Mandatory Fields in the ÜTS Device Registration
The first time you open the ÜTS device registration screen, dozens of fields appear, and it is not clear at a glance which are mandatory, which are conditional and which are filled in by the system
ÜTS Device Registration Mandatory Fields: List of Regulation, Manual and System Fields
When you open the ÜTS device registration screen for the first time, dozens of fields appear, and at first glance it is unclear which are mandatory, which are conditional and which are filled in by the system. Leaving a field empty keeps the record in draft; filling one in wrongly leads to a negative outcome for the application.
Based on Articles 25, 27 and 30 of the TİTCK Manual, this page gives the complete field lists of the registration screens for the old regulations (MDD/AIMDD/IVDD), the MDR and the IVDR. The four appendices below contain all the fields mentioned in the Manual, in the order of the source.
[APPENDIX: D-1] Old Regulations (MDD/AIMDD/IVDD) Mandatory Field List - ARTICLE 25
Article 25(1) of the Manual states that, during device registration under the old regulations, the information on the ÜTS registration screen is under nine headings: "Identifying Information" ("Tanımlayıcı Bilgiler"), "Import/Domestic Information" ("İthal/İmal Bilgileri"), "Classification Information" ("Sınıflandırma Bilgileri"), "Product Documents" ("Ürün Belgeleri"), "Product Images" ("Ürün Görselleri"), "Features" ("Özellikler"), "Dimension Information" ("Ölçü Bilgileri"), "Storage/Use Condition Information" ("Saklama/Kullanım Koşulu Bilgileri") and "Application Status" ("Başvuru Durumu"). The list below contains all the fields under these nine headings.
Identifying Information heading (paragraph 2 of the article):
• Primary Barcode Authority ("Birincil Barkod Kuruluşu"): Mandatory. The standard to which the barcode is allocated is selected; GS1, HIBCC and ICCBBA are authorised.
• Primary Product Number ("Birincil Ürün Numarası"): Mandatory. The barcode is entered exactly as it is, without any change; if it is not entered as it appears in the technical documents, the application ends with a negative result.
• Secondary Barcode Authority ("İkincil Barkod Kuruluşu"): If the device has barcodes from two different standards, the other standard, the one not entered as primary, is selected.
• Secondary Product Number ("İkincil Ürün Numarası"): Mandatory if a Secondary Barcode Authority has been selected. Again entered exactly; a mismatch leads to a negative outcome for the application.
• Brand Name ("Marka Adı"): Mandatory. The brand that appears on the label or documents is written; a device whose brand is not declared is not registered.
• Product Name ("Ürün Adı"): Mandatory. The specific name identifying the device is written in Turkish; a name that is missing, unintelligible or does not match the label leads to a negative outcome for the application.
• Version/Model ("Versiyon/Model"): Mandatory. Entered as it is written in the documents or on the label; if it is not expressly declared, "YOK" is written. The barcode is not entered in this field.
• Reference/Catalogue Number ("Referans/Katalog Numarası"): If it appears in the documents or on the label, it is entered as written; if there is more than one, one is written outside the parentheses and the others inside. The barcode is not entered.
• Product Description ("Ürün Açıklaması"): Optional. For configurable devices, the name, variant and reference number of each configuration must be written in this field.
• Number of Products in the Content ("İçerikteki Ürün Sayısı"): Mandatory. The quantity in the smallest packaging is entered; 1 for kits, and the number of pairs for devices used in pairs.
Import/Domestic Information heading (paragraph 3):
• Import/Domestic Information ("İthal/İmal Bilgisi"): Mandatory. If the legal manufacturer is established in Turkey, "Domestic" ("İmal") is selected; otherwise "Imported" ("İthal").
• Country of Import ("İthal Edilen Ülke"): Mandatory if "Imported" is selected. The country or countries from which the device is imported are selected.
• GTİP Number ("GTİP Numarası"): Mandatory if "Imported" is selected. If it is not in the system, a "GTİP Kodu Ekleme" (add GTİP code) request is opened.
• Country of Origin ("Menşei Ülke"): Mandatory. The country where the device is physically produced is selected; for contract manufacturing, the country of the manufacturer producing it is entered.
Classification Information heading (paragraph 4):
• Class ("Sınıf"): Mandatory. The suitable one among ten options is selected; for AIMDD devices "Class-III" is marked.
• GMDN: Mandatory. Each device has a single GMDN code; if the code description conflicts with the sterility and single-use information, the application ends with a negative result.
• Branch Type Code ("Branş Türü Kodu"): Mandatory. If there is no code that fully describes the device, a "Branş Türü Ekleme" (add branch type) request is opened.
Product Documents heading (paragraph 5):
• Documents ("Belgeler"): Mandatory. According to the document business rule suited to the class, documents in "Registered" ("Kayıtlı") status in ÜTS are selected; the class must have been selected before the selection.
Product Images heading (paragraph 6):
• Product Images ("Ürün Görselleri"): Mandatory. A photograph or technical drawing clearly showing the device is uploaded; at most 10. An unclear, unreadable image or one belonging to another device leads to a negative outcome for the application.
Features heading (paragraph 7):
• Is the Original Label in Turkish? ("Orijinal Etiket Türkçe mi?"): Mandatory. "Yes" if the label is in Turkish or in multiple languages including Turkish, "No" if it is in a foreign language.
• Original Label ("Orijinal Etiket"): Mandatory. The current, original label as placed on the market is uploaded, together with the inner/outer sales packaging if required.
• Turkish Label ("Türkçe Etiket"): Not filled in by the firm. If "Is the Original Label in Turkish?" is "No", it is created automatically by ÜTS after the application.
• MRI Safety Information ("MRG Güvenlik Bilgisi"): Mandatory. "MR Safe", "MR Conditional", "MR Unsafe" or "No Information Available" is selected; if there is no information, the last option is marked.
• Does it contain latex? ("Lateks içeriyor mu?"): Mandatory. If "Yes", a warning on the label and in the instructions for use is required.
• Does it contain phthalate/DEHP? ("Ftalat/DEHP içeriyor mu?"): Mandatory. If "Yes", a warning is sought on the label and in the instructions for use.
• Does it contain ionising radiation? ("İyonize radyasyon içerir mi?"): Mandatory. If the device emits or produces ionising radiation, "Yes" is selected and the information is in the instructions.
• Does it contain nanomaterials? ("Nanomateryal içerir mi?"): Mandatory. If "Yes", this information must be in the instructions for use.
• Can it be implanted in the body? ("Vücuda implante edilebilir mi?"): Mandatory. When "Yes" is selected, the single-use and limited-use fields fill in automatically and are locked.
• Is it single use? ("Tek kullanımlık mı?"): Mandatory. If "Yes", the information must be on both the label and the instructions; the field triggers auto-fill.
• Is there a limited number of uses? ("Sınırlı kullanım sayısı var mı?"): Mandatory. If "Yes", the Limited Number of Uses field opens; if "No", the single patient question opens.
• Limited Number of Uses ("Sınırlı Kullanım Sayısı"): Mandatory if the previous field is "Yes". For implantable or single-use devices, ÜTS writes "1" automatically and it cannot be changed.
• Can it be used on a single patient? ("Tek bir hastaya mı kullanılabilir?"): Mandatory if the single-use and limited-use questions are "No". It is different information from the single-use information.
