Obligations After ÜTS Registration
When your device reaches "Registered" status in ÜTS, the work is not finished; the real follow-up starts at that point. Linking multi-pack barcodes, updating the record when a certificate is revised, answering data change requests from other companies registered with the same barcode on time, and continuing unique movement notifications are the jobs that keep a record alive
Obligations After ÜTS Registration: Notification, Updating and Follow-Up
When your device reaches "Registered" status in ÜTS, the work is not finished; the real follow-up starts at that point. Linking multi-pack barcodes, updating the record when a certificate is revised, answering data change requests from other companies registered with the same barcode on time, and continuing unique movement notifications are the jobs that keep a record alive.
Based on the TİTCK Guide, the Medical Device Regulation and the Regulation on Sales, Advertising and Promotion of Medical Devices, this page covers all post-registration obligations in order: when to update, which operations require an application, how a record is deleted and how many years which documents are kept.
What Changes After "Registered" Status?
"Registered" status is not just a label showing that the device is visible in the system; it is a precondition for certain operations. The first paragraph of Article 36 of the Guide says that adding market placement information can be done only for devices that were registered under Articles 25, 27 and 30 of the Guide and then reached "Registered" status after the application made to the Agency under Article 39 concluded positively. You cannot add a multi-pack barcode before the registration is complete.
The same condition applies to the free sale certificate. Sub-item (p) of the second paragraph of Article 42 of the Guide states that, after the application fee is paid, the free sale certificate is issued only for devices that are included in the relevant application and are in "Registered" status in ÜTS on the date the certificate is issued, and that it is sent by post to the company's address registered in ÜTS. That is, you need to be registered not at the time of application but on the date the certificate is issued; a device that falls into revision in between drops out of the certificate's scope.
Market Placement Information: Linking Multi-Pack Barcodes
ÜTS registration is made with the barcode of the smallest packaging; box and carton barcodes are added afterwards. Under the sixth sub-item of sub-item (b) of the second paragraph of Article 25 of the Guide, as a rule, device registration in ÜTS is made with the barcode of the smallest packaging placed on the market (unit packaging or single packaging); the barcodes of sales packaging containing more than one product, formed by combining several devices registered with the barcode allocated to the smallest packaging, are added, together with their content quantities, to the "Market Placement Information" area on the screen of the relevant device after the ÜTS device registration is complete. The same provision is repeated for MDR registrations in Article 27 and for IVDR registrations in the sixth sub-item of sub-item (b) of the second paragraph of Article 30.
The first paragraph of Article 36 of the Guide describes this operation at screen level: linking box/packaging and container barcodes containing multiple products to the unique product barcode of the device (Primary Product Number or UDI-DI Number) is done through the "Add Market Placement Information" screen in the "Market Placement Information" area on the "Medical Device Information" screen of the relevant device. The second to fourth paragraphs of the same article explain that this operation can also be done in bulk with a template Excel file by entering the "Bulk Add Market Placement Information" screen from the "Bulk Operations" area on the "List My Medical Devices" screen.
1 Week for Data Changes, 1 Year and 2 Years for Confirmation
When the information in your economic operator registration changes, you need to update it without waiting. The fourth paragraph of Article 31 of the Medical Device Regulation says that economic operators update the data in the electronic system referred to in Article 30 within 1 week from the occurrence of any change to the information referred to in the first paragraph of the same article. This information is the manufacturer, authorized representative and importer registration information listed in item 1 of Annex VI Part A of the Regulation.
Even if there is no change, there is a periodic confirmation obligation. The fifth paragraph of the same article states that economic operators confirm the accuracy of the data within 1 year of submitting the information and every 2 years thereafter; if this is not done within 6 months from the end of those periods, the Agency may take appropriate corrective measures until the economic operator complies with this obligation. These are the periods the legislation gives to the parties; they are not a commitment about how long a transaction will take.
1 week: the period allowed for updating when there is a change in the registered information.
1 year: the period allowed for the first accuracy confirmation after the first submission of the information.
2 years: the periodic confirmation interval repeated after the first confirmation.
6 months: the threshold after which the Agency can take corrective measures after the periods end.
What Is an Imported Product Data Change Request?
If the same device is imported into Türkiye by more than one company, there are multiple records with the same barcode in ÜTS. If there are data differences between these records, a new device application cannot be made. The first paragraph of Article 43 of the Guide says that where there are records of different companies for a device registered with the same barcode in ÜTS and there are data differences between these records, the relevant product parameters must be made the same and consistent in ÜTS before the first device application to be made for that device under Article 40.
The request is opened through the barcode number of the device. The second paragraph of Article 44 of the Guide states that the company enters the "List Imported Product Data Change Applications" screen through the "Imported Product Data Change Requests" area under the "Medical Device" tab on the ÜTS company screen; the fourth paragraph states that the new application screen is accessed using the "Add Data Change Request" area. The fifth paragraph says that in the "Barcode Number" field, the Primary Product Number is entered if the device is under the Old Medical Device Regulations and the UDI-DI Number if it is under the New Medical Device Regulations.
23 Data Fields for Which a Change Request Can Be Made
A change request cannot be opened for an arbitrary field; the Guide gives a closed list. The seventh paragraph of Article 44 of the Guide states that a change request application can currently be made in the "Field to Be Changed" area for the data fields listed below and for the data fields that are linked to these fields and assessed together with them. The paragraph lists a total of 23 fields, from sub-item (a) to sub-item (ş).
Number of Products in the Contents: the number of products in the smallest packaging contents.
Class: the risk class of the device.
GMDN: the GMDN code assigned to the device.
Branch Type Code: the branch type identifying the device.
Is it implantable in the body?: implantability information.
Is it single-use?: whether it is single-use.
Is there a limited number of uses?: whether there is a limited number of uses.
Is there a shelf life?: whether there is a shelf life.
Is it subject to calibration?: the requirement for testing, control and calibration.
Is it subject to maintenance?: the requirement for periodic maintenance.
Was it packaged sterile?: whether it is supplied packaged sterile.
Is sterilization required before use?: whether sterilization is required before use.
MRI Safety Information: the option MR Safe, MR Conditional, MR Unsafe or No Information.
Does it contain latex?: latex content.
Does it contain phthalate/DEHP?: phthalate or DEHP content.
Does it contain ionizing radiation?: whether it emits or produces ionizing radiation.
Does it contain nanomaterial?: nanomaterial content.
Is it within the scope of Annex-3 of the Regulation on Sales, Advertising and Promotion of Medical Devices?: coverage by the Annex-3 list.
Is it a component/accessory of a device?: whether it is an accessory or component.
Product Surface Area (in cm2): the value entered for devices reimbursed by SGK on the basis of surface area.
Dimension: dimension type and unit information.
Storage Condition: storage and/or usage condition records.
EMDN: the EMDN term assigned to the device.
The eighth paragraph of the same article adds that the data field options in this list can be changed or updated by the Agency when deemed necessary.
A Separate Application for Each Data Field
More than one field cannot be corrected in a single request. The ninth paragraph of Article 44 of the Guide states that only one of the data fields listed in the area mentioned in the seventh paragraph will be selected; and if a change request is also wanted for different data fields, a separate application will be made for each data field listed there for which a change is requested.
In practice this means that a company that wants to correct four fields opens four separate requests, and each goes through its own opinion process. Identifying one by one which fields are really inconsistent before opening a request is therefore the most practical way of not branching the process unnecessarily.
Evidence Document and Image Area
Whether the other companies and the Agency accept your request depends on the evidence you submit. The tenth paragraph of Article 44 of the Guide says that after the data field is selected, the "Submitted Documents and Images Area" will appear on the screen, together with the relevant data field and, if any, the other data fields linked to it, where documents and images of an evidence nature, which are submitted to the other companies and the Agency if needed in the assessment of the request, are uploaded.
What the Agency's assessment is based on is also defined. The eighth paragraph of Article 43 of the Guide states that the assessment will be made on the basis of the relevant provisions mentioned in the regulation covering the device and in the Guide, the documents and device notification records of the device in ÜTS, the guides and announcements prepared by the Commission and the Agency, the data on the relevant official query sites for fields such as "GMDN" and "EMDN", and the evidence documents and image records submitted in the relevant request.
The 5-Day Opinion Period of Other Companies
The request goes first not to the Agency but to the other companies holding records with the same barcode. The third paragraph of Article 43 of the Guide states that the request application created according to the work steps in Article 44 is first submitted for opinion to the other company or companies that have records with the same barcode in "Registered" or "Unique Product Exists in the System" status in ÜTS, so that they assess the data field or fields requested to be corrected; at this stage the request is in "Awaiting Opinions of Other Companies" status.
