Medical Device Registration

You have a product, you know it has to be registered, but you do not know where to register it. In Turkey this question has no single answer: two products on the same shelf can fall under two different authorities

Medical Device Registration by Product Type: Registration System (ÜTS/EUDAMED) and Competent Authority

You have a product, you know it has to be registered, but you do not know where to register it. In Turkey this question has no single answer: two products on the same shelf can go to two different authorities, under two different pieces of legislation and through two entirely different types of procedure.

A hand sanitiser and a device disinfectant sit side by side; one requires a biocidal licence, the other a medical device registration. Toothpaste goes to a cosmetic notification, while toothpaste for desensitisation is on the medical device list. A condom looks like a simple consumer product, yet the classification rules make it Class IIb. You draw these distinctions not by looking at what the product is, but at its intended purpose, its claim and its mechanism of action.

This page does three things at once. First, it lets you test whether your product is a medical device by the criteria of the Medical Device Regulation (Tıbbi Cihaz Yönetmeliği) itself. Second, if it is not a medical device, it shows which system and which authority it goes to, with the regulation references. Third, it lists 290 product names with their framework and probable class, so you can find the product you are looking for directly.

There are two kinds of information on the page. Provisions of the legislation are given with their source: you will see within the sentence which article of which regulation it is. Notes based on practice are marked separately. One warning from the start: the class information below is not a binding classification decision. The final class is determined by the declared intended purpose of your product, and responsibility lies with the manufacturer.

Determining Whether a Product Is a Medical Device

At the centre of the definition of a medical device is not the physical characteristics of the product but the medical purpose declared by the manufacturer. Article 3 of the Medical Device Regulation, the definitions article, defines a device as a product intended to be used for one of specific medical purposes such as diagnosis, prevention, monitoring, treatment or alleviation of disease. So the first question is not "what is this product" but "what do you claim with this product".

Answer the following five questions in order.

• 1. Is the declared purpose of the product medical: Diagnosis, prevention, monitoring, treatment, alleviation, compensation for an injury or disability, investigation or modification of the anatomy or of a physiological process, control or support of conception. If one applies, you are on the device side.

• 2. By what means does it achieve its principal action: This is the only real criterion for the boundary with medicines; it is explained under a separate heading below.

• 3. It has no medical purpose but is it on the list: Products within the list of groups of products without a medical purpose are also considered "devices".

• 4. Does it fall into an excluded category: Article 1(4) of the Regulation excludes nine categories one by one.

• 5. Does it have both a medical and a non-medical purpose: If it does, the two sets of requirements are met together.

The principal mechanism of action test

Article 1(6) of the Medical Device Regulation says this: when deciding whether a product falls under the Regulation on Licensing of Medicinal Products for Human Use or the Medical Device Regulation, particular account is taken of the principal mechanism of action of the product. The device definition in Article 3 of the same Regulation completes this: a device does not achieve its intended principal function in or on the human body by pharmacological, immunological or metabolic means, but may be assisted in its function by such means.

The decision rule has three levels. If the product achieves its principal action by physical, mechanical or physicochemical means, it is a device. If it achieves its principal action by pharmacological, immunological or metabolic means, it is a medicine. If the pharmacological action is only supportive, the product can remain a device. The same paragraph also regulates the last case: devices that incorporate, as an integral part, a substance which, if used separately, would be considered a medicinal product and which assists the function of the device are subject to the Medical Device Regulation; if that substance does not assist but has a principal action, the whole integral product moves to the medicines legislation.

Products without a medical purpose are also considered devices

Article 1(2) of the Regulation is explicit: medical devices, medical device accessories and the products listed in Annex XVI to which this Regulation applies are hereafter referred to as "devices". Point (b) of paragraph 3 of the same article says that the Regulation will apply to these groups of products without a medical purpose from the date of application of the common specifications adopted under Article 9.

The practical consequence is this: a product that carries no medical claim and is entirely aesthetic in purpose can also be a "device" and give rise to a registration obligation. The scope of this group is defined in the annex to the Medical Device Regulation; you need to determine whether your product falls within it from the current text of the annex or by obtaining an opinion from the Agency. This page does not enumerate the contents of the list by product name; our sources name the heading of the list and do not count the products in it one by one.

One connection point is verified. Item 16 of Annex-3 of the Medical Device Sales, Advertising and Promotion Regulation has put products within the list of groups of products without a medical purpose, except those whose places of sale are restricted under the relevant legislation and those manufactured exclusively for use by healthcare professionals, on the "list of devices that can be sold outside sales centres". The exemption here concerns the place of sale; it does not remove the product's status as a device or the registration obligation.

Cumulative requirement for products with two purposes

The key provision for aesthetic lasers, light-based skin devices and similar products is Article 9(6) of the Medical Device Regulation: where a device has both a medical and a non-medical intended purpose, these devices cumulatively fulfil the requirements applicable to devices with a medical intended purpose and those applicable to devices without a medical intended purpose.

The word cumulative has a technical meaning here: you cannot choose one of the two regimes, you meet both. If the intended purpose form of a laser device says both "hair removal" and "treatment of vascular lesions", the product must carry both sets of requirements.

What is left out of scope

Article 1(4) of the Medical Device Regulation excludes nine categories from its scope. They are outside not because they are not medical devices, but because they are the subject of other legislation.

• (a) In vitro diagnostic devices: Subject to the In Vitro Diagnostic Medical Device Regulation. This is the provision that ends the assumption that "the Medical Device Regulation covers everything".

• (b) Medicinal products: Medicinal products as defined in the Regulation on Licensing of Medicinal Products for Human Use.

• (c) Advanced therapy medicinal products: Gene therapy, somatic cell therapy and tissue engineering products.

• (ç) Human blood and blood products: Human blood, blood products, plasma or blood cells of human origin, and devices incorporating them; except for the devices referred to in paragraph 6.

• (d) Cosmetic products: Products within the scope of the Cosmetics Regulation.

• (e) Transplants, tissues and cells of animal origin: Except for devices manufactured utilising non-viable or rendered non-viable ones.

• (f) Transplants, tissues and cells of human origin: Except for devices manufactured utilising derivatives of non-viable or rendered non-viable ones.

• (g) Those containing live biological material: Products containing or consisting of live biological material or live organisms, including living micro-organisms, bacteria, fungi or viruses.

• (ğ) Food: Food within the scope of Veterinary Services, Plant Health, Food and Feed Law No. 5996.

The other paragraphs of the same article resolve the overlaps. Under paragraph 5, a device incorporating an in vitro diagnostic device as an integral part is subject to the Medical Device Regulation, but the provisions of the In Vitro Diagnostic Medical Device Regulation apply to that part. Paragraph 7 subjects a device intended to administer a medicinal product to the Medical Device Regulation; but if the device and the medicine are placed on the market only in the given combination and as a single integral product not intended for reuse, that product moves to the medicines legislation. Paragraph 10 requires devices that fall within the definition of machinery also to meet the Annex I requirements of the Machinery Safety Regulation.

Does Your Product Go to ÜTS or to Another System?

The list below gathers, in seventeen rows, the registration, notification, approval and licensing routes you may meet when placing a product on the market in Turkey. Each row gives the name of the procedure, the competent authority and the Official Gazette reference of the underlying legislation. Your product may fall into more than one row: an importer selling medical devices to a public hospital is in three separate rows at the same time.

• Medical device: Registration notification and application in ÜTS · Ministry of Health, Turkish Medicines and Medical Devices Agency (TİTCK) · Medical Device Regulation, Official Gazette No. 31499 (1st repeated) of 2/6/2021

• In vitro diagnostic device: Registration notification and application in ÜTS, separate screen · TİTCK · In Vitro Diagnostic Medical Device Regulation, Official Gazette No. 31499 (repeated) of 2/6/2021

• Cosmetic product: Notification through the cosmetic product management functions of ÜTS; not a licence or approval · TİTCK · Cosmetic Products Regulation, Official Gazette No. 32184 (repeated) of 8/5/2023, Article 16; based on Cosmetics Law No. 5324 of 30/3/2005

• Biocidal, Product Type 1 (human hygiene) and Product Type 19 (repellents and attractants): Licence · TİTCK · Biocidal Products Regulation, Official Gazette No. 27449 (4th repeated) of 31/12/2009; transfer of authority by the Presidential Decree published in Official Gazette No. 30887 of 13/9/2019

• Biocidal, other product types: Licence; follow-up through the Biocidal Product Tracking System · Ministry of Health General Directorate of Public Health (HSGM) · Biocidal Products Regulation, same reference

• Biocidal applicator firm (pest control company, responsible manager): Permit · Provincial Health Directorate and HSGM · Regulation on the Procedures and Principles of Use of Biocidal Products, Official Gazette No. 30821 of 4/7/2019

• Food supplement: Food supplement approval and business registration/approval procedures; queries through the Food Safety Information System (GGBS) · Ministry of Agriculture and Forestry General Directorate of Food and Control, application to the provincial or district directorate · Regulation on the Import, Production, Processing and Placing on the Market of Food Supplements, Official Gazette No. 28635 of 2/5/2013; Regulation on the Registration and Approval Procedures of Food Businesses, Official Gazette No. 28145 of 17/12/2011

• Health claim on food or food supplement: Health claim permit · TİTCK · Regulation on the Use of Health Claims on Foods and Food Supplements, Official Gazette No. 32169 of 20/4/2023

• Medicinal product for human use (medicine): Licence · TİTCK · Regulation on Licensing of Medicinal Products for Human Use, Official Gazette No. 31686 of 11/12/2021

• Detergent and general cleaning product: No notification or registration obligation before placing on the market; keeping a technical file and a safety data sheet is mandatory · Ministry of Trade · Regulation on Detergents, Official Gazette No. 30314 of 27/1/2018

• Borderline product decision: Classification and Borderline Products Opinion Application, electronically · TİTCK · explained under a separate heading below

• SGK reimbursement: Application for inclusion in the Healthcare Implementation Communiqué lists, through its own portal · Social Security Institution · Healthcare Implementation Communiqué, Official Gazette No. 28597 of 24/3/2013; Directive on Working Procedures and Principles Concerning Medical Supply Applications, 17/10/2024

• Sales to public institutions: Supplier pre-registration and catalogue or framework agreement application · State Supply Office (DMO) · DMO Articles of Association, Presidential Decree No. 1899, Official Gazette No. 30989 of 25/12/2019; DMO Procurement Regulation, Official Gazette No. 27054 of 14/11/2008

• Optician establishment: Licence certificate · Provincial Health Directorate · Regulation on Optician Establishments, Official Gazette No. 28886 of 18/1/2014

• Custom-made prosthesis-orthosis centre and hearing aid centre: Licence certificate and responsible manager certificate · Provincial Health Directorate · Regulation on Custom-Made Prosthesis and Orthosis Centres and Hearing Aid Centres, Official Gazette No. 28064 of 24/9/2011

• Medical device sales centre (except the three types of establishment above): Sales centre authorisation certificate · Provincial Health Directorate · Medical Device Sales, Advertising and Promotion Regulation, Official Gazette No. 29001 of 15/5/2014; last amended by Official Gazette No. 33299 of 3/7/2026

• Firm and institution identity (precondition): ÇKYS institution code and TİTCK Electronic Application and Process Management System (EBS) firm registration · Ministry of Health and TİTCK

The five points in this list that are most often confused are dealt with separately below. These are the misunderstandings that cost the most time and money in the registration process.

