How Is the Class of a Medical Device Determined? Medical Device Classification and ÜTS
CLASSIFICATION RULES
Current guide: For a rule-by-rule, up-to-date explanation of the MDR and IVDR classification rules, see our guide Medical Device Classification Rules: MDR, IVDR and ÜTS Class. For your device’s classification decision, you can review our medical device classification consultancy page.
CHAPTER I
DEFINITIONS SPECIFIC TO THE CLASSIFICATION RULES
1. DURATION OF USE
1.1. ‘Transient’ means normally intended for continuous use for less than 60 minutes.
Contents
1.2. ‘Short term’ means normally intended for continuous use for between 60 minutes and 30 days.
1.3. ‘Long term’ means normally intended for continuous use for more than 30 days.
2. INVASIVE AND ACTIVE DEVICES
2.1. ‘Body orifice’ means any natural opening in the body, as well as the external surface of the eyeball, or any permanent artificial opening, such as a stoma.
2.2. ‘Surgically invasive device’ means:
(a) an invasive device which penetrates inside the body through the surface of the body, including through mucous membranes of body orifices, with the aid or in the context of a surgical operation; and
(b) a device which produces penetration other than through a body orifice.
2.3. ‘Reusable surgical instrument’ means an instrument intended for surgical use in cutting, drilling, sawing, scratching, scraping, clamping, retracting, clipping or similar procedures, without a connection to an active device, and which is intended by the manufacturer to be reused after appropriate procedures, such as cleaning, disinfection and sterilisation, have been carried out.
2.4. ‘Active therapeutic device’ means any active device used, whether alone or in combination with other devices, to support, modify, replace or restore biological functions or structures with a view to treatment or alleviation of an illness, injury or disability.
2.5. ‘Active device intended for diagnosis and monitoring’ means any active device used, whether alone or in combination with other devices, to supply information for detecting, diagnosing, monitoring or treating physiological conditions, states of health, illnesses or congenital deformities.
2.6. ‘Central circulatory system’ means the following blood vessels: arteriae pulmonales, aorta ascendens, arcus aortae, aorta descendens to the bifurcatio aortae, arteriae coronariae, arteria carotis communis, arteria carotis externa, arteria carotis interna, arteriae cerebrales, truncus brachiocephalicus, venae cordis, venae pulmonales, vena cava superior and vena cava inferior.
2.7. ‘Central nervous system’ means the brain, meninges and spinal cord.
2.8. ‘Injured skin or mucous membrane’ means an area of skin or a mucous membrane presenting a pathological change or change following disease or a wound.
CHAPTER II
IMPLEMENTING RULES
3.1. Application of the classification rules shall be governed by the intended purpose of the devices.
3.2. If the device in question is intended to be used in combination with another device, the classification rules shall apply separately to each of the devices. Accessories for a medical device and for a product listed in Annex XVI shall be classified in their own right separately from the device with which they are used.
3.3. Software, which drives a device or influences the use of a device, shall fall within the same class as the device. If the software is independent of any other device, it shall be classified in its own right.
3.4. If the device is not intended to be used solely or principally in a specific part of the body, it shall be considered and classified on the basis of the most critical specified use.
3.5. If several rules, or if, within the same rule, several sub-rules, apply to the same device based on the device’s intended purpose, the strictest rule and sub-rule resulting in the higher classification shall apply.
3.6. In calculating the duration referred to in Section 1, continuous use shall mean:
(a) the entire duration of use of the same device without regard to temporary interruption of use during a procedure or temporary removal for purposes such as cleaning or disinfection of the device. Whether the interruption of use or the removal is temporary shall be established in relation to the duration of the use prior to and after the period when the use is interrupted or the device removed; and
(b) the accumulated use of a device that is intended by the manufacturer to be replaced immediately
with another of the same type.
3.7. A device is considered to allow direct diagnosis when it provides the diagnosis of the disease or condition in question by itself or when it provides decisive information for the diagnosis.
