6, paragraph two, subparagraph (g) of the Guide states that the EC certificate, and Article 7, paragraph two, subparagraph (g) that the EU certificate, must contain the annex followed. We do not give an opinion without seeing the class that the document in your hands claims.

Once these four arrive, we tie the class claim to a rule and provide the applied rule number and its rationale in writing. If your claim conflicts with the class on the document, we write that too; we do not gloss over it.

The role of duration definitions in the classification rules

Most of the rules on the MDD and MDR side depend on duration, and the duration definitions are written to the minute in the Guide. These definitions are not open to debate.

  • Transient use: Article 51, paragraph two, subparagraph (a) of the Guide states that the device is intended for continuous use for less than 60 minutes under normal conditions.
  • Short-term use: Subparagraph (b) of the same paragraph states that continuous use is intended for a period between 60 minutes and 30 days under normal conditions.
  • Long-term use: Subparagraph (c) states that continuous use is intended for more than 30 days under normal conditions.

How the duration is counted is also defined, and this is where companies often make mistakes. Article 52, paragraph six, subparagraph (a) of the Guide states that continuous use means the entire duration of use of the same device, regardless of temporary interruption of use during a procedure or temporary removal from use for purposes such as cleaning or disinfection. Subparagraph (b) of the same paragraph states that, for a device intended by the manufacturer to be replaced promptly by another device of the same type, the total use is taken as the basis.

In other words, the argument “it is replaced every eight hours” does not reset the duration. In replacements with a device of the same type, the total duration of use is calculated.

Software and accessories are evaluated separately

For companies with software or accessories in their portfolio, the assumption we correct most often is that these automatically take over the class of the main device. The Guide divides this into three.

  • Software that is itself a device is an active device: Article 51, paragraph two, subparagraph (g) of the Guide states at the end of the definition of active medical device that software that is itself a medical device is considered an active medical device.
  • Software attached to a device is in the same class as the device: Article 52, paragraph three of the Guide states that software that drives a device or influences the use of a device falls in the same class as the device; if the software is independent of any device, it is classified in its own right.
  • Accessories are classified in their own right: Paragraph two of the same article states that accessories of a medical device are classified in their own right, separately from the device with which they are used. Paragraph eight repeats the same provision for IVD device accessories.

Paragraph two of the same article sets one more point: if the device in question is intended to be used in combination with another device, the classification rules apply separately to each of the devices.

Effects of class selection in ÜTS

The consequences of the class decision are not theoretical; they show on screen immediately. It is a direct trigger in three places.

  • It opens document selection: Article 25, paragraph five, subparagraph (a), sub-item 3 of the Guide states that it is mandatory for the class of the device to have been selected on the ÜTS registration screen before document selection, and that the selection will be made from the documents that can be selected according to the class of the device.
  • It determines which document is valid: Articles 63 to 67 of the Guide define a separate document business rule annex for each regulation and each class; these are Annex-81 to Annex-100 of the Guide. A wrong class produces a wrong document list on screen.
  • It determines the instructions for use requirement: Article 17, paragraph two of the Guide lists the classes for which uploading instructions for use is mandatory; paragraph three states that it is optional for Class-I Other, Class-Im, Class-Is, Class-Ir, Class A, Class A Sterile and IVD Other devices, but that the Agency may request it where it deems necessary.

The classes that can be selected in ÜTS vary by regime

The class field is not free text; each regime has its own list, and a class claim that is not on that list cannot be entered in ÜTS.

  • Former Medical Device Directives: Article 25, paragraph four, subparagraph (a), sub-item 3 of the Guide lists the options: “Class-I Other (sterile and without measuring function)”, “Class-Im (with measuring function)”, “Class-Is (with sterile function)”, “Class-IIa”, “Class-IIb”, “Class-III”, “Annex-2 List-A”, “Annex-2 List-B”, “Self-Test Devices” and “IVD Other”.
  • MDR: Article 27, paragraph four, subparagraph (a), sub-item 3 of the Guide adds the option “Class-Ir (Reusable surgical instruments)” to the MDR list. The MDR list consists of “Class-I Other”, “Class-Ir”, “Class-Im”, “Class-Is”, “Class-IIa”, “Class-IIb” and “Class-III”.
  • IVDR: Article 30, paragraph four, subparagraph (a), sub-item 3 of the Guide places the sub-classes Self-Test Device (Class B/C/D), Near-Patient Test Device (Class B/C/D) and Companion Diagnostic Device (Class C/D) alongside “Class A Sterile”, “Class A”, “Class B”, “Class C” and “Class D”.
  • AIMDD: Article 25, paragraph four, subparagraph (a), sub-item 4 of the Guide states that “Class-III” will be selected in the relevant field for devices under the AIMDD. Article 64, paragraph three of the Guide confirms this: all devices under the AIMDD are classified in ÜTS in the same risk group as devices classified as Class-III under the MDD.

Classes that do not require a notified body

The answer to “do I need a notified body certificate?” also depends on the class. We do not repeat all conformity assessment routes here; we only summarize the narrow area where the notified body does not come into play.

  • On the MDD side: Article 58, paragraph two of the Guide states that manufacturers of Class-I devices will follow the procedures in MDD Annex VII and draw up a declaration of conformity. The same paragraph states that an additional procedure will be followed if the device is supplied sterile (Class-Is) or has a measuring function (Class-Im), but that the involvement of the notified body is limited to the sterility and metrological requirements respectively.
  • On the MDR side: Article 61, paragraph four of the Guide sets the same structure and adds Class-Ir devices to the list; in these three sub-classes, the involvement of the notified body is again framed with a limited scope.
  • On the IVDD side: Article 60, paragraph two of the Guide states that, except for performance evaluation devices and the devices specified in IVDD Annex-II, manufacturers of devices under the IVDD will carry out the procedures in IVDD Annex-III and draw up a declaration of conformity.
  • On the IVDR side: Article 62, paragraph four of the Guide states that manufacturers of Class A devices will draw up technical documentation and a declaration of conformity; that an additional procedure will be followed if the devices are supplied in sterile condition (Class A Sterile); and that the involvement of the notified body is limited to sterile conditions.
  • On the AIMDD side: Article 64, paragraph four of the Guide states that the document business rules in Annex-84 will be applied to devices under the AIMDD. The combination in Annex-84 consists of EC Certificates labeled “ANNEX-2 FULL QUALITY ASSURANCE SYSTEM” and “ANNEX-2 ARTICLE 4, DESIGN EXAMINATION”; no route without a notified body is defined.

The declaration of conformity and CE marking article explains the declaration step.

A class increase changes your transition timeline

A class moving up does not just mean more documents; it also changes the date until which you can stay on the market. The most common case we encounter is a device that is Class-I Other under the MDD but requires a notified body under the MDR.

Article 45, paragraph six of the Guide defines this group: devices for which the MDD conformity assessment procedure did not require the involvement of a notified body, for which a declaration of conformity was drawn up before 26 May 2021, and for which the MDR conformity assessment procedure requires the involvement of a notified body. The same paragraph states that, provided the conditions in the relevant Regulation are met, these devices may be placed on the market or put into service until 31.12.2028.

This date does not apply by itself. Article 46, paragraph four of the Guide requires an application through EBS for these devices and looks for two thresholds: an application to a designated notified body under the MDR before 26.05.2024, or a contract with such a body before 26.09.2024. Subparagraph (c) of the same paragraph states that, if only the application condition is met, the registered period will be extended until 26.09.2024; subparagraph (ç) states that, if the contract condition is also met, it will be extended until 31.12.2028.

This is the concrete return of determining the class early: without knowing whether your device’s class will go up under the MDR, you cannot know which application timeline you are subject to.

A class change has a chain effect

Class is not a field that can be changed on its own in ÜTS. Article 43, paragraph seven of the Guide states that data change requests concerning fields requiring Agency evaluation, such as “Class”, “GMDN” and “EMDN”, will in every case be submitted to the Agency’s evaluation and approval. Even if other companies with records under the same barcode unanimously accept the request, it is not sufficient for these fields.

When the change is approved, the effect is not limited to a single record. Article 43, paragraph eleven of the Guide states that, if, as a result of updating the “Class” data field, the documents linked to the relevant device records become incompatible with the device class, revisions will be issued to the devices with the relevant barcode.

This means that correcting the class later may require rebuilding the document side as well. The cost of obtaining the class opinion before registration is lower than the cost of correcting it after registration.

The class decision on the IVD side

The rule set is different on the IVD side, and three special provisions help us most when making the decision.

  • Calibrators: Article 52, paragraph nine of the Guide states that calibrators intended to be used with a device under the IVDR will fall in the same class as the device.
  • Control materials: Paragraph ten of the same article states that control materials with quantitative or qualitative assigned values for a specific analyte or multiple analytes under the IVDR will fall in the same class as the device. Controls without an assigned value are classified as Class B under Rule 7 in Article 57 of the Guide.
  • Types of analysis: Article 52, paragraph eleven of the Guide states that, under the IVDR, each of the classification rules will be applied separately to primary level analyses, confirmation analyses and supporting analyses.

There is one more definitional difference, and it is reflected directly in the sub-class selection in ÜTS. Article 4, subparagraph (r) of the Guide defines a near-patient test device as an in vitro diagnostic device that is not intended for self-testing and is generally used outside the laboratory environment, near or at the side of the patient, for the test to be performed by a health professional. Subparagraph (ee) of the same article defines a self-test device as an in vitro diagnostic device intended by the manufacturer to be used by lay persons.

How decisions are made on borderline products

Whether a product is a medical device is not always clear. For products on the borderline with cosmetics, biocides, foods and medicinal products for human use, the decision mechanism is separate, and it needs to be described correctly.

Article 4, paragraph one of the Medical Device Regulation states that the Agency will submit to the Commission, where necessary, its duly reasoned request to determine whether a specific product, or a category or group of products, falls within the definition of “medical device” or “medical device accessory”, or to clarify the regulatory status of products on the borderline between medicinal products, human tissues and cells, cosmetic products, biocidal products or food products and medical devices. In other words, the decision is not considered made at a single desk; the process is an application and evaluation process.

Some products are not even borderline. Article 1, paragraph four of the Medical Device Regulation lists the areas to which the Regulation does not apply: in vitro diagnostic devices, medicinal products for human use, advanced therapy medicinal products, human blood and blood products, products within the scope of the Cosmetics Regulation, certain transplants, tissues and cells of animal and human origin, products containing viable biological material, and foods. If your product is on this list, the question is not one of classification but one of legislation.

In practice, in such files we prepare a scope opinion instead of a class opinion: we state in writing which regulation the product falls under, which definition it relies on and under which articles it may be considered out of scope.

Putting the class opinion in writing

A verbal class opinion has no counterpart in the technical file. The Guide also describes where the rationale is to be written, and we prepare the output in this form.

Article 10, paragraph six of the Guide states that, if deemed necessary for any of the devices in the declaration of conformity, the technical document will be uploaded to the “Technical Document of the Product(s) in the Document” field and lists the information this document must contain. Subparagraph (c) lists “the class of the medical device, the classification rule and its rationale”. The other items are the effect and mechanism of action of the medical device, intended purpose, place of use, duration of use and target population, product content and percentage component information by form, and label information and instructions for use.

The list is more detailed on the side of the new regulations. Article 11, paragraph five of the Guide lists, for the same field, the name and general description of the device, the manufacturer’s name and address, the UDI-DI number and the EU certificate number if any, intended purpose, place of use, manner of contact, duration of use, population to be used on, mechanism of action, class and classification rule, device content and percentage component information, and label information and instructions for use. The same paragraph also states that these technical documents must be in the device’s technical file and that, if deemed necessary, the entire technical file will be kept ready to be submitted to the Agency.

The class opinion we deliver complies with this form: the applied rule number, the text of the rule, which data of the device places it under that rule, and the alternative rules that were eliminated, if any. We deliver the document in a form you can place directly in your technical file. UDI data are handled within the same discipline; see UDI-DI and Basic UDI-DI.

Scope of service: the responsibilities of the consultant and the company

We tie the class opinion to a rule and give its rationale in writing. If you also want us to carry out the registration on the ÜTS side, you define a user of the “Medical Device Registration Officer” type for our staff from your company page; under Article 39, paragraph eleven and Article 40, paragraph five, subparagraph (f) of the Guide, only users of this type can perform the “Approve” action in document and device applications.

Two things are not delegated. The electronic signature stays with your company’s signatory; Article 39, paragraph thirteen and Article 40, paragraph five, subparagraph (g) of the Guide state this explicitly. The accuracy of the data also rests with the company; Article 31, paragraph four of the Regulation on Medical Device Sales, Advertising and Promotion keeps the accuracy of the information on devices recorded in the system as the responsibility of the sales center that makes the product registration.

One more limit, stated up front: a classification opinion does not replace conformity assessment. The class decision is the manufacturer’s responsibility, and the certification obligation does not disappear in classes that require a notified body. What we do is base the decision on the rules and document its rationale; we give no commitment about the decision the Agency or a notified body will make.

To get started, you can request a quote for ÜTS registration consultancy.

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara