ÜTS Registration Guide 2026

Your device has cleared customs, it is sitting in your warehouse and your buyer is ready; but if the product does not appear as registered in the Product Tracking System, you cannot take the sales step

ÜTS Registration Guide 2026: The End-to-End Process Map for Medical Device Registration

Your device has cleared customs, it is sitting in your warehouse and your buyer is ready; but if the product does not appear as registered in the Product Tracking System, you cannot take the sales step. ÜTS registration is also not a form-filling job that finishes on a single screen in a single session: first the company, then the document, and finally the product are registered, and this order cannot be changed.

This page brings the whole process together on a single map, based on the “Guide on Registration Processes of Medical Devices to the Product Tracking System” published by the Turkish Medicines and Medical Devices Agency, together with the provisions of the Medical Device Regulation and the Medical Device Sales, Advertising and Promotion Regulation. Each heading links to the sub-page that carries its details. The aim is for you to see which step gets stuck where before you start, not in the middle of the registration.

What Is ÜTS (the Product Tracking System) and What Does It Record?

ÜTS is the national product tracking system for medical devices. Article 1 of the Guide defines the purpose of the document as “determining the procedures and principles for the implementation of registration processes in line with the provisions of the medical device regulations”. In other words, ÜTS is not a statistics pool but the official environment in which the registration obligation set by the legislation is fulfilled.

The scope definition in Article 2 of the Guide groups what the system records under three headings. The scope covers medical device registration transactions, document registration transactions relating to medical devices, registration transactions of companies operating in this field, and other related registration notification transactions. This three-part structure is no coincidence; the whole process moves forward by linking these three records to each other.

  • Company registration: Defines who will enter data into the system. Article 5 of the Guide regulates these transactions.

  • Document registration: Introduces the certificates, declarations, user manuals and similar documents showing the conformity of the device into the system. The document types are regulated one by one between Articles 6 and 24 of the Guide.

  • Product (device) registration: Registers the device to be placed on the market itself, with its barcode, class, name, dimensions and images. Articles 25 to 33 of the Guide regulate this transaction.

  • Other related registration notifications: Free sale certificate applications, data correction requests and similar transactions are also carried out through the same system.

Taking ÜTS and EUDAMED to be the same system is a common confusion, especially among importers who work with foreign manufacturers. In the definitions in Article 4 of the Guide, EUDAMED is described as “the European database system on medical devices”. Article 34 of the Medical Device Regulation, for its part, opens the national obligation with the words “without prejudice to the obligations relating to EUDAMED set out in this Regulation”. The two systems do not replace each other; having a registration on the European side does not remove your ÜTS obligation in Türkiye.

The Legal Basis of ÜTS Registration

The registration obligation is not an implementation preference but a direct provision of the legislation. Article 3 of the Guide ties its basis to three sources: Articles 506, 508 and 796 of Presidential Decree No. 4 of 15/7/2018 on the Organisation of Ministries’ Affiliated, Related and Associated Institutions and Other Institutions; Article 34 of the Medical Device Regulation published in Official Gazette No. 31499 (repeated) of 2/6/2021; and Article 32 of the In Vitro Diagnostic Medical Device Regulation published in the same Official Gazette.

In practice this chain of basis means the following: the Guide does not create an obligation on its own; it shows how the obligation set by the regulation is to be fulfilled. Paragraph 2 of Article 34 of the Medical Device Regulation says this explicitly: the obligations relating to ÜTS are determined by the regulations published by the Agency. The reading that “the Guide is only a recommendation” is therefore wrong.

The legal answer to the question “Is ÜTS registration mandatory?” is in paragraph 1 of Article 34 of the Medical Device Regulation. The article provides, with the word “shall be fulfilled”, that the obligations relating to ÜTS apply to the listed parties. This is not an option but an imposition.

Who Must Register in ÜTS?

Paragraph 1 of Article 34 of the Medical Device Regulation lists the obligated parties one by one: economic operators established in Türkiye, notified bodies, healthcare institutions, healthcare professionals and other relevant persons. According to the definition in Article 4 of the Guide, an economic operator means a manufacturer, authorised representative, importer, distributor or system/procedure pack producer. A company that brings devices from abroad falls within this definition.

Beyond this general obligation there is one more precondition. According to paragraph 1 of Article 5 of the Guide, the registration to ÜTS of companies that carry out the sale and/or application of medical devices is made if they are authorised under the relevant legislation. That is, not every company can register with the system on its own; you must first be authorised in accordance with the relevant legislation. This is the step that newly established companies overlook most often.