A Medical Device Registration Is Not the Same as a Cosmetic Notification

ÜTS is not a single door; it is a set of separate registration lines running under one roof. The scope provision of the Manual establishes this with the wording "without prejudice to the registration procedures of biocidal products or cosmetic products": the medical device line is subject to this Manual, while the cosmetic line has its own legislation and its own function. The sentence "I entered ÜTS, I'm done" means nothing unless you say which line you entered.

The difference between the two lines lies in the legal nature of the procedure. On the medical device side, the Agency examines documents such as the certificate and the declaration of conformity and approves or rejects the application; the result is a registered device. On the cosmetic side, under Article 16 of the Cosmetic Products Regulation, the responsible person notifies the Agency through the national electronic database before placing the product on the market. This is a notification; the Agency issues no prior permit or approval certificate.

An expression commonly used in the market is therefore invalid: there is no such document as a "Ministry of Health approved cosmetic". In cosmetics there is a notification that has been recorded, not an approval certificate. Using this expression on a label or in an advertisement makes a claim with nothing behind it.

The structure of responsibility also differs. For medical devices the obliged party is the manufacturer, the authorised representative or the importer; for cosmetics it is the responsible person who holds the product information file. In practice, for a cosmetic notification the fee accrues separately for each product barcode, whereas for a medical device it is charged per application. This difference directly affects the budget for product families with many variants.

There is also a reverse direction. CPNP, the European Commission's cosmetic product notification portal, is mandatory for placing products on the EU market; since Turkey is not an EU member, CPNP is not required for a product sold in Turkey. Likewise, your notification in Turkey is not valid in the EU. The two systems do not replace each other; you may need both.

For Biocidal Products, Authority Is Split in Two

The most commonly misunderstood feature of the biocidal side is that there is no single agency. Biocidal products that come into direct contact with the human body, namely Product Type 1 (human hygiene biocides) and Product Type 19 (repellents and attractants), were placed within the remit of TİTCK by the Presidential Decree published in Official Gazette No. 30887 of 13/9/2019. TİTCK set out the procedures and principles for these products in its Circular No. E.70938 of 19/3/2020. All other product types remain with the General Directorate of Public Health.

In practice: hand antiseptic, hand disinfectant gel and skin antiseptic, as well as mosquito repellent lotions and sprays, go to TİTCK. Surface disinfectant, instrument disinfectant, insecticide, rodenticide and water disinfectant go to HSGM. Biocidal products are divided into four main groups and twenty-two product types; unless you determine which type your product falls into, you cannot know which agency to apply to. The type of procedure also differs from cosmetics: a biocidal product is licensed and cannot be placed on the market by notification.

The line between a detergent and a biocide depends on a single word: the claim. The moment a cleaning product carries a claim of microbiological efficacy such as "disinfects" or "kills bacteria", it becomes a biocide and requires a licence. The same formulation can be sold as a detergent, with no notification before placing on the market, if it only claims to "clean". What brings the product into the system is not the formula but the sentence on the label.

The name of the tracking system on the HSGM side is the Biocidal Product Tracking System. We do not give the system's current login address on this page; consultancy sites publish conflicting addresses, and none of them can be confirmed from an official source. Confirm the address from HSGM's current announcement before applying.

ÇKYS Is Not a Biocidal System

This is the most frequent confusion in the sector and it is easy to correct. ÇKYS is the abbreviation of the Core Resource Management System (Çekirdek Kaynak Yönetim Sistemi). It is the system in which the Ministry of Health keeps its institution and personnel records. It has nothing to do with biocidal product licensing, biocidal product tracking or biocidal applicator permits.

ÇKYS reaches you from another direction: the institution code. When a healthcare institution, optician establishment, custom-made prosthesis-orthosis centre, hearing aid centre or medical device sales centre obtains its licence or authorisation certificate, an institution code is assigned through this system in practice, and that code is a precondition for some procedures on the ÜTS side. ÇKYS is not a product registration system; it is an institution identity system.

The result of the confusion is concrete lost time: a firm that looks for a biocidal licence application in ÇKYS reaches the right door weeks later. For the product licence, HSGM or TİTCK; for institution identity, ÇKYS.

ÜTS Registration and SGK Reimbursement Are Separate Matters

ÜTS registration is for being able to place the product on the market legally in Turkey. Entering the Healthcare Implementation Communiqué lists, on the other hand, is for having the product's price paid by the Social Security Institution. These two procedures run at different agencies, on different portals and under different criteria.

The Healthcare Implementation Communiqué was published in Official Gazette No. 28597 of 24/3/2013 and is amended very frequently; in 2026 alone, amendments dated 17/1/2026, 8/4/2026 and 23/5/2026 have been published. The application procedure is regulated by the Directive on Working Procedures and Principles Concerning Medical Supply Applications dated 17/10/2024. SGK states that applications are not accepted through ÜTS or manually, and can be made only through its own portal.

Under the Directive, applications are grouped into three categories and evaluation proceeds through the Preliminary Review Commission, the Scientific and Academic Advisory Commission and the Medical Supply Evaluation Commission, in that order; one right of objection is granted against each commission decision. The order matters too: first ÜTS, then the Healthcare Implementation Communiqué. A device that is not on the list can still be sold legally in Turkey; its price is simply paid by the patient.

A DMO Registration Does Not Replace a ÜTS Registration

The State Supply Office is the central procurement organisation of public institutions. Its legal status was determined by Presidential Decree No. 1899, published in Official Gazette No. 30989 of 25/12/2019; its purchasing procedures are subject to the DMO Procurement Regulation published in Official Gazette No. 27054 of 14/11/2008. The Office makes purchases below the statutory exemption limit under its own legislation, and those above it under the provisions of Public Procurement Law No. 4734.

The firm's route here begins with supplier pre-registration and continues with a catalogue or framework agreement application. However, this is a commercial channel, not a licence or permit, and it comes after ÜTS registration. Appearing in the DMO catalogue does not mean the product is registered in ÜTS; the reverse is not true either.

What Is and Is Not Registered in ÜTS

The answer to this section lies directly in the text of the Manual and leaves no room for guessing. The Manual on the Registration Processes of Medical Devices into the Product Tracking System states that ÜTS handles only the registration of medical devices, in vitro diagnostic devices and parts defined as medical device accessories or in vitro diagnostic device accessories that fall within the definition of "Device" and are within the scope of the Medical Device Regulations. The rest of the same sentence is also clear: parts defined as spare parts are not registered in ÜTS. Article 40, paragraph 3 of the Manual lists custom-made medical devices and system and procedure packs as separate application types; these two are registered in a different flow from standard device registration.

• Medical devices: Devices within the scope of the Old Medical Device Regulations and devices within the scope of the Medical Device Regulation.

• In vitro diagnostic devices: Those within the scope of the old and new regulations, on a separate screen.

• Accessories: Parts defined as medical device accessories or in vitro diagnostic device accessories.

• System and procedure packs (SİP): Listed as a separate application type in Article 40 of the Manual.

• Custom-made medical devices (ITC): Devices made specifically for an individual, with their own application type.

What is not registered falls under four headings, and each has its own reasoning in a separate provision.

• Spare part: Article 23, paragraph 1 of the Medical Device Regulation covers a part intended to replace a defective or worn part in order to maintain or restore the function of a device without changing its performance, safety characteristics or intended purpose. Such a part is not a device in itself and is not registered in ÜTS. However, paragraph 2 of the same article establishes the opposite: a part that significantly changes the performance, safety characteristics or intended purpose is considered a device and must meet the requirements of the regulation.

• Component or part without CE marking: The definitions article of the Manual describes a component or part as an integral part or component of a device that is not defined as a device on its own and does not bear a CE marking. Since it is not regarded as a device on its own, it is not registered separately.

• Change of hands on the second-hand market: Article 1, paragraph 12 of the Medical Device Regulation states that the regulation does not cover the rules on making devices available on the market by changing hands after they have already been put into service, such as second-hand sales.

• Out-of-scope products: The nine categories listed above. You cannot open a medical device registration for your cosmetic product; a food supplement does not enter the medical device line of ÜTS.

The cost of this distinction is concrete. Creating a medical device registration notification for an out-of-scope product ends in rejection of the application, and during that time your product has never applied to the right door.

290 Products: Which Framework, Which Class

The list below gathers 290 product names that people search for in Turkish with questions such as "how is it registered", "is it a medical device" and "which class", together with their framework and probable class. When reading the list, you need to distinguish three things.

First, some of the product names appear directly in the source texts. Products such as contact lens solution, blood bag, specimen container, condom and surgical mask are named in the classification rules or in the annex of the sales regulation. In these rows the link between the rule and the product is direct.

Second, most of the products are not named in the source; they were added under the names used in everyday Turkish, and the class hint was derived from the logic of the relevant classification rule. Do not read these rows as "the regulation says so"; read them as "the logic of this rule leads this product here".

Third, and most important: the class information here is a hint, not a decision. The application of the classification rules is based on the intended purpose of the devices. The same product changes class when the intended purpose statement changes; when several rules apply, the strictest rule, which gives the highest classification, prevails. The final class and the responsibility for it rest with the manufacturer.

IVD: Infection and Microbiology Tests (19 products)

• HIV test (HIV 1 and 2 infection marker screening/confirmation kit): IVDR · Art. 57/Rule 1-1 and 1-2 → Class D; under the IVDD, Annex II List A

• Hepatitis B test (HBsAg, hepatitis B marker): IVDR · Art. 57/Rule 1 → Class D

• Hepatitis C test (Anti-HCV): IVDR · Art. 57/Rule 1 → Class D

• Hepatitis D test: IVDR · Art. 57/Rule 1 → Class D

• HTLV I/II test: IVDR · Art. 57/Rule 1 → Class D

• Cytomegalovirus (CMV) test: IVDR · Art. 57/Rule 3-3 → Class C; under the IVDD, Annex II List B

• Chlamydia test: IVDR · Art. 57/Rule 3-1 (sexually transmitted agent) → Class C

• Rubella test: IVDR · Art. 57/Rule 3-4 (prenatal immune status screening) → Class C

• Toxoplasma test: IVDR · Art. 57/Rule 3-4 → Class C

• PCR test kit (molecular diagnostic kit): IVDR · Art. 57/Rule 1 (D) or Rule 3 (C) depending on the target agent; Rule 6 (B) if neutral (not named in the source)

• Rapid test kit: IVDR · Unless for self-testing or near-patient use, the "exclusion" statement is mandatory on the label; class depends on the target

• COVID-19 test kit: IVDR · Art. 57/Rule 1-2 (agent with high propagation risk, life-threatening) → Class D assessment (not named in the source)

• Influenza test kit: IVDR · Art. 57/Rule 3 (C) or Rule 6 (B) depending on the propagation risk assessment (not named in the source)

• Syphilis test: IVDR · Art. 57/Rule 3-1 (sexually transmitted agent) → Class C (not named in the source)

• Tuberculosis test (TB IGRA / mycobacteria test): IVDR · Art. 57/Rule 3 or Rule 6 (not named in the source)

• Helicobacter pylori test: IVDR · Art. 57/Rule 6 → Class B (not named in the source)

• Faecal occult blood test: IVDR · If used in cancer screening, Art. 57/Rule 3-8 → Class C (not named in the source)

• Throat/urine culture medium (general culture medium): IVDR · Art. 57/Rule 5-1 → Class A

• Antibiogram disc / susceptibility test: IVDR · Art. 57/Rule 6 → Class B (not named in the source)

IVD: Blood Bank, Transfusion and Tissue Typing (9 products)

