EUDAMED Actor Registration: Who Registers, How to Get an SRN, Steps and LAA

SUMMARY
EUDAMED actor registration means that the manufacturer, authorised representative and importer enter their information into the electronic system established by the European Commission and, after verification, receive a single registration number (SRN). The basis is Articles 30 and 31 of the Medical Device Regulation and Annex VI Part A.
In this article we explain who registers, when registration becomes mandatory, how the SRN is issued and the steps of the registration form, with quotations from the regulation, the MDCG Q&A document and TİTCK's Turkish EUDAMED user guide. Device registration is covered in a separate article.
For your EUDAMED actor registration and ÜTS registration: Medical Device ÜTS Registration Consultancy
Contents
1. What Is EUDAMED Actor Registration?
EUDAMED consists of six modules: actors, UDI/devices, notified bodies and certificates, clinical investigations and performance studies, vigilance and post-market surveillance, and market surveillance. Actor registration is done in the first module. The purpose of the system is defined in Article 30 of the Medical Device Regulation:
“ARTICLE 30 -(1) The details of the information to be entered by economic operators into the electronic system established and managed by the Commission to collect and process the information necessary and proportionate to generate the single registration number (SRN) referred to in the second paragraph of Article 31 and to identify the manufacturer and, where applicable, the authorised representative and the importer, are set out in point 1 of Annex VI Part A.”
For a general description of the modules, see our page on What is EUDAMED.
2. Who Registers as an Actor?
The registration obligation is in the first paragraph of Article 31:
“ARTICLE 31 -(1) Before placing a device other than a custom-made device on the market, manufacturers, authorised representatives and importers that are not already registered in accordance with this Article shall enter the information referred to in point 1 of Annex VI Part A into the electronic system referred to in Article 30 in order to register.”
According to the same paragraph, where the involvement of a notified body is required in the conformity assessment, this information is entered before applying to the notified body. The MDCG Q&A document (Q4) states that manufacturers of systems or procedure packs, manufacturers supplying only custom-made devices and clinical investigation sponsors also register in the actor module and receive an Actor ID. For distributors (Q5):
“Distributors are not required to register in EUDAMED.”
3. When Is Actor Registration Mandatory?
The EUDAMED modules enter into operation gradually under Article 34 of the MDR as amended by Regulation (EU) 2024/1860. For the actor module, the MDCG Q&A document (Q3) states:
“Registration in the ACT module under MDR Article 31 and IVDR Article 28 will become mandatory 6 months after the publication in the OJEU of the notice confirming its functionality.”
According to the answer to Q6 of the same document, manufacturers, importers, authorised representatives and system and procedure pack producers have been able to register in the actor module since December 2020.
4. What Is the SRN and How Is It Issued?
The TİTCK ÜTS Registration Guideline defines the SRN as follows:
“rr) Single Registration Number (SRN): the single registration number generated by EUDAMED to be allocated to the manufacturer, authorised representative or importer,”
The Authority conveys the number after verifying the data (Article 31):
“(2) After verifying the data entered in accordance with the first paragraph, the Authority conveys the single registration number obtained from the electronic system referred to in Article 30 to the manufacturer, authorised representative or importer.”
“(3) The manufacturer uses the single registration number when applying to a notified body for conformity assessment and when accessing EUDAMED to fulfil its obligations under Article 29.”
The SRN is also among the information requested when an EU declaration of conformity is registered in ÜTS (Guideline, Article 11):
“c) Single Registration Number (SRN) of the manufacturer, if already issued (entered in the "Münferit Kimlik No" field in ÜTS),”
5. What Information Is Entered in the Registration?
Point 1 of Annex VI Part A lists the information on the economic operator:
“1.1. Type of economic operator (manufacturer, authorised representative or importer),”
“1.2. Name, address and contact details of the economic operator,”
“1.3. Where the information is submitted by another person on behalf of any of the economic operators referred to in point 1.1 of this Part (Amended phrase: RG-2/4/2023-32151), the name, address and contact details of that person,”
“1.4. Name, address and contact details of the person or persons responsible for regulatory compliance referred to in Article 15.”
In the "Actor identity" step of the user guide, the following identifiers are requested:
“5. Enter all information needed to identify your economic operator: VAT, EORI and national trade number.”
For the person responsible for regulatory compliance, see our page on PRRC: Medical Device Regulation Article 15.
6. Actor Registration Step by Step
According to TİTCK's Turkish EUDAMED user guide (except for non-EU manufacturers), the flow is as follows:
Log in to EUDAMED with an EU Login account linked to the business e-mail address.
On the "User and Actor Registration" page, select ACTOR REGISTRATION and read and confirm the disclaimer.
Select the role and country and enter the organisation name.
Enter the actor identity (VAT, EORI, national trade number), actor address and contact details.
Enter the details of the regulatory person (person responsible for regulatory compliance) and the areas they are responsible for.
Confirm that you will be the Local Actor Administrator; the signed declaration is uploaded in PDF format only.
Select the competent authority that will approve the registration request and, after the preview, submit the request with "Submit actor registration".
After submission, the request is recorded in the "Pending requests" list with the status "Submitted" and a unique application ID is given. The general section of the guide contains the following sentence on the SRN:
“A Single Registration Number (SRN) is generated by EUDAMED and allocated by the competent authority.”
7. Local Actor Administrator (LAA)
The role of the person who makes the registration is described in the guide as follows:
“The person who first enters an actor's information in EUDAMED automatically becomes the LAA of that actor once the registration is accepted.”
“It is good practice for each actor to have at least 2 LAAs (and at least one of them should not belong to a subcontractor).”
The profiles are hierarchical: the LAA is the highest profile; below it are the Local User Administrator (LUA) and the Viewer. The authorised representative has a "Verifier" profile, the non-EU manufacturer a "Mandate administrator" profile and the importer a "Binder" profile. Access requests of users who later join a registered actor are approved by the LAA or LUA.
8. Registration of Non-EU Manufacturers
Article 11 of the Medical Device Regulation:
“ARTICLE 11 -(1) Where the manufacturer is not established in Türkiye or in an EU member state, the device may be placed on the market only if the manufacturer appoints an authorised representative.”
According to the user guide, a non-EU manufacturer identifies, during registration, the authorised representative holding the mandate; where there is more than one representative, the main one is indicated with the SRN. The request first goes to the authorised representative for verification and then to the competent authority for approval:
“Your authorised representative will verify your registration details before the request is sent to the competent authority for approval. For this, it must have registered in EUDAMED beforehand.”
For the tasks of the authorised representative, see our page on authorised representative tasks: Article 11.
9. Multiple Roles, Updates and Rejection
If an organisation takes on more than one role, the guide says:
“you must make a separate actor registration for each of them.”
For changes after registration, Article 31:
“(4) Economic operators shall update the data in the electronic system referred to in Article 30 within 1 week of any change to the information referred to in the first paragraph.”
“(5) Economic operators shall confirm the accuracy of the data within 1 year of submitting the information under the first paragraph and every 2 years thereafter.”
“(7) The data entered into the electronic system referred to in Article 30 under the first paragraph shall be publicly available.”
According to the guide, after the SRN is received only the Local Actor Administrator can change the registration details. A request that is not approved remains on the "My pending requests" page; a request with the status "Correction requested" is edited and resubmitted, while a request with the status "Rejected" cannot be edited or resubmitted.
Entering devices into the UDI/device module after actor registration is a separate task; the details are in our article on EUDAMED device registration. For these tasks, see our page on EUDAMED actor and device registration service.
For the registration steps on the Turkish side, see our ÜTS registration guide and, for having your EUDAMED and ÜTS registrations carried out on your behalf, our page on ÜTS consultancy.
10. Frequently Asked Questions
Who makes the EUDAMED actor registration?
According to Article 31 of the Medical Device Regulation, manufacturers, authorised representatives and importers. According to the MDCG Q&A document, system or procedure pack producers also register and receive an Actor ID; distributors do not need to register.
Who issues the SRN?
EUDAMED generates the SRN; after the competent authority verifies the data entered, it conveys it to the manufacturer, authorised representative or importer (Article 31, second paragraph).
I am both a manufacturer and an importer; is one registration enough?
No. According to the EUDAMED user guide, if an organisation has more than one role, a separate actor registration is made for each role.
How is a non-EU manufacturer registration approved?
The request is first verified by the manufacturer's authorised representative and then goes to the competent authority for approval. The authorised representative must already be registered in EUDAMED.
What happens if the actor information changes?
According to the fourth paragraph of Article 31, the data in EUDAMED is updated within 1 week of the change. In addition, the information is confirmed within 1 year of first submission and every 2 years thereafter.
For your EUDAMED actor registration and ÜTS transactions: Medical Device ÜTS Registration Consultancy