The same paragraph also sets the period: the period for the company or companies concerned to give an opinion on the imported product data correction request is 5 days. This is the response period the legislation gives to other companies; it is not the time in which the request will be concluded.
If You Do Not Answer a Request Submitted for Your Opinion
Silence here means approval. The last sentence of the third paragraph of Article 43 of the Guide clearly states that the company or companies that do not give an opinion within the stated period are deemed to have accepted the data field change request mentioned in the request. A company that does not follow the ÜTS screen gives approval, without knowing it, to a field in its own record being changed at another company's request.
The result of an accepted request is processed automatically. Sub-item (d) of the fourth paragraph of the same article states that if all the companies concerned approve the request within the period and the request concerns fields other than those requiring the Agency's assessment, such as "Class", "GMDN" and "EMDN", the request will move to "Accepted" status with the status description "Request was accepted unanimously by the companies" and the fields requested to be changed will be corrected automatically by ÜTS in all the device records of the relevant companies.
Rejecting or staying silent does not stop the process either. The fifth paragraph of the same article states that if at least one of the relevant companies does not approve the request, after all companies have given their opinion or at the latest at the end of the relevant period, the request will be submitted to the Agency's assessment in "Under Review" status; the sixth paragraph says that if no company gives an opinion, the request will also go to the Agency's assessment after the period.
Why Do Class, GMDN and EMDN Changes Go to the Agency's Approval?
Three fields are kept outside this mechanism. Sub-item (e) of the fourth paragraph of Article 43 of the Guide states that if the request concerns the change of fields requiring the Agency's assessment, such as "Class", "GMDN" and "EMDN", the request will be submitted to the Agency's assessment in "Under Review" status.
The seventh paragraph of the same article makes the matter definite: data change requests concerning these fields are submitted to the Agency's assessment and approval in every case mentioned in the fourth, fifth and sixth paragraphs of the article. Unanimous acceptance by the other companies does not change the result. The tenth paragraph says that if the request is not found appropriate as a result of the Agency's assessment, it will be rejected with its reasons and will move to "Rejected" status in ÜTS; the ninth paragraph states that if it is found appropriate, the fields will be corrected automatically in all the records of the relevant companies.
If Documents Fall Into Revision When the Class Changes
A class correction does not remain alone; it also affects the documents linked to the device. The eleventh paragraph of Article 43 of the Guide states that if, as a result of the update of the "Class" data field based on the provisions in the same article, the documents linked to the relevant device records become inconsistent with the device class, revision will be given to the devices with the relevant barcode.
The practical consequence is that a request you did not open can put your record into revision. If a class change is accepted on a device you are registered with under the same barcode, the document combination in your record is also reviewed again according to the document business rule of the new class. Assessing the requests submitted for your opinion on time therefore creates less work than completing documents afterwards.
How Is the Record Updated When a Document Is Updated?
When your certificate is revised, the document record in ÜTS also needs to be updated. The fourth paragraph of Article 6 of the Guide states that if the document has been revised, the revision date is entered as the document start date; if the document has not been revised and the document start date is not clearly stated on the certificate, the most recent issue (signature) date on the certificate is entered. The fifth paragraph of the same article states that where the revision number is stated on the EC certificate and shown together with the certificate number, the relevant certificate is recorded with its revision number, giving the example XXXXX.Rev.1. The fourth and fifth paragraphs of Article 7 of the Guide repeat the same rules for EU certificates.
For instructions for use updates, the date and information fields come into play. The third paragraph of Article 18 of the Guide says that if the instructions for use have been updated, entering at least one of the update date and/or update information of that document into ÜTS is mandatory. The fourth paragraph states that where the original document is prepared in Turkish by its manufacturer and therefore two different original instructions for use are uploaded to ÜTS, the print date and update date and/or update information of the more recent original document are entered in the "Print/Update Information" field.
The ninth paragraph of Article 17 of the Guide resolves a small but common detail: if the print date and/or update date in the instructions for use is entered as month and year but the day is not stated, "1" is entered as the day of that date in ÜTS. For example, if the date in the instructions is 05/2020, it is entered in ÜTS as 01/05/2020.