• Blood group determination kit (ABO and Rh reagents): IVDR · Art. 57/Rule 2 → Class D (ABO and Rhesus markers are listed); under the IVDD, Annex II List A

• Anti-Kell reagent: IVDR · Art. 57/Rule 2-3 [Kel1 (K)] → Class D

• Anti-Kidd reagent (JK1/JK2): IVDR · Art. 57/Rule 2-4 → Class D; under the IVDD, Annex II List B

• Anti-Duffy reagent (FY1/FY2): IVDR · Art. 57/Rule 2-5 → Class D

• Irregular antibody screening test (anti-erythrocyte antibodies): IVDR · Art. 57/Rule 2 → Class C; under the IVDD, Annex II List B

• HLA tissue typing kit (DR, A, B): IVDR · Art. 57/Rule 2 (tissue typing) → Class C

• Cross-match kit: IVDR · Art. 57/Rule 2 → Class C (not named in the source)

• Foetal-maternal blood incompatibility test (Rh incompatibility): IVDR · Art. 57/Rule 2 → Class C/D

• Transplant compatibility test (organ/tissue donation screening): IVDR · Art. 57/Rule 1-1 → Class D

IVD: Oncology, Genetic and Screening Tests (10 products)

• PSA test (prostate tumour marker): IVDR · Art. 57/Rule 3-8 (cancer screening/diagnosis/staging) → Class C; under the IVDD, Annex II List B

• Tumour marker test kit (CEA, CA-125, etc.): IVDR · Art. 57/Rule 3-8 → Class C (not named in the source)

• HPV DNA test: IVDR · Art. 57/Rule 3-1 and 3-8 → Class C (not named in the source)

• Genetic test kit (human genetic test): IVDR · Art. 57/Rule 3-9 → Class C

• NIPT / Trisomy 21 risk assessment test: IVDR · Art. 57/Rule 3-12 (screening for congenital disorder in the foetus) → Class C; under the IVDD, Annex II List B

• Double/triple screening test (pregnancy screening panel): IVDR · Art. 57/Rule 3-12 → Class C (not named in the source)

• Newborn heel-prick screening test: IVDR · Art. 57/Rule 3-13 → Class C

• Phenylketonuria (PKU) test kit: IVDR · Art. 57/Rule 3-13 → Class C; under the IVDD, Annex II List B

• Companion diagnostic: IVDR · Art. 57/Rule 3-6 → Class C; separate class in ÜTS: "Destek Tanı Cihazı (Sınıf C)" (Companion Diagnostic Device, Class C)

• Drug level monitoring test (therapeutic drug monitoring): IVDR · Art. 57/Rule 3-10 → Class C

IVD: Self-Test and Near-Patient Test Devices (12 products)

• Blood glucose meter (blood glucose measurement system, glucometer): IVDR · Self-test → Art. 57/Rule 4-1 main provision Class C; *blood* glucose is not in the exception list. Under the IVDD, Annex II List B

• Blood glucose test strip: IVDR · Art. 52/9-10: calibrators and control materials with assigned values are in the same class as the device (not named in the source)

• Continuous glucose monitoring sensor (CGM): unclear · Depending on the intended purpose, IVDR (Art. 57/Rule 4) or the MDR active monitoring rule; a dual assessment is needed (not named in the source)

• Pregnancy test: IVDR · Art. 57/Rule 4-1 exception (detection of pregnancy) → Class B

• Ovulation test (fertility test): IVDR · Art. 57/Rule 4-1 exception → Class B

• Cholesterol measurement test (home): IVDR · Art. 57/Rule 4-1 exception → Class B

• Urine test strip (glucose, erythrocytes, leukocytes, bacteria in urine): IVDR · Art. 57/Rule 4-1 exception → Class B

• Rapid test kit for home use (self-test device): IVDR · Art. 57/Rule 4-1 → Class C (except for the four exceptions); under the IVDD, "Self-Test Devices" as a separate class

• Near-patient test device (POCT): IVDR · Art. 57/Rule 4-2: classified in its own right; in ÜTS "Hastabaşı Test Cihazı (Sınıf B/C/D)" (Near-Patient Test Device, Class B/C/D)

• INR / coagulation measurement device (self-test): IVDR · Art. 57/Rule 4-1 → Class C (not named in the source)

• Home urinary tract infection test: IVDR · Art. 57/Rule 4-1 exception (leukocytes/bacteria in urine) → Class B

• Alcohol / drug test kit: unclear · If there is a medical intended purpose, Art. 57/Rule 3-10 or Rule 6; if for forensic or workplace purposes, scope is debatable (not named in the source)

IVD: Laboratory Consumables, Reagents and Equipment (15 products)

• Specimen container (urine container, stool container, swab tube): IVDR · Art. 57/Rule 5-3 → Class A

• Blood collection tube (vacuum tube): IVDR · Art. 57/Rule 5-3 "specimen receptacles" logic → Class A (not named in the source)

• Buffer solution: IVDR · Art. 57/Rule 5-1 → Class A

• Wash solution (wash buffer): IVDR · Art. 57/Rule 5-1 → Class A

• Histological stain (pathology stain): IVDR · Art. 57/Rule 5-1 → Class A

• Product intended for general laboratory use: IVDR · Art. 57/Rule 5-1 → Class A

• IVD instrument / autoanalyser (instrument for in vitro diagnostic procedures): IVDR · Art. 57/Rule 5-2 → Class A

• Automated blood coagulation analyser: IVDR · Art. 57/Rule 5-2 (instrument) → Class A; its reagent is classified separately

• Calibrator: IVDR · Art. 52/9: same class as the device it is used with

• Control material (with assigned qualitative/quantitative values): IVDR · Art. 52/10: same class as the device

• Control without an assigned value: IVDR · Art. 57/Rule 7 → Class B

• Reagent / reagent product: IVDR · Art. 57/Rule 1-7 depending on the intended purpose

• Test kit (set of components): IVDR · Art. 4, definition (ff); class depends on the intended purpose of the kit

• IVD device accessory (pipette tip, rack, adapter): IVDR · Art. 52/8: classified in its own right, separately from the device it is used with

• IVD device for performance studies: IVDR · A separate indication on the label is mandatory; the placing-on-the-market regime is different

Implants (non-active) (25 products)

• Breast implant (silicone breast prosthesis): MDR · Art. 56/Rule 8-7 → Class III; under the MDD, Art. 53/Rule 8-5 → Class III

• Surgical mesh (hernia patch): MDR · Art. 56/Rule 8-7 → Class III (under the MDD, IIb by the general provision of Rule 8)

• Hip prosthesis (total hip replacement): MDR · Art. 56/Rule 8-8 → Class III; under the MDD, Art. 53/Rule 8-6 → Class III

• Knee prosthesis: MDR · Art. 56/Rule 8-8 → Class III

• Shoulder prosthesis: MDR · Art. 56/Rule 8-8 → Class III

• Spinal disc prosthesis (disc replacement implant): MDR · Art. 56/Rule 8-9 → Class III

• Implant in contact with the spine (pedicle screw system): MDR · Art. 56/Rule 8-9 → Class III; however, components such as screws, wedges, plates and instruments are excluded from this provision

• Bone plate and screw (osteosynthesis set): MDD transition · Art. 56/Rule 8 → Class IIb; under the MDR, "screws, plates" are on the well-established technologies (WET) list → with an MDD certificate until the end of 31.12.2027

• Bone nail / pin: MDD transition · Art. 56/Rule 8 → Class IIb; on the WET list

• Surgical clip: MDD transition · Art. 56/Rule 8 → Class IIb; on the WET list

• Surgical connector: MDD transition · Art. 56/Rule 8 → Class IIb; on the WET list

• Bone wire (cerclage wire): MDD transition · Art. 56/Rule 8 → Class IIb; "wires" are on the WET list

• Surgical wedge: MDD transition · Art. 56/Rule 8 → Class IIb; on the WET list

• Suture: MDD transition · Art. 56/Rule 8 → Class IIb; for absorbable ones, a Class III assessment arises under Rule 8-3

• Surgical staple (stapler cartridge): MDD transition · Art. 56/Rule 8 → Class IIb; on the WET list

• Heart valve (bioprosthetic / mechanical valve): MDR · Art. 56/Rule 8-2 (direct contact with the heart / central circulatory system) → Class III (not named in the source)

• Coronary stent: MDR · Art. 56/Rule 8-2 → Class III (not named in the source)

• Peripheral stent / vascular graft: MDR · Art. 56/Rule 8-2 → Class III (not named in the source)

• Intraocular lens (IOL, cataract lens): MDR · Art. 56/Rule 8 (long-term implant) → Class IIb (not named in the source)

• Bone cement: MDR · Art. 56/Rule 8-4 (undergoes chemical change in the body) → Class III (not named in the source)

• Bone graft / bone substitute material: MDR · Art. 56/Rule 8-3 (biological effect / absorbed) → Class III; if of animal origin, Rule 18 → Class III (not named in the source)

• Implant made from animal tissue (pericardial patch, collagen membrane): MDR · Art. 56/Rule 18 → Class III (except those that only contact intact skin); under the MDD, Art. 53/Rule 17

• Device containing a human blood derivative: MDR · Art. 56/Rule 14 → Class III; under the MDD, Art. 53/Rule 13-2

• Device containing a medicinal substance (drug-coated stent, antibiotic cement): MDR · Art. 56/Rule 14 (integral medicinal product) → Class III

• Implant containing nanomaterial: MDR · Art. 56/Rule 19: high/medium internal exposure → III, low → IIb, negligible → IIa

Active Implants (AIMDD / MDR) (7 products)

• Pacemaker: MDR · Art. 54: AIMDD devices are Class III in ÜTS; under the MDR, Art. 56/Rule 8-6 → Class III (not named in the source)

• Implantable cardioverter defibrillator (ICD): MDR · Art. 56/Rule 8-6 → Class III (not named in the source)

• Cochlear implant: MDR · Art. 56/Rule 8-6 → Class III (not named in the source)

• Deep brain stimulator (DBS): MDR · Art. 56/Rule 8-6 → Class III (not named in the source)

• Implantable drug pump: MDR · Art. 56/Rule 8-5 (administration of a medicinal product) and 8-6 → Class III (not named in the source)

• Programmer / external controller of an active implant: MDR · Art. 56/Rule 9-4 (controlling/monitoring the performance of an active implant) → Class III

• Active implant accessory (charging unit, telemetry head): MDR · Art. 52/2: an accessory is classified in its own right; Art. 56/Rule 8-6 → Class III

Consumables and Single-Use Devices (29 products)

• Syringe: MDR · When supplied with its needle, surgically invasive for transient use → Art. 56/Rule 6 → Class IIa

• Needle tip (hypodermic needle): MDR · Art. 56/Rule 6 → Class IIa

• Insulin pen needle: MDR · Art. 56/Rule 6-6 (administration of a medicinal product) → Class IIa/IIb (not named in the source)

• Infusion set (gravity infusion set): MDR · Art. 56/Rule 2: Class IIa if connectable to a Class IIa or higher active device, otherwise Class I (not named in the source)

• Infusion pump set: MDR · Art. 56/Rule 2-1 → Class IIa (not named in the source)

• Blood administration set (transfusion set): MDR · Art. 56/Rule 2-2 (transfusion/storage of blood) → Class IIa (not named in the source)

• Blood bag: MDR · Art. 56/Rule 2-3 → Class IIb; under the MDD, expressly Class IIb by Art. 53/Rule 18

• Intravascular cannula (IV catheter, branule): MDR · Art. 56/Rule 7 (short-term surgically invasive) → Class IIa (not named in the source)

• Central venous catheter: MDR · Art. 56/Rule 7-2 (contact with the heart / central circulatory system) → Class III (not named in the source)

• Urinary catheter (Foley catheter): MDR · Art. 56/Rule 5: short-term IIa, long-term IIb (not named in the source)

• Urine bag: MDR · Art. 56/Rule 2 → Class I/IIa (not named in the source)

• Nasogastric tube (feeding tube): MDR · Art. 56/Rule 5: short-term IIa, long-term IIb (not named in the source)

• Intubation tube (endotracheal tube): MDR · Art. 56/Rule 5 last paragraph (connected to a Class IIa or higher active device) → Class IIa (not named in the source)

• Tracheostomy cannula: MDR · Art. 56/Rule 5 → Class IIa/IIb (not named in the source)

• Aspiration catheter: MDR · Art. 56/Rule 5 → Class IIa (not named in the source)

• Oxygen mask and nasal cannula: MDR · Art. 56/Rule 5 → Class I/IIa (not named in the source)

• Surgical glove: MDR · Art. 56/Rule 1 (non-invasive) → Class I; if packaged sterile, Class Is in ÜTS

• Latex examination glove: MDR · Art. 56/Rule 1 → Class I; latex information is mandatory on the label and in ÜTS

• Non-sterile single-use glove: MDR · Art. 56/Rule 1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Surgical mask: MDR · Art. 56/Rule 1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Surgical gown and sterile drape: MDR · Art. 56/Rule 1 → Class I (Class Is if sterile) (not named in the source)

• Trocar: MDR · Art. 56/Rule 6 → Class IIa (not named in the source)

• Biopsy needle: MDR · Art. 56/Rule 6 → Class IIa (not named in the source)

• Scalpel blade (surgical blade): MDR · Art. 56/Rule 6 → Class IIa; reusable handle Rule 6-2 → Class Ir (not named in the source)

• Dialyser (haemodialysis filter): MDR · Art. 56/Rule 3-1 (modifying the chemical composition of blood) → Class IIb (not named in the source)

• Leukocyte filter / blood filter: MDR · Art. 56/Rule 3-1 exception (filtration) → Class IIa (not named in the source)

• Organ and tissue storage solution / container: MDR · Art. 56/Rule 2-2 (storage of organs, tissues, cells) → Class IIa

• Embryo/cell culture medium (ART medium): MDR · Art. 56/Rule 3-2 (in vitro contact with human cells, embryos) → Class III

• Single-use device (general): MDR · Art. 4, definition (fff); the "Is it single-use" field is mandatory in ÜTS

Wound Care and Dressings (13 products)

• Plaster (adhesive bandage): MDR · Art. 56/Rule 4-1 (mechanical barrier) → Class I; within the scope of Annex 3 of the Sales Regulation

• Adhesive tape (plaster tape): MDR · Art. 56/Rule 4-1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Cotton wool (hydrophilic cotton): MDR · Art. 56/Rule 4-1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Sterile gauze (swab, compress): MDR · Art. 56/Rule 4-1 → Class I (not named in the source)

• Modern wound dressing (hydrocolloid, foam dressing): MDR · Art. 56/Rule 4-3 (management of the micro-environment) or 4-4 → Class IIa (not named in the source)

• Burn dressing: MDR · Art. 56/Rule 4-2 (destruction of the dermis, healing by secondary intent) → Class IIb (not named in the source)

• Pressure ulcer (decubitus) dressing: MDR · Art. 56/Rule 4-2 or 4-3 → Class IIb/IIa (not named in the source)

• Silver-containing antibacterial wound dressing: MDR · Art. 56/Rule 4 + Rule 14 assessment (Class III if it contains a medicinal substance) (not named in the source)

• Elastic bandage / wrapping cloth: MDR · Art. 56/Rule 4-1 or Rule 1 → Class I (not named in the source)

• Hot/cold compress (non-invasive): MDR · Art. 56/Rule 1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Self-adhesive anti-pressure tape for blisters: MDR · Art. 56/Rule 4-1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Toe separator intended to relieve corns and blisters: MDR · Art. 56/Rule 1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Tourniquet (haemostatic tourniquet): MDR · Art. 56/Rule 1 → Class I (not named in the source)

Active Devices: Therapeutic (27 products)

• Haemodialysis machine: MDR · Art. 56/Rule 9-1 (active therapeutic, potentially hazardous energy) and Rule 12 → Class IIb (not named in the source)

• Peritoneal dialysis device: MDR · Art. 56/Rule 12 → Class IIb (not named in the source)

• Infusion pump: MDR · Art. 56/Rule 12 (administration of substances) → Class IIa, IIb if risky; under the MDD, Art. 53/Rule 11 (not named in the source)

• Syringe pump (injectomat): MDR · Art. 56/Rule 12 → Class IIa/IIb (not named in the source)

• Insulin pump: MDR · Art. 56/Rule 12 → Class IIb; if closed-loop, Rule 22 → Class III

• Ventilator (mechanical respirator): MDR · Art. 56/Rule 9-1 → Class IIb (not named in the source)

• CPAP / sleep apnoea device: MDR · Art. 56/Rule 9-1 → Class IIa/IIb (not named in the source)

• Nebuliser: MDR · Art. 56/Rule 20 (administration of a medicinal product by inhalation) → Class IIa, IIb in life-threatening situations

• Oxygen concentrator: MDR · Art. 56/Rule 12 → Class IIa/IIb (not named in the source)

• Aspirator (vacuum / secretion aspirator): MDR · Art. 56/Rule 12 (removal of substances from the body) → Class IIa (not named in the source)

• Electrosurgical unit (cautery, electrocautery): MDR · Art. 56/Rule 9-1 → Class IIb (not named in the source)

• Surgical laser device: MDR · Art. 56/Rule 9-1 → Class IIb (not named in the source)

• TENS device (electrotherapy device): MDR · Art. 56/Rule 9-1 → Class IIa (not named in the source)

• Ultrasound therapy device (physiotherapy ultrasound): MDR · Art. 56/Rule 9-1 → Class IIa (not named in the source)

• Magnetic field therapy device (PEMF): MDR · Art. 56/Rule 9-1 → Class IIa (not named in the source)

• Automated external defibrillator (AED): MDR · Art. 56/Rule 22 (active therapeutic device with an integrated diagnostic function) → Class III

• Manual defibrillator: MDR · Art. 56/Rule 9-1 → Class IIb

• Incubator (neonatal incubator): MDR · Art. 56/Rule 9-1 → Class IIb (not named in the source)

• Phototherapy device (neonatal jaundice lamp): MDR · Art. 56/Rule 9-1 → Class IIa/IIb (not named in the source)

• Radiotherapy device (linear accelerator): MDR · Art. 56/Rule 9-3 (therapeutic ionising radiation) → Class IIb

• Brachytherapy device: MDR · Art. 56/Rule 9-3 → Class IIb (not named in the source)

• Heart-lung machine (extracorporeal circulation pump): MDR · Art. 56/Rule 9-1 and Rule 3-1 → Class IIb/III (not named in the source)

• Blood warmer: MDR · Art. 56/Rule 3-1 exception (heat exchange) → Class IIa (not named in the source)

• Patient bed (motorised hospital bed): MDR · Art. 56/Rule 13 (all other active devices) → Class I (not named in the source)

• Operating table: MDR · Art. 56/Rule 13 → Class I (not named in the source)

• Operating light (scialytic): MDR · Art. 56/Rule 10-1 exception (illumination in the visible spectrum) → Class I

• Massage / compression device (lymphoedema pump): MDR · Art. 56/Rule 9-1 → Class IIa (not named in the source)

Active Devices: Diagnosis, Monitoring and Imaging (25 products)

• X-ray device: MDR · Art. 56/Rule 10 last paragraph (diagnostic ionising radiation) → Class IIb (not named in the source)

• Computed tomography (CT) device: MDR · Art. 56/Rule 10 last paragraph → Class IIb (not named in the source)

• Mammography device: MDR · Art. 56/Rule 10 last paragraph → Class IIb (not named in the source)

• Angiography device (interventional radiology): MDR · Art. 56/Rule 10 last paragraph: "interventional radiology devices" → Class IIb

• C-arm fluoroscopy device: MDR · Art. 56/Rule 10 last paragraph → Class IIb (not named in the source)

• Dental X-ray / panoramic X-ray: MDR · Art. 56/Rule 10 last paragraph → Class IIb (not named in the source)

• Bone density measurement device (DEXA): MDR · Art. 56/Rule 10 last paragraph → Class IIb (not named in the source)

• X-ray image recording device (digital detector, cassette): MDR · Art. 56/Rule 17 → Class IIa; under the MDD, Art. 53/Rule 16

• X-ray film: MDR · Art. 56/Rule 17 → Class IIa

• MRI (magnetic resonance) device: MDR · Art. 56/Rule 10-1 (energy absorbed by the body) → Class IIa (not named in the source)

• Ultrasound device (USG): MDR · Art. 56/Rule 10-1 → Class IIa (not named in the source)

• Doppler device: MDR · Art. 56/Rule 10-1 → Class IIa (not named in the source)

• Gamma camera / SPECT / PET device: MDR · Art. 56/Rule 10-2 (imaging the in vivo distribution of radiopharmaceuticals) → Class IIa

• ECG device (electrocardiograph): MDR · Art. 56/Rule 10-3 → Class IIa; IIb if monitoring immediate danger (not named in the source)

• Bedside monitor (vital signs monitor): MDR · Art. 56/Rule 10-3 second sentence (cardiac, respiratory, CNS parameters) → Class IIb (not named in the source)

• Pulse oximeter: MDR · Art. 56/Rule 10-3 → Class IIa/IIb (not named in the source)

• Holter device (rhythm / blood pressure Holter): MDR · Art. 56/Rule 10-3 → Class IIa/IIb (not named in the source)

• EEG device (electroencephalograph): MDR · Art. 56/Rule 10-3 (CNS activity) → Class IIb (not named in the source)

• Endoscope (gastroscope, colonoscope): MDR · Art. 56/Rule 5 last paragraph → Class IIa; Rule 6/7 for surgical endoscopes (not named in the source)

• Blood pressure monitor (blood pressure measuring device): MDR · Art. 56/Rule 10 for electronic ones → Class IIa; because of the measuring function, Class Im under the MDD (not named in the source)

• Fever thermometer (clinical thermometer): MDR · Measuring function → Class Im in ÜTS; a Rule 10 assessment is also made under the MDR (not named in the source)

• Stethoscope: MDR · Art. 56/Rule 1 → Class I (not named in the source)

• Otoscope / ophthalmoscope: MDR · Art. 56/Rule 10-1 exception (illumination) → Class I (not named in the source)

• Spirometer (pulmonary function testing device): MDR · Art. 56/Rule 10-3 → Class IIa (not named in the source)

• Scales / patient weighing scale (medical scale): MDR · Measuring function → Class Im; otherwise Rule 13 → Class I (not named in the source)

Software and Digital Health (11 products)

• Medical device software (standalone software, SaMD): MDR · Art. 52/3: if independent of a device, it is classified in its own right; Art. 56/Rule 11

• Mobile health application: MDR · Art. 56/Rule 11 → IIa if it feeds a diagnostic/therapeutic decision, IIb for serious deterioration/surgery, III for death/irreversible deterioration; Class I for the others (not named in the source)

• Diagnostic support software (AI-based decision support): MDR · Art. 56/Rule 11-1 and 11-2 → Class IIb/III