The sanction on the sales side is found in paragraph 2 of Article 26 of the Medical Device Sales, Advertising and Promotion Regulation: sales centres cannot sell devices that are not registered in the Agency’s registration and information management system. Even if the device is in your hands and an invoice can be issued, sale is legally not possible before the registration is completed.

  • Importer company: If the legal manufacturer that issues the declaration of conformity is abroad, the device is registered as “imported” and the registration obligation lies with the importer in Türkiye.

  • Domestic manufacturer: Makes the document and device registration for the device it produces in its own name; it uses an application type specific to it in document applications.

  • Dealer and distributor: A company registration is required; having the dealership relationship defined in the system is mandatory, especially in system and procedure pack registrations.

  • The foreign manufacturer’s authorised representative in Türkiye: For transactions such as a free sale certificate, a registration in ÜTS under the manufacturer/importer/dealer/exporter activity area type is required.

The Registration Chain: Why the Company → Document → Product Order Cannot Change

The whole process is a three-link chain, and the order of the links is imposed by the system. If you have no company registration, you cannot reach the screen where you would create a document application; your document cannot be linked to a device until it reaches “Registered” status; and even once a device registration is completed, it is not evaluated by the Agency without a separate application.

The first link is the company. Article 5 of the Guide regulates company registration transactions, and no transaction can be made in the system without a company registration. The second link is the document: according to paragraph 2 of Article 39 of the Guide, a document application is made from the “Create Document Application” area on the “Document Application Transactions” screen under the “Document” tab in ÜTS. The third link is the device, and device registration rests on documents registered before it.

The last step of the chain is where most companies slip. Paragraph 11 of Article 25, paragraph 12 of Article 27 and paragraph 13 of Article 30 of the Guide say the same thing: for the conformity of a device registration or update notification with the legislation to be evaluated, a device application must be made to the Agency as set out in Article 40. Filling in the record does not mean making an application. Paragraph 14 of Article 39 of the Guide sums up this distinction in one sentence: applications that do not reach “Under Review” status cannot be evaluated by the Agency’s review specialist.

A Document Cannot Be Linked to a Device Unless It Is “Registered”

The “Product Documents” block on the device registration screen is not a free-text field. According to subparagraph (a) of paragraph 5 of Article 25 of the Guide, the documents to be added are selected “from among the documents registered in ÜTS as a result of an application made under Article 39”. Your document may be sitting in the system; if its application has not been completed and has not reached “Registered” status, it is not listed on the device screen.

This precondition is repeated in the same way in all five legislative frameworks. Article 63 of the Guide regulates the document business rules under MDD, Article 64 under AIMDD, Article 65 under IVDD, Article 66 under MDR and Article 67 under IVDR; paragraph 1 of all five requires the documents to be linked to be in “Registered” status in ÜTS as a result of an application made under Article 39. Whichever framework you work under, the rule is the same.

A Document Cannot Be Selected Until the Class Is Selected

There is also a micro-order within the chain. According to the third sub-subparagraph of subparagraph (a) of paragraph 5 of Article 25 of the Guide, the class of the relevant device must be selected on the registration screen before the document selection. The system filters the documents that can be selected according to the class of the device; when the class is empty, the document list does not make sense.

On the MDR side this dependency extends one step further. According to subparagraph (c) of paragraph 5 of Article 27 of the Guide, before a selection is made from the “Select Basic UDI-DI” screen, the EU Declaration of Conformity and the EU Certificate related to the device must have been selected according to the document business rule; the Basic UDI-DI list consists only of numbers associated with the selected EU Declaration of Conformity. In practice this means a one-way path of class → document → Basic UDI-DI.

What Does the 2026 Guide (TCKKD-KLVZ-03) Cover?

The document that draws the current framework of the registration processes is the “Guide on Registration Processes of Medical Devices to the Product Tracking System”, with document number TCKKD-KLVZ-03. In its identification block the first publication date appears as 13/05/2026, the revision number as 00, and there is no revision date. The Guide is 316 pages long.

Structurally the Guide consists of 70 articles, and its annexes are numbered from Annex-1 to Annex-100. The annexes carry the real operational value of the Guide: tables showing which document is uploaded with which approval in which situation, petition templates, conformity assessment routes and class-based document business rules are found there. For example, the petition template to be used in the company registration of an ordinary partnership is in Annex-1, and the certificate upload rules for importer companies are in Annex-3.

  • First and second parts: Purpose, scope, basis, definitions and company registration transactions.

  • Third part: Document types and document registration and update notification transactions; each document type is regulated separately from Article 6 to Article 24.

  • Fourth and following parts: Device, system/procedure pack and custom-made medical device registration notifications under the old and new regulations.

  • Articles 39 and 40: Document application and medical device application transactions; the step at which the registration is submitted to the Agency is described here.

  • Eighth part and after: Free sale certificate, transitional provisions, data correction and conformity assessment procedures.

  • Articles 63 to 67: ÜTS document business rules for the five legislative frameworks.

Provisions Entering into Force on 13.05.2026 and Those Left to the ÜTS Version Update

The whole Guide does not come into operation at the same time. According to subparagraph (a) of paragraph 1 of Article 69 of the Guide, the provisions mentioned in Articles 8, 12 and 15 and the provisions mentioned in paragraph 4 of Article 39 enter into force together with the ÜTS Version Update, with which these procedures and principles become technically functional in ÜTS. According to subparagraph (b) of the same paragraph, the provisions in the other articles enter into force on the date the Guide is published.

The practical consequence of this distinction is as follows: the application of paragraph 4 of Article 39, which lists the document application types, and of Articles 8, 12 and 15, which regulate importer and domestic manufacturer document notifications, depends on the technical update of the system. The Guide does not state the date on which this update will take place. The options you see on the screen and the options you read in the text may therefore differ for a period; the current state of the screen should be confirmed before you start the registration.

The Repealed TCKKD-KLVZ-01 Guide

Article 68 of the Guide provides that the guide with document number TCKKD-KLVZ-01, published on 15/11/2022 and titled “Guide on Registration of Medical Devices to the Product Tracking System (Rev.1)”, has been repealed.

For companies that keep a copy of the old guide in their quality file and wrote their procedures according to it, this is a point at which the internal documentation needs to be reviewed. In practice the place that causes the most problems is checklists prepared according to the old text: document type names, application types and field names may have changed in the new Guide. Working with a checklist prepared according to the old Guide is one of the silent reasons why applications go in incomplete.

Company Roles in ÜTS: Manufacturer, Importer, Dealer, Exporter and Others

The “company” definition in Article 4 of the Guide shows whom the system works with. According to the definition, a company means a company that carries out the sale and/or application of medical devices in fields such as medical device sales centres, manufacturers/importers/dealers/exporters, opticianry establishments, hearing aid application centres, custom-made orthosis-prosthesis application centres, pharmacies and pharmaceutical warehouses.

This list is not merely a count; the role you choose determines your entire registration flow. According to paragraph 5 of Article 5 of the Guide, companies that will make a company registration in ÜTS under the “Manufacturer/Importer/Dealer/Exporter” type must hold a medical device sales centre licence, and the ÇKYS number information must be entered as mandatory during company registration. The same article defines ÇKYS as the Core Resource Management System.

The role selection also determines which applications you can make later. According to paragraph 1 of Article 29 of the Guide, companies or institutions of the types “Healthcare Service Provider”, “Manufacturer/Importer/Dealer/Exporter” and “Custom-Made Orthosis-Prosthesis Application Centre” in ÜTS can make a custom-made medical device application under MDR. A similar restriction exists for the free sale certificate: according to paragraph 3 of Article 42 of the Guide, for devices of manufacturers whose EU authorised representative is established in Türkiye, the application is made by the authorised representative registered in ÜTS under the manufacturer/importer/dealer/exporter activity area type. The wrong role shows up as a closed screen in the later steps of the registration.

The Five Legislative Frameworks: MDD, AIMDD, IVDD, MDR, IVDR

According to the definitions in Article 4 of the Guide, the Old Medical Device Regulations are the repealed Medical Device Regulation (MDD), the repealed Regulation on Active Implantable Medical Devices (AIMDD) and the repealed Regulation on In Vitro Diagnostic Medical Devices (IVDD). The New Medical Device Regulations are the Medical Device Regulation (MDR) published in Official Gazette No. 31499 (repeated) of 2/6/2021 and the In Vitro Diagnostic Medical Device Regulation (IVDR). The framework under which your device is certified directly determines from which screen and with which document type the registration will be made.

Document types are divided in two according to the framework. According to paragraph 1 of Article 6 of the Guide, the document of type “EC Certificate” is used only for certificates prepared by notified bodies under the Old Medical Device Regulations; for certificates under the new regulations, the “EU Certificate (MDR/IVDR)” type regulated in Article 7 is used. There is the same distinction on the declaration side: the “Declaration of Conformity” in Article 10 is for the old regulations and the “EU Declaration of Conformity (MDR/IVDR)” in Article 11 is for the new regulations.

Device registration also runs through three separate articles. Article 25 of the Guide regulates device registration under the Old Medical Device Regulations, Article 27 under MDR and Article 30 under IVDR. Although the block structure of the screens looks similar, the fields are not the same; for example, the screens in Articles 27 and 30 have a “CMR and/or Endocrine-Disrupting Substance Information” block, while the screen in Article 25 does not.

Who Makes the Registration? The In-House Team or an Authorised Consultant?

Who actually creates the registration in ÜTS is a matter of authority, and the legislation ties this authority not to the person but to the user type. According to paragraph 11 of Article 39 of the Guide, the “Approve” action on the “Document Application Summary” screen can be performed only by users of the relevant company of the “Medical Device Registration Officer” type. The same rule applies on the device side: according to subparagraph (f) of paragraph 5 of Article 40, the “Approve” action on the “Medical Device Application Summary” screen can again be performed only by users of this type.

This user type is defined through the company’s own ÜTS account. Paragraph 6 of Article 5 of the Guide, by stating that the signatories and users defined in ÜTS must be updated for ÜTS activities to continue in company transfers and mergers, shows that these definitions are kept in the company’s own record. In practice the consulting relationship is established at exactly this point: the customer opens the “Medical Device Registration Officer” user authority for the consultant from its own company page in ÜTS, and the consultant creates and approves the company, document and product records with this authority. The authority is in the company’s hands and can be removed whenever desired.

The electronic signature step, on the other hand, cannot be delegated. According to paragraph 13 of Article 39 of the Guide, an application in “Awaiting Signature” status is electronically signed by the company’s applicant signatory. For a device application, subparagraph (g) of paragraph 5 of Article 40 says the same thing: for the application to reach “Under Review” status, it must first be electronically signed by the company’s signatory. In practice this means that your e-signature stays with your company, and that you give the final approval of the application the consultant prepared in every case.

  • Medical Device Registration Officer: Creates the record, edits the application and performs the “Approve” action. Paragraph 11 of Article 39 and subparagraph (f) of paragraph 5 of Article 40 of the Guide give this authority to this type.

  • Signatory: Signs the approved application electronically. This step rests with the company’s own signatory.

  • Company: Is responsible for the accuracy of the data entered; this responsibility is not transferred by granting authority.

Training Requirement for Personnel Making Registrations in ÜTS

There is also a qualification requirement for the person who will enter data into the system. According to paragraph 10 of Article 26 of the Medical Device Sales, Advertising and Promotion Regulation, persons who make medical device registration transactions in the Agency’s registration and information management system on behalf of a sales centre must successfully complete the training determined by the Agency.

This requirement has a reverse side. According to paragraph 2 of Article 28/A of the same regulation, if it is determined that transactions have been made repeatedly in the registration and information management systems contrary to the procedures and principles determined by the Agency, the training received by the persons making these transactions under paragraph 10 of Article 26 is cancelled. In other words, whoever works in the system, that person’s qualification is also a subject of inspection.

Responsibility Always Remains with the Company Making the Registration

Whether the registration is made by an in-house employee or by an externally authorised consultant, the accuracy of the data remains with the company. Paragraph 4 of Article 31 of the Medical Device Sales, Advertising and Promotion Regulation states this explicitly: the accuracy of the information about devices recorded in the system is the responsibility of the sales centre that makes the product registration. The same paragraph also provides that the Agency issues guides on the registration of devices in the registration and information management system, and that sales centres will set up the infrastructure in line with these guides within the specified periods and register the devices in the system.

A good consulting relationship is therefore not “we will take care of it, you need not look”. The original of the document, its translation, the label image and the technical file come from the company; the consultant places them in the fields the legislation asks for in the correct form and identifies deficiencies before registration. The final declaration still belongs to the company.

Paid and Free ÜTS Applications

Not every application is subject to a fee, and which ones are paid is determined by the document type. According to subparagraph (a) of paragraph 13 of Article 39 of the Guide, among the applications signed electronically, those containing document types mentioned in Articles 6, 7, 10, 11 and 14 of the Guide move to “Awaiting Payment” status. The document types regulated in these articles are, respectively, the EC Certificate, EU Certificate (MDR/IVDR), Declaration of Conformity, EU Declaration of Conformity (MDR/IVDR) and SPP Declaration (MDR) (Article 22) documents. For documents with a payment requirement, the application moves to “Under Review” status after the application fee is paid and approved.

Subparagraph (b) of the same paragraph separates the other document types: documents other than these document types move directly to “Under Review” status as soon as the document application is electronically signed. Applications containing a user manual, intended use form, ISO 13485 quality management document, domestic goods certificate and authorised distributorship document are in this group.

On the device side the payment step is standard. According to subparagraph (g) of paragraph 5 of Article 40 of the Guide, the device application is first electronically signed by the signatory and brought to “Awaiting Payment” status, and then moves to “Under Review” status when the application fee is paid. The same subparagraph is repeated in subparagraph (ç) of paragraph 6, which regulates custom-made medical device and system/procedure pack applications.