• Does it have a shelf life? ("Raf ömrü var mı?"): Mandatory. If "Yes", the Shelf Life field opens.
• Shelf Life ("Raf Ömrü"): Mandatory if the previous field is "Yes". Entered in Days, Months or Years; the label of a device with a shelf life must carry an expiry date.
• Is it subject to calibration? ("Kalibrasyona tabi mi?"): Mandatory. "Yes" is selected if testing, checking and calibration are required.
• Calibration Period (months) ("Kalibrasyon Periyodu (ay)"): Mandatory if the previous field is "Yes". If the manufacturer has not specified a particular period and the instructions do not state one, 12 is entered.
• Is it subject to maintenance? ("Bakıma tabi mi?"): Mandatory. "Yes" is selected if maintenance is required at certain intervals.
• Maintenance Period (months) ("Bakım Periyodu (ay)"): Mandatory if the previous field is "Yes". The period stated in the instructions for use is entered in months.
• Was it packaged sterile? ("Steril paketlendi mi?"): Mandatory. If "Yes", the information must be on the label and in the instructions and the next field opens.
• Is the sterilisation method used in packaging known? ("Paketlemede kullanılan sterilizasyon yöntemi belli mi?"): Mandatory if sterile packaging is "Yes". If the method is known, "Yes" is selected.
• Sterilisation Method (Used in Packaging) ("Sterilizasyon Metodu (Paketlemede Kullanılan)"): Mandatory if the previous field is "Yes". All the methods used are selected from the list of fourteen methods.
• Is Sterilisation Required Before Use? ("Kullanım Öncesinde Sterilizasyon Gerekli mi?"): Mandatory. If "Yes", the information must be in the instructions for use; it is not sought on the label.
• Sterilisation Method (Before Use) ("Sterilizasyon Metodu (Kullanım Öncesi)"): Mandatory if the previous field is "Yes". A selection is made from the same list of fourteen methods.
• Is it within the scope of Annex-3 of the Medical Device Sales, Advertising and Promotion Regulation? ("Tıbbi Cihaz Satış, Reklam ve Tanıtım Yönetmeliği Ek-3 Kapsamında mı?"): Mandatory. If the device is among the products in the Annex-3 list, "Yes" is selected.
• Extra Information Link ("Ekstra Bilgi Linki"): Optional. A web address related to the device can be entered.
• SUT Matching Status ("SUT Eşleşme Durumu"): Mandatory. "Match Exists", "Match to Be Made" or "No Match" is selected; if a matching application will be made, uploading a Turkish catalogue is mandatory.
• Will it be sold in the health market / Will SGK reimbursement be made? ("Sağlık Marketinde Satılacak Mı / SGK Geri Ödemesi Yapılacak Mı?"): Mandatory. If "Yes", additional standards apply to the product images.
• Is it a component/accessory of a device? ("Bir cihazın bileşeni/aksesuarı mı?"): Mandatory. If "Yes", the Branch Types of the Main Device field opens and no instructions-for-use requirement is sought for Class IIa.
• Branch Types of the Main Device ("Ana Cihazın Branş Türleri"): Mandatory if the previous field is "Yes". At least 1 branch type code must be selected.
• Product Surface Area (in cm2) ("Ürün Yüzey Alanı"): Entered for devices that SGK pays for on the basis of surface area; may be left empty for other devices.
Dimension Information heading (paragraph 8):
• Dimension List ("Ölçü Listesi"): If there is any, dimension information is added with units; eleven dimension types are defined, and more than one dimension of the same type cannot be added.
Storage/Use Condition Information heading (paragraph 9):
• Is there a storage and/or use condition? ("Saklama ve/veya Kullanım Koşulu var mı?"): Mandatory. If there is a condition, "Yes" is selected and the list field opens.
• Storage and/or Use Condition List ("Saklama ve/veya Kullanım Koşulu Listesi"): Mandatory if the previous field is "Yes". For each condition, the type, minimum value, maximum value, unit and special condition are entered.
Application Status heading (paragraph 10):
• Is it ready for application? ("Başvuruya hazır mı?"): Mandatory. If "No" is selected the record is saved as "Draft" ("Taslak"); if "Yes", as "Ready for Application" ("Başvuruya Hazır").
[APPENDIX: D-2] MDR Mandatory Field List - ARTICLE 27
Article 27(1) of the Manual states that, during device registration under the MDR, the information is under ten headings. In addition to the nine headings on the old regulation screen, there is a "CMR and/or Endocrine-Disrupting Substance Information" ("CMR ve/veya Endokrin Bozucu Madde Bilgileri") block.
Identifying Information heading (paragraph 2):
• UDI Issuing Entity ("UDI Tahsis Kuruluşu"): Mandatory. GS1, HIBCC, ICCBBA and IFA are authorised.
• UDI-DI Number ("UDI-DI Numarası"): Mandatory. The barcode is entered exactly; if it is not entered as it appears in the technical documents, the application ends with a negative result.
• Did it transition to the MDR with a different product number? ("MDR'a farklı ürün numarası ile mi geçti?"): Mandatory. "Yes" is selected if the device was previously registered under the MDD/AIMDD with a different barcode.
• Product Number under 93/42/EEC or 90/385/EEC ("93/42/EEC veya 90/385/EEC'deki Ürün Numarası"): Mandatory if the previous field is "Yes". A single barcode is selected from the old records.
• Products with the Same Barcode ("Aynı Barkodlu Ürünler"): A display field. Other firms' records registered with the same UDI-DI are listed.
• Brand Name ("Marka Adı"): Mandatory. A device without a brand is not registered; title suffixes such as GmbH, Inc, A.Ş. are not written.
• Product Name / Trade Name ("Ürünün Adı / Ticari Adı"): Mandatory. Written in Turkish; a name that is missing, unintelligible or does not match the label leads to rejection of the application.
• Additional trade names of the product, if any ("Ürünün ilave ticari adları"): Entered if the device is also supplied under other trade names; a change of name resulting from translation into EU languages requires a new UDI-DI.
• Version/Model ("Versiyon/Model"): Mandatory if Reference/Catalogue Number has been left empty. If it is not expressly declared, it is left empty; "YOK" is not written.
• Reference/Catalogue Number ("Referans/Katalog Numarası"): Mandatory if Version/Model has been left empty. If there is more than one, one is written outside the parentheses and the others inside.
• Product Description ("Ürün Açıklaması"): Optional. For configurable devices, the information on each configuration must be entered in this field.
• Number of Products in the Content ("İçerikteki Ürün Sayısı"): Mandatory. The quantity in the smallest packaging is entered.
Import/Domestic Information heading (paragraph 3):
• Import/Domestic Information ("İthal/İmal Bilgisi"): Mandatory. "Domestic" or "Imported" is selected according to the country where the legal manufacturer is established.
• Country of Origin ("Menşei Ülke"): Mandatory. The country where the device is physically produced; for contract manufacturing, the country of the producing firm is selected.
• Country of Import ("İthal Edilen Ülke"): Mandatory if "Imported" is selected.
• GTİP Number ("GTİP Numarası"): Mandatory if "Imported" is selected. If it is not in the system, a request is opened.
Classification Information heading (paragraph 4):
• Class ("Sınıf"): Mandatory. Selected among Class-I Other, Class-Ir, Class-Im, Class-Is, Class-IIa, Class-IIb and Class-III.