• Image processing and archiving software (PACS): MDR · Art. 56/Rule 11 → Class IIa (if it provides information for a diagnostic decision) (not named in the source)

• Radiotherapy planning software: MDR · Art. 56/Rule 11-1 → Class III (not named in the source)

• Patient monitoring software (remote monitoring platform): MDR · Art. 56/Rule 11: monitoring of physiological processes IIa, IIb for immediate danger in vital parameters

• Embedded software running a device: MDR · Art. 52/3: same class as the device

• Trisomy 21 risk assessment software: IVDR · mentioned together with "software" in IVDD Annex II List B; under the IVDR, Art. 57/Rule 3-12 → Class C

• Laboratory information management system (LIS): unclear · May not be a device if it does not interpret results; if it does, Art. 57 and Art. 56/Rule 11 are assessed together (not named in the source)

• Hospital information management system (HIS): not a medical device · Software for purely administrative/record-keeping purposes does not fall within the MDR definition of a medical device (Art. 4/ııı) (not named in the source)

• Wearable health tracker (smart watch, wristband): unclear · If a medical intended purpose is declared, Art. 56/Rule 10 and Rule 11; if only "wellness", it is not a device (not named in the source)

Disinfection and Sterilisation (9 products)

• Contact lens solution (lens care and disinfection solution): MDR · Art. 56/Rule 16-1 → Class IIb; under the MDD, Art. 53/Rule 15-1 → Class IIb

• Medical device disinfectant: MDR · Art. 56/Rule 16-2 → Class IIa

• Invasive device disinfectant solution (high-level disinfectant): MDR · Art. 56/Rule 16-2 second sentence → Class IIb; under the MDD, Art. 53/Rule 15-2

• Washer-disinfector: MDR · Art. 56/Rule 16-2 → Class IIb

• Autoclave (steam steriliser): MDR · Art. 56/Rule 16-2 (sterilising medical devices) → Class IIa (not named in the source)

• Endoscope washing and disinfection device: MDR · Art. 56/Rule 16-2 → Class IIb (not named in the source)

• Sterilisation pouch and indicator: unclear · Rule 16 does not apply to products that clean by physical action only; assessed as an accessory under Art. 52/2 (not named in the source)

• Surgical instrument washing machine: MDR · Art. 56/Rule 16-2 → Class IIa/IIb (not named in the source)

• Hand disinfectant (biocidal): not a medical device · Subject to biocidal legislation if not specific to device disinfection; Art. 4/1 of the Medical Device Regulation, borderline product provision

Dental Products (12 products)

• Dental implant: MDR · Art. 56/Rule 8-1 (placed in the teeth) → Class IIa; under the MDD, Art. 53/Rule 8-1

• Dental filling (composite/amalgam filling): MDD transition · Art. 56/Rule 8-1 → Class IIa; on the WET list, with an MDD certificate until the end of 31.12.2027

• Dental bracket (orthodontic bracket): MDD transition · Art. 56/Rule 8-1 → Class IIa; on the WET list

• Dental crown: MDD transition · Art. 56/Rule 8-1 → Class IIa; on the WET list

• Orthodontic wire: MDD transition · "Wires" are on the WET list; Art. 56/Rule 8-1 → Class IIa

• Impression material (dental impression material): MDR · Art. 56/Rule 5 (body orifice, transient use) → Class I (not named in the source)

• Toothpaste for relieving sensitivity: MDR · Art. 56/Rule 21-3 (substance-based device applied to the oral cavity) → Class IIa assessment; within the scope of Annex 3 of the Sales Regulation

• Denture care product for individual use: MDR · Art. 56/Rule 1 or Rule 21-3; within the scope of Annex 3 of the Sales Regulation

• Mouthwash: MDR · Art. 56/Rule 21-3 → Class IIa assessment; for the cosmetic borderline, Art. 4/1 of the Medical Device Regulation; within the scope of Annex 3 of the Sales Regulation

• Dental unit: MDR · Art. 56/Rule 13 → Class I; integrated components are classified separately (Art. 52/2) (not named in the source)

• Complete / partial denture: MDR · If made specifically for an individual, ITC (Art. 4/t); dental prosthesis laboratories are outside the scope of the Sales Regulation

• Root canal file and filling paste: MDR · Art. 56/Rule 5 / Rule 8-1 assessment → Class IIa (not named in the source)

Orthoses-Prostheses, Custom-Made Devices and Mobility Aids (16 products)

• Custom-made orthopaedic prosthesis (leg prosthesis, arm prosthesis): MDR · Registered as ITC; instead of classification, ITC registration under Art. 29 and the Annex XIII ITC Statement

• Custom-made orthosis: MDR · ITC; Art. 29/3-a: only a "Custom-Made Orthosis-Prosthesis Application Centre" or a healthcare institution applies

• Custom-made medical device (ITC, device made specifically for an individual): MDR · Art. 29: prescription, patient and doctor/dentist information are mandatory; ITC regime instead of risk class

• Orthopaedic insole: MDR · ITC if custom-made (Art. 29); if mass-produced, Art. 56/Rule 1 → Class I (not named in the source)

• Corset (lumbosacral orthosis, back brace): MDR · If mass-produced, Art. 56/Rule 1 → Class I (not named in the source)

• Neck collar (cervical orthosis): MDR · Art. 56/Rule 1 → Class I (not named in the source)

• Knee brace / ankle brace: MDR · Art. 56/Rule 1 → Class I (not named in the source)

• Arm sling (shoulder sling): MDR · Art. 56/Rule 1 → Class I (not named in the source)

• Varicose stocking (medical compression stocking): MDR · Art. 56/Rule 1 → Class I (not named in the source)

• Wheelchair (manual and powered): MDR · Manual: Art. 56/Rule 1 → Class I; powered: Rule 13 → Class I (not named in the source)

• Walker (walking frame): MDR · Art. 56/Rule 1 → Class I (not named in the source)

• Crutch: MDR · Art. 56/Rule 1 → Class I (not named in the source)

• Walking stick (medical cane): MDR · Art. 56/Rule 1 → Class I (not named in the source)

• Patient lift (transfer lift): MDR · Art. 56/Rule 13 → Class I (not named in the source)

• Air mattress (pressure ulcer prevention mattress): MDR · Art. 56/Rule 13 → Class I (not named in the source)

• Breast prosthesis (external prosthesis after mastectomy): MDR · Art. 56/Rule 1 → Class I (not an implant; must be distinguished from a breast implant) (not named in the source)

Optical Products (7 products)

• Contact lens (prescription, daily/monthly): MDR · The outer surface of the eyeball is regarded as a "body orifice" (Art. 51/2-ç) → Art. 56/Rule 5: short-term IIa, long-term IIb

• Prescription spectacle lens: MDR · Art. 56/Rule 1 → Class I (not named in the source)

• Spectacle frame: unclear · May have no medical purpose on its own; when supplied with lenses, assessed as an accessory under Art. 52/2 (not named in the source)

• Reading glasses (ready-made prescription glasses): MDR · Art. 56/Rule 1 → Class I (not named in the source)

• Intraocular lens (IOL): MDR · Art. 56/Rule 8 → Class IIb (not named in the source)

• Artificial tears / eye lubricant drops: unclear · A device if the mechanism of action is not pharmacological; the principal mode of action test of Art. 1/6 of the Medical Device Regulation applies (not named in the source)

• Coloured / cosmetic contact lens (plano): MDR · Assessed under the list of product groups without a medical purpose; the product name is not listed in the source (not named in the source)

Hearing Products (6 products)

• Hearing aid (behind-the-ear / in-the-ear): MDR · Art. 56/Rule 9-1 (active therapeutic, energy delivery) → Class IIa

• Bone-conduction hearing aid: MDR · Art. 56/Rule 9-1 → Class IIa; if there is an implanted component, Rule 8-6 → III (not named in the source)

• Hearing aid ear mould: MDR · ITC if custom-made (Art. 29); otherwise an accessory (Art. 52/2) (not named in the source)

• Audiometer: MDR · Art. 56/Rule 10-1 → Class IIa (not named in the source)

• Tympanometer: MDR · Art. 56/Rule 10 → Class IIa (not named in the source)

• Hearing aid battery and accessory: unclear · Art. 4/sss spare part / Art. 4/iii accessory distinction; if an accessory, classified in its own right under Art. 52/2 (not named in the source)

Contraception and Reproductive Health (5 products)

• Condom: MDR · Art. 56/Rule 15 → Class IIb; under the MDD, Art. 53/Rule 14 → Class IIb; within the scope of Annex 3 of the Sales Regulation

• Intrauterine device (IUD, coil): MDR · Art. 56/Rule 15 second sentence (implantable or long-term invasive) → Class III (not named in the source)

• Diaphragm (barrier contraceptive): MDR · Art. 56/Rule 15 → Class IIb (not named in the source)

• Lubricant gel (with a medical device claim): unclear · If there is a contraception/STD prevention claim, Rule 15 → IIb; otherwise a Rule 21-4 assessment (not named in the source)

• Device intended to support conception (ART consumable, IVF catheter): MDR · The definition of a medical device covers devices intended for the prevention or support of conception (Art. 4/ııı-2)

Incontinence, Hygiene and Home Care (mainly Annex 3 of the Sales Regulation) (9 products)

• Adult diaper: MDR · Art. 56/Rule 1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Incontinence pad: MDR · Art. 56/Rule 1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Bladder pad: MDR · Art. 56/Rule 1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Urinal (urine bottle): MDR · Art. 56/Rule 1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Bedpan (bed-type toilet container): MDR · Art. 56/Rule 1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Nasal dilator strip: MDR · Art. 56/Rule 1 → Class I; within the scope of Annex 3 of the Sales Regulation

• Paediatric nasal aspirator: MDR · Art. 56/Rule 5-1 (transient use, body orifice) → Class I; within the scope of Annex 3 of the Sales Regulation

• Stoma bag and care product: MDR · A stoma is regarded as a "permanent artificial opening" (Art. 51/2-ç); Art. 56/Rule 5 and Rule 4 assessment (not named in the source)

• Pacifier / feeding bottle (with a medical claim): unclear · Not a device unless a medical intended purpose is declared (Art. 4/ııı, definition of a medical device) (not named in the source)

Group of Products Without a Medical Purpose (assessed under MDR Annex XVI) (9 products)

• Filler (dermal filler, hyaluronic acid filler): MDR · Product group without a medical purpose; compliance with common specifications is mandatory (Art. 9/4 of the Medical Device Regulation) (not named in the source)

• Hair removal laser: MDR · Product group without a medical purpose; Art. 9/4-5 of the Medical Device Regulation (not named in the source)

• IPL device (intense pulsed light): MDR · Same group; if a medical purpose is also declared, both sets of requirements are met together under Art. 9/6 of the Medical Device Regulation (not named in the source)

• Lipolysis / liposuction equipment (fat removal device): MDR · Product group without a medical purpose (not named in the source)

• Cryolipolysis device (body contouring device): MDR · Product group without a medical purpose (not named in the source)

• Transcranial brain stimulation device (tDCS/TMS, non-medical): MDR · Product group without a medical purpose (not named in the source)

• Implant for cosmetic purposes (hip/calf silicone): MDR · Product group without a medical purpose (not named in the source)

• Fractional/RF skin device for cosmetic purposes: MDR · Product group without a medical purpose (not named in the source)

• Product from the group without a medical purpose (general): MDR · Art. 1/2 and 1/3-b of the Medical Device Regulation: these products are also called "devices" and the Regulation applies to them; they are within the scope of Annex 3 of the Sales Regulation

Surgical Instruments (6 products)

