Periodic Inspection of Weighing Instruments: Law No. 3516, Stamp, Penalties, Medical Device Measuring Function and ÜTS Calibration Fields

SUMMARY
The periodic inspection obligation for weighing instruments is based on Law No. 3516 on Measures and Adjustment. The Law covers the inspection, adjustment and stamping of weight measures. A weighing instrument placed on the market as a medical device is also subject to the Medical Device Regulation and ÜTS registration; the obligations under the two sets of legislation are fulfilled separately.
In this article we give, with verbatim quotations (unofficial translation), the inspection, stamp and penalty provisions of the measures and adjustment legislation, the provision on medical devices with a measuring function, the TİTCK regulation on testing, control and calibration, and the calibration fields on the ÜTS registration screen.
For ÜTS document and product registration of your devices with a measuring function: Medical Device ÜTS Registration Consultancy
Contents
1. Scope of Law No. 3516
2. Types of Inspection: Article 9
3. Periodic Inspection Application: Regulation on Inspection of Measures and Measuring Instruments
4. Stamp and Prohibitions
5. Administrative Fines: Article 15
6. Measuring Function in a Weighing Instrument That Is a Medical Device
7. Testing, Control and Calibration Regulation
8. Calibration Fields in the ÜTS Record
9. Frequently Asked Questions
1. Scope of Law No. 3516
Article 2 of the Law (unofficial translation):
“The inspection, adjustment and stamping of length, area, volume and weight measures, areometers, grain testing instruments, electricity, water, town gas, natural gas and fuel meters, taximeters, naklimeters, current and voltage measuring transformers, and railway freight and tank wagons shall be carried out in accordance with the provisions of this Law.”
Article 3 excludes some measuring instruments from the scope. The subparagraph relating to weighing instruments:
“f) Dynamometric weighing instruments manufactured exclusively for use in bathrooms and kitchens,”
Definitions in the Law (Article 4):
“Measuring instrument: The tools, meters and equipment used in the measuring process,”
“Stamp: The general adjustment mark placed on measures and measuring instruments found to be compliant at the end of inspection,”
“Dynamometric weighing instrument: A spring weighing instrument,”
2. Types of Inspection: Article 9
“Measures and measuring instruments are subject to the following inspections.”
“a) Initial inspection: Applied to newly made measures and measuring instruments, or those formed by assembling parts, before they are put on sale or into use; to imported measures and measuring instruments at the time of their entry into the country; after the repair and adjustment of measures and measuring instruments whose stamps have been cancelled at the end of periodic, sudden, complaint and stock inspections; or, for measures and measuring instruments whose adjustment depends on their location, upon every change of location.”
“b) Periodic inspection: A general inspection carried out at set intervals for measures and measuring instruments within the scope of this Law.”
“c) Sudden inspection: An inspection carried out without notice, where they are located, by officials of the Ministry's central and provincial measures and adjustment organizations on their own initiative or upon a report.”
“e) Stock inspection: An inspection to which measures and measuring instruments that bear the initial inspection stamp but have not been sold, or that are kept in stock in warehouses, workshops, production or sales premises because they are not yet needed for use, are subjected again within the periodic inspection periods.”
The Law leaves the procedures and principles of the inspections to a regulation:
“The procedures and principles of these inspections, by whom and in what manner they shall be carried out, the duties and responsibilities of the owners or users of measures and measuring instruments subject to inspection, and the prohibitions they are subject to shall be set out in a regulation to be issued by the Ministry.”
3. Periodic Inspection Application: Regulation on Inspection of Measures and Measuring Instruments
Article 6 of the Regulation sets the general rule for periodic inspection applications:
“In calculating the periodic inspection period, the year of stamping is taken as the basis, and whatever date in the year the stamp was applied, the year of stamping is counted as the first year.”
“(Amended paragraph: OG-27/2/2019-30699) Those concerned must apply to the Ministry's Measures and Adjustment Organization or the Group Center Municipality Measures and Adjustment Offices for periodic inspection of measures and measuring instruments by the last day of February of the year following the year in which the stamp period expires. Applications are made through the Ministry's website or the e-government system.”
“Administrative and criminal action is taken under Law No. 3516 on Measures and Adjustment, for the act of using a measuring instrument whose stamp period has expired, against those who do not apply on time for periodic inspection, even if they apply later, where the instrument is found to be in use.”
Information to be included in the application and the initial inspection condition:
“The application must include the type, brand, number and capacity information of the measure and measuring instrument.”
“Article 12 - Periodic inspection is not carried out for measures and measuring instruments whose initial inspection has not been carried out.”
The periodic inspection periods are given in Articles 9 and 11 of the Regulation; these articles contain no period specific to weighing instruments. Articles 196-216 titled "Weighing Instruments" and Article 224 of the Regulation were repealed by the amendment dated 12/12/2013. For instruments within the scope of regulations harmonized from European Union legislation, Article 22/A provides:
“For measures and measuring instruments within the scope of regulations harmonized from European Union legislation, the provisions of this Regulation apply on matters not specified in the relevant technical regulation.”
4. Stamp and Prohibitions
Article 13 of the Law:
“Measures and measuring instruments found to be accurate at initial, periodic and stock inspections are stamped if they are suitable for stamping. For those not suitable for stamping, a document is issued in place of the stamp.”
Of the prohibitions in Article 14 of the Law, those relevant to weighing instruments:
“b) Offering for sale or selling unstamped measures and measuring instruments, mediating their sale, using them in trade, or keeping them in places of trade,”
“d) Using, offering for sale or selling measures and measuring instruments whose stamp is broken or damaged, for which no application was made at the time of periodic inspection, whose stamp period has passed or whose stamps have been cancelled; keeping them in places of trade; or delivering to their owners, after repair, measures and measuring instruments whose stamps have been cancelled without having them re-inspected and stamped,”
“e) Having the inspection, repair or adjustment of measures and measuring instruments within the scope of this Law carried out by persons who do not hold the Ministry's authorization certificate,”
5. Administrative Fines: Article 15
The amounts below are those in the text of the Law:
“c) A person who uses a measuring instrument whose stamp is broken or damaged or whose stamp period has passed shall be given an administrative fine of five hundred Turkish Liras to ten thousand Turkish Liras, depending on the type of the instrument and the nature of the work in which it is used. In addition, these instruments shall be confiscated and a decision shall be made to transfer their ownership to the public.”
“j) (Added: 14/2/2019-7164/25 art.) A person who sells, offers for sale or uses in commercial relations unstamped measures and measuring instruments shall be given an administrative fine of one thousand Turkish Liras to ten thousand Turkish Liras. In addition, these instruments shall be confiscated and a decision shall be made to transfer their ownership to the public.”
“These administrative sanctions are decided by the Provincial Director of the Ministry of Industry and Trade.”
The annual increase of administrative fines is regulated in Article 17/7 of the Misdemeanors Law No. 5326:
“(7) Administrative fines are applied for each calendar year, effective from the beginning of that year, increased by the revaluation rate determined and announced for that year under repeated Article 298 of the Tax Procedure Law No. 213 dated 4.1.1961.”
For the penalties under the medical device legislation, see our article on medical device administrative fines.
6. Measuring Function in a Weighing Instrument That Is a Medical Device
Article 52/6 of the Medical Device Regulation provides a notified body procedure for Class I devices with a measuring function:
“If these devices are placed on the market in sterile condition, have a measuring function or are reusable surgical instruments, the manufacturer applies the procedures specified in Chapters I and III of Annex IX or in Part A of Annex XI. However, the involvement of the notified body in these procedures is limited to:”
“b) in the case of devices with a measuring function, aspects relating to the conformity of the devices with metrological requirements,”
General safety and performance requirements of Annex I:
“15.1. Diagnostic devices and devices with a measuring function shall be designed and manufactured in such a way as to provide sufficient accuracy, precision and stability for their intended purpose, based on appropriate scientific and technical methods. The limits of accuracy shall be indicated by the manufacturer.”
For classification, see our page on medical device classification rules and, for registration by class, our page on ÜTS registration by device class.
7. Testing, Control and Calibration Regulation
The TİTCK Regulation on Testing, Control and Calibration of Medical Devices is a separate regulation from the measures and adjustment legislation. Its scope (Article 2):
“(1) This Regulation covers the procedures and principles on the application, authorization, monitoring and inspection of organizations that will perform the testing, control and calibration of medical devices requiring testing, control and calibration during their use, and on the personnel to be employed in these organizations, their qualifications and their training.”
“(2) The testing, control and calibration activities carried out by manufacturers, importers, technical services and users on medical devices are outside the scope of this Regulation.”
A device within the scope of the Regulation is a device listed under the authorization groups in the Agency's guidance (Article 4/ç). Among the authorization groups in Annex-1 is the group "Flow, Weight, Length, Volume, Temperature, Pressure, Speed". The period and the obligation to have testing performed:
“(2) If no period is specified in national and international protocols, guidelines and standards on the periods of testing, control and calibration of devices, or in the manufacturer's criteria, testing, control and calibration are carried out at least once a year.”
“(2) Healthcare service providers are required to have the testing, control and calibration services of devices performed by the organizations announced under the first paragraph of this article;”
A label is affixed to a device that has passed testing (Article 12/2):
“(2) A green label stating that the device has passed calibration is affixed to the device that has passed testing, control and calibration, in a legible manner.”
8. Calibration Fields in the ÜTS Record
According to Article 27 of TCKKD-KLVZ-03, the record of a device within the scope of the MDR contains calibration fields under the "Properties" heading:
“n) Subject to calibration?:”
“1) This field is mandatory to select in ÜTS.”
“2) This field is selected as "Yes" if testing, control and calibration activities must be applied to the device concerned, and "No" if they need not be applied.”
“o) Calibration Period (months):”
“2) The calibration period specified in the device's instructions for use is entered in this field, in "Months".”
“3) If the calibration period of a device has been determined by the manufacturer, it is mandatory that this information be in the instructions for use and entered in this field.”
“4) If the calibration period of a device has not been specifically determined by the manufacturer and is not stated in the instructions, the value 12 (12 'Months') is entered in this field.”
“5) Calibration period information is not sought on the label or sales packaging.”
For other mandatory fields in the ÜTS record, see our pages on ÜTS device registration mandatory fields and ÜTS product registration.
9. Frequently Asked Questions
Are weighing instruments within the scope of Law No. 3516?
Article 2 of the Law covers the inspection, adjustment and stamping of weight measures. Subparagraph (f) of Article 3 excludes only dynamometric weighing instruments manufactured for use in bathrooms and kitchens.
When is the application for periodic inspection made?
Under Article 6 of the Regulation on Inspection of Measures and Measuring Instruments, the application is made through the Ministry's website or e-government by the last day of February of the year following the year in which the stamp period expires.
What happens if a weighing instrument with an expired stamp is used?
Article 14/d of the Law prohibits it. Under Article 15/c, an administrative fine from five hundred Turkish Liras to ten thousand Turkish Liras, at the amounts in the Law's text, is imposed and the instrument is confiscated.
How is the calibration period entered in ÜTS?
Under Article 27 of TCKKD-KLVZ-03, the period in the instructions for use is entered in months. If the manufacturer has not determined one and the instructions do not contain it, the value 12 is entered.
Is a notified body required for a Class I device with a measuring function?
Under Article 52/6 of the Medical Device Regulation, the manufacturer applies the procedures of Annex IX or Annex XI Part A; the notified body's role is limited to conformity with metrological requirements.
For ÜTS document and product registration of your devices with a measuring function: Medical Device ÜTS Registration Consultancy
Sources