Updates Not Requiring an Application
Not every post-registration change means a new device application. The first paragraph of Article 35 of the Guide states that device information for which updates without application are allowed, such as linking a new catalogue to the relevant device or uploading a new product image in ÜTS, can also be entered in bulk through the "Bulk Update Product Without Application" screen in the "Bulk Operations" area on the medical device registration screen. The ninth paragraphs of Articles 31, 32 and 33 of the Guide repeat that fields that can be updated without application can be updated through this screen mentioned in Article 35.
There are two technical rules to watch on this screen. The third paragraph of Article 35 of the Guide says that in the template Excel file, red columns indicate mandatory fields and yellow columns indicate fields that are mandatory depending on the entry of another field. The eighth paragraph states that Turkish characters (Ç/ç, Ğ/ğ, İ/ı, Ö/ö, Ş/ş and Ü/ü) cannot be used in file names. In addition, the fifth paragraph says that when the field "Will It Be Sold in Health Market / Will SGK Reimbursement Be Made?" is marked "Yes", there must be at least 3 images in total for the device.
Deleting a Device Record: What Is Really Deleted and "Deleted" Status
"Deletion" in ÜTS is not a single operation; it has two different results depending on the status of the record. The sixth paragraph of Article 34 of the Guide states that the device records requested to be deleted in ÜTS must be in "Draft", "Ready for Application", "Registered" or "Rejected" status. Records outside these four statuses cannot be deleted.
The result differs by status. The seventh paragraph of the same article says that device records in "Draft" and "Ready for Application" status will be deleted from ÜTS. The eighth paragraph states that device records in "Registered" and (except those with a Reject-3 decision) "Rejected" status will be set to "Deleted" status without being deleted from ÜTS. That is, the trace of a device that has once become registered is not removed from the system entirely.
The second to fifth paragraphs of Article 34 of the Guide describe the bulk deletion route: the "Bulk Delete Device" screen is entered from the "Bulk Operations" area on the "List My Medical Devices" screen, the template Excel file is downloaded, the primary product numbers and deletion reasons of all devices requested to be deleted are entered, and the Bulk Device Deletion File in Excel format is created.
Records With a Reject-3 Decision
Not all rejected records are treated the same. While defining the rule for setting to "Deleted" status, the eighth paragraph of Article 34 of the Guide explicitly leaves those with a Reject-3 decision out of scope. Records in "Rejected" status with a Reject-3 decision therefore fall into the group that is deleted directly from ÜTS, like "Draft" and "Ready for Application" records.
The Agency Does Not Delete Device Records
The authority to delete is one-sided. The ninth paragraph of Article 34 of the Guide states that device records can be deleted only by the relevant company and that the Agency does not carry out device record deletion in ÜTS. This means the records of devices you no longer supply remain in the system unless you clean them up. In practice, closing the records of products you remove from your portfolio prevents confusion in later free sale certificate and data confirmation operations.
Unique Movement Notification and Traceability
Registration is the definition of the product in the system; movement notification is the tracking of where that product is. The fourth paragraph of Article 8 of the Regulation on Sales, Advertising and Promotion of Medical Devices states that sales centers must register the medical devices they place on the market or make available on the market in the record and information management system created by the Agency, keep the recorded information up to date, and perform unique movement notification operations in order to ensure the traceability of these devices.
Both ends of the supply chain are recorded. The second paragraph of Article 25 of the Medical Device Regulation says that economic operators must be able to identify to the Agency, for the period referred to in the eighth paragraph of Article 10, each economic operator to whom they have directly supplied a device, each economic operator who has directly supplied a device to them, and each healthcare institution or healthcare professional to whom they have directly supplied a device. The third paragraph of Article 31 of the Sales Regulation states that sales centers will keep, in a regular manner and on the basis of parameters such as batch, serial, lot, brand and model, the name, trade name or brand and contact information of the previous and, if any, next economic operator in the supply chain and other information that will facilitate tracking of the devices.
A Device Not Registered in ÜTS Cannot Be Sold
The commercial counterpart of registration shows here. The second paragraph of Article 26 of the Regulation on Sales, Advertising and Promotion of Medical Devices clearly states that sales centers cannot sell devices that are not registered in the Agency's record and information management system. A device whose registration is not complete cannot enter the sales chain in Türkiye.