• Reusable surgical instrument (scissors, forceps, retractor): MDR · Art. 56/Rule 6-2 → Class I, Class Ir in ÜTS; under the MDD, Art. 53/Rule 6-2 → Class I. Art. 61/4: the notified body's involvement is limited to the aspects of reuse

• Surgical saw and motor system: MDR · Art. 56/Rule 9-1 → Class IIa (not named in the source)

• Laparoscopy hand instrument set: MDR · Art. 56/Rule 6 → Class IIa; reusable ones Rule 6-2 → Class Ir (not named in the source)

• Orthopaedic instrument set (rasp, impactor): MDR · Art. 56/Rule 6-2 → Class Ir; if sold together with implant screws, Art. 52/2 applies to each separately

• Surgical instrument sterilisation box (container): unclear · Art. 52/2 accessory assessment (not named in the source)

• Dental hand instrument (probe, mirror, curette): MDR · Art. 56/Rule 6-2 → Class Ir (not named in the source)

Product Categories as Registration Types (9 products)

• Procedure pack (surgical set, procedure kit): MDR · Not classified separately; a SİP registration is made. If it contains a device without CE marking or the combination is incompatible, it is regarded as a device in its own right

• System (assembled product combination): MDR · Art. 28: if at least one device it contains has moved to the MDR, the SİP registration is made under the MDR

• SİP placed on the market sterilised: MDR · Art. 22/3 of the Medical Device Regulation: notified body involvement for the sterility procedure

• Medical device accessory: MDR · Art. 52/2: classified in its own right, separately from the device it is used with

• Spare part: not a medical device · Art. 23/1 of the Medical Device Regulation: not regarded as a device if it does not change performance/safety/intended purpose. If it does, it is a device under Art. 23/2 and bears its own CE marking

• Component or part (without CE marking): not a medical device · Art. 4/ç: an integral part of a device; not regarded as a device on its own

• Configurable device: MDR · Art. 4/ğğ: components that combine in more than one configuration, each of which may be a device in its own right

• Investigational device (clinical investigation device): MDR · Art. 3/c of the Medical Device Regulation: a device assessed within a clinical investigation; the placing-on-the-market regime is different

• Device manufactured within a healthcare institution (in-house): unclear · Art. 5/4-5 of the Medical Device Regulation: regarded as put into service; if the conditions are met, the requirements of the Regulation do not apply

The Next Step After the Product Is Determined

The information in this list is not enough to make an application; it only shows the direction. Once the framework and probable class of your product are clear, the real work begins: your firm's ÜTS registration, which document type is required for that class and framework, whether the document from the manufacturer is in a form ÜTS accepts, and filling in the mandatory fields correctly. The order in this chain does not change; no document without a firm registration, and no product registration until the document reaches "Kayıtlı" (Registered) status.

Your firm's own staff can do this work, or you can have an outside consultant do it. The "Who Does the ÜTS Registration" section below explains in detail how the authority is set up, why the e-signature stays with you, and who bears the responsibility.

14 Product Groups That Are Thought to Be Medical Devices but Are Not

The fourteen headings below gather products and situations that go to the wrong door because they are mistaken for medical devices. For each item we give the provision that settles the matter together with the correct route.

• 1. Cosmetic products: Article 1, paragraph 4, sub-paragraph (d) of the Medical Device Regulation states that the regulation does not apply to products within the scope of the Cosmetics Regulation. The correct route is the cosmetic product notification.

• 2. Foods and food supplements: Sub-paragraph (ğ) of the same paragraph states that the regulation does not apply to foods within the scope of Veterinary Services, Plant Health, Food and Feed Law No. 5996. The correct route is food supplement approval and business registration/approval procedures at the Ministry of Agriculture and Forestry.

• 3. Medicines: Sub-paragraph (b) places medicinal products defined in the Regulation on Licensing of Medicinal Products for Human Use outside the scope. The test for the distinction is the principal mode of action criterion in paragraph 6 of the same article. The correct route is a TİTCK licence.

• 4. Gene therapy, somatic cell therapy and tissue engineering products: Sub-paragraph (c) places advanced therapy medicinal products covering these three outside the scope.

• 5. Human blood, blood products, plasma and blood cells: Sub-paragraph (ç) states that the regulation does not apply to these and to devices containing them, except for the devices referred to in paragraph 6. If a human blood derivative is present in a device as an ancillary substance to its function, the device is within the scope and the classification rules move it to Class III.

• 6. Products containing live biological material: Sub-paragraph (g) places outside the scope products containing live biological material or live organisms, including living micro-organisms, bacteria, fungi or viruses, in order to achieve or support the intended purpose of the product.

• 7. Transplants, tissues and cells of animal and human origin: Sub-paragraphs (e) and (f) keep these outside the scope. Do not miss the reverse: devices manufactured using non-viable or rendered non-viable tissue or cell derivatives are within the scope under sub-paragraphs (c) and (ç) of paragraph 3 of Article 1, and go to Class III.

• 8. Spare parts: Article 23, paragraph 1 does not regard a spare part that does not change performance, safety characteristics or intended purpose as a device. Paragraph 2 states that a part that significantly changes them will be considered a device and must meet the requirements of the regulation. The distinction lies in the effect of the part, not in its name.

• 9. Components and parts without CE marking: The definitions article of the Manual describes this as an integral part and component of a device that is not defined as a device on its own and does not bear a CE marking. It is not registered separately.

• 10. Sale of second-hand devices: Article 1, paragraph 12 states that the regulation does not cover the rules on making devices available on the market by changing hands after they have been put into service, such as second-hand sales. This does not remove the obligations at the first supply of the device; it only leaves the subsequent change of hands outside the regulation.

• 11. Devices manufactured and used within a public healthcare institution: Article 2, paragraph 3 of the Medical Device Sales, Advertising and Promotion Regulation keeps these devices outside the scope, provided that they comply with the other relevant technical regulations. Paragraphs 4 and 5 of Article 5 of the Medical Device Regulation also establish a conditional exemption for devices manufactured within a healthcare institution.

• 12. Healthcare institutions, optician establishments, custom-made prosthesis-orthosis centres, hearing aid centres and dental prosthesis laboratories: Article 2, paragraph 2 of the Sales Regulation keeps these places outside the scope. But what follows is critical: trade, advertising and promotion of the devices placed on the market in these places by means of distance communication, and registration procedures for the record and information management systems determined by the Agency, are within the scope of the regulation. You do not obtain a sales centre authorisation certificate, but the ÜTS registration obligation continues.

• 13. In vitro diagnostic devices are not within the scope of the Medical Device Regulation: sub-paragraph (a) of Article 1, paragraph 4 says this expressly; the applicable one is the In Vitro Diagnostic Medical Device Regulation. Registration is still made in ÜTS, but the framework, classification rules and fields are different.

• 14. Electromagnetic Compatibility Regulation: Article 1, paragraph 9 states that the provisions of the Electromagnetic Compatibility Regulation are not sought for devices subject to conformity assessment under this regulation. A buyer's request for a separate certificate under that regulation for an electrical medical device can be rejected on the basis of this provision.

The product groups listed in TİTCK's announcement of 3/11/2023 can also be added to this list: office and furniture items, patient and baby wristbands, examination screens, medicine and device transport trolleys, general-purpose disinfectants not specific to a device, air filtration and air conditioning units, tooth whitening products, sunscreen creams, menstrual cups, chemotherapy drug preparation units, personal protective equipment, general laboratory equipment not intended for in vitro diagnosis, and products for research purposes only. Of these, general-purpose disinfectants go to a biocidal licence, tooth whitening products and sunscreens to cosmetic notification, and personal protective equipment such as masks and gloves to its own legislation.

16 Products That Are Not Thought to Be Medical Devices but Are

This section does the opposite. The sixteen headings below gather products whose registration is never opened because they are taken for ordinary consumer goods or cleaning materials, yet which the source texts treat directly as devices.

• 1. Products without a medical purpose: Under paragraphs 2 and 3 of Article 1 of the Medical Device Regulation, products within the list of product groups without a medical purpose are also called "devices". Not declaring a medical purpose does not remove the registration obligation.

• 2. Condom: The classification rules place all devices used for contraception or for the prevention of sexually transmitted diseases in Class IIb. It is also listed at item 3 of Annex 3 of the Sales Regulation among the devices that can be sold outside sales centres. Being in Annex 3 does not mean it is not a device; the title of that annex is already "list of devices".

• 3. Surgical mask: It is at item 11 of Annex 3 of the Sales Regulation. The exemption concerns the place of sale, not the status as a device. On the classification side it is Class I under the non-invasive device rule.

• 4. Cotton wool, plaster and adhesive tape: Items 7 and 8 of Annex 3. The same logic applies: the place of sale is free, while the status as a device and the ÜTS registration stay in place.

• 5. Mouthwash and toothpaste for relieving sensitivity: Items 9 and 1 of Annex 3. These two products, thought to be cosmetics, appear in the device list when a medical intended purpose is declared. The boundary between cosmetic and device is resolved here by looking at the claim of the product.

• 6. Adult diaper, incontinence pad, urinal and bedpan: Items 4 and 5 of Annex 3. Incontinence care products are listed together in the annex.

• 7. Non-sterile single-use glove: Item 14 of Annex 3. When presented as sterile, the class shown in ÜTS changes because of sterility.

• 8. Software: The definition of active device in the Manual covers software; software that is a medical device on its own is regarded as an active medical device. The general rule is clear: software that drives a device or influences the use of a device falls in the same class as the device; if the software is independent of any device, it is classified in its own right. There is a separate classification rule for software.

• 9. Contact lens solution: The classification rule names it: devices intended specifically for disinfecting, cleaning, rinsing or hydrating contact lenses are Class IIb. This product, taken for a cleaning product, is one of the few products named in the rules.

• 10. Blood bag: The rule in the classification of the old regulation settles this in one sentence: unlike the other rules, blood bags are classified as Class IIb. In the new regulation too, the rule on the transfusion and storage of blood gives the same result. It looks like a simple plastic bag; its class is IIb.

• 11. Specimen receptacles: The in vitro diagnostic classification rules list specimen receptacles as Class A. Urine containers, stool containers and similar products are in vitro diagnostic devices; their registration is also made under the in vitro diagnostic framework.

• 12. Buffer solutions, wash solutions, general culture media and histological stains: The same group of rules also treats these as Class A in vitro diagnostic devices. This is the most crowded product family left unregistered on the grounds that it is "laboratory consumables".

• 13. Reusable surgical instruments: The classification rule places them in Class I; it is not "no class". The Manual also states that the notified body's involvement in these devices is limited to aspects relating to the reuse of the device, chiefly cleaning, disinfection, sterilisation, maintenance, functional testing and instructions for use.

• 14. Suture, staple, dental filling, dental bracket, dental crown, screw, wedge, plate, wire, pin, clip and connector: Article 18, paragraph 4 of the Medical Device Regulation exempts these products only from the implant card obligation. It does not exempt them from being devices. When this distinction is confused, the conclusion is drawn that since no implant card is needed, no registration is needed either; the provision does not say that.

• 15. Calibrators and control materials: The general rules say that calibrators intended to be used with a device and controls with assigned values are in the same class as the device. Controls without an assigned value are Class B under a separate rule.

• 16. Accessories: The general rule is clear: accessories of a medical device are classified in their own right, separately from the device they are used with. The same provision is established for in vitro diagnostic device accessories. The assumption that "the accessory is included in the registration of the main device" is wrong; an accessory is both classified separately and registered separately.