• GMDN: Mandatory. Each device has a single GMDN code.
• Branch Type Code ("Branş Türü Kodu"): Mandatory. If there is no code, a "Branş Türü Ekleme" (add branch type) request is opened.
• EMDN Term ("EMDN Terim"): Optional until the Agency takes a new decision and if the code does not appear in the documents; if it appears in the documents, selecting the most detailed terminal term is mandatory.
• Selected EMDN Term ("Seçilen EMDN Terim"): The information on the selected EMDN code appears in this field; a wrong selection is removed with "Remove" ("Çıkar").
Product Documents heading (paragraph 5):
• Documents ("Belgeler"): Mandatory. Documents in "Registered" status are selected according to the document business rule suited to the class; the class must be selected first.
• Selected Documents ("Seçilen Belgeler"): The information on the selected documents appears; a wrong selection is removed with "Remove".
• Basic UDI-DI ("Temel UDI-DI"): Mandatory. To make a selection, the relevant EU Declaration of Conformity must be selected first; more than one selection cannot be made.
Product Images heading (paragraph 6):
• Product Images ("Ürün Görselleri"): Mandatory. At most 10 images are uploaded; an unclear or unreadable image leads to a negative outcome for the application.
Features heading (paragraph 7):
• Is the Original Label in Turkish? ("Orijinal Etiket Türkçe mi?"): Mandatory. If "No" is selected, the Turkish Label template is created automatically by ÜTS.
• Original Label and Sales Packaging ("Orijinal Etiket ve Satış Ambalajı"): Mandatory. The current, original label as placed on the market is uploaded, together with the inner/outer sales packaging if required.
• Turkish Label ("Türkçe Etiket"): Not filled in by the firm. It is created automatically by ÜTS after the application and does not replace the original label.
• MRI Safety Information ("MRG Güvenlik Bilgisi"): Mandatory. One of four options is marked; if "MR Conditional", a detailed explanation is sought in the instructions.
• Does it contain latex? ("Lateks içeriyor mu?"): Mandatory. If the word latex appears in the product name or documents, it must also appear on the label and in the instructions.
• Does it contain phthalate/DEHP? ("Ftalat/DEHP içeriyor mu?"): Mandatory. If "Yes", a warning on the label and in the instructions is required.
• Does it contain ionising radiation? ("İyonize radyasyon içerir mi?"): Mandatory. If "Yes", the information must be in the instructions for use.
• Does it contain nanomaterials? ("Nanomateryal içerir mi?"): Mandatory. If "Yes", the type field opens and the information appears in the instructions.
• Nanomaterial Type ("Nanomateryal Türü"): Mandatory if nanomaterial is "Yes". "Particle", "Aggregate", "Agglomerate" or "Unknown" is selected.
• Can it be implanted in the body? ("Vücuda implante edilebilir mi?"): Mandatory. When "Yes" is selected, the linked fields fill in automatically and are locked.
• Is it single use? ("Tek kullanımlık mı?"): Mandatory. If "Yes", the information must be on both the label and the instructions.
• Is there a limited number of uses? ("Sınırlı kullanım sayısı var mı?"): Mandatory. If "Yes", the number field opens; if "No", the single patient question opens.
• Limited Number of Uses ("Sınırlı Kullanım Sayısı"): Mandatory if the previous field is "Yes". For implantable or single-use devices, "1" is written automatically.
• Can it be used on a single patient? ("Tek bir hastaya mı kullanılabilir?"): Mandatory if the single-use and limited-use questions are "No".
• Does it have a shelf life? ("Raf ömrü var mı?"): Mandatory. If "Yes", the Shelf Life field opens.
• Shelf Life ("Raf Ömrü"): Mandatory if the previous field is "Yes". Days, Months or Years is entered; the expiry date must be on the label, the sales packaging and/or the UDI carrier.
• Is it subject to calibration? ("Kalibrasyona tabi mi?"): Mandatory. If "Yes", the period field opens.
• Calibration Period (months) ("Kalibrasyon Periyodu (ay)"): Mandatory if the previous field is "Yes". If the manufacturer has not specified one and the instructions do not state it, 12 is entered.
• Is it subject to maintenance? ("Bakıma tabi mi?"): Mandatory. If "Yes", the maintenance period field opens.
• Maintenance Period (months) ("Bakım Periyodu (ay)"): Mandatory if the previous field is "Yes". The period must be stated in the instructions for use.
• Was it packaged sterile? ("Steril paketlendi mi?"): Mandatory. If "Yes", the sterilisation method field opens directly.
• Sterilisation Method (Used in Packaging) ("Sterilizasyon Metodu (Paketlemede Kullanılan)"): Mandatory if sterile packaging is "Yes". All the methods used are selected from the list of fourteen methods.
• Is Sterilisation Required Before Use? ("Kullanım Öncesinde Sterilizasyon Gerekli mi?"): Mandatory. If "Yes", the information must be in the instructions for use.
• Sterilisation Method (Before Use) ("Sterilizasyon Metodu (Kullanım Öncesi)"): Mandatory if the previous field is "Yes". All the methods are selected from the same list.
• Does it contain a substance considered a medicinal product? ("Tıbbi ürün olarak kabul edilen bir madde ihtiva ediyor mu?"): Mandatory. If "Yes", the device is considered Class III and the name of the substance appears on the label and in the instructions.
• Does it contain a medicinal product derived from human blood or plasma? ("İnsan kanı veya plazmasından elde edilen bir tıbbi ürün ihtiva ediyor mu?"): Mandatory. If "Yes", the device is Class III and the substance name is on the label and in the instructions.
• Does it contain tissues or cells of human origin or their derivatives? ("İnsan kaynaklı dokular veya hücreler veya onların türevlerini ihtiva ediyor mu?"): Mandatory. If "Yes", the device is Class III.
• Does it contain tissues or cells of animal origin or their derivatives, according to (EU) 722/2012? ("(AB) 722/2012'ye göre, Hayvan kaynaklı dokular veya hücreler veya onların türevlerini ihtiva ediyor mu?"): Mandatory. If "Yes", the device is Class III, except for those that contact only intact skin.
• Is it within the scope of Annex-3 of the Medical Device Sales, Advertising and Promotion Regulation? ("Tıbbi Cihaz Satış, Reklam ve Tanıtım Yönetmeliği Ek-3 Kapsamında mı?"): Mandatory. If the device is among the products in the Annex-3 list, "Yes" is selected.
• Extra Information Link ("Ekstra Bilgi Linki"): Optional. A web address related to the device can be entered.
• SUT Matching Status ("SUT Eşleşme Durumu"): Mandatory. One of three options is marked; if a matching application will be made, a Turkish catalogue is mandatory.
• Will it be sold in the health market / Will SGK reimbursement be made? ("Sağlık Marketinde Satılacak Mı / SGK Geri Ödemesi Yapılacak Mı?"): Mandatory. If "Yes", the image standards clause comes into effect.
• Product Image Standards ("Ürün Görseli Standartları"): Applies if the previous field is "Yes". For each device at least 3 photographs, at least 600x600 pixels, square, at most 2 MB.
• Is it a component/accessory of a device? ("Bir cihazın bileşeni/aksesuarı mı?"): Mandatory. If "Yes", the Branch Types of the Main Device field opens.