The places where sales can be made are also restricted. The seventh paragraph of the same article states that sales centers are prohibited from selling devices to places not registered in the Agency's record and information management system, except for the places stated in the guide published by the Agency and for devices listed in Annex-3. The eighth paragraph says that in the implementation of administrative sanction decisions, all sales centers that place on the market and make available the product subject to the sanction are jointly and severally liable.
What to Do When a Device Is Declared Unsafe?
If a device is declared unsafe as a result of market surveillance, the obligation cannot be postponed. The second paragraph of Article 31 of the Regulation on Sales, Advertising and Promotion of Medical Devices states that sales centers must immediately stop the sale of devices that are found to be unsafe or contrary to technical regulations within the scope of market surveillance and inspection and the alert system and that are announced by the Agency, return these devices to the manufacturer or importer, and, if requested, share with the Agency all information and documents about the company, the user and the activities carried out on the matter relating to the device sold.
The first paragraph of the same article says that sales centers will immediately notify the Agency and the manufacturer or importer of the relevant device of the adverse events they encounter during their activities regarding their own devices or other devices.
Other Post-Registration Obligations: PMS, PSUR and Vigilance
An ÜTS registration does not fulfil your post-market surveillance obligation; it runs alongside it. The first paragraph of Article 81 of the Medical Device Regulation states that manufacturers will plan, establish, document, implement, maintain and update a post-market surveillance system for each device, proportionate to the risk class and appropriate for the type of device, and that this system is an integral part of the manufacturer's quality management system. Article 82 says this system will be based on a post-market surveillance plan whose requirement is stated in item 1 of Annex III.
Reporting differs by risk class, and this is a point that is often mixed up.
Class I: under the first paragraph of Article 83 of the Regulation, manufacturers of Class I devices prepare a post-market surveillance report, not a PSUR; this report is updated when necessary and submitted to the Agency upon request.
Class IIa: under sub-item (c) of the first paragraph of Article 84, manufacturers of Class IIa devices update the periodic safety update report when necessary and at least every 2 years.
Class IIb and Class III: under sub-item (b) of the same paragraph, manufacturers of these devices update the periodic safety update report at least once a year.
Custom-made devices: under sub-item (ç) of the same paragraph, for custom-made devices the PSUR becomes part of the documentation referred to in item 2 of Annex XIII.
The vigilance side is regulated in Articles 85 to 89 of the Regulation. The first paragraph of Article 85 states that manufacturers of devices on the market, except for investigational devices, will report to the Agency and the other relevant competent authorities any serious adverse event involving devices on the market and any field safety corrective action relating to devices lawfully on the market in Türkiye or EU member states; these reports will be submitted through the electronic system referred to in Article 89.
Document and Record Retention Periods
Retention periods vary with the type of device and the nature of the document. The list below brings together the periods in the sources.
Technical documentation, EU declaration of conformity and copies of certificates: under sub-item (a) of the eighth paragraph of Article 10 of the Medical Device Regulation, at least 10 years after the last device covered by the EU declaration of conformity has been placed on the market.
The same documents for implantable devices: under the same sub-item, at least 15 years after the last device has been placed on the market.
Custom-made device statement (Annex XIII ITC Statement): under the sixth paragraph of Article 29 of the Guide, at least 10 years after the device is placed on the market; at least 15 years for implantable devices.
Sales center activity documents: under the sixth paragraph of Article 8 of the Regulation on Sales, Advertising and Promotion of Medical Devices, five years; presented upon the request of the Agency or the directorate.
Supply chain records: under the third paragraph of Article 31 of the same Regulation, at least ten years from the date they start to make the device available on the market.
What post-registration obligations have in common is that they are tied not to a date but to an event: a certificate is revised, a packaging changes, a request lands for your opinion, a device leaves the portfolio. Companies can also carry out this follow-up by opening user authorization to a consultant from their own ÜTS company page; the consultant creates and approves the update records with the "Medical Device Registration Officer" authorization, while the e-signature remains with the company's authorized signatory and responsibility for the accuracy of the data remains with the company.
The initial setup of the registration is covered on the ÜTS Product (Medical Device) Registration page and the tracking of application statuses on the ÜTS Application Tracking page. Barcode and UDI rules are covered on the UDI-DI and Barcode Rules page, the exact list of mandatory fields on the ÜTS Device Registration Mandatory Fields page, and the GMDN and EMDN distinction on the GMDN and EMDN Code page.