CHAPTER III
CLASSIFICATION RULES
4. NON-INVASIVE DEVICES
4.1. Rule 1
All non-invasive devices are classified as class I, unless one of the rules set out below applies.
4.2. Rule 2
All non-invasive devices intended for channelling or storing blood, body liquids, cells or tissues, liquids or gases for the purpose of eventual infusion, administration or introduction into the body are classified as:
- class IIa if they may be connected to a class IIa, class IIb or class III active device;
- class IIa if they are intended for use for channelling or storing blood or other body liquids or for storing organs, parts of organs or body cells and tissues, except for blood bags; blood bags are classified as class IIb.
In all other cases, such devices are classified as class I.
4.3. Rule 3
All non-invasive devices intended for modifying the biological or chemical composition of human tissues or cells, blood, other body liquids or other liquids intended for implantation or administration into the body are classified as class IIb, unless the treatment for which the device is used consists of filtration, centrifugation or exchanges of gas or heat, in which case they are classified as class IIa.
All non-invasive devices consisting of a substance or a mixture of substances intended to be used in vitro in direct contact with human cells, tissues or organs taken from the human body or used in vitro with human embryos before their implantation or administration into the body are classified as class III.
4.4. Rule 4
All non-invasive devices which come into contact with injured skin or injured mucous membrane are classified as:
- class I if they are intended to be used as a mechanical barrier, for compression or for absorption of exudates;
- class IIb if they are intended to be used principally for injuries to skin which have breached the dermis or mucous membrane and can only heal by secondary intent;
- class IIa if they are principally intended to manage the micro-environment of injured skin or mucous membrane; and
- class IIa in all other cases.
This rule also applies to the invasive devices that come into contact with injured mucous membrane.
5. INVASIVE DEVICES
5.1. Rule 5
All invasive devices with respect to body orifices, other than surgically invasive devices, which are not intended for connection to an active device or which are intended for connection to a class I active device:
- are classified as class I if they are intended for transient use;
- are classified as class IIa if they are intended for short-term use, except if they are used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or in a nasal cavity, in which case they are classified as class I; and
- are classified as class IIb if they are intended for long-term use, except if they are used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or in a nasal cavity and are not liable to be absorbed by the mucous membrane, in which case they are classified as class IIa.
All invasive devices with respect to body orifices, other than surgically invasive devices, intended for connection to a class IIa, class IIb or class III active device are classified as class IIa.
5.2. Rule 6
All surgically invasive devices intended for transient use are classified as class IIa unless they are:
- intended specifically to control, diagnose, monitor or correct a defect of the heart or of the central circulatory system through direct contact with those parts of the body, in which case they are classified as class III;
- reusable surgical instruments, in which case they are classified as class I;
- intended specifically for use in direct contact with the heart or central circulatory system or the central nervous system, in which case they are classified as class III; - intended to supply energy in the form of ionising radiation, in which case they are classified as class IIb;
- intended to have a biological effect or to be wholly or mainly absorbed, in which case they are classified as class IIb; or
- intended to administer medicinal products by means of a delivery system, if such administration of a medicinal product is done in a manner that is potentially hazardous taking account of the mode of application, in which case they are classified as class IIb.
5.3. Rule 7
All surgically invasive devices intended for short-term use are classified as class IIa unless they are:
- intended specifically to control, diagnose, monitor or correct a defect of the heart or of the central circulatory system through direct contact with those parts of the body,
in which case they are classified as class III;
- intended specifically for use in direct contact with the heart or central circulatory system or the central nervous system, in which case they are classified as class III;
- intended to supply energy in the form of ionising radiation, in which case they are classified as class IIb;
- intended to have a biological effect or to be wholly or mainly absorbed, in which case they are classified as class III;
- intended to undergo chemical change in the body, except if the devices are placed in the teeth, in which case they are classified as class IIb; or
- intended to administer medicines, in which case they are classified as class IIb.
5.4. Rule 8
All implantable devices and long-term surgically invasive devices are classified as class IIb unless they are:
- intended to be placed in the teeth, in which case they are classified as class IIa;
- intended to be used in direct contact with the heart, the central circulatory system or the central nervous system, in which case they are classified as class III;
- intended to have a biological effect or to be wholly or mainly absorbed, in which case they are classified as class III;
- intended to undergo chemical change in the body, except if the devices are placed in the teeth, in which case they are classified as class III;
- intended to administer medicinal products, in which case they are classified as class III;
- active implantable devices or their accessories, in which cases they are classified as class III;
- breast implants or surgical meshes, in which cases they are classified as class III;
- total or partial joint replacements, with the exception of ancillary components such as screws, wedges, plates and instruments, in which case they are classified as class III; or
- spinal disc replacement implants or implantable devices that come into contact with the spinal column, with the exception of components such as screws, wedges, plates and instruments, in which case they are classified as class III.
6. ACTIVE DEVICES
6.1. Rule 9
All active therapeutic devices intended to administer or exchange energy are classified as class IIa unless their characteristics are such that they may administer energy to or exchange energy with the human body in a potentially hazardous way, taking account of the nature, the density and site of application of the energy, in which case they are classified as class IIb.
All active devices intended to control or monitor the performance of active therapeutic class IIb devices, or intended directly to influence the performance of such devices, are classified as class IIb.
All active devices intended to emit ionizing radiation for therapeutic purposes, including devices which control or monitor such devices, or which directly influence their performance, are classified as class IIb.
All active devices that are intended to control, monitor or directly influence the performance of active implantable devices are classified as class III.
6.2. Rule 10
Active devices intended for diagnosis and monitoring are classified as class IIa:
- if they are intended to supply energy which will be absorbed by the human body, except for devices intended to illuminate the patient’s body, in the visible spectrum, in which case they are classified as class I;
- if they are intended to image in vivo distribution of radiopharmaceuticals; or
- if they are intended to allow direct diagnosis or monitoring of vital physiological processes, unless they are specifically intended for monitoring of vital physiological parameters and the nature of variations of those parameters is such that it could result in immediate danger to the patient, for instance variations in cardiac performance, respiration, activity of the central nervous system, or are specifically intended for diagnosis in clinical situations where the patient is in immediate danger, in which cases they are classified as class IIb.
Active devices intended to emit ionizing radiation and intended for diagnostic or therapeutic radiology, including interventional radiology devices and devices which control or influence such devices, are classified as class IIb.
6.3. Rule 11
Software intended to provide information which is used to take decisions with diagnosis or therapeutic purposes is classified as class IIa, except if such decisions have an impact that may cause:
- death or an irreversible deterioration of a person’s state of health, in which case it is classified as class III; or
- a serious deterioration of a person’s state of health or a surgical intervention, in which case it is classified as class IIb.
Software intended to monitor physiological processes is classified as class IIa, except if it is intended for monitoring of vital physiological parameters, where the nature of variations of those parameters is such that it could result in immediate danger to the patient, in which case it is classified as class IIb.
All other software is classified as class I.
6.4. Rule 12
All active devices intended to administer and/or remove medicinal products, body liquids or other substances to or from the body are classified as class IIa, unless this is done in a manner that is potentially hazardous, taking account of the nature of the substances involved, of the part of the body concerned and of the mode of application, in which case they are classified as class IIb.
6.5. Rule 13
All other active devices are classified as class I.
7. SPECIAL RULES
7.1. Rule 14
All devices incorporating, as an integral part, a substance which, if used separately, can be considered to be a medicinal product, as defined in point 2 of Article 1 of Directive 2001/83/EC, including a medicinal product derived from human blood or human plasma, as defined in point 10 of Article 1 of that Directive, and that has an action ancillary to that of the devices, are classified as class III.
7.2. Rule 15
All devices used for contraception or the prevention of the transmission of sexually transmitted diseases are classified as class IIb, unless they are implantable or long-term invasive devices, in which case they are classified as class III.
7.3. Rule 16
All devices intended specifically to be used for disinfecting, cleaning, rinsing or, where appropriate, hydrating contact lenses are classified as class IIb.
All devices intended specifically to be used for disinfecting or sterilising medical devices are classified as class IIa, unless they are disinfecting solutions or washer-disinfectors intended specifically to be used for disinfecting invasive devices, as the end point of processing, in which case they are classified as class IIb.
This rule does not apply to devices that are intended to clean devices other than contact lenses by means of physical action only.
7.4. Rule 17
Devices specifically intended for recording of diagnostic images generated by X-ray radiation are classified as class IIa.
7.5. Rule 18
All devices manufactured utilising tissues or cells of human or animal origin, or their derivatives, which are non-viable or rendered non-viable, are classified as class III, unless such devices are manufactured utilising tissues or cells of animal origin, or their derivatives, which are non-viable or rendered non-viable, and are intended to come into contact with intact skin only.
7.6. Rule 19
All devices incorporating or consisting of nanomaterial are classified as:
- class III if they present a high or medium potential for internal exposure;
- class IIb if they present a low potential for internal exposure; and
7.7. Rule 20
All invasive devices with respect to body orifices, other than surgically invasive devices, intended to administer medicinal products by inhalation are classified as class IIa, unless their mode of action has an essential impact on the efficacy and safety of the administered medicinal product or they are intended to treat life-threatening conditions, in which case they are classified as class IIb.
7.8. Rule 21
Devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by or locally dispersed in the human body are classified as:
- class III if they, or their products of metabolism, are systemically absorbed by the human body in order to achieve the intended purpose;
- class III if they achieve their intended purpose in the stomach or lower gastrointestinal tract and they, or their products of metabolism, are systemically absorbed by the human body;
- class IIa if they are applied to the skin or if they are applied in the nasal cavity or in the oral cavity as far as the pharynx, and achieve their intended purpose on those cavities; and
- class IIb in all other cases.
Özet Olarak ;
Medical Device Classification: How Is It Done?
The classification of medical devices is vital so that products can be placed on the market safely and effectively. This classification is determined according to the device’s intended purpose, duration of patient contact and potential risk factors. Medical device classification carried out in accordance with the European Union regulation (MDR 2017/745) and the legislation of the TİTCK (Turkish Medicines and Medical Devices Agency) in Türkiye is also one of the basic stages of the ÜTS (Product Tracking System) registration process.
The Importance of Medical Device Classification
Proper classification of medical devices:
Ensures that products are subject to the relevant regulations.
Determines the clinical evaluation and technical file preparation processes.
Shows which procedures the CE certification process is subject to.
Guides the creation of the documents required for ÜTS registration and market access.
How Are Medical Devices Classified?
Medical device classification is determined according to the European Union’s medical device regulation MDR 2017/745 and the relevant guidance documents. Devices are divided into four main classes depending on their risk level:
1. Class I (Low Risk)
Intended for transient use on the patient.
Has no penetration into the body or invasive effect.
Examples: Spectacle frames, bandages, patient support devices.
2. Class IIa (Medium-Low Risk)
Includes short-term invasive devices.
Used to monitor or support physiological processes.
Examples: Dental fillings, contact lenses, ultrasound devices.
3. Class IIb (Medium-High Risk)
Invasive devices requiring long-term use.
Devices that support or modify vital functions.
Examples: Ventilators, infusion pumps, X-ray devices.
4. Class III (High Risk)
Devices that are placed in the body or that may be life-threatening.
They must go through special clinical evaluation and certification processes.
Examples: Pacemakers, implants, medical products containing biological tissue.
How Is Medical Device Classification Determined?
The following factors must be evaluated in order to determine which class a medical device falls into:
Intended Purpose: The area of use and functionality stated by the manufacturer must be examined.
Duration of Patient Contact: How long the device is in contact with the body is determined.
Invasiveness: Whether the device is in contact with skin, mucous membranes or body cavities is evaluated.
Active or Passive: It must be determined whether the device has an electrical or mechanical function.
Risk Level: The potential risks that may arise from the use of the device are taken into account.
The Effect of Medical Device Classification on the CE Certificate
Medical device classification directly affects the CE certification process. While low-risk devices (Class I) can be placed on the market with a manufacturer declaration, Class IIa, IIb and III devices must undergo the supervision of a notified body.
Points to pay attention to in the CE certification process:
Complete preparation of the technical file,
Carrying out clinical evaluation and appropriate tests,
Providing the documents required for ÜTS registration.
Conclusion
Medical device classification is a critical step for ensuring the legal compliance of products, guaranteeing patient safety and completing the CE certification process. Manufacturers and importers who manage this process correctly gain an advantage when entering the market.
To get professional consultancy on medical device classification and ÜTS registration, you can contact Medex Kurumsal Danışmanlık.
Would You Like to Benefit from Our ÜTS Consultancy Service?
Just leave your e-mail address.
After you fill in the form, we will deliver our price quote to your e-mail address as soon as possible.
Medical Device Classification Rules – Table Format
A. Non-Invasive Devices
Rule No | Definition / Scope | Class |
1 | All non-invasive devices | I |
2 | Intended for carrying/storing body fluid, tissue and organ samples – Contact with the body: | IIa (if there is contact) IIb (if cell/organism viability is affected) |
3 | Devices that prepare a substance for analysis by filtering, centrifuging or modifying it | IIa / IIb / III |
4 | Devices applied to injured tissue | I (does not go below the epidermis) IIb (if it affects tissue regeneration) |
B. Invasive Devices
Rule No | Definition / Scope | Class |
5 | Invasive devices with respect to body orifices - Transient or short-term | I / IIa / IIb |
6 | Surgical transient invasive devices | IIa (general use) IIb (contact with heart/circulation/nerves) |
7 | Surgical short-term invasive devices | IIa (general) IIb (vital organs, hazardous areas) |
8 | Implantable / long-term invasive devices | IIa / IIb / III |
C. Active Devices
Rule No | Definition / Scope | Class |
9 | Active therapeutic devices (those that apply energy) | IIa / IIb |
10 | Active diagnostic / monitoring devices | I / IIa / IIb |
11 | Software – diagnostic or therapeutic use | I / IIa / IIb / III |
12 | Active substance/injection devices | IIa / IIb |
13 | Other active devices | I |
D. Special Rules
Rule No | Definition / Scope | Class |
14 | Devices containing a medicinal product | III |
15 | Contraception / prevention of sexually transmitted infections | IIb / III |
16 | Devices used for device disinfection | IIa / IIb |

Medical Device Classification Rules
A. Non-Invasive Devices
Rule No | Definition | Class |
1.1 | Device applied to the body surface that does not cross the anatomical barrier | Class I |
1.2 | Device applied to the body surface that may come into contact with injured tissue | Class IIa |
1.3 | Device with a healing effect on injured tissue | Class IIb |
1.4 | Devices applied to body cavities for filling purposes | Class IIa or IIb (depending on scope) |
B. Invasive Devices
Rule No | Definition | Class |
2.1 | Transient invasive device (≤60 minutes), entering through natural body orifices | Class I |
2.2 | Short-term invasive device (>60 min ≤30 days) | Class IIa |
2.3 | Long-term invasive device (>30 days) | Class IIb |
2.4 | Surgically invasive device, transient use | Class IIa |
2.5 | Surgically invasive device, short-term use | Class IIb |
2.6 | Surgically invasive device, long-term use | Class III |
C. Active Devices
Rule No | Definition | Class |
3.1 | Active devices for diagnostic purposes | Class IIa |
3.2 | Devices that deliver energy to the body (electricity, heat, light, etc.) | Class IIb |
3.3 | Devices that support or sustain vital functions | Class III |
3.4 | Software-based devices | Class IIa / IIb / III (depending on function) |
D. Special Rules
Rule No | Definition | Class |
4.1 | Bags for body fluids, blood bags | Class IIb |
4.2 | Safety-critical accessories | Class III |
4.3 | Implants, permanent devices | Class III |
4.4 | Devices containing human-origin derivatives | Class III |