• Branch Types of the Main Device ("Ana Cihazın Branş Türleri"): Mandatory if the previous field is "Yes". At least 1 branch type code is selected.
• Product Surface Area (in cm2) ("Ürün Yüzey Alanı"): Entered for devices that SGK pays for on the basis of surface area; may be left empty for others.
• Was the product manufactured by another manufacturer on behalf of its legal manufacturer? ("Ürün, yasal imalatçısı tarafından başka bir imalatçıya mı ürettirildi?"): Mandatory. "Yes" is selected if there is contract design or manufacturing.
• Manufacturing Firm/Institution ("Üretimi Yapan İmalatçı Firma/Kurum"): Mandatory if the previous field is "Yes". If the firm is not on the list, a record is opened with a "Secondary Firm No" ("İkincil Firma No") by entering the name, address, country and e-mail.
Dimension Information heading (paragraph 8):
• Dimension List ("Ölçü Listesi"): If there are any, dimensions are added with units; eleven dimension types are defined and more than one dimension of the same type cannot be added.
Storage/Use Condition Information heading (paragraph 9):
• Is there a storage and/or use condition? ("Saklama ve/veya Kullanım Koşulu var mı?"): Mandatory. If "Yes", the condition list field opens.
• Storage and/or Use Condition List ("Saklama ve/veya Kullanım Koşulu Listesi"): Mandatory if the previous field is "Yes". The type, minimum value, maximum value, unit and special condition are entered.
• Are there other special conditions? ("Diğer özel koşullar var mı?"): Mandatory. "Yes" is selected if there is a warning, precaution or special condition that should draw the user's attention.
• Other Special Conditions ("Diğer Özel Koşullar"): Mandatory if the previous field is "Yes". Several warnings and conditions can be selected from the list; the information must be in the instructions for use.
CMR and/or Endocrine-Disrupting Substance Information heading (paragraph 10):
• Does it contain CMR and/or endocrine-disrupting substances as specified in Annex-I 10.4.1? ("Ek-I 10.4.1'de belirtildiği şekilde, CMR ve/veya endokrin bozucu madde ihtiva ediyor mu?"): Mandatory. "Yes" is selected if there is a substance within the scope of point 10.4.1 of MDR Annex-I.
• CMR and/or Endocrine-Disrupting Substance Information ("CMR ve/veya Endokrin Bozucu Madde Bilgileri"): Mandatory if the previous field is "Yes". Substances are selected from a list based on ECHA data; for a missing substance a "CMR / ED Madde Ekleme" (add CMR / ED substance) request is opened.
Application Status heading (paragraph 11):
• Is it ready for application? ("Başvuruya hazır mı?"): Mandatory. If "No", the record is saved as "Draft"; if "Yes", as "Ready for Application".
[APPENDIX: D-3] IVDR Mandatory Field List - ARTICLE 30
Article 30(1) of the Manual states that on the registration screen under the IVDR the information is under the same ten headings as for the MDR. The field contents are different, however: the questions on implantability, latex, phthalate and nanomaterials are not on this screen, but fields specific to IVDs are added.
Identifying Information heading (paragraph 2):
• UDI Issuing Entity ("UDI Tahsis Kuruluşu"): Mandatory. GS1, HIBCC, ICCBBA and IFA are authorised.
• UDI-DI Number ("UDI-DI Numarası"): Mandatory. The barcode is entered exactly; if it is not entered as it appears in the documents, the application ends with a negative result.
• Did it transition to the IVDR with a different product number? ("IVDR'a farklı ürün numarası ile mi geçti?"): Mandatory. "Yes" is selected if the device was previously registered under the IVDD with a different barcode.
• Product Number under 98/79/EC ("98/79/EC'deki Ürün Numarası"): Mandatory if the previous field is "Yes". A single barcode is selected from the old IVDD records.
• Products with the Same Barcode ("Aynı Barkodlu Ürünler"): A display field. Other firms' records registered with the same UDI-DI are listed.
• Brand Name ("Marka Adı"): Mandatory. A device without a brand is not registered; abbreviations of commercial status are not written.
• Product Name / Trade Name ("Ürünün Adı / Ticari Adı"): Mandatory. Written in Turkish; a name that is missing or does not match the label leads to rejection of the application.
• Additional trade names of the product, if any ("Ürünün ilave ticari adları"): Entered if it is also supplied under other trade names; a change resulting from translation into EU languages requires a new UDI-DI.
• Version/Model ("Versiyon/Model"): Mandatory if Reference/Catalogue Number has been left empty. "YOK" is not written; the model name in the Domestic Product Certificate is entered.
• Reference/Catalogue Number ("Referans/Katalog Numarası"): Mandatory if Version/Model has been left empty. If there is more than one, one is written outside the parentheses and the others inside.
• Product Description ("Ürün Açıklaması"): Optional. Additional packaging and label explanations that the review expert needs to know are written here.
• Number of Products in the Content ("İçerikteki Ürün Sayısı"): Mandatory. The quantity in the smallest packaging is entered.
Import/Domestic Information heading (paragraph 3):
• Import/Domestic Information ("İthal/İmal Bilgisi"): Mandatory. Selected according to the country where the legal manufacturer is established.
• Country of Origin ("Menşei Ülke"): Mandatory. The country or countries where the device is physically produced are selected.
• Country of Import ("İthal Edilen Ülke"): Mandatory if "Imported" is selected.
• GTİP Number ("GTİP Numarası"): Mandatory if "Imported" is selected.
Classification Information heading (paragraph 4):
• Class ("Sınıf"): Mandatory. The suitable one is selected from a list of thirteen options.
• GMDN: Mandatory. Each device has a single GMDN code.
• Branch Type Code ("Branş Türü Kodu"): Mandatory. If there is no code describing the device, a request is opened.
• EMDN Term ("EMDN Terim"): Optional until the Agency takes a new decision and if the code does not appear in the documents; if it appears in the documents, the most detailed term is selected.
• Selected EMDN Term ("Seçilen EMDN Terim"): The information on the selected code appears in this field.
Product Documents heading (paragraph 5):
• Documents ("Belgeler"): Mandatory. The selection is made according to the document business rule suited to the class; the class must be selected first.
• Selected Documents ("Seçilen Belgeler"): The information on the selected documents appears; a wrong selection is removed with "Remove" ("Çıkar").
• Basic UDI-DI ("Temel UDI-DI"): Mandatory. The relevant EU Declaration of Conformity and EU Certificate must be selected first; more than one selection cannot be made.
Product Images heading (paragraph 6):
• Product Images ("Ürün Görselleri"): Mandatory. A photograph or technical drawing clearly showing the device is uploaded.
Features heading (paragraph 7):
• Is the Original Label in Turkish? ("Orijinal Etiket Türkçe mi?"): Mandatory. If "No" is selected, the Turkish Label template is created by ÜTS.
• Original Label and Sales Packaging ("Orijinal Etiket ve Satış Ambalajı"): Mandatory. The current original label and, if required, the inner/outer sales packaging are uploaded.
• Turkish Label ("Türkçe Etiket"): Not filled in by the firm; it is created automatically by ÜTS and does not replace the original label.
• MRI Safety Information ("MRG Güvenlik Bilgisi"): Mandatory. One of four options is marked.