  • If the payment step is skipped: The application stays in “Awaiting Payment” status and does not reach the review specialist.

  • If the signature is not made: The application waits in “Awaiting Signature” status; being approved is not enough.

  • In both cases: Under paragraph 14 of Article 39 of the Guide, applications that do not reach “Under Review” status cannot be evaluated.

What Happens If ÜTS Registration Is Not Made?

The most direct consequence is on the sales side. Under paragraph 2 of Article 26 of the Medical Device Sales, Advertising and Promotion Regulation, sales centres cannot sell devices that are not registered in the Agency’s registration and information management system. Even if the product is physically in your hands, a placing-on-the-market chain cannot be built before the registration is completed.

Administrative sanctions relating to the registration system are regulated in a separate article. According to paragraph 1 of Article 28/A of the same regulation, if the Agency determines that applications have been made to the registration and information management systems with information and documents contrary to the truth, it may restrict the access of the persons or organisations making these transactions to the relevant systems. Restriction of access means that all of that company’s transactions in progress in ÜTS stop.

It is not enough for data merely to have been entered; it must be kept up to date. According to paragraph 5 of Article 31 of the Medical Device Regulation, economic operators confirm the accuracy of the data within 1 year of submitting the information and every 2 years thereafter; if this is not done within 6 months of the end of these periods, the Agency may take appropriate corrective measures until the economic operator complies with this obligation. The periods here are confirmation periods that the legislation grants to the parties; they are not a commitment on how long a job will take. Paragraph 4 of the same article also provides that the data must be updated within 1 week if there is a change in the information.

The Map of This Guide: Which Topic Is on Which Page

This page shows the whole chain; the details of each link are on its own page. The guidance below lets you go directly to the right heading according to the question you come with.

  • Company registration: The ÜTS Company Registration page, for the MERSİS and VEDOP distinction, the documents required for an ordinary partnership, the ÇKYS number and company type selection.

  • Document registration: The ÜTS Document Registration page, for the four document application types, the original and Turkish document fields, and having a notified body defined.

  • Product registration: The ÜTS Product Registration page, for the blocks on the registration screen, the “Is it ready to apply?” question, the single-declaration rule and product limits.

  • Catalogue of document types: The ÜTS Document Types page, showing for which of the eleven document types are mandatory and which optional under each framework.

  • Classification: The Medical Device Classification Rules page, for the full list of MDR and IVDR classification rules.

  • Rejection and correction: The Why Is a ÜTS Application Rejected page, for why an application ends negatively and how to correct it.

  • Authorisation: The Granting a Consultant Authority in ÜTS page, for how the registration officer authority is defined and the model of working with a consultant.

  • Service side: If you are curious how running the process externally works, the Having ÜTS Registration Made and ÜTS Registration Consultancy Service pages.

Frequently Asked Questions About ÜTS Registration

The questions below are the ones most often asked by companies starting registration for the first time. All answers rest on the relevant articles of the Guide.

  • “If there is a CE mark, is ÜTS registration still needed?” According to the definition in Article 4 of the Guide, the CE mark is the mark that enables the manufacturer to show that the device conforms to the applicable requirements. This mark does not replace the national registration obligation established by Article 34 of the Medical Device Regulation. The CE mark is a precondition of registration, not an alternative to it.

  • “How many documents can be added to one application?” According to paragraph 15 of Article 39 of the Guide, a document application contains at most 5 document notifications. The same limit is repeated in paragraph 7 of Article 8.

  • “How many products can be added to one application?” According to subparagraph (ğ) of paragraph 5 of Article 40 of the Guide, applications for optical products, active implantable medical devices, class 3 products, other products and products under 2023/KK-1 and 2023/KK-2 contain at most 500 medical device notifications. According to subparagraph (d) of paragraph 6, this limit is 5 for custom-made medical device and system/procedure pack applications.

  • “Is the Authorised Distributorship Document mandatory?” According to paragraph 9 of Article 23 of the Guide, the Authorised Distributorship Document is not a document that must be uploaded to ÜTS; it is an optional document. Even if the registration of such a document falls into rejected status for any reason, the device or devices linked to that document continue to remain in registered status.

  • “Do ÜTS and EUDAMED replace each other?” No. Article 34 of the Medical Device Regulation regulates the national product tracking system obligation without prejudice to the EUDAMED obligations. The two systems run in parallel.

For the counterpart on the EU side, our EU authorised representative guide page (in English) is a guide.

ÜTS Registration

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara