What Is Done If an ÜTS Document Registration Is Rejected?

There are two ways. If the decision is thought to be contrary to the relevant legislation and the provisions of the Guide, a re-evaluation of the decision is requested through the "Submit Request" screen in ÜTS before reapplying…

ÜTS frequently asked questions: What Is Done If an ÜTS Document Registration Is Rejected?

There are two ways. If the decision is thought to be contrary to the relevant legislation and the provisions of the Guide, a re-evaluation of the decision is requested through the "Submit Request" screen in ÜTS before reapplying. If the decision is considered justified, the deficiency is remedied, the document notification is updated and a new document application is created.

How to open a re-evaluation request

  • Go to the "Submit Request" screen in ÜTS.
  • Select "Medical Device Document and Product Registration Transactions" as the "Request Area".
  • Enter the Document Application Number assigned to the document application in question in the "Medical Device/Document Application Number" field.
  • In the request, refer to the legislation and/or Guide provisions relating to the point objected to.

These two conditions are mandatory: requests in which the Document Application Number is not entered, or in which no reference is made to the legislation or the Guide on the point objected to, are not taken into consideration. The purpose of the request is to have the decision corrected for an application that was wrongly evaluated, without reapplying. The Agency is not held responsible for applications that are sent for review by reapplying without the correction being made through the request.

The same route is open for an "Request Update" decision

If the Agency gives a 3-day update period for one or more of the document notifications and this decision is thought to be contrary to the legislation, re-evaluation can be requested in the same way through the "Submit Request" screen before the update is made. Here too, the Document Application Number must be entered and reference must be made to the legislation. The Agency is not held responsible for applications for which the update was made without the correction being made through the request.

When preparing the new application

  • Whichever field the rejection or revision reason relates to, that field in the document notification is corrected first. The information entered in ÜTS must be the same as the information on the document.
  • The document is uploaded again in full, with all its annexes and additional pages.
  • If the approval route is missing, one of the accepted routes is completed: an apostilled version of the document, an embassy/consulate-certified version, an e-Apostille-certified original or an e-signed original.
  • If the translation is missing, the Turkish translation made by a sworn translator is uploaded to the "Turkish Document" field.
  • A new application is created, the "Confirm" action is performed and the application is electronically signed. A document application contains at most 5 document notifications.

2026/KKB-2 update: With TİTCK's Announcement No. 2026/KKB-2 dated 02.11.2026, a new route was introduced in place of the option of uploading the manufacturer's e-signed original document for a declaration of conformity and SPP declaration belonging to a foreign manufacturer: TİTCK'nın 02.11.2026 tarihli 2026/KKB-2 Sayılı Duyurusu ile yabancı imalatçıya ait uygunluk beyanı ve SİP beyanında, imalatçının e-imzalı belge aslını yükleme seçeneğinin yerine yeni yol getirildi:

“it has been decided to apply the option of uploading the document in the relevant field in ÜTS, together with an undertaking electronically signed by the authorised person of the importer company authorised by the foreign manufacturer, with an e-signature (secure e-signature) created on the basis of a qualified electronic certificate obtained from an electronic certificate service provider recognised by Türkiye.”

Source: TİTCK Announcement No. 2026/KKB-2. Required documents and steps: ÜTS document registration guide.ÜTS belge kaydı rehberi.

Decisions arising from the certificate query

After an application containing an EC Certificate or EU Certificate is concluded positively, the validity and currency of the certificates are queried by the Agency. A 45-day revision process is applied to certificates found not to be current; only for EU certificates under the MDR and IVDR, and only for certificates revised because of an extension of the product scope, a 120-day extension process is applied. Certificates found invalid for reasons other than losing currency are rejected directly. These decisions are also carried out for other records linked to the same document number.

Effect on device registration

Quality Management Certificate – ISO-13485, Domestic Goods Certificate and Authorised Distributorship Certificate are documents whose upload is not mandatory. Even if their ÜTS registration is rejected for any reason, the device or devices linked to the relevant document remain registered.

Common mistakes

  • Not entering the Document Application Number when opening a request; these requests are not taken into consideration.
  • Writing the objection without stating grounds; reference to the legislation or the Guide is required.
  • Creating successive new applications with the same document without examining the decision.
  • Correcting only the field information when the rejection reason is a missing translation or approval.

Who it concerns: Importer and domestic manufacturer companies whose document application has had a negative outcome, and registration staff. Belge başvurusu olumsuz sonuçlanan ithalatçı ve yerli imalatçı firmalar ile kayıt personeli.

Common mistake: Opening a new application instead of objecting to the rejection decision, and receiving a negative result again for the same reason. Ret kararına itiraz yerine doğrudan yeni başvuru açıp aynı gerekçeyle tekrar olumsuz sonuç almak.

Basis: Guide on Registration Processes of Medical Devices in the Product Tracking System (TCKKD-KLVZ-03, Rev.00, effective 13.05.2026) ARTICLE 21, ARTICLE 22, ARTICLE 23 and ARTICLE 39. Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz (TCKKD-KLVZ-03, Rev.00, yürürlük 13.05.2026) MADDE 21, MADDE 22, MADDE 23 ve MADDE 39.

Document Registration

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara