Device Manufactured within a Healthcare Institution: Article 5(5) of the Tıbbi Cihaz Yönetmeliği, Conditions and ÜTS

SUMMARY
Under the Tıbbi Cihaz Yönetmeliği (Medical Devices Regulation), a device manufactured within a healthcare institution is considered to have been put into service. If all of the conditions in the fifth paragraph of Article 5 are met, the requirements of the regulation do not apply to these devices, except for the general safety and performance requirements in Annex I.
In this article we give the fourth, fifth and sixth paragraphs of Art. 5, the definition of healthcare institution, the differences in the IVD Regulation and the ÜTS provision as direct quotations.
For the ÜTS registration of your medical device: Medical Device ÜTS Registration Consultancy
Contents
1. Definition of Healthcare Institution
Article 3, point (vv) of the Tıbbi Cihaz Yönetmeliği:
“vv) Healthcare institution: an organisation the primary purpose of which is the care or treatment of patients or the promotion of public health,”
The definition looks at the primary purpose of the organisation. A short explanation of the concept is on our page what is a healthcare institution.
2. Article 5(4): Considered Put into Service
“(4) Devices manufactured and used within healthcare institutions shall be considered as having been put into service.”
The definition of putting into service is in Art. 3, point (r):
“r) Putting into service: the stage at which a device, other than an investigational device, has been made available to the final user as being ready for use on the Turkish market for the first time for its intended purpose,”
Even if the device is not placed on the market, it is considered put into service when it is used within the institution. For the concept, see our page on what is putting into service.
3. Article 5(5): Conditions of the Exemption
“(5) With the exception of the relevant general safety and performance requirements set out in Annex I, the requirements of this Regulation shall not apply to devices manufactured and used only within healthcare institutions, provided that all of the following conditions are met:”
“a) The devices are not transferred to another legal entity.”
“b) The manufacture and use of the devices occur under appropriate quality management systems.”
“c) The healthcare institution justifies in its documentation that the target patient group's specific needs cannot be met, or cannot be met at the appropriate level of performance, by an equivalent device available on the market.”
“ç) The healthcare institution provides information to the Agency, upon request, on the use of such devices, which includes a justification of their manufacture, modification and use.”
“e) The healthcare institution draws up documentation that makes it possible to have an understanding of the manufacturing facility, the manufacturing process, and the design and performance data of the devices, including the intended purpose, in sufficient detail to enable the Agency to ascertain that the general safety and performance requirements set out in Annex I are met.”
“f) The healthcare institution takes all necessary measures to ensure that all devices are manufactured in accordance with the documentation referred to in point (e).”
“g) The healthcare institution reviews experience gained from clinical use of the devices and takes all necessary corrective actions.”
The paragraph says "provided that all of the following conditions are met": if one condition is not met, the exemption does not apply. The exemption does not cover Annex I; the general safety and performance requirements apply to these devices as well. Point (d) is given under a separate heading below.
4. Public Declaration: Point (d)
“d) The healthcare institution draws up a public declaration, which shall include:”
“1) the name and address of the manufacturing healthcare institution,”
“2) the details necessary to identify the devices,”
“3) a declaration that the devices meet the general safety and performance requirements set out in Annex I and, where applicable, information on which requirements are not fully met, with a reasoned justification.”
The declaration is public and contains these three items. It is not the same document as the EU declaration of conformity that the manufacturer draws up under Art. 19. For the EU declaration of conformity, see our article on the EU declaration of conformity and CE marking.
5. Article 5(6): The Agency's Power and Industrial Scale
“(6) The Agency may request from healthcare institutions further relevant information about such devices manufactured and used in Türkiye. The Agency has the right to inspect the activities of healthcare institutions and to restrict the manufacture and use of any specific type of such devices under this Article. The fifth paragraph and this paragraph shall not apply to devices manufactured on an industrial scale.”
The three elements of the paragraph:
The Agency may request further information about these devices from healthcare institutions.
The Agency may inspect the activities of healthcare institutions and restrict the manufacture and use of a specific type of device.
The fifth and sixth paragraphs do not apply to devices manufactured on an industrial scale.
6. Reprocessing of a Single-Use Device in a Healthcare Institution
The third paragraph of Art. 17 refers to the conditions of Art. 5(5) for single-use devices reprocessed within a healthcare institution:
“1) The safety and performance of the reprocessed device are equivalent to those of the original device, and the requirements in points (a), (b), (ç), (d), (e), (f) and (g) of the fifth paragraph of Article 5 are complied with.”
The other conditions of reprocessing are in our article on reprocessing of a single-use device.
7. Differences in the IVD Regulation
In the In Vitro Tanı Amaçlı Tıbbi Cihaz Yönetmeliği (In Vitro Diagnostic Medical Devices Regulation), the same provision is in Art. 6. The fourth paragraph excludes devices intended for performance studies:
“(4) Devices manufactured and used within healthcare institutions, other than devices intended for performance studies, shall be considered as having been put into service.”
The fifth paragraph contains a separate condition concerning the laboratory:
“c) The laboratory of the healthcare institution complies with the relevant legislation, with the EN ISO 15189 standard or, where applicable, with the accreditation provisions determined by the Turkish Accreditation Agency.”
The detailed documentation condition is written for class D devices in the IVD regulation:
“f) In respect of class D devices in accordance with the rules set out in Annex VIII, the healthcare institution draws up documentation that makes it possible to have an understanding of the manufacturing facility, the manufacturing process, and the design and performance data of the devices, including the intended purpose, in sufficient detail to enable the Agency to ascertain that the general safety and performance requirements set out in Annex I are met. The Agency may also apply these provisions to class A, B or C devices in accordance with the rules set out in Annex VIII.”
The fifteenth paragraph of Provisional Article 1 of the IVD regulation sets dates for certain conditions:
“a) those specified in points (b), (c) and (d) to (ğ) of the fifth paragraph of Article 6, until 26/5/2024 (inclusive),”
“b) (Amended: Official Gazette 17/8/2024-32635)(5) those specified in point (ç) of the fifth paragraph of Article 6, until 31/12/2030,”
In the IVD regulation, point (ç) is the justification that the need cannot be met by an equivalent device on the market. This condition is not required until 31/12/2030.
8. The Position under ÜTS and the Sales Regulation
Article 34 of the Tıbbi Cihaz Yönetmeliği also mentions healthcare institutions when listing the ÜTS obligations:
“(1) Without prejudice to the obligations relating to EUDAMED laid down in this Regulation, economic operators, notified bodies, healthcare institutions, health professionals and other relevant persons established in Türkiye shall fulfil their obligations relating to the national product tracking system for medical devices (ÜTS).”
“(2) Those obligations relating to ÜTS shall be determined by the regulations published by the Agency.”
The ÜTS Kayıt Kılavuzu (ÜTS Registration Guide) TCKKD-KLVZ-03 mentions these devices separately in the conformity assessment under the IVDR (Art. 62):
“(3) Before putting into service a device that is not placed on the market, other than devices manufactured within healthcare institutions in accordance with the fifth paragraph of Article 6 of the IVDR, manufacturers shall carry out the assessment of the conformity of that device in accordance with the relevant conformity assessment procedures set out in IVDR Annex IX to IVDR Annex XI.”
The third paragraph of Art. 2 of the Tıbbi Cihaz Satış, Reklam ve Tanıtım Yönetmeliği (Medical Device Sales, Advertising and Promotion Regulation) excludes production within public healthcare institutions from its scope:
“(3) This Regulation does not cover devices manufactured and used within public healthcare institutions and organisations, provided that they comply with the other relevant technical regulations.”
For a device that does not meet the exemption conditions, the conformity assessment, declaration and ÜTS registration follow the general provisions. For the ÜTS process, see our article on ÜTS product registration.
9. Difference from a Custom-Made Device
The definition of a custom-made device is in Art. 3, point (s); the procedure to be followed is in Art. 52(7) and Annex XIII. The details are in our article on the custom-made device and Annex XIII.
10. Frequently Asked Questions
Does Annex I apply to a device manufactured within a healthcare institution?
Yes. Art. 5(5) disapplies the requirements of the regulation "with the exception of the relevant general safety and performance requirements set out in Annex I"; Annex I remains valid.
Can a device manufactured in the institution be given to another hospital?
Point (a) requires that the devices are not transferred to another legal entity. If this condition is not met, the exemption does not apply.
What does the public declaration contain?
The name and address of the manufacturing healthcare institution, the details necessary to identify the devices, a declaration that the requirements of Annex I are met, and the justification for any requirements that are not fully met.
Does the exemption apply to industrial-scale production?
No. Under the last sentence of Art. 5(6), the fifth and sixth paragraphs do not apply to devices manufactured on an industrial scale.
Can the Agency restrict the production of these devices?
Yes. Art. 5(6) gives the Agency the right to inspect the activities of healthcare institutions and to restrict the manufacture and use of a specific type of these devices.