• Does it contain ionising radiation? ("İyonize radyasyon içerir mi?"): Mandatory. "Yes" is selected if the device emits or produces ionising radiation.
• Is it single use? ("Tek kullanımlık mı?"): Mandatory. If "Yes", the linked fields fill in automatically and are locked.
• Is there a limited number of uses? ("Sınırlı kullanım sayısı var mı?"): Mandatory. If "Yes", the number field opens; if "No", the single patient question opens.
• Limited Number of Uses ("Sınırlı Kullanım Sayısı"): Mandatory if the previous field is "Yes". For single-use devices "1" is written automatically.
• Can it be used on a single patient? ("Tek bir hastaya mı kullanılabilir?"): Mandatory if the single-use and limited-use questions are "No".
• Does it have a shelf life? ("Raf ömrü var mı?"): Mandatory. If "Yes", the Shelf Life field opens.
• Shelf Life ("Raf Ömrü"): Mandatory if the previous field is "Yes". Days, Months or Years is entered; the expiry date must be on the label and/or the UDI carrier.
• Is it subject to calibration? ("Kalibrasyona tabi mi?"): Mandatory. If "Yes", the period field opens.
• Calibration Period (months) ("Kalibrasyon Periyodu (ay)"): Mandatory if the previous field is "Yes". If the manufacturer has not specified one and the instructions do not state it, 12 is entered.
• Is it subject to maintenance? ("Bakıma tabi mi?"): Mandatory. If "Yes", the maintenance period field opens.
• Maintenance Period (months) ("Bakım Periyodu (ay)"): Mandatory if the previous field is "Yes". The period must be stated in the instructions for use.
• Was it packaged sterile? ("Steril paketlendi mi?"): Mandatory. If "Yes", the sterilisation method field opens.
• Sterilisation Method (Used in Packaging) ("Sterilizasyon Metodu (Paketlemede Kullanılan)"): Mandatory if sterile packaging is "Yes". A selection is made from the list of fourteen methods.
• Is Sterilisation Required Before Use? ("Kullanım Öncesinde Sterilizasyon Gerekli mi?"): Mandatory. If "Yes", the information must be in the instructions for use.
• Sterilisation Method (Before Use) ("Sterilizasyon Metodu (Kullanım Öncesi)"): Mandatory if the previous field is "Yes". All the methods are selected from the same list.
• Does it contain cells or substances of microbial origin? ("Mikrobiyal kaynaklı hücreler veya maddeler ihtiva ediyor mu?"): Mandatory. If "Yes", the information must be on the label and/or the sales packaging and in the instructions.
• Does it contain tissues or cells of human origin or their derivatives? ("İnsan kaynaklı dokular veya hücreler veya onların türevlerini ihtiva ediyor mu?"): Mandatory. If "Yes", the information appears on the label and in the instructions.
• Does it contain tissues or cells of animal origin or their derivatives, according to (EU) 722/2012? ("(AB) 722/2012'ye göre, Hayvan kaynaklı dokular veya hücreler veya onların türevlerini ihtiva ediyor mu?"): Mandatory. If "Yes", the information appears on the label and in the instructions.
• Is it a new type of device? ("Yeni bir tür cihaz mıdır?"): Mandatory. "Yes" is selected if it is considered a new device under the last 3 years criterion.
• Is it a device for self-testing or near-patient testing? ("Kişisel test veya hastabaşı test amaçlı cihaz mıdır?"): Mandatory. If "Yes", the indication must be on the label and/or the sales packaging and in the instructions.
• Is it within the scope of Annex-3 of the Medical Device Sales, Advertising and Promotion Regulation? ("Tıbbi Cihaz Satış, Reklam ve Tanıtım Yönetmeliği Ek-3 Kapsamında mı?"): Mandatory. Since there are no devices under the IVDR in Annex-3, selecting "No" is mandatory.
• Extra Information Link ("Ekstra Bilgi Linki"): Optional. A web address related to the device can be entered.
• SUT Matching Status ("SUT Eşleşme Durumu"): Mandatory. "Match Exists", "Match to Be Made" or "No Match" is selected.
• Will it be sold in the health market / Will SGK reimbursement be made? ("Sağlık Marketinde Satılacak Mı / SGK Geri Ödemesi Yapılacak Mı?"): Mandatory. If "Yes", the image standards apply.
• Product Image Standards ("Ürün Görseli Standartları"): Applies if the previous field is "Yes". For each device at least 3 photographs and the specified size requirements are sought.
• Is it a component/accessory of a device? ("Bir cihazın bileşeni/aksesuarı mı?"): Mandatory. If "Yes", the Branch Types of the Main Device field opens.
• Branch Types of the Main Device ("Ana Cihazın Branş Türleri"): Mandatory if the previous field is "Yes". At least 1 branch type code is selected.
• Product Surface Area (in cm2) ("Ürün Yüzey Alanı"): Entered for devices that SGK pays for on the basis of surface area.
• Was the product manufactured by another manufacturer on behalf of its legal manufacturer? ("Ürün, yasal imalatçısı tarafından başka bir imalatçıya mı ürettirildi?"): Mandatory. "Yes" is selected if there is contract design or manufacturing.
• Manufacturing Firm/Institution ("Üretimi Yapan İmalatçı Firma/Kurum"): Mandatory if the previous field is "Yes". If the firm is not on the list, a record is opened with a "Secondary Firm No" ("İkincil Firma No").
Dimension Information heading (paragraph 8):
• Dimension List ("Ölçü Listesi"): If there are any, dimensions are added with units; more than one dimension of the same type cannot be added.
Storage/Use Condition Information heading (paragraph 9):
• Is there a storage and/or use condition? ("Saklama ve/veya Kullanım Koşulu var mı?"): Mandatory. If "Yes", the condition list field opens.
• Storage and/or Use Condition List ("Saklama ve/veya Kullanım Koşulu Listesi"): Mandatory if the previous field is "Yes". The type, minimum and maximum values, unit and special condition are entered.
• Are there other special conditions? ("Diğer özel koşullar var mı?"): Mandatory. "Yes" is selected if there is a warning, precaution or special condition.
• Other Special Conditions ("Diğer Özel Koşullar"): Mandatory if the previous field is "Yes". Several conditions can be selected from the list.
CMR and/or Endocrine-Disrupting Substance Information heading (paragraph 10):
• Does it contain CMR and/or endocrine-disrupting substances as specified in Annex-I 10.3? ("Ek-I 10.3'te belirtildiği şekilde, CMR ve/veya endokrin bozucu madde ihtiva ediyor mu?"): Mandatory. "Yes" is selected if there is a substance within the scope of point 10.3 of IVDR Annex-I.
• CMR and/or Endocrine-Disrupting Substance Information ("CMR ve/veya Endokrin Bozucu Madde Bilgileri"): Mandatory if the previous field is "Yes". Substances are selected from a list based on ECHA data.
Application Status heading (paragraph 12):
• Is it ready for application? ("Başvuruya hazır mı?"): Mandatory. If "No", the record is saved as "Draft"; if "Yes", as "Ready for Application".
[APPENDIX: D-4] Comparison of Field Differences Across the Three Frameworks
The differences between the three screens are easy to miss when read one by one. The list below shows the fields that exist in one framework and not in another, and how the same field behaves differently depending on the framework.
• Basic UDI-DI: A mandatory field in the Product Documents block in MDR and IVDR records; it does not exist at all on the old regulation screen.
• EMDN Term and Selected EMDN Term: Found in the Classification block of the MDR and IVDR; not on the old regulation screen. GMDN is separately mandatory in all three frameworks.
• Nanomaterial Type: Only on the MDR screen; if nanomaterial is "Yes", Particle, Aggregate, Agglomerate or Unknown is selected.
• CMR and/or Endocrine-Disrupting Substance Information block: A separate heading on the MDR and IVDR screens; absent from the old regulation screen.
• Contract manufacturing fields: The fields "Was the product manufactured by another manufacturer on behalf of its legal manufacturer?" and "Manufacturing Firm/Institution" exist on the MDR and IVDR screens; not on the old regulation screen.
• Cells or substances of microbial origin: A mandatory field found only on the IVDR screen.
• Is it a new type of device?: Only on the IVDR screen and based on the last 3 years criterion.
• Is it a device for self-testing or near-patient testing?: Found only on the IVDR screen.
• Is the sterilisation method used in packaging known?: An intermediate question only on the old regulation screen; in the MDR and IVDR, when sterile packaging is "Yes" the method field opens directly.
• Does it contain latex?: Mandatory on the old regulation and MDR screens; not on the IVDR screen.
• Does it contain phthalate/DEHP?: Mandatory on the old regulation and MDR screens; not on the IVDR screen.
• Does it contain nanomaterials?: Mandatory on the old regulation and MDR screens; not on the IVDR screen.
• Can it be implanted in the body?: Mandatory on the old regulation and MDR screens; not on the IVDR screen.
• Medicinal product, human blood/plasma derivative and human-origin tissue fields: Four separate mandatory fields on the MDR screen; on the IVDR there are only the human- and animal-origin tissue fields, and none exists on the old regulation screen.
• Secondary Barcode Authority and Secondary Product Number: Only on the old regulation screen; the MDR and IVDR screens have a single UDI-DI field.
• Leaving Version/Model empty: Under the old regulations the field is mandatory and "YOK" is written if there is no declaration; under the MDR and IVDR it can be left empty if Reference/Catalogue Number is filled in, and "YOK" is not written.
• Are there other special conditions? / Other Special Conditions: In the Storage/Use Condition block of the MDR and IVDR; not on the old regulation screen.
• Is it within the scope of Annex-3?: Mandatory in all three frameworks; in IVDR records, selecting "No" is mandatory.
Conditionally Mandatory Fields: Which Answer Opens Which Field?
Not seeing a field on the screen does not mean it does not exist; most fields appear depending on the previous answer. Point (ı) of Article 25(7) of the Manual says that if the "Is there a limited number of uses?" field is selected as "Yes", the Limited Number of Uses field in point (i) appears and the number stated on the label and in the instructions is entered; if "No" is selected, the single patient question in point (j) opens. The same structure is set up on the MDR side in points (i) and (j) of Article 27(7).
• "Is there a limited number of uses?" - Yes: The Limited Number of Uses field opens and filling it in becomes mandatory.
• "Is it single use?" and "Is there a limited number of uses?" - No: The "Can it be used on a single patient?" field opens and a selection is mandatory.
• "Does it have a shelf life?" - Yes: The Shelf Life field opens; points (k) and (l) of Article 25(7).
• "Is it subject to calibration?" - Yes: The Calibration Period field opens; points (m) and (n) of the same paragraph.
• "Is it a component/accessory of a device?" - Yes: The Branch Types of the Main Device field opens and at least 1 code must be selected; points (z) and (aa) of Article 25 and points (dd) and (ee) of Article 27.
Fields That Fill in Automatically and Lock
Some fields are not filled in by you; the system fills them in and does not allow you to change them. The third sub-point of point (ğ) of Article 25(7) of the Manual says that when the "Can it be implanted in the body?" field is selected as "Yes", ÜTS automatically selects "Yes" for the single-use and limited number of uses fields, sets the limited number of uses to "1" and selects "No" for the calibration and maintenance fields, and that the fields concerned cannot be changed. The same auto-fill is applied, under the fourth sub-point of point (h), when the "Is it single use?" field is selected as "Yes"; the MDR counterparts are points (h) and (ı) of Article 27(7), and the IVDR counterpart is point (e) of Article 30(7).
The Turkish Label field is also not filled in by the firm. Point (c) of Article 25(7) of the Manual states that when the "Is the Original Label in Turkish?" field is selected as "No", this field is created automatically by ÜTS after the application is made, that firms will not upload a Turkish label separately, and that the template created does not replace the original label; the same point of Article 27 repeats the provision for the MDR.
Even the Class field is not visible on the screen in some records. Point (a) of Article 28(6) of the Manual says that in an MDR SİP record the Class field is not filled in by the firm and is not visible on the screen; because a Basic UDI-DI must be assigned to the SİP, the class of the highest-risk device among the MDR devices making up the SİP is selected automatically by ÜTS.
Version/Model and Reference/Catalogue Number Cannot Both Be Left Empty
On the MDR and IVDR screens these two fields are linked to each other. The seventh sub-points of points (ğ) and (h) of Article 27(2) of the Manual say that the Version/Model and Reference/Catalogue Number fields cannot both be left empty, that "YOK" is not written in these fields, and that information must be entered in at least one of the two fields. On the IVDR side the same provision is in the seventh sub-point of point (ğ) of Article 30(2).
The barcode is not entered in these fields under any condition. The sixth sub-point of point (f) of Article 25(2) of the Manual says that the Primary or Secondary Product Number is not entered in this field; the same points of Articles 27 and 30 impose the same prohibition for the UDI-DI Number. For devices with more than one reference number, under the fourth sub-point of point (g) of Article 25(2) one of them is written outside the parentheses and the others inside; the MDR and IVDR counterparts are the fifth sub-point of point (h) of Articles 27 and 30.
What Is Written If There Is No Model Information?
The answer depends on the framework, and this difference is often missed. In an old regulation record, under the fifth sub-point of point (f) of Article 25(2) of the Manual, if the version/model of the device is not expressly declared in the documents and/or on its label, "YOK" is written in this field. In an MDR record, under the fifth sub-point of point (ğ) of Article 27(2), the field is left empty; the seventh sub-point of the same point expressly prohibits writing "YOK". An exception applies in both new frameworks: the sixth sub-point of point (ğ) of Articles 27 and 30 says that if the model name appears in the Domestic Product Certificate mentioned in Article 22 of the Manual, entering this name is mandatory.
Brand Name Field: A Device Without a Brand Is Not Registered
The brand field cannot be skipped. The second sub-point of point (d) of Article 25(2) of the Manual says that, for the sake of product safety, the ÜTS registration of devices for which no brand is declared will not be carried out. The second sub-point of point (e) of Article 27(2) repeats the same provision for the MDR.
If the brand information is on the label or in the documents, it is entered as written; if not, a third route is defined. The fourth sub-point of the same points says that if the brand information is not on the label and in at least one of the documents linked to the device, the name of the manufacturer firm is entered in this field, without expressions indicating its commercial status.
Do Not Write the Company Title and Company Suffix in the Brand Field
The same sub-point also sets a prohibition: abbreviations expressing the commercial status of the firm, such as GmbH, Inc, A.Ş., found in company titles, are not included in the brand field. This provision is word for word the same in the fourth sub-points of point (d) of Article 25(2), point (e) of Article 27(2) and point (e) of Article 30(2) of the Manual. In practice, "XYZ Medical" is written instead of "XYZ Medical GmbH".
Product Name and Trade Name: Writing Rules
The name of the field varies by framework. The old regulation screen has a single "Product Name" ("Ürün Adı") field. On the MDR and IVDR screens the field is defined as "Product Name / Trade Name" ("Ürünün Adı / Ticari Adı"); the third and fourth sub-points of point (f) of Article 27(2) of the Manual describe the two separately: the product name is the name that helps identify the device and is used in common for devices of the same type; the trade name is the branded, commercialised or registered name that describes the type, model or version of the device.
A name error is a direct ground for rejection. The fourth sub-point of point (e) of Article 25(2) of the Manual says that the application for a device whose product name does not make clear what the device is, is incomplete, is not in Turkish or is not entered in a way that matches what is shown on the label will end with a negative result; the sixth sub-points of point (f) of Articles 27 and 30 state that in the same situation the application will be rejected.
Branded Trade Names Are Not Translated
The requirement to render names in Turkish has a limit. The ninth sub-point of point (f) of Article 27(2) and Article 30(2) of the Manual says that the branded, commercialised or registered parts of the trade names of the device cannot be translated into Turkish and that whether these parts are in Turkish will not be taken into consideration; as an example, the Turkish trade name of a device named "DoX-GloveTM Surgical Gloves" is entered as "DoX-GloveTM Cerrahi Eldivenler". If the device is also supplied under other trade names, under point (g) of the same articles these names are written in the "Additional trade names of the product, if any" field.
Branch Type Code and Branch Types of the Main Device
The branch type code is a field that cannot be skipped in any of the three frameworks. Point (c) of Article 25(4), point (c) of Article 27(4) and point (c) of Article 30(4) of the Manual state that selecting this field in ÜTS is mandatory and that, if there is no option fully describing the device, the branch type that describes it most appropriately is selected.
If there is no suitable code in the list at all, a route is defined. The third sub-point of the same points says that if there is no branch type describing the device, a request is opened through the "Branş Türü Ekleme" (add branch type) field using the "Report Request" screen in ÜTS, and if the request is found suitable the code concerned will be defined. If the device is an accessory, a second branch type selection is made: at least 1 code must be selected in the Branch Types of the Main Device field.
Features Block: Sterility, Shelf Life, Calibration and Maintenance
On the sterility side there are two separate questions, and one does not replace the other. Point (p) of Article 25(7) of the Manual asks whether the device is supplied in sterile packaging, and point (ş) asks whether it must be sterilised before use; the MDR counterparts are points (r) and (ş) of Article 27(7). It is possible for a device packaged sterile to require sterilisation before use.
Shelf life is directly tied to the label check. Point (l) of Article 25(7) of the Manual says that shelf life is entered as "Days", "Months" or "Years" and that the original label or sales packaging of devices with a shelf life must carry an expiry date; point (m) of Article 27 and point (ı) of Article 30 repeat the same rule in a way that also covers the UDI carrier. "Single use" and "can be used on a single patient" are different pieces of information: the third sub-point of point (j) of Article 25(7) says that in the single patient question a selection is made only for devices that are not single use, and that if the device is single use this field does not appear on the screen at all.
If No Calibration Period Is Stated, 12 Months
This field cannot be left empty and the Manual defines a default value. The fourth sub-point of point (n) of Article 25(7) of the Manual says that if the calibration period of a device has not been specifically determined by the manufacturer and is not stated in the instructions, the value 12 (12 "Months") is entered in this field. Point (o) of Article 27 on the MDR side and point (j) of Article 30 on the IVDR side give the same default; the same points also state that the calibration period is not sought on the label or sales packaging.
The 14 Options in the Sterilisation Method List
If there is more than one method, marking all of them is mandatory. The fourth sub-point of point (s) of Article 25(7) of the Manual gives the list one by one: Aseptic, Ethylene oxide, Hydrogen peroxide, Chlorine dioxide, Dry heat, Microwave radiation, Moist heat or steam, Ozone, Peracetic acid, Radiation, Sound waves, Liquid chemical sterilants, Ultraviolet light and High-intensity light. The same list is repeated in point (s) of Article 27 and point (n) of Article 30. If the method is not on the list, a request is opened through the "Medical Device Document and Product Registration Procedures" ("Tıbbi Cihaz Belge Ve Ürün Kayıt İşlemleri") field on the "Report Request" screen, and detailed information about the method and reference documents are uploaded as attachments to the request.
Substance Content Fields: Latex, Phthalate/DEHP, Ionising Radiation, Nanomaterial
These four questions are mandatory on the old regulation and MDR screens. Points (d), (e), (f) and (g) of Article 25(7) of the Manual and points (d), (e), (f) and (g) of Article 27(7) state that selecting all four fields in ÜTS is mandatory. If the answer is "Yes", the relevant information is required to be in the instructions for use, and for latex and phthalate/DEHP a warning is additionally required on the label and/or sales packaging.
The MDR adds one more field. Point (ğ) of Article 27(7) of the Manual states that if the "Does it contain nanomaterials?" field is selected as "Yes", the Nanomaterial Type field becomes mandatory and the suitable one among "Particle", "Aggregate", "Agglomerate" and "Unknown" is selected.
The definition itself is in Article 4 of the Manual. The definition describes a nanomaterial as a material containing particles, in an unbound state or as an aggregate or agglomerate, including fullerenes, graphene flakes and single-wall carbon nanotubes with one or more external dimensions below 1 nm, where, for 50% or more of the particles in the number size distribution, one or more external dimensions is in the size range 1-100 nm.
Blocks Specific to the MDR: CMR/Endocrine-Disrupting Substances and Contract Manufacturing
The CMR and endocrine-disrupting substance block is a separate heading on the MDR screen. Point (a) of Article 27(10) of the Manual asks whether the device contains substances as specified in point 10.4.1 of MDR Annex-I; point (b) makes selection from the substance list mandatory if "Yes" is selected. The third sub-point of the same point states that the list is based on substances obtained from the website of the European Chemicals Agency (ECHA) and that if the substance to be selected is not on the list, a request can be created through the "CMR / ED Madde Ekleme" (add CMR / ED substance) request field.
A threshold is defined for display on the label. Item (ğ) of the label information list in the second sub-point of point (b) of Article 27(7) of the Manual says that if the device contains CMR and/or endocrine-disrupting substances, and parts of the device or the materials used contain these substances at a concentration above 0.1% by weight (w/w), information showing the presence of these substances and a list of them will be on the label.
On the contract manufacturing side, two fields are linked. Point (gg) of Article 27(7) of the Manual makes the field "Was the product manufactured by another manufacturer on behalf of its legal manufacturer?" mandatory; if "Yes" is selected, the manufacturing firm is selected from the "Manufacturing Firm/Institution" field in point (ğğ). The third sub-point of the same point says that if the firm is not on the list, a record is created in ÜTS with a "Secondary Firm No" by entering the name, address, country and e-mail details; the IVDR counterpart is point (dd) of Article 30(7).
MDR-Specific Substance Content Fields
The MDR screen asks three more separate questions about the content of the device, and the answer to each can affect the class of the device.
• Does it contain a substance considered a medicinal product?: Mandatory under point (u) of Article 27(7) of the Manual; if "Yes", the device is Class III under the MDR Annex VIII classification rule and the name of the substance is on the label and in the instructions.
• Does it contain a medicinal product derived from human blood or plasma?: Mandatory under point (ü) of the same paragraph; if "Yes", the device is considered Class III and the conformity assessment is carried out by a notified body.
• Does it contain tissues or cells of animal origin according to (EU) 722/2012?: Mandatory under point (y) of the same paragraph; if "Yes", the device is Class III, except for devices intended to contact only intact skin. The counterpart on the IVDR side is point (s) of Article 30(7).
These fields are not just checkboxes; the answer you give changes the risk class of the device and therefore the set of documents that can be linked. For this reason the content questions must be answered together with the technical file.
IVDR-Specific Fields
The IVDR screen brings three fields that do not exist in the MDR, and all three are mandatory.
• Does it contain cells or substances of microbial origin?: Mandatory under point (p) of Article 30(7) of the Manual; if "Yes", the information must be on the label and/or sales packaging and in the instructions for use.
• Is it a new type of device?: Mandatory under point (ş) of the same paragraph. The second sub-point gives the definition: a device is considered a "new device" if, for the analyte or other parameters concerned, a device of this type has not been continuously available in Turkey and the EU member states during the last 3 years, or if the procedure involves an analytical technology not continuously used in connection with a particular analyte or other parameter during the last three years.
• Is it a device for self-testing or near-patient testing?: Mandatory under point (t) of the same paragraph; if "Yes", this indication must be on the label and/or sales packaging and in the instructions for use.
Dimension and Storage/Use Condition Blocks
The dimension block works from a closed list. The second sub-point of point (a) of Article 25(8) of the Manual says that the "Add Dimension" ("Ölçü Ekle") screen lists eleven dimension types, namely Angle, Weight, Area, Pressure, Circumference, Depth, Outer Diameter, Width, Total Volume, Length and Height, and that more than one dimension of the same type cannot be added; the same provision is repeated in paragraph 8 of Articles 27 and 30.
The storage and use condition is entered with its parameters. The second sub-point of point (b) of Article 25(9) of the Manual says that for each condition the Storage/Use Condition Type, Minimum Value, Maximum Value, Unit and Special Storage/Use Condition information are entered. The condition type list includes Storage Environment Temperature, Humidity, Altitude and Pressure, Use Environment Temperature, Humidity, Altitude and Pressure, and Special Storage Conditions for other conditions. Point (ç) of the same sub-point lists ten units of measurement: ATM, Degree Fahrenheit, Degree Kelvin, Degree Celsius, Hectopascal, Kilopascal, mbar, metre, mmHg and Percent Relative Humidity (%).
On the MDR and IVDR screens two more fields are added to this block. Points (c) and (ç) of Article 27(9) and Article 30(9) of the Manual make mandatory the "Are there other special conditions?" field and, if "Yes" is selected, the "Other Special Conditions" list that opens; the list includes options such as "Keep away from / protect from sunlight", "Store at room temperature", "Store in a dry place" and "Protect from heat and radioactive sources".
SUT Matching, Health Market and Product Surface Area Fields
SUT Matching Status is filled in even for devices outside reimbursement. Point (v) of Article 25(7) of the Manual says that this field is mandatory; "Match Exists" is selected if the device is within SGK reimbursement, "Match to Be Made" if an application will be made to bring it within the scope, and "No Match" if it is not within the scope. The fifth sub-point of the same point states that uploading a Turkish Catalogue to ÜTS is mandatory for devices for which a SUT matching application will be made.
The Health Market field, on the other hand, triggers the image standards. Point (y) of Article 25(7) says that "Yes" is selected if the device will be sold within the Health Market established under DMO or SGK reimbursement will be made; in that case at least 3 photographs for each device, at least 3 photographs for each packaging type for devices with inner and outer packaging, and images of spare parts, consumables and accessories, if any, are required, and each photograph must be at least 600x600 pixels, square and at most 2 MB; on the MDR side these rules are split into points (cc) and (çç) of Article 27(7). Product Surface Area is for a narrow group: under point (bb) of Article 25 and point (ff) of Article 27, it is entered only for devices that SGK pays for in terms of surface area, and may be left empty for others.
The "Is it within the scope of Annex-3?" Field and the Current List
This field is mandatory in all three frameworks. Point (u) of Article 25(7), point (z) of Article 27(7) and point (u) of Article 30(7) of the Manual require a selection of whether the device is among the products within the scope of Annex-3 of the Medical Device Sales, Advertising and Promotion Regulation.
The current list is enumerated one by one in the Manual. The third sub-point of point (u) of Article 25(7) and the third sub-point of point (z) of Article 27(7) say that, with the amendment published in Official Gazette No. 32202 of 26/5/2023, the current Annex-3 list includes toothpaste for desensitisation, denture care products for individual use, condoms, bed pads, incontinence care products, hot/cold compresses, adhesive bandages and plasters, cotton, mouthwash, nasal breathing strips, surgical masks, self-adhesive pressure-relief tape for blisters, toe separators for relieving corns and blisters, non-sterile single-use gloves and pediatric nasal aspirators, and the groups of products without a medical purpose in MDR Annex XVI.
In IVDR records the answer is fixed. The fourth sub-point of point (u) of Article 30(7) of the Manual states that there is no device under the IVDR in the current Annex-3 and that, as long as there is no change in this direction in Annex-3 of the Regulation concerned, it is mandatory for this field to be selected as "No".
Field Filling Order: The Flow That Prevents Lock-ups
Filling in the fields in the order they appear on the screen is not always the most efficient way; some answers lock later fields. The third sub-point of point (a) of Article 25(5) of the Manual says that the class of the device must have been selected before documents are selected. In MDR records the chain is one link longer: the third and fourth sub-points of point (c) of Article 27(5) state that the relevant EU Declaration of Conformity and EU Certificate must have been selected before the Basic UDI-DI is selected, and that only the Basic UDI-DI numbers associated with the selected EU Declaration of Conformity will appear in the list.
The implantability and single-use questions should also be filled in early. The third sub-point of point (h) of Article 27(7) of the Manual says that when the "Can it be implanted in the body?" field is selected as "Yes", the linked fields are filled in automatically and cannot be changed. Setting up the order as class, documents, Basic UDI-DI, the implantable and single-use questions, and then the remaining features prevents lock-ups.
The field lists are long but closed, and every field has a corresponding piece of evidence (the label, the declaration of conformity, the certificate or the instructions for use). Firms can also have this filling-in done by opening a user authority for a consultant from their own ÜTS firm page; the consultant fills in the fields and approves the record with the "Medical Device Registration Operator" authority, the e-signature stays with the firm's signatory, and responsibility for the accuracy of the data rests with the firm.
The screen-by-screen flow of the registration is covered on our ÜTS Product (Medical Device) Registration page and the list of rejection reasons on our Why Is a ÜTS Application Rejected page. The barcode and UDI rules are explained on our UDI-DI and Barcode Rules page, the distinction between GMDN and EMDN on our GMDN and EMDN Code page, bulk upload on our ÜTS Bulk Product Registration page, and the image and label rules on our ÜTS Product Image and Label Rules page.
For the counterpart on the EU side, our IVDR classification tool page (in English) will guide you.