Distance Sales Rule for Medical Devices: Information Society Services, Remote Diagnosis and Treatment Services, Obligation to Provide the EU Declaration of Conformity

SUMMARY
The distance sales rule for medical devices is set out in Article 6 of the Tıbbi Cihaz Yönetmeliği (Medical Device Regulation), titled “Uzaktan/Mesafeli satışlar” (Distance Sales). The Article brings within the scope of the regulation a device offered through information society services and a device used to provide a remote diagnostic or therapeutic service without being placed on the market.
In this article we give the four paragraphs of Article 6, Article 7 of the In Vitro Tanı Amaçlı Tıbbi Cihaz Yönetmeliği (In Vitro Diagnostic Medical Device Regulation), which is its counterpart, and the text of Article 6 of EU Regulation 2017/745, with word-for-word quotations.
For the ÜTS registrations of your devices offered at a distance: Medical Device ÜTS Registration Consultancy
Contents
1. Article 6/1: A Device Offered through Information Society Services
“ARTICLE 6 - (1) A device offered by means of information society services to a natural or legal person shall comply with the provisions of this Regulation.”
The provision says that all provisions of the regulation apply where a device is offered through information society services. The addressee is a natural or legal person.
2. Article 6/2: A Service Provided with a Device Not Placed on the Market
“(2) A device not placed on the market but used in the context of a commercial activity, whether in return for payment or free of charge, to provide a diagnostic or therapeutic service offered by means of information society services or by other means of communication, directly or through intermediaries, to a natural or legal person, shall comply with the provisions of this Regulation.”
The elements of the paragraph are listed in the text as follows:
The device has not been placed on the market.
The service is offered within a commercial activity.
The service may be in return for payment or free of charge.
The service is offered through information society services or other means of communication.
The service is offered directly or through intermediaries.
The device is used to provide a diagnostic or therapeutic service.
The concepts of placing on the market and making available on the market are defined in Article 3 of the regulation:
“(ss) Making available on the market: any supply of a device, other than an investigational device, for distribution, consumption or use on the Turkish market in the course of a commercial activity, whether in return for payment or free of charge,”
“(tt) Placing on the market: the first making available of a device, other than an investigational device, on the Turkish market,”
For the details of the concepts, see our pages on what is making available on the market and what is placing on the market.
3. Article 6/3: Providing a Copy of the EU Declaration of Conformity
“(3) Any natural or legal person who offers a device in accordance with the first paragraph or provides a service in accordance with the second paragraph shall, upon request by the Agency, make available a copy of the EU declaration of conformity of the device concerned.”
The obligation falls on both the person who offers the device and the person who provides the service, and arises upon the Agency's request. What the EU declaration of conformity means is set out in Article 19 of the regulation:
“ARTICLE 19 - (1) The EU declaration of conformity shall state that the requirements specified in this Regulation have been fulfilled in relation to the device covered. The manufacturer shall keep the EU declaration of conformity up to date. The EU declaration of conformity shall contain, as a minimum, the information set out in Annex IV, and, for devices placed on the market in Türkiye, where the declaration is drawn up in other languages, its Turkish translation shall be attached.”
For the content of the declaration, see our article on EU declaration of conformity and the CE marking.
4. Article 6/4: Request to Terminate the Activity
“(4) The Agency may, on grounds of protection of public health, require an information society services provider to cease its activity.”
The addressee of the paragraph is the information society services provider; the ground is the protection of public health.
5. Article 7 of the IVD Regulation
The same provision is in Article 7 of the In Vitro Tanı Amaçlı Tıbbi Cihaz Yönetmeliği (In Vitro Diagnostic Medical Device Regulation):
“ARTICLE 7 - (1) A device offered by means of information society services to a natural or legal person shall comply with the provisions of this Regulation.”
“(2) A device not placed on the market but used in the context of a commercial activity, whether in return for payment or free of charge, to provide a diagnostic or therapeutic service offered by means of information society services or by other means of communication, directly or through intermediaries, to a natural or legal person, shall comply with the provisions of this Regulation.”
The definition of self-test device in Article 3 of the IVD Regulation also refers to information society services:
“(bb) Self-test device: a device intended by the manufacturer to be able to be used by lay persons, including devices used for testing services offered to lay persons by means of information society services,”
6. Article 6 of EU Regulation 2017/745
The counterpart of Article 6 of the Tıbbi Cihaz Yönetmeliği in the EU text is Article 6, titled “Distance sales”:
“1. A device offered by means of information society services, as defined in point (b) of Article 1(1) of Directive (EU) 2015/1535, to a natural or legal person established in the Union shall comply with this Regulation.”
“3. Upon request by a competent authority, any natural or legal person offering a device in accordance with paragraph 1 or providing a service in accordance with paragraph 2 shall make available a copy of the EU declaration of conformity of the device concerned.”
“4. A Member State may, on grounds of protection of public health, require a provider of information society services, as defined in point (b) of Article 1(1) of Directive (EU) 2015/1535, to cease its activity.”
The EU text defines information society services by reference to Directive (EU) 2015/1535. In the EU text, the second paragraph begins with the wording “Without prejudice to national law regarding the exercise of the medical profession”.
7. Internet Sales in Türkiye Are Regulated Separately
Who may sell medical devices over the internet in Türkiye is regulated separately in Article 26 of the Tıbbi Cihaz Satış, Reklam ve Tanıtım Yönetmeliği (Medical Device Sales, Advertising and Promotion Regulation); the details are in our article on medical device internet sales.
For the registration of devices in ÜTS, see our article on ÜTS product registration and our page on ÜTS product registration service.
8. Frequently Asked Questions
Which article of the Tıbbi Cihaz Yönetmeliği contains the distance sales rule?
It is in Article 6, titled “Uzaktan/Mesafeli satışlar” (Distance Sales). In the IVD Regulation the same title is in Article 7.
I provide a remote diagnostic service without selling the device; does the regulation apply?
Under Article 6/2, a device that is not placed on the market but is used to provide a diagnostic or therapeutic service offered within a commercial activity must comply with the provisions of the regulation.
Does it change if the service is provided free of charge?
No. Article 6/2 uses the wording “in return for payment or free of charge”.
Which document can the Agency request?
Under Article 6/3, a copy of the EU declaration of conformity of the device concerned is provided upon the Agency's request.
Can the Agency stop the activity?
Under Article 6/4, the Agency may, on grounds of protection of public health, require an information society services provider to cease its activities.
For the ÜTS registrations of your devices offered at a distance: Medical Device ÜTS Registration Consultancy
Sources