How Is the Decision Made for Borderline Products?

If your product falls between two sets of legislation, writing your own decision and proceeding is the most expensive option. The way to follow is first to read correctly what the provision says, and then to use the application route that actually operates in Turkey.

Article 4 of the Medical Device Regulation Is Not a Domestic Application Route

There is a common mistake here and it has to be corrected. Article 4 of the Medical Device Regulation is titled "Regulatory status of products" and its first paragraph says this: the Agency submits its duly reasoned request to the Commission, where necessary, to determine whether a specific product, or a category or group of products, falls within the definition of "medical device" or "medical device accessory", or to clarify the regulatory status of products that are borderline between medicinal products, human tissues and cells, cosmetic products, biocidal products or food products on the one hand and medical devices on the other. The second paragraph provides that the Agency will share expertise with the Commission on products within its own field of duty.

The critical detail is in the definitions article of the same regulation: "Commission" means the European Commission. In other words, the wording of Article 4 is not a provision regulating firms' applications to TİTCK; it describes the relationship between the Agency and the European Commission. The sentence "apply to TİTCK for a borderline product on the basis of Article 4 of the Regulation" does not match the text of the article.

The value of the article for you lies elsewhere: it is the provision that officially lists the categories of borderline products. Medicinal products, human tissues and cells, cosmetic products, biocidal products and food products. If your product stands between one of these five areas and medical devices, you are in the borderline product discussion.

The Route That Actually Operates in Turkey

The route for a firm to obtain an official classification opinion in Turkey is TİTCK's Classification and Borderline Products Opinion Application procedure. The Agency has announced that these applications are to be made electronically; the application channel is TİTCK's Electronic Application and Process Management System, and the deciding authority is TİTCK.

We do not give the documents required in the application, the evaluation period or the fee on this page; these items could not be confirmed from a verifiable official source and they change over time. Check the Agency's current announcement before applying.

Evaluation Criteria: Six Criteria, One Decision

TİTCK's Guide on Cosmetic Products and Borderline Products, in its 5th edition dated 14/10/2024, lists the criteria looked at in a borderline decision. The Guide deals with the boundary of cosmetics with medicines, medical devices, biocides, toys, textiles and food.

• Composition: Which substances are in the product, and in what concentration.

• Intended purpose: What the product is intended to be used for.

• Site and method of application: Does it stay on the skin surface, is it applied to a body orifice, is it injected.

• Presentation: The packaging, the form and the perception created by the presentation.

• Mode of action: By what route the product produces its effect.

• Product claims: What is said on the label, in promotion and in advertising.

The point the Guide underlines is that these criteria are assessed together. No single criterion decides. The reasoning "its composition is a cosmetic composition, therefore it is a cosmetic" or "it is in cream form, therefore it is a cosmetic" runs against the approach of the Guide.

The concrete examples that come out of the Guide are indicative. An anti-dandruff shampoo containing a corticosteroid, a skin lightener containing hydroquinone or mercury, and a shampoo containing minoxidil are not cosmetics because of a prohibited or pharmacologically active ingredient. Applications injected beneath the epidermis cannot be cosmetics. Oral slimming products are not cosmetics because they are ingested. Products carrying a claim to remove or treat acne, papules and pustules fall outside cosmetics because of the therapeutic purpose. Leave-on antiseptic or antibacterial hand gels go to the biocidal side.

There are surprising results in the opposite direction as well. According to the same Guide, face and body paints applied by airbrush technique, temporary transfer tattoos applied to the skin after wetting, toothbrushes that release a cosmetic substance, and garments that release a cosmetic substance to clean or deodorise the skin are regarded as cosmetics. Foundations and concealers that work only by masking or covering a colour disorder are also cosmetics; the moment a therapeutic claim is added to the product, it loses this status.

Decision Flow

In practice the following order shortens the work. First write down the claim of the product: does it clean, protect, beautify; does it kill germs; does it diagnose or treat disease. Then determine the principal mode of action: pharmacological, immunological or metabolic, or physical or mechanical. Then clarify the route of application: is it taken orally and absorbed, does it stay on the skin surface, is it injected into the body. If uncertainty remains after these three steps, ask TİTCK for an opinion instead of writing your own decision.

Sanctions Applied if There Is No ÜTS Registration

The first consequence of a missing registration is not a penalty but the stopping of sales. Article 26, paragraph 2 of the Medical Device Sales, Advertising and Promotion Regulation says in a single sentence that sales centres cannot sell devices that are not registered in the Agency's record and information management system. In other words, if your product is not registered, a sales centre that wants to buy from you will not buy. The commercial chain breaks at the very first link.

Paragraph 1 of the same article closes the other end of the chain: devices cannot be sold by natural or legal persons who are not authorised under the provisions of the regulation.

Registration is an obligation, not a choice. Article 34, paragraph 1 of the Medical Device Regulation provides that, without prejudice to obligations relating to EUDAMED, economic operators established in Turkey, notified bodies, healthcare institutions, health professionals and other relevant persons will fulfil their obligations relating to the national product tracking system. Paragraph 2 says that these obligations will be determined by regulations published by the Agency. The Manual is the regulation in question.

Registration is also not a job done once and left. Article 31, paragraph 4 of the Medical Device Regulation requires economic operators to update the data in the system within 1 week of any change in their registration information. This is a period the regulation gives to the firm. Paragraph 5 of the same article also establishes a verification obligation: the accuracy of the data is confirmed within 1 year of submission and every 2 years thereafter; if this is not done within 6 months of the end of these periods, the Agency may take appropriate corrective measures until the economic operator complies.

The accuracy of the data is also yours. Article 31, paragraph 4 of the Medical Device Sales, Advertising and Promotion Regulation says that the Agency will issue guidelines on the registration of devices in the record and information management system, and that sales centres will set up an infrastructure in line with these guidelines within the stated periods and register the devices in the system; the sentence that follows is clear: the accuracy of the information on devices registered in the system is the responsibility of the sales centre that made the product registration. Even if you work with a consultant, this responsibility stays with you.

This page does not include administrative sanction amounts; the amounts could not be confirmed from a verifiable current source and they change during the year.

The Registration Order Does Not Change: Firm, Document, Product

Registration in ÜTS has three stages and the order is fixed. First the firm registration is made, then the document registration, and last the product registration. This order is not a preference but the working logic of the system: the declaration of conformity and the certificate you will select in the product registration must already be registered in the system at that stage. Without a document registration you will find nothing to select on the product application screen.

The firm registration defines who will enter data into the system; the importer, manufacturer, authorised representative and distributor roles each have their own requirements. The document registration is the introduction to the system of the certificates, declarations of conformity, instructions for use and intended purpose form showing the conformity of the device; Article 39 of the Manual regulates document application procedures. The product registration is the registration of the device with its barcode, name, class, dimensions and images; Article 40 of the Manual lists the application types as optical products, active implantable medical devices, Class 3 products, other products, custom-made medical devices, system and procedure packs, and products within the scope of an announcement.

The Manual also sets volume limits. A document application contains at most 5 document notifications. For standard device application types, one application contains at most 500 medical device notifications, while for custom-made medical device and system and procedure pack applications this limit is 5. For firms with a wide portfolio these limits directly affect application planning.

Each of these three stages is explained on separate pages: see the firm registration, document registration and product registration headings. The most frequent deadlock is noticing, once product registration has begun, a document that was left incomplete on the document side.

Who Does the ÜTS Registration: You or a Consultant?

Most firms that reach this page get stuck in the same place. It is understood that the product is a medical device and its class becomes clear, but when it comes to sitting down at the Product Tracking System, the work stops. The reason is not ignorance; ÜTS has its own logic, and learning that logic for the first time takes much longer than registering a product. The firm registration, document registration and product registration are linked to one another; a document cannot be attached to a device until it reaches "Kayıtlı" (Registered) status, a document cannot be selected until the class is selected, and when a single field is entered incorrectly the application is returned.

For this reason ÜTS registration in Turkey is carried out in two ways: the firm does it with its own staff, or has an outside consultant do it. The authority structure of the system is open to both. What matters is knowing correctly how the authority is set up, because there is a lot that is wrongly believed about this in the sector.

How Authority Is Given to a Consultant

The authority is not something the consultant gives you; it is a user you open from your own firm page in ÜTS. After your firm registration is made, users are defined in the system on behalf of your firm and roles are assigned to these users. The role of the person who will carry out the registration procedures is "Tıbbi Cihaz Kayıt Elemanı" (Medical Device Registration Operator).

Article 39, paragraph 11 of the Manual says that the approval step on the document application summary screen can be performed only by users of the relevant firm of the "Tıbbi Cihaz Kayıt Elemanı" type. Article 40, paragraph 5, sub-paragraph (f) establishes the same provision for the device application. In other words, the consultant works as a user defined on behalf of your firm; it is not an independent account connecting from outside.

In practice this means: you open the authority and you close it. The consultant's presence in the system is tied to your firm page, and you can remove that authority at any moment. When the relationship ends, no access remains in the consultant's hands.

Your E-Signature Does Not Pass to the Consultant

This is the point firms worry about most, and the worry is misplaced. The consultant prepares and approves the registration, but the signature that sends the application to the Agency stays with you.

Article 39, paragraph 13 of the Manual says that a document application in the "İmza Bekliyor" (Awaiting Signature) status will be signed electronically by the firm's applicant signatory. Sub-paragraph (g) of Article 40, paragraph 5 also states that for a device application to be forwarded to the Agency, it must first be signed electronically by the firm's signatory.

Read together, these two provisions make the picture clear: the consultant carries out the preparation and approval stage, and the signature stage rests with the firm's signatory. You do not need to hand your e-signature to anyone, and you should not. If a consultant asks you for your e-signature, they are asking for something the Manual does not provide for.

Who Keeps the Responsibility

Delegating authority is not delegating the obligation. The accuracy of the information on devices registered in the system is the responsibility of the firm that made the registration. Article 34 of the Medical Device Regulation places the registration obligation on the economic operator; the consultant does not assume this obligation, but sees that it is fulfilled.

This distinction is not abstract; it determines the daily work. The consultant does not invent the technical data of the product; they read the document coming from the manufacturer, write it correctly into the field the Manual asks for, and tell you before the application if there is anything that does not comply with the business rules. If a document you hold is incomplete, the right way is to have the document completed, not to fill in the field and move on.

Which Part of the Work Is Yours and Which Is the Consultant's

• Stays with you: the e-signature of the signatory, the signature stage, the commercial relationship with the manufacturer, obtaining the originals of the documents, the legal responsibility for the registration obligation

• Done by the consultant: determining which regulatory framework the product falls into, applying the classification rule, determining which document type is needed, checking whether the document is in the form ÜTS requires, setting up the firm-document-product order, filling in the mandatory fields, preparing and approving the application, and working out the reasons if a rejection or revision comes

• Done together: producing the product list, identifying missing documents and requesting them from the manufacturer, following up post-registration changes

The Real Work Starts Before the Registration

Entering data on the screen is the easiest part of the job. The reason applications are returned is usually not pressing the wrong key but starting with the wrong decision: the framework of the product was chosen wrongly from the start, the class was determined wrongly, the selected document type is not valid for that class in that framework, or the document from the manufacturer is not in the form ÜTS accepts.

These decisions are made before the application, and going back is expensive when they are made wrongly. The class of a device registered with the wrong class can only be corrected with a data change request approved by the Agency. For this reason, at Medex the first step of the work is not registration but determination: the products and documents you hold are read one by one, and which framework they fall into, which classification rule applies and which document is missing are set out in writing before the application is opened.

Medex's Working Method

Medex Kurumsal Danışmanlık does only this work: carrying out firm, document and product registrations in the Product Tracking System of the Turkish Medicines and Medical Devices Agency. Firms that bring medical devices from abroad must have their ÜTS registration made in order to place their products on the market in Turkey; firms that do not want to carry out this work in-house, or that meet it for the first time, have us do the registration.

The process is this: from your own firm page in ÜTS, you open the "Tıbbi Cihaz Kayıt Elemanı" (Medical Device Registration Operator) authority for us; with that authority we create and approve the registrations, and the signature stage stays with you. The authority is under your control, your e-signature stays with you, and the registration obligation legally rests with you - we make sure that obligation is fulfilled correctly.

Frequently Asked Questions

The questions below answer what is most often searched for on product registration. The answers rest on the provisions of the legislation; notes based on practice are stated separately.

My Product Has a CE Certificate; Do I Still Need to Register It?

Yes. The CE marking shows that the product has completed conformity assessment under European legislation. The registration obligation in Turkey is a separate obligation and arises from Article 34 of the Medical Device Regulation. Your CE certificate is a precondition for ÜTS registration; it does not replace it. Indeed, in the registration process your certificate and declaration of conformity are introduced to the system at the document registration stage.

I Am Bringing a Sample; Is Registration Needed?

It depends on what the sample is brought for. If the product will not be placed on the market, sold or put into service, the placing-on-the-market obligations do not arise. But if the sample is left with a customer, given for trial or made available for use, even free of charge, the situation changes. Selling on the strength of sample status does not remove the registration obligation.

Is Registration Required for a Class I Device Too?

It is. Being Class I does not remove the registration obligation; it only changes the conformity assessment route and the involvement of the notified body. This is the most frequent mistake with Class I products such as wheelchairs, crutches, patient beds and examination gloves. The scope provision of the Manual makes no distinction by class; every product that falls within the definition of a device is registered.

Do I Need a Medical Device Registration in ÜTS for My Cosmetic Product?

No. Sub-paragraph (d) of Article 1, paragraph 4 of the Medical Device Regulation places cosmetic products outside the scope. The procedure for your cosmetic product is the notification under Article 16 of the Cosmetic Products Regulation, and this is not an approval. Although the cosmetic notification and the medical device registration are under the same roof, they are separate lines.

Which System Does a Hand Disinfectant Go Into?

It depends on what the product is for. A hand antiseptic and hand disinfectant gel applied to human skin belong to the group of human hygiene biocides and fall within TİTCK's licensing field. Surface and instrument disinfectants go to HSGM. Products specific to the disinfection of medical devices are medical devices under the classification rules. All three doors are different; the intended purpose on your product's label decides which one you go to.

Is Registration Needed for a Wheelchair?

Yes. A wheelchair is a medical device; the manual one is assessed under the non-invasive device rule and the powered one under the general rule on active devices, and both arrive at Class I. Being Class I does not make registration unnecessary. The same misconception is seen with home care products such as patient beds, patient lifts and air mattresses.

Is My Software Registered?

Software can be a medical device on its own. The general rules state that software that drives a device is in the same class as the device, and that software independent of any device is classified in its own right. What is decisive is whether the software serves a medical purpose. Hospital information management systems for purely administrative or record-keeping purposes do not fall within the definition of a medical device.

I Sell Second-Hand Devices; Is Registration Needed?

Article 1, paragraph 12 of the Medical Device Regulation states that the regulation does not cover the rules on making devices available on the market by changing hands after they have been put into service, such as second-hand sales. This provision does not remove the obligations at the first placing of the device on the market. The situation is different for refurbished or reprocessed devices; those products are subject to their own regime.

Is a Spare Part Registered?

No. The scope provision of the Manual says so expressly: parts defined as spare parts are not registered in ÜTS. However, note paragraph 2 of Article 23 of the Medical Device Regulation: a part that significantly changes the performance, safety characteristics or intended purpose of the device is considered a device. In that case the part is no longer a spare part, bears its own CE marking and is registered.

My Product Has Both Cosmetic and Medical Device Characteristics; What Should I Do?

First write down the six criteria for your own product: composition, intended purpose, site and method of application, presentation, mode of action and product claims. TİTCK's Guide on Cosmetic Products and Borderline Products requires them to be assessed together. Do not decide on the basis of a single criterion. If the result is still unclear, request a classification and borderline product opinion from TİTCK for the specific product.

If I Bring It From Abroad Just Once, Is Registration Needed?

The frequency of bringing it in does not determine the obligation; whether the product is placed on the market does. Even if it is a single batch, if the product is to be sold or put into service in Turkey, the registration obligation arises. Personal use, clinical investigation and similar special regimes are subject to separate provisions and follow their own application routes.

Food Supplement or Medical Device?

The distinction looks at how the product produces its effect. Products that are taken orally and digested, absorbed or metabolised are on the food or medicine side; sub-paragraph (ğ) of Article 1, paragraph 4 of the Medical Device Regulation places foods outside the scope. Products that are not absorbed even when taken orally and that work by physical action can be assessed as devices; this is not a generalisation, and a separate opinion is needed for each product. Even if the product stays a food, if you will use a health claim, the competent authority is TİTCK.

Is Registration Needed for a Mask?

A surgical mask is a medical device and is listed at item 11 of Annex 3 of the Sales Regulation. Being in Annex 3 is only an exemption concerning the place of sale; it does not remove the product's status as a device or the registration obligation. On the other hand, masks placed on the market as personal protective equipment are not medical devices and are subject to their own legislation. What decides which regime you are in is the declared intended purpose of the product.

I Sell Supplies to a Dental Clinic; Who Registers?

The importer that brings the product to Turkey is responsible for the registration; the clinic is the buyer. Under Article 26, paragraph 2 of the Sales Regulation, sales centres cannot sell unregistered devices, so without registration the sales chain cannot be built. Dental prosthesis laboratories are outside the scope of the regulation, but paragraph 2 of Article 2 keeps registration procedures within the scope for these places as well.

Is an Accessory Registered Separately?

It is registered separately and classified separately. The general classification rules of the Manual say that accessories of a medical device are classified in their own right, separately from the device they are used with; the same provision exists for in vitro diagnostic device accessories. The scope provision of the Manual also expressly lists accessories among those registered in ÜTS. The assumption "I registered the main device, it covers the accessory too" is wrong.

Can I Promote at a Trade Fair Without Registration?

Promotion and sale are separate acts, but both are within the scope of the same regulation: the Medical Device Sales, Advertising and Promotion Regulation regulates advertising, information, clinical support, promotion and marketing activities. If you also make sales at the fair, the prohibitions in paragraphs 1 and 2 of Article 26 apply directly. In practice, having the registration completed before the fair keeps you in a safe position on both sides.

What Is Needed First to Enter SGK?

First the ÜTS registration, then the Healthcare Implementation Communiqué application. The two run at different agencies and on different portals, and SGK does not accept applications through ÜTS or manually; it uses its own portal. The application procedure is regulated by the Directive on Working Procedures and Principles Concerning Medical Supply Applications dated 17/10/2024. A device that is not on the list can also be sold legally in Turkey; its price is simply paid by the patient.

Is ÜTS Required to Enter DMO?

A DMO registration is a separate commercial channel and does not replace a ÜTS registration. For a medical device to be sold to a public institution, it must be possible to place it on the market legally in Turkey, which requires ÜTS registration; supplier pre-registration on the DMO side and the catalogue or framework agreement application come on top of that as a separate step. The order is: first the legal status of the product, then the commercial channel.

Can I Do the Registration Myself?

You can. The Manual ties the approval of the application to users of the firm of the "Tıbbi Cihaz Kayıt Elemanı" (Medical Device Registration Operator) type; this user can be the firm's own staff. Working with an outside consultant is not an obligation but a matter of preference. What to look at when deciding is the size of your product portfolio, the complexity of the classification and how often gaps on the document side come up.

If I Give Authority to a Consultant, Does My E-Signature Pass to Them?

No. Article 39, paragraph 13 of the Manual says that an application in the "İmza Bekliyor" (Awaiting Signature) status will be signed electronically by the firm's applicant signatory; sub-paragraph (g) of Article 40, paragraph 5 establishes the same order for device applications. The consultant creates and approves the registrations with the registration operator authority; the signature step stays with the firm. You do not need to hand over your e-signature.

Are ÜTS and EUDAMED the Same System?

No. EUDAMED is the European database system for medical devices. ÜTS is Turkey's national product tracking system. Article 34, paragraph 1 of the Medical Device Regulation establishes the national obligation with the words "without prejudice to obligations relating to EUDAMED". Having a registration on the European side does not remove your obligation in Turkey.

My Product Is Made to Order; Will I Make a Normal Registration?

No, custom-made medical devices are registered in a separate flow. Article 40, paragraph 3 of the Manual lists the "Custom-Made Medical Device Application" as a separate application type, and such applications contain at most 5 device notifications. For orthoses and prostheses made to order, the party that can apply is also limited. If you make individual-specific products, you are in this flow, not on the standard device registration screens.

My Product Is Sold as a Set; Will I Register Each Part Separately?

It depends on the nature of the set. Product combinations intended to be put together to achieve a specific medical purpose are assessed as a system or procedure pack and are registered in ÜTS with the "Sistem İşlem Paketi Başvurusu" (System and Procedure Pack Application) type; these applications contain at most 5 notifications. If the combination is incompatible, or the pack contains a device without CE marking, the product is assessed as a device in its own right.

I Have No Firm Registration; Can the Consultant Open the Firm Registration on My Behalf?

The firm registration is the first link of the chain, and at this step you do not yet have a firm page from which to grant authority. In practice the process runs like this: first your firm's ÜTS registration is made, and once the registration is complete, a user is defined through your firm page and the registration operator authority is opened. At this first step the consultant shows you what is needed, which documents will be prepared and in what order to proceed; they carry out the procedures in the system after the authority is opened.

Can I Remove the Authority I Gave Later?

Yes. The authority is a user defined on your firm page, not an independent account connecting from outside. You define the user, you assign the role and you remove it at any moment. When the work ends, no access remains in the consultant's hands. Giving authority therefore does not mean handing over control of your system.

Even If I Get Consultancy, Does the Registration Obligation Stay With Me?

Yes, it does. Article 34 of the Medical Device Regulation places the registration obligation on the economic operator, and this obligation cannot be transferred by contract. The accuracy of the information entered in the system is also the responsibility of the firm that made the registration. The consultant's job is not to assume the obligation but to see that it is fulfilled correctly. We say this openly, because a consultancy relationship that promises otherwise does not comply with the legislation.

I Have Hundreds of Products; Will I Register Them One by One?

No. The Manual has set aside a separate section for bulk procedures: bulk device addition and update, bulk deletion, bulk addition of placing-on-the-market information and bulk addition of storage/use conditions are defined. For catalogues with many products, the work is not entering data one by one but preparing the data in the catalogue correctly in the template the Manual requires. Most of the errors also arise at this preparation stage.

My Application Was Rejected; Can the Consultant Fix It?

In a rejected application, first it is worked out which rule the reason for rejection rests on; then the correction is made in line with that rule. The reason is sometimes a single field, sometimes a wrong decision taken earlier - for example, the framework or class of the product was chosen wrongly from the start. In the second case the correction starts not from the application screen but from the decision itself. So the first job after a rejection is not to reapply but to read the reason correctly.

Our EUDAMED registration service page (in English) is a guide for its counterpart on the EU side.

ÜTS Registration

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara