Technical File Pre-Review and Preparation: Before ÜTS Document Registration

We read your technical file section by section against Annexes I, II and III and give you a written gap report; if you wish, we create the file from your documents.

Technical File Pre-Review and Preparation: Before ÜTS Document Registration

Technical File Preparation and the Technical Document to Be Submitted to ÜTS

When you register a document in ÜTS, you come across a field called “Technical Document of the Product(s) in the Document”, and you try to upload the manufacturer’s entire technical file there. These two are not the same thing: one is the full technical documentation the manufacturer keeps against the legislation, the other is a summary file that ÜTS asks for under certain headings.

This page separates the two, lists which headings go into the ÜTS field with the Guide article, and shows what content the technical file should be requested from the manufacturer with. The sources are the Guide on the Registration Processes of Medical Devices in the Product Tracking System and the Medical Device Regulation published in the Official Gazette dated 2/6/2021, repeated issue no. 31499.

Technical file pre-review service

We read your technical file section by section against Annex I, Annex II and Annex III of the Medical Device Regulation and give you a written gap report. The report states, for each gap, which section it is in, what it is based on and how it can be closed. If you wish, we create the technical file from the documents coming from the manufacturer. To start, send the device list and the table of contents of the file; you can use our price quote form.

Who it is for

  • Domestic manufacturers of Class I and Class A devices: for these devices, no notified body reads the file before placing on the market.
  • Manufacturers who want their file read by an independent eye before applying to a notified body.
  • Manufacturers who are preparing or updating a declaration of conformity: every piece of information in the declaration rests on the technical file.
  • Importers bringing in a product from a foreign manufacturer: the declaration, label and instructions for use entered in the ÜTS document registration must be consistent with the file.

Legal basis

Preparing and keeping the technical file up to date is the manufacturer’s obligation:

“Manufacturers of devices, other than custom-made devices, shall draw up technical documentation for those devices and keep it up to date. The technical documentation shall be such as to allow the assessment of the conformity of the device with the requirements of this Regulation. The technical documentation shall include the elements set out in Annex II and Annex III.” (Medical Device Regulation Art. 10/4)

Annex II also states how the file is to be presented:

“The technical documentation to be drawn up by the manufacturer and, where applicable, the summary thereof shall be presented in a clear, organised, readily searchable and unambiguous manner and shall include in particular the elements listed in this Annex.” (Medical Device Regulation Annex II)

The post-market surveillance file is also part of the technical file:

“The technical documentation on post-market surveillance to be drawn up by the manufacturer in accordance with Articles 81 to 84 shall be presented in a clear, organised, readily searchable and unambiguous manner and shall include in particular the elements described in this Annex.” (Medical Device Regulation Annex III)

For in vitro diagnostic devices, the same structure is found in Annex II and Annex III of the IVD Regulation. We explain the general content of the technical file on our technical file glossary page.

What we review

We read the file in the order of Annex II, then with Annex III.

  • Device description and specifications: product name, intended purpose, intended user and patient group, Basic UDI-DI, risk class and the rationale for the classification rule applied, variants and accessories.
  • Information supplied by the manufacturer: label and instructions for use, in the required languages.
  • Design and manufacturing information: whether the design stages and manufacturing sites are described.
  • General safety and performance requirements: whether each applicable requirement in Annex I is tied to evidence.
  • Benefit-risk analysis and risk management: completeness of the risk file and its link to the rest of the file.
  • Product verification and validation: whether the reports you hold are cited in the file and match the device, including clinical evaluation.
  • Post-market surveillance: the plan and, depending on your class, the surveillance report or periodic safety update report.

GSPR checklist

At the center of the review is a table based on Annex I. For each general safety and performance requirement, the table holds four pieces of information: whether it applies to the device, in which document the evidence is, which standard or common specification was used, and whether the evidence is present in the file. A requirement that applies but has no evidence is a gap. A requirement marked “not applicable” without justification is a gap. A standard cited in one document with one version and in another with a different version is a gap.

Depth by class

For Class I and Class A devices we read the whole file, because no notified body reads it before placing on the market and the first external reader of the file may be the Agency. For Class I devices that are sterile, have a measuring function or are reusable surgical instruments, we also mark the parts the notified body will look at. For higher classes, the review is a reading before the notified body application or certificate renewal; the table and report are the same.

The difference between the technical file and the technical document field in ÜTS

Article 11, paragraph five of the Guide states that, where deemed necessary for any of the devices in the declaration of conformity, the technical document of the relevant device will be uploaded to the “Technical Document of the Product(s) in the Document” field so as to contain the information listed. The same paragraph draws the limit in two sentences: the technical documents in question must be in the technical file of the device, and where deemed necessary the entire technical file is kept ready to be submitted to the Agency.

In other words, the ÜTS field is an extract, and the technical file is the source. The summary you upload to ÜTS must come from within the technical file; writing information in the ÜTS field that is not in the technical file puts you in a difficult position at the first inspection.

The limit of the technical file itself is drawn by the Medical Device Regulation. Article 10, paragraph four states that manufacturers of devices other than custom-made devices will draw up technical documentation for these devices and keep it up to date; that the documentation will allow the assessment of the conformity of the device with the requirements of the Regulation; and that it will include the elements set out in Annex II and Annex III. Article 10, paragraph eight, subparagraph (b) also states that, upon request of the Agency, the whole of the technical documentation or a summary thereof will be submitted.

The five headings required in declarations under the old regulations

For declarations of conformity of devices under the MDD, AIMDD and IVDD, Article 10, paragraph six of the Guide lists five subparagraphs. Where deemed necessary, the technical document is uploaded so as to contain the following information:

  • Effect and mechanism of action: subparagraph (a) requires the effect and mechanism of action of the medical device.
  • Framework of use: subparagraph (b) requires the intended purpose, place of use, duration of use and target population.
  • Classification: subparagraph (c) requires the class of the medical device, the classification rule and its rationale.
  • Component information: subparagraph (ç) requires the product content and percentage component information according to the form of the product.
  • Label and instructions: subparagraph (d) requires the label information and the instructions for use.

This list has one more use. Article 10, paragraph five of the Guide requires that the device scope in a declaration of conformity be able to define the product, and states that general expressions are not sufficient; as an example, detailing as forceps, retractor and dilator is given instead of “surgical hand instruments”. The same paragraph states that, for imported products, for documents drawn up contrary to this, a document listing the branch types or product descriptions that contain the detail of products defined with a general expression will also be uploaded to the same technical document field.

Twelve headings to collect from your manufacturer for the technical file

For devices under the MDR and IVDR, the list is expanded. Article 11, paragraph five of the Guide lists the following in twelve subparagraphs, and these subparagraphs are also the backbone of the request list you will send to your manufacturer:

  • Device identity: subparagraph (a): the name / trade name and general description of the device; subparagraph (b): the name and address of the manufacturer.
  • Numbers: subparagraph (c): the UDI-DI number (barcode) and the EU certificate number, if any.
  • Framework of use: subparagraph (ç): intended purpose; (d): place of use; (e): manner of contact; (f): duration of use; (g): population to be used on.
  • Technical rationale: subparagraph (ğ): mechanism of action; (h): class and classification rule.
  • Content and information: subparagraph (ı): device content and percentage component information according to the form of the device; (i): label information and instructions for use.

These twelve headings do not sit in a single section of the manufacturer’s technical file; they are compiled from different sections. In practice, what we do is draw up the list in English to fit the manufacturer’s own technical file structure, reorganize the response according to the headings ÜTS asks for, and ask back for the missing items one by one. The manufacturer gives the answer; we prepare the right question and the right format. See also GMDN and EMDN codes and UDI-DI and Basic UDI-DI.

Classification rationale: rule and rationale, not just class

Both Article 10, paragraph six, subparagraph (c) and Article 11, paragraph five, subparagraph (h) of the Guide require not the class but the classification rule and its rationale. It is not enough for the file to say only “Class IIa”; the number of the rule and its rationale must also be written. For the rules themselves, see classification.

Article 52 of the Guide shows how this rationale is built:

  • The basis is the intended purpose: According to paragraph one, the application of the classification rules for all Medical Device Regulations is based on the intended purpose of the devices.
  • The strictest rule applies: According to paragraph five, if several rules, or several sub-rules within the same rule, apply to the same device based on its intended purpose, the strictest rule and sub-rule resulting in the higher classification applies.
  • The most critical use is the basis: According to paragraph four, if a device is not intended to be used solely or specifically in a particular part of the body, it is evaluated and classified on the basis of the most critical specified use.
  • Accessories are classified separately: According to paragraph two, accessories of a medical device are classified in their own right, separately from the device with which they are used; paragraph eight repeats the same rule for IVD device accessories.
  • Software: According to paragraph three, software that drives a device or influences its use falls in the same class as the device; if the software is independent of any device, it is classified in its own right.

The classification rationale is not just a part of the file; it also determines which certificate combination the document business rules will ask for. A wrong class later breaks the whole document chain.

What must appear in the file if there is contract manufacturing

If the product is manufactured by another manufacturer, this information must be visible both in the document and in ÜTS. Article 10, paragraph four of the Guide states that, if the manufacturer has its devices within the scope of the declaration of conformity manufactured by another manufacturer, the title and address information of that manufacturer is also expected to be in the declaration of conformity or the EC certificate. The same paragraph states that, if this information is not in the document and also not on the device label or sales packaging, the Agency will, where it deems necessary, request that the technical file pages containing this information be uploaded to the technical document field.

There is a counterpart on the ÜTS side as well. Article 27, paragraph seven, subparagraph (gg) of the Guide states that the field “Was the product manufactured for its legal manufacturer by another manufacturer?” is a mandatory field to be selected in ÜTS. Sub-item two of the same subparagraph states that “Yes” is selected if the legal manufacturer has the device designed and/or manufactured by another natural or legal person, and “No” if it manufactures in its own facilities. According to sub-item three, if “Yes” is selected, the field in subparagraph (ğğ) opens, and it is mandatory to select the manufacturing company from the “Select Manufacturer Company/Institution” screen.

Read together, these three provisions give a preparation item: if there is contract manufacturing, before the application we check whether the title and address of the manufacturing company appear in the declaration or certificate; if not, we keep the relevant technical file pages ready.

Clinical evaluation and post-market surveillance

The part of the technical file that demands the most work is the clinical side, and the manufacturer is responsible for it. Article 61, paragraph one of the Medical Device Regulation states that the manufacturer determines and justifies the level of clinical evidence necessary to demonstrate conformity with the general safety and performance requirements; that this level will be appropriate in view of the characteristics of the device and its intended purpose; and that manufacturers will plan, conduct and document a clinical evaluation in accordance with this Article and Annex XIV Part A.

  • Expert panel: According to paragraph two, for all Class III devices and for Class IIb devices referred to in Article 54, paragraph one, subparagraph (b), the manufacturer may consult the expert panel established by the Commission on its Clinical Development Strategy and proposals for clinical investigation before its clinical evaluation and/or investigation. The same paragraph states that the manufacturer will give due consideration to the views expressed by the expert panel and cannot make any claim in relation to those views.
  • Equivalence route: According to paragraph three, subparagraph (a), the clinical evaluation is based on a critical evaluation of the relevant scientific literature, provided that equivalence of the device under evaluation to the device to which the data relate is demonstrated in accordance with Section 3 of Annex XIV in terms of intended purpose, and that the data adequately demonstrate compliance with the relevant requirements.
  • PMS plan: According to Article 82, paragraph one, the post-market surveillance system is based on a post-market surveillance plan, the requirements of which are set out in Section 1 of Annex III, and for devices other than custom-made devices this plan is part of the technical documentation specified in Annex II.

The common result of these provisions is this: a consultancy firm cannot produce the clinical evaluation report and the PMS plan in place of the manufacturer. What we can do is identify the points the file in hand leaves incomplete from the perspective of ÜTS and the document application and ask the manufacturer about them in writing.

Who must have an SSCP

Article 32, paragraph one, subparagraph (a) of the Medical Device Regulation states that the summary of safety and clinical performance will be drawn up by the manufacturer for implantable devices and Class III devices, other than custom-made or investigational devices.

Subparagraph (c) of the same paragraph also regulates the fate of the document: the draft of the summary forms a part of the documentation to be submitted to the notified body involved in the conformity assessment and is validated by that body; after validating it, the notified body uploads the summary to EUDAMED, and the manufacturer indicates on the label or instructions for use where the summary is available. Subparagraph (b) states that the summary will be written in a way that is clear to the intended user and, where relevant, to the patient, and will be publicly available via EUDAMED.

Retention periods

The technical file continues to live after the product is withdrawn from the market.

  • Ten years, fifteen for implants: According to Article 10, paragraph eight, subparagraph (a) of the Medical Device Regulation, manufacturers keep the technical documentation, the EU declaration of conformity and, where applicable, a copy of any relevant certificate, including amendments and supplements, for a period of at least 10 years after the last device covered by the declaration has been placed on the market, and submit it to the Agency upon request. For implantable devices, this period is at least 15 years after the last device has been placed on the market.
  • Custom-made device statement: According to Article 29, paragraph six of the Guide, the relevant Annex XIII ITC Statement is kept for at least 10 years after the device has been placed on the market; for implantable devices, this period is at least 15 years.
  • Authorized representative’s file: According to Article 10, paragraph eight, subparagraph (c), a manufacturer whose registered place of business is outside Türkiye or the EU member states ensures that its authorized representative has the necessary documentation permanently available.

Is there a document type called “Technical File” in ÜTS?

There is a point in the source that is not reconciled, and we convey it as it is.

On one side are the document business rules regulated in Articles 63 to 67 of the Guide. For example, Article 66, paragraph one states that, for a device registration notification under the MDR, the suitable one of the documents mentioned in Articles 7, 11, 17, 20, 22, 23 and 24 of the Guide will be linked to the device. There is no “Technical File” document type in this list of articles.

On the other side, in the conformity annexes of the Guide, that is, in the document business rule tables running from Annex-81 to Annex-100, the line “Technical File (maximum 1)” is listed in the “Optional Documents” column. The Catalog is also in the same column.

The source does not reconcile these two statements. In practice this means that the technical file is not designed as a document registration type of its own in ÜTS, but it can appear as an optional document in the device registration. The mandatory technical document entry, on the other hand, runs through the “Technical Document of the Product(s) in the Document” field in Articles 10 and 11.

Instructions for use are part of the technical file

Instructions for use are the most visible link between the technical file and ÜTS, and the Guide states this explicitly.

Article 18, paragraph one, subparagraph (a), first sub-item of the Guide states that, in instructions-for-use notifications of importing companies concerning a foreign manufacturer, the most current original instructions for use found in the technical file of the relevant medical device are uploaded to the “Original Document” field; the second sub-item states that its Turkish translation is uploaded to the “Turkish Document” field. Subparagraphs (b) and (c) of the same paragraph repeat the same “most current in the technical file” criterion for multilingual instructions and instructions prepared in Turkish.

Paragraph two states that, in notifications of domestic manufacturers concerning their own products, the most current Turkish instructions for use found in the technical file are uploaded in full, with the option “Is the Original Document Turkish?” selected as “Yes”.

Paragraph three adds one more requirement: if the instructions for use have been updated, at least one of the update date and/or update information of the document in question must be entered in ÜTS. The revision in the technical file and the record in ÜTS are aligned through this field.

The relationship between the quality management system and the technical file

The technical file does not stand alone; it is produced within the manufacturer’s quality management system. Article 10, paragraph nine of the Medical Device Regulation states that, with the exception of investigational devices, manufacturers will establish, document, implement, maintain, keep up to date and continually improve a quality management system proportionate to the risk class and type of the device. The same paragraph lists the minimum content of the system in thirteen subparagraphs: a strategy for regulatory compliance, identification of general safety and performance requirements, responsibility of the management, resource management, risk management, clinical evaluation, product realisation, verification of the UDI assignment, setting-up of a post-market surveillance system, communication with stakeholders, processes for reporting vigilance, management of corrective and preventive actions, and monitoring and measurement of output.

Uploading the quality certificate to ÜTS, on the other hand, is a separate matter. Article 21, paragraph nine of the Guide explicitly states that uploading the document named “Quality Management Certificate - ISO-13485” to ÜTS is not a mandatory document but an optional one.

The role of standards is also defined in the Regulation. Article 8, paragraph one of the Medical Device Regulation states that devices that are in conformity with the harmonised standards, or the relevant parts of those standards, the references of which have been published in the Official Journal of the European Union, will be presumed to be in conformity with the requirements of the Regulation covered by those standards. The same paragraph states that this presumption also applies to quality management systems, risk management, post-market surveillance systems, clinical investigations, clinical evaluation and post-market clinical follow-up.

Link to the ÜTS document registration

The declaration of conformity, label and instructions for use uploaded to ÜTS are like a summary of the technical file. The product name, brand name, risk class and Basic UDI-DI in the declaration must be the same as in the technical file; the label and instructions must not contradict the intended purpose in the file. Checking this consistency before registration reduces the risk of a document rejection and of a product application being returned. Our declaration of conformity preparation service is a continuation of this step. See also the article on the declaration of conformity and CE marking.

EMDN and GMDN codes

In the review, we also check the device’s EMDN code; if the code is in the document, the same code is used in the ÜTS registration. We explain the difference between the two codes on our GMDN and EMDN code page.

Process

  • You send the device list and the table of contents of the file. We clarify the scope and give a quote per device family.
  • You share the file. We read it against Annexes I, II and III.
  • We send the gap report: section, gap or contradiction, the requirement it is based on and what is needed to close it.
  • You close the gaps; if you wish, we create the technical file from the documents you send.
  • We re-read the updated file and close each item in the report.

What we deliver

  • A section-by-section gap report, with the basis of each finding.
  • A GSPR checklist for the device family.
  • A summary showing which section is complete, which is incomplete and which is missing altogether.
  • A list of changes that the gaps require in the declaration of conformity, label or instructions for use.
  • If requested, a technical file built from the documents you send, in the Annex II and Annex III layout.

What we ask from you

  • Device list: risk class, intended purpose, Basic UDI-DI.
  • The current state of the technical file, or at least its table of contents.
  • Label and instructions for use.
  • The test and verification reports you hold. We read these reports for consistency with the file; we do not evaluate the test results themselves.
  • The notified body certificate, if any.

Scope and limits of the service

The output we deliver in technical file consultancy is concrete: a section-by-section gap report against Annexes I, II and III; if requested, a technical file compiled from the documents coming from the manufacturer; a request list to be sent to the manufacturer; the incoming documents reorganized according to the headings in Articles 10 and 11; the classification rule and its rationale built in writing on Article 52; comparison of areas such as contract manufacturing and instructions-for-use revision with the document and the ÜTS record; preparation of the file to be uploaded to the “Technical Document of the Product(s) in the Document” field; and creation of the record.

What we cannot do is equally clear. The obligation to prepare and keep the technical documentation up to date belongs to the manufacturer under Article 10, paragraph four of the Medical Device Regulation; the manufacturer plans, conducts and documents the clinical evaluation under Article 61, paragraph one; and the manufacturer establishes the quality management system under Article 10, paragraph nine. We can compile the technical file from the documents the manufacturer sends, but the owner of and the party responsible for the file is the manufacturer. We do not produce the clinical evaluation or the quality management system on behalf of the manufacturer, we do not determine the level of clinical evidence in the manufacturer’s place, and we do not promise the outcome of a notified body’s assessment.

Our way of working on the ÜTS side operates within the same limit: the company opens user authority for the consultant from its own ÜTS company page, the consultant creates document and product records with the “Medical Device Registration Officer” authority, and the person who signs the application electronically remains the company’s signatory. The company that makes the record is ultimately responsible for the information entered into the system matching the document; we take on the preparation, comparison and registration work. See ÜTS registration.

Out of scope

  • Notified body conformity assessment and CE certification.
  • Test and laboratory processes, including laboratory selection and arranging tests.
  • Interpreting or changing your technical specifications.
  • Finding an importer or distributor.

Related services

Frequently asked questions

Do you prepare the technical file, or only review it?

We do both. The core service is the pre-review. If you wish, we create the technical file from the documents coming from the manufacturer. Design, tests and data are yours.

Can you tell me which tests need to be done?

No. Test and laboratory processes are outside our scope. We read the reports you hold and check whether they match the device and whether they are cited in the file.

Is a pre-review needed for a Class I device?

That is where it helps most. A Class I device is placed on the market without a notified body reading the file; the first external reader of the file may be the Agency.

Do you also review the files of IVD devices?

Yes. The IVD Regulation sets up the same structure in its own Annex II and Annex III; we read the file against Annex I of that regulation.

Is the post-market surveillance file included in the review?

Yes. Annex III is part of the technical file; the plan and, depending on your class, the report are included in the review.

How is the fee determined?

Per device family, after seeing the device list. A family is the devices under the same Basic UDI-DI.

To get started, you can request a quote for ÜTS registration consultancy.

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara