IVDR Transition Period and IVD Registration
IVDR Transition Period and ÜTS Registration Consultancy for IVD Devices

IVDR Transition Period and ÜTS Registration Consultancy for IVD Devices
On the IVD side, the question "when does the transition period end" has no single answer. Depending on which class your device falls into under IVDR, the deadline varies as 31.12.2027, 31.12.2028 or 31.12.2029, and benefiting from these dates depends on two separate applications to be made, first through EBS and then through ÜTS.
On this page we explain which work we take over, which information and documents we ask from you, and what we hand over to you. All dates are taken from the TİTCK "Guide on Registration Processes of Medical Devices to the Product Tracking System" (TCKKD-KLVZ-03); each is given with its source article number.
The transition calendar for IVD devices is divided into three according to class
Article 48 of the Guide divides IVD devices into two major groups and divides the second group into three within itself. The first group consists of those placed on the market with a notified body certificate under IVDD. The second group consists of devices that did not require a notified body under IVDD, whose declaration of conformity was issued before 26 May 2022, but that will require a notified body under IVDR. The calendar in the second group varies according to the class the device will fall into under IVDR.
The dates below are subject to the conditions specified in the relevant Regulation being fulfilled. That is, the calendar does not run by itself; you need to document that you meet the conditions.
Placed on the market under IVDD as Annex-2 List-A, Annex-2 List-B and Self-Testing Devices: may be placed on the market or put into service until 31.12.2027. Third paragraph of Article 48 of the Guide.
Devices that will be Class D, Self-Testing Device (Class D), Near-Patient Testing Device (Class D) and Companion Diagnostic Device (Class D) under IVDR: 31.12.2027. Subparagraph (a) of the fourth paragraph of the same article.
Devices that will be Class C, Self-Testing Device (Class C), Near-Patient Testing Device (Class C) and Companion Diagnostic Device (Class C) under IVDR: 31.12.2028. Subparagraph (b) of the fourth paragraph.
Devices that will be Class B, Near-Patient Testing Device (Class B) and Class A Sterile under IVDR: 31.12.2029. Subparagraph (c) of the fourth paragraph.
It is normal for your portfolio to contain all three groups. For this reason, there is no such thing as "the company's transition date"; every barcode has its own date.
The transition calendar is determined by device class
The three dates above depend on the class the device is claimed to fall into under IVDR. If you determine the class wrongly, you plan according to the wrong date, and your extension application is also evaluated on the wrong date. For this reason, the first thing we do in IVD files is to bring the class claim under rules.
On the ÜTS side, the class is a field selected from a list, not free text. According to the fourth paragraph of Article 30 of the Guide, the classes that can be selected under IVDR are Class A Sterile, Class A, Class B, Class C, Class D, as well as Self-Testing Device (Class B/C/D), Near-Patient Testing Device (Class B/C/D) and Companion Diagnostic Device (Class C/D).
The intended purpose moves the class up: Rule 1 in Article 57 of the Guide moves the assessment of suitability for transfusion or transplantation and the detection of infectious agents with a high risk of propagation to Class D. Rule 2 moves blood grouping and tissue typing to Class C, and certain ABO, Rhesus, Kell, Kidd and Duffy markers to Class D. Rule 3 moves companion diagnostic use, cancer screening and staging, and human genetic tests to Class C.
Calibrators are not considered separately: According to the ninth paragraph of Article 52 of the Guide, calibrators intended to be used with a device are in the same class as the device.
For control materials the distinction depends on the assigned value: According to the tenth paragraph of the same article, control materials with a quantitative or qualitative assigned value for a specific analyte or multiple analytes are in the same class as the device. Controls without an assigned value are Class B under Rule 7 in Article 57.
IVD devices that cannot get an extension
Not every IVDD certificate can benefit from an extension. Subparagraph (a) of the second paragraph of Article 49 of the Guide draws the limit in a single sentence: no extension is applied to devices covered by an EC certificate issued under IVDD by an authorized notified body whose document expiry date is before 26.05.2022.
The practical meaning of this subparagraph is this: starting application preparation without looking at the expiry date of the certificate you hold wastes the effort from the start. Before opening a file, we check the certificate expiry date, we do not prepare an application for devices that fall under this subparagraph, and we notify you of the reason in writing. For such devices the solution is not an extension but a new certification under IVDR.
Three scenarios for those with an IVDD certificate
The second paragraph of Article 49 of the Guide builds a structure divided into scenarios for IVDD certificates valid on and after 26.05.2022. Which scenario you are in depends on two things: the expiry date of your certificate and the stage your manufacturer has reached in its relationship with the IVDR notified body.
Document expiry date before 09.07.2024, the manufacturer applied and signed a contract before the document expiry date: Subparagraph (b) of the second paragraph applies. If the application is found appropriate, the validity date of the certificate is extended until 31.12.2027 under subparagraph (f) of the same paragraph.
Document expiry date before 09.07.2024, but the contract was signed after the document expiry date: Subparagraph (c) applies. In this scenario, in addition to the Manufacturer Declaration and the notified body document, either an authorization document obtained from the Agency under the first paragraph of Article 56 of IVDR or a document given by the Agency to the relevant economic operator under the first paragraph of Article 91 of IVDR is requested. If found appropriate, the period is again extended until 31.12.2027.
Document expiry date on or after 09.07.2024, the manufacturer has not yet applied to an IVDR notified body: Subparagraph (ç) applies. If found appropriate, under subparagraph (g) the validity date of the certificate is extended only until 26.05.2025.
In all three scenarios the application is made through EBS, by selecting "2023/KK-1 Numbered Extension Application" as the document type, and without physical documents. Determining the scenario is up to us; we read the certificate and the manufacturer's correspondence and notify you in writing, with reasons, which subparagraph you fall under.
Application and contract deadlines by class
The third paragraph of Article 49 of the Guide requires the manufacturer to meet two separate thresholds for the devices in the fourth paragraph of Article 48: the date of application to the notified body and the date of the contract with the notified body. Both vary by class, and both are in the manufacturer's hands, not yours.
Class D group: Application to the notified body before 26.05.2025, contract before 26.09.2025. Sub-items 1 of subparagraph (a) and of subparagraph (b) of the third paragraph.
Class C group: Application before 26.05.2026, contract before 26.09.2026. Sub-items 2 of the same subparagraphs.
Class B, Near-Patient Testing Device (Class B) and Class A Sterile: Application before 26.05.2027, contract before 26.09.2027. Sub-items 3 of the same subparagraphs.
If only the application condition is met, under subparagraph (c) of the third paragraph the registered period and the validity of the declaration of conformity are extended until 26.09.2025, 26.09.2026 or 26.09.2027 depending on class. If the contract condition is also met, under subparagraph (ç) the period is extended until 31.12.2027, 31.12.2028 or 31.12.2029 according to the risk class of the device. This distinction is where most companies slip: the application alone does not bring the longest period.
Surveillance responsibility document: the third document on the IVD side
While two documents are enough on the MDR side, a third document comes into play on the IVD side. Several subparagraphs of the second paragraph of Article 49 of the Guide require a document stating that responsibility for surveillance of the existing EC certificate has been taken over. This document is issued either by the notified body that issued the existing certificate or, if it assumes the responsibility, by the IVDR notified body with which a contract has been made.
In subparagraphs (b) and (e): a document covering the surveillance responsibility until 26.09.2025 is requested.
In subparagraph (ç): a document covering the surveillance responsibility until 26.05.2025 is requested.
In subparagraph (d): a document covering the surveillance responsibility until 26.09.2025 is requested.
A date commitment is also required in the Manufacturer Declaration itself. Subparagraph (ç) of the fourth paragraph of Article 49 of the Guide requires the declaration to include a statement that the manufacturer will put in place, or has put in place, a quality management system under Article 10(8) of IVDR no later than 26.05.2025. Subparagraph (f) of the same paragraph requires the contact details, including the e-mail address, of the manufacturer's company official who issued the declaration; the confirmation correspondence is carried out through this address.
The format in which the IVD portfolio is submitted
What speeds up the file is not the number of documents, but the list arriving in good order. We ask you to start with a single table; we determine the columns, and you or your manufacturer fill it in.
Barcode column: The same logic as the distinction in the fifth paragraph of Article 44 of the Guide applies. If the device is under the Old Medical Device Regulations, the Primary Product Number is written; if it is under the New Medical Device Regulations, the UDI-DI Number is written.
Product name and analyte/parameter column: For the classification rules to be applied, we need to know which analyte or parameter the device targets.
Claimed IVDR class column: It must be a value selected from the list in the fourth paragraph of Article 30 of the Guide. You can leave it blank; in that case we form the class opinion and give our reasoning in writing.
Self-test and near-patient test marking: Rule 4 in Article 57 of the Guide classifies devices intended for self-testing as Class C; those intended for pregnancy detection, fertility testing, cholesterol level, and detection of glucose, erythrocytes, leukocytes and bacteria in urine are placed in Class B. Near-patient testing devices are classified on their own. Mark the products that carry this claim separately in the list.
Document sets: For each product, the existing IVDD certificate and the declaration of conformity are required. The first paragraph of Article 12 of the Guide requires that, in declaration of conformity notifications of an importer belonging to the foreign manufacturer, the original of an apostilled, embassy or consulate of the Republic of Türkiye certified, e-Apostille certified or e-signed document be uploaded in full with its annexes to the "Original Document" field.
ÜTS after EBS: if the second stage is skipped, the device registration falls
The extension is in two stages and the second stage is not voluntary. The fifth paragraph of Article 48 of the Guide requires, in order to benefit from the transition provisions, that an application first be made through EBS under Article 49 and, if the application is found appropriate, that a registration or update application be made in ÜTS under Article 50 for the relevant devices and their linked documents.
Article 50 of the Guide states the consequence clearly: if no application is made in EBS under Article 49 and an extension is not permitted by the Agency, the document registration or update application to be made in ÜTS is rejected. That is, an application that breaks the sequence results negatively from the very beginning.
Regulation field: According to subparagraph (b) of the second paragraph of Article 50 of the Guide, the field "Is it a Document Registration/Update Notification Application under the 2023/KK-1 Numbered Announcement?" does not appear on the screen unless "98/79/EC" is selected in the "Regulation It Belongs To" field on the "Update Document" screen.
Evidence document field: Under subparagraph (c), when this field is set to "Yes", a mandatory field named "Evidence Documents for IVDR Transition Period Extension" opens. Subparagraph (ç) requires the Confirmation Letter of the Agency and all of the statements and documents in the application to be uploaded in a single file with all pages.
Document expiry date field: According to subparagraph (e), the original expiry date of the EC certificate is entered in this field. If the application is found appropriate, under subparagraph (ğ) the Agency updates the date to the date in the Confirmation Letter.
Application type: Under subparagraph (g), after the notification is created, "Document Application under the 2023/KK-1 Numbered Announcement" is selected as the "Application Type", and there is no need to physically send the uploaded documents to the Agency.
Fields specific to ÜTS in IVD device registration
In an IVDR registration, some fields arise not from the device itself but from the design of the system. Companies that do not know them come with the right document and get a negative result because of the wrong field.
Sales, Advertising and Promotion Regulation Annex-3 field: Subparagraph (u) of the seventh paragraph of Article 30 of the Guide determines that there is no device under IVDR in the current Annex-3 list, and states that, unless there is a change to this effect in Annex-3 of the relevant Regulation, it is mandatory for this field to be selected as "No".
"Is it a new type of device?" field: Subparagraph (ş) of the same paragraph states that a device will be considered a "new device" if it has not been continuously available in Türkiye or EU member countries for the last 3 years for the analytes or parameters concerned, or if it includes an analytical technology that has not been continuously used in the procedure for the last three years.
GMDN field: Under subparagraph (b) of the fourth paragraph of the same article, GMDN is a field that must be selected in ÜTS in an IVDR registration as well, and each device has only one GMDN code. GMDN and EMDN are two separate fields in ÜTS; both are required.
In which classes the instructions for use are mandatory on the IVD side
The obligation to upload instructions for use varies by class on the IVD side, and this is one of the first questions we ask in file preparation. Sending the instructions incomplete in a class where they are mandatory holds up the entire application.
Mandatory under IVDR: Under subparagraph (b) of the second paragraph of Article 17 of the Guide, uploading the instructions for use issued by the manufacturer is mandatory for Class B, Class C and Class D devices, Self-Testing Devices (Class B/C/D), Near-Patient Testing Devices (Class B/C/D) and Companion Diagnostic Devices (Class C/D).
Mandatory under IVDD: Under subparagraph (c) of the same paragraph, instructions for use are mandatory for devices under Annex-II List-A or Annex-II List-B and for Self-Testing Devices.
Optional: The third paragraph of the same article makes uploading instructions for use optional for Class A, Class A Sterile and IVD Other devices. However, the same paragraph states that the Agency may request instructions for use for devices in this risk group as well where it deems necessary.
There is one more case where the obligation does not apply: under the fourth paragraph of the same article, no obligation to upload instructions for use is sought for Class IIa devices for which "Yes" is selected in the "Is it a component/accessory of a device?" field in the device registration.
An IVD-specific error in certificate labelling
One of the most frequent reasons for rejection we encounter on the IVD side is the right certificate arriving with the wrong label. ÜTS holds the document not only by its number but by the triple of class, regulation and annex under which it is conducted. If this triple does not match the combination allowed for the class of the device, the document cannot be linked to the device.
Article 67 of the Guide defines a separate document business rule annex for each class under IVDR. Annex-97 applies to Self-Testing Device (Class B) and Near-Patient Testing Device (Class B), Annex-98 to Self-Testing Device (Class C/D), Annex-99 to Near-Patient Testing Device (Class C/D), and Annex-100 to Companion Diagnostic Device (Class C/D). In all four annexes, the expected document is the EU Certificate labelled "ANNEX IX CHAPTER II (INCLUDING ARTICLE 5) TECHNICAL DOCUMENTATION ASSESSMENT".
In contrast, for plain Class B and Class C devices in Annex-95 and for Class D devices in Annex-96, the expected label is of the form "EXCLUDING ARTICLE 5". The source of the distinction is Article 62 of the Guide: its fifth, sixth and eighth paragraphs state that for near-patient testing, self-testing and companion diagnostic devices the manufacturer will carry out the additional technical documentation assessment procedure. If the certificate you hold is labelled "EXCLUDING ARTICLE 5" and your device carries a self-test, near-patient test or companion diagnostic claim, this document is not the right document for that device.
Translation, apostille and physical submission
The translation and approval rules are not the same on the EBS and ÜTS sides; companies that do not know this difference either incur unnecessary costs or send an incomplete file.
Translation in EBS applications: The eleventh paragraph of Article 49 of the Guide requires that sworn-translator-approved Turkish translations of all documents whose original is in a foreign language and that are to be submitted to the Agency also be added to the application file.
Evidence documents uploaded to ÜTS: Subparagraph (d) of the second paragraph of Article 50 of the Guide states that there is no need for an apostille or notary approval for the documents uploaded to the "Evidence Documents for IVDR Transition Period Extension" field. However, Turkish translations of documents whose original is in a foreign language must be approved by a sworn translator.
One document per application: The seventeenth paragraph of Article 49 of the Guide states that in an application to be made under this scope, a transition period can be requested for only one document. If more than one document appears in the petition, the application is not taken into evaluation and ends negatively.
Petition requirement: The seventh paragraph of the same article states that applications without a petition are not accepted, and that the petition must declare the document for which a transition period is requested and indicate any devices that are left out of scope.
We set up the plan for splitting applications according to the number of your documents, and give you in advance in writing which document will go with which petition.
Scope of service: responsibilities of the consultant and the company
Our way of working is simple. From your own company page in ÜTS you define our staff as users of the "Medical Device Registration Officer" type; under the eleventh paragraph of Article 39 and subparagraph (f) of the fifth paragraph of Article 40 of the Guide, only users of this type can perform the "Approve" action in document and device applications. We prepare the class opinion, the scenario determination, the EBS petition, the evidence document set and the ÜTS notifications.
There are two things you do not hand over. The first is the e-signature: under the thirteenth paragraph of Article 39 and subparagraph (g) of the fifth paragraph of Article 40 of the Guide, an application in the "Awaiting Signature" status is signed electronically by the company's authorized signatory. The second is data accuracy: the fourth paragraph of Article 31 of the Medical Device Sales, Advertising and Promotion Regulation keeps the accuracy of the information about devices registered in the system under the responsibility of the sales center that makes the product registration. We fill in the fields correctly and show their source; the accuracy of the information coming from your manufacturer stays with your company. We do not give a commitment on when and how an application will conclude; the Guide also contains no such period or result commitment in any of its articles.