When you register a surgical set, a dialysis kit or a prosthesis made specifically for one patient in ÜTS, you do not get the familiar product registration screen

System/Procedure Pack (SİP) and Custom-Made Medical Device (ITC) ÜTS Registration

When you register a surgical set, a dialysis kit or a prosthesis made specifically for one patient in ÜTS, you do not get the familiar product registration screen. The system/procedure pack (SİP) and the custom-made medical device (ITC) are two special registration routes in ÜTS, each with its own application type, mandatory fields and document rules. Below we explain both routes based on the Guide on Registration Processes of Medical Devices to the Product Tracking System (the TİTCK Guide).

What Is a SİP? Definition of System and Procedure Pack

SİP is the abbreviation of "system or procedure pack" in the definitions article of the Guide. The two concepts are close but not identical; in ÜTS registration both go through the same application type.

  • System: a combination of products, packaged together or not, that are intended to be connected or combined to achieve a specific medical purpose.

  • Procedure pack: a combination of products packaged together and placed on the market to be used for a specific medical purpose.

  • SİP manufacturer: the natural or legal person or company that assembles the SİP and issues the SİP Declaration for it; the Guide counts it as a separate role within the definition of economic operator.

Under Article 14 of the Guide, the SİP manufacturer combines devices bearing the CE mark within the scope of MDR together with other CE-marked devices, in vitro diagnostic devices bearing the CE mark in accordance with IVDR and, only where their use in a medical procedure or their presence in the pack is separately justified, other products. The combination must be made in line with the intended purpose of the devices and within the limits of use set by their manufacturers.

Will You Register the SİP Under MDD or MDR?

This is the first and most decisive decision in a SİP registration. If you choose the wrong framework, both the on-screen fields and the required documents change. The rule looks at what is inside the pack, not at the preference of the registering company.

  • Registration under the old directives (MDD): under the first paragraph of Article 26 of the Guide, a SİP registration notification under the old directives is made only if all devices in the pack are placed on the market with a CE mark under MDD.

  • Registration under the new regulation (MDR): under the first paragraph of Article 28 of the Guide, if at least one of the devices in the pack has completed its transition to MDR, the SİP is registered under MDR.

During the transition period the framework of the same pack can change over time: when a single device in the pack moves to MDR, the whole pack moves to the MDR side. Before registering, it is worth checking the ÜTS framework of every device in the pack.

SİP Declaration: MDD Article 12 and MDR Article 22

The document of a SİP is a declaration, not a certificate, and it is issued by the SİP manufacturer itself, not by a notified body. In both frameworks the declaration rests on three commitments.

Under the second paragraph of Article 14 of the Guide, in its declaration the SİP manufacturer states that it has verified the mutual compatibility of the devices and, where applicable, the other products in accordance with the manufacturers' instructions, that it has packaged the system or procedure pack and supplied the relevant information to users, including the information provided by the manufacturers, and that the assembly activity was subject to appropriate methods of internal monitoring, verification and validation.

  • Under the old directives: under the eighth paragraph of Article 26 of the Guide, the SİP Declaration prepared according to Article 12 of MDD is uploaded in the "Original Declaration" field; it is issued by the SİP manufacturer that assembles the CE-marked devices.

  • Under MDR: the document type in ÜTS is called "SİP Declaration (MDR) (Article-22)". Under the eighth paragraph of Article 28 of the Guide, selecting a declaration of this document type is mandatory.

  • Link between the declaration and the registration: Article 26 of the Guide requires the SİP Declaration to contain at least one piece of identifying information on the SİP and the devices in it (such as primary product number, brand name, product name, model/version, reference/catalogue number), and this information must match what is entered in ÜTS.

If the Original SİP Declaration Is Not in Turkish

If the original declaration is prepared in Turkish, or in several languages including Turkish, the "Is the original declaration in Turkish?" field is set to "Yes" and no additional upload is requested. If it is not in Turkish, the field is set to "No" and the "Turkish Declaration" field opens.

Under the eighth paragraph of Article 26 of the Guide, the complete Turkish translation of the original SİP Declaration made by a sworn translator is uploaded to this field, and only the Turkish part of the document is placed there. The document must carry the sworn translator's name and surname, stamp, signature or seal, and the translator's statement that the translation was made in accordance with the original.

The Devices in a SİP Must Be "Registered"

The point where SİP registrations most often get stuck is not the pack itself but its contents. ÜTS does not stop at a warning at this point; it blocks the application outright.

  • All devices must be registered: under the eleventh paragraphs of Articles 26 and 28 of the Guide, for a SİP to be registered in ÜTS, all medical devices that make up the SİP must be registered in ÜTS.

  • Additional condition for an MDR SİP: under the eleventh paragraph of Article 28 of the Guide, for a SİP to be registered under MDR, at least one of the devices that make up the SİP must be registered in ÜTS as a medical device under MDR.

  • Automatic block by the system: if devices that are not in "Registered" status are selected on the device-adding screens, ÜTS automatically blocks the SİP application. If there are records in "Rejected" or "Deleted" status in the contents, those devices must first be brought to "Registered" status.

  • Quantity entry: the quantities of the selected devices in the SİP are selected correctly in the list.

If You Are Not the Importer of a Device in the Pack: Dealership Record

The company setting up the pack is often not the importer of every device in it. The Guide attaches a clear condition to this situation: if the company registering the SİP in ÜTS is neither the importer nor the manufacturer of the devices that make up the SİP, an ÜTS dealership record for those devices is mandatory. The rule is the same for SİP registrations under the old directives and under MDR. The dealership is granted by the importer or manufacturer of the device in ÜTS; until the chain is established, the device does not appear on the "Add Device You Are a Dealer Of" screen.

How Is the Class of a SİP Determined?

In an MDR SİP registration, the risk class is not a field the company chooses. Under the sixth paragraph of Article 28 of the Guide, the "Class" field is not filled in by the applicant company and does not appear on the SİP registration screen. The system determines the class: the class of the highest-risk device among the MDR devices that make up the SİP is selected automatically by ÜTS as the class of that SİP. The reason is stated in the same paragraph: a Basic UDI-DI must be assigned to a SİP registered under MDR.

  • Basic UDI-DI requirement: under the eighth paragraph of Article 28 of the Guide, Basic UDI-DI is a mandatory selection field in an MDR SİP registration; only numbers associated with the selected SİP Declaration are listed, and more than one selection cannot be made.

  • Not visible on the label: under the same paragraph, the Basic UDI-DI does not appear on the SİP's market-placement label and/or sales packaging.

  • Issuing entities: the authorized Basic UDI-DI issuing entities and standards for SİPs are GS1, HIBCC, ICCBBA and IFA; the code value consists of at most 25 characters.

Shelf Life and Sterility Information of a SİP

These two fields are the ones most often filled in wrongly in a SİP registration. In both, the error comes from attributing the characteristics of the devices that make up the pack to the pack itself.

  • Shelf-life ceiling: under the seventh paragraphs of Articles 26 and 28 of the Guide, the shelf life of the SİP cannot be longer than that of the device with the shortest shelf life among the devices that make up the SİP and have a shelf life. A SİP's shelf life can be at most that of the shortest-lived device inside it.

  • Sterility: under the same paragraphs, the sterility information requested in this field concerns whether the SİP manufacturer additionally sterilized the pack before placing it on the market. This information is independent of whether the devices inside are supplied sterile on their own.

  • Sterility and document consistency: if the SİP itself is packaged sterile, this information must appear on the SİP label and/or sales packaging, and in the instructions for use if the SİP manufacturer has prepared a separate one, and the relevant field must be set to "Yes".

  • Exception: if the SİP manufacturer has no separate instructions and an instruction document made up of all the devices' instructions has been uploaded, SİP sterility information is not required in these documents.

EU Certificate Required for a Sterile SİP

If an MDR SİP is sterilized, the declaration alone is not enough. Under the seventh and eighth paragraphs of Article 28 of the Guide, if the answer to the sterility question is "Yes", an EU Certificate must be selected on the document selection screen together with the SİP Declaration (MDR). The certificates that can be selected are:

  • EU Quality Management Certificate: the certificate issued after carrying out the conformity assessment procedure "Annex IX (Chapters I & III) Quality Management System" under MDR.

  • EU Production Quality Assurance Certificate: the certificate issued after carrying out the conformity assessment procedure "Annex XI (Part A) Production Quality Assurance" under MDR.

This certificate must first have been registered in ÜTS as a document, and the document application under Article 39 of the Guide must have been completed. Under Articles 14 and 61 of the Guide, the application of these procedures and the involvement of the notified body is limited to securing sterility until the sterile pack is opened or damaged; the notified body does not assess the whole pack from scratch. Article 28 of the Guide also states that Basic UDI-DI information is not requested for these two certificate types.

A SİP Does Not Carry an Additional CE Mark

This is a frequently asked and frequently misanswered question in SİP registration. Under the fifth paragraph of Article 14 of the Guide, a SİP does not bear an additional CE mark. Instead, it bears the name, registered trade name or registered trademark of the person or company, together with an address at which the SİP manufacturer can be reached. The SİP is also accompanied by the information referred to in item 23 of MDR Annex I.

The same provision is repeated in the labelling rules. The tenth paragraph of Article 26 of the Guide lists the information in item 13 of MDD Annex I as the minimum label content for a SİP under the old directives, and the tenth paragraph of Article 28 lists the information in item 23 of MDR Annex I for a SİP under MDR. The MDR list also adds the UDI carrier.

When Is a SİP Considered a Device in Its Own Right?

A SİP does not always remain a SİP. The fourth paragraph of Article 14 of the Guide lists three cases in which the pack is treated as a device in its own right:

  • Containing a device without a CE mark: the pack contains a device that does not bear a CE mark.

  • Incompatible combination: the chosen combination of devices is not compatible in view of their original intended purposes.

  • Sterilization not carried out per instructions: the sterilization is not carried out in accordance with the manufacturer's instructions.

When one of these occurs, the SİP becomes subject to the appropriate one of the conformity assessment procedures referred to in Article 52 of MDR and Article 61 of the Guide (except its ninth paragraph), and the SİP manufacturer assumes the obligations of manufacturers. Article 26 of the Guide carries a similar provision for the old directives. This means the registration route changes: you must now proceed not with a declaration but with a certified conformity assessment.

Non-Medical-Device Products Inside a SİP

An MDR SİP registration has a separate heading that does not exist in registrations under the old directives. Under the fourteenth paragraph of Article 28 of the Guide, the field "Does the SİP contain products outside the scope of medical devices?" under the heading "Non-Medical-Device Products in the Contents" is mandatory.

  • If the field is set to "Yes": the Non-Medical-Device Products List opens, and adding the relevant products to the system becomes mandatory.

  • Product-adding details: if the product you are looking for is not in the list, you enter "Barcode Organization", "GTIN No", "Manufacturer Institution/Company", "Brand Name" and "Product Name", save the product to the list and then select it. If the manufacturer company is not registered, the company is registered under the "Add Manufacturer Institution/EU Authorized Representative" section on the same screen.

  • Quantity and declaration consistency: the quantities of the listed products in the SİP are entered correctly. These products must both appear in the SİP Declaration and be selected in this field, and the information on both sides must be consistent.

Instructions for Use for a SİP

The instructions-for-use rule for a SİP differs from that of a single device and is regulated in the sixth paragraph of Article 17 of the Guide. For a SİP under MDR and MDD the order is as follows:

  • If the SİP has its own instructions: if the SİP manufacturer has prepared separate instructions for use, the SİP's own instructions are uploaded to ÜTS.

  • If it has no instructions of its own: all the instructions for use of the devices in the SİP that are within the scope of the medical device regulations are uploaded.

  • Exemption: if all devices in the SİP are in the risk groups listed in the third paragraph of Article 17 of the Guide that are not required to upload instructions for use, no instructions upload is required for that SİP.

  • Selection screen and quantity: under Articles 26 and 28 of the Guide, at most one instructions for use and at most one catalogue can be selected for a SİP. If at least one of the devices in the SİP is in a risk class that requires uploading instructions, selecting instructions is mandatory.

For the instructions for use to be selected in a SİP registration, they must first have been registered in ÜTS as a document and the document application under Article 39 of the Guide must have concluded.

SİP Application and the 5-Product Limit

After completing the registration notification, an application must be made to the Agency for the device to reach "Registered" status. Among the device application types listed in Article 40 of the Guide, "System Procedure Pack Application" and "Custom-Made Medical Device Application" appear as separate types.

  • Product count limit: under the sixth paragraph of Article 40 of the Guide, custom-made medical device and system procedure pack applications contain at most 5 medical device notifications.

  • Approval authority: only the company's users of the "Medical Device Registration Officer" type can perform the "Approve" action.

  • Status flow: the application is first created in "Awaiting Signature" status; once the authorized signatory signs electronically it moves to "Awaiting Payment", and once the fee is paid it moves to "Under Review". Applications that do not reach "Under Review" cannot be assessed by the review expert.

What Is a Custom-Made Medical Device (ITC)?

The ITC is defined in detail in the definitions article of the Guide. Excluding mass-produced devices that need to be adapted to meet the specific requirements of any professional user, and devices mass-produced by industrial manufacturing processes in accordance with a written prescription, it is a device that is specifically made according to a written prescription stating its specific design characteristics, issued by a person authorized by virtue of their professional qualifications under Law No. 1219, under the responsibility of that person, and intended for the sole use of one particular patient to meet that patient's individual condition and needs.

The Guide also uses two further definitions on the custom-made orthosis and prosthesis side:

  • Orthopedic prostheses and orthoses produced and/or fitted as custom-made: orthopedic prostheses and orthoses that must be produced and/or fitted with measurement and trial for use on a specific patient.

  • Custom-made prosthesis and orthosis center: the workplaces where these products are manufactured, sold, fitted, maintained and repaired.

Under the eighth paragraph of Article 61 of the Guide, ITC manufacturers follow the procedure set out in MDR Annex XIII and prepare the statement specified in that annex before placing these devices on the market. The same paragraph states that ITCs are placed on the market without a CE mark.

Who Can Apply for an ITC?

ÜTS does not open ITC applications to every company. Under the first paragraph of Article 29 of the Guide, the company and institution types that can submit an ITC application under MDR are "Healthcare Service Provider", "Manufacturer/Importer/Dealer/Exporter" and "Custom-Made Orthosis-Prosthesis Application Center". Under the second paragraph of the same article, temporarily, only institutions of the "Healthcare Service Provider" type have the option to submit an ITC application under MDD or AIMDD. The third paragraph divides the authority to apply by product group:

  • Custom-made orthopedic prostheses and orthoses: the application is made only by centers that produce the product concerned and are registered in ÜTS with the field of activity "Custom-Made Orthosis-Prosthesis Application Center", or by healthcare institutions that have an orthopedic prosthesis and orthosis manufacturing service and whose field of activity is "Healthcare Service Provider".

  • Other ITCs: for ITCs other than custom-made orthopedic orthoses-prostheses, the application is made only by companies that manufacture or import the ITC concerned and are registered in ÜTS with the field of activity "Manufacturer/Importer/Dealer/Exporter".

No Dealership for Custom-Made Orthoses and Prostheses

The fourth paragraph of Article 29 of the Guide states that, with the UTS-v7.20.0 release published on 07.02.2019, granting dealerships through ÜTS became possible for ITCs other than custom-made orthopedic orthoses-prostheses. Companies that manufacture or import these ITCs can grant dealerships to other companies whose field of activity is "Manufacturer/Importer/Dealer/Exporter". By contrast, the same paragraph clearly states that custom-made orthosis-prosthesis centers and healthcare institutions with in-house manufacturing services cannot grant dealerships for the custom-made orthopedic prostheses and orthoses they produce. In this product group the distribution chain cannot be set up in ÜTS; the registration is made by the producing center.

Annex XIII ITC Statement and Retention Period

The fifth paragraph of Article 29 of the Guide lists the minimum information the statement prepared pursuant to MDR Annex XIII must contain:

  • Manufacturing information: the name and address of the manufacturer and of all manufacturing sites.

  • Authorized representative: where applicable, the name and address of the EU authorized representative.

  • Device description: information allowing identification of the device: brand name, product name, and, where available, reference/catalogue number and/or primary product number.

  • Single-patient statement: a statement that the device is intended for exclusive use by a certain patient or user, identified by name, an acronym or a numerical code.

  • Prescriber: the name of the physician or dentist, authorized by virtue of their professional qualifications, who issued the prescription, and, where available, the name of the healthcare institution concerned.

  • Specific characteristics: the specific characteristics of the product as indicated in the prescription.

  • General safety and performance: a statement that the device conforms to the general safety and performance requirements in MDR Annex I and, where applicable, an indication, with reasons, of which requirements are not entirely met.

  • Special substance indicator: where applicable, an indication that the device incorporates a medicinal substance, including human blood or plasma derivatives, or tissues or cells of human or animal origin.

Under the sixth paragraph of the same article, this statement is kept for at least 10 years after the device is placed on the market; for implantable devices this period is at least 15 years.

For an ITC Marked "Manufactured", ÜTS Creates the Statement

The seventh and eighth paragraphs of Article 29 of the Guide link how the statement enters ÜTS to the choice in the "Imported/Manufactured Information" field. If the company or institution issuing the ITC Statement is established in Türkiye, this field is set to "Manufactured".

  • If "Imported" is selected: the applicant company uploads the original ITC Statement and, if the original is not in Turkish, its Turkish translation made by a sworn translator to ÜTS.

  • If "Manufactured" is selected: the company does not upload the statement to ÜTS. The statement is generated automatically by ÜTS in light of the information and explanations entered under the headings on the registration screen. At signing, this statement file is signed with an e-signature by the company's or institution's responsible manager, signatory or hospital officer and sent with the application.

On the "Manufactured" route, the accuracy of the statement equals the accuracy of the data you enter on screen; if you fill in the fields wrongly, the statement generated by the system will be wrong too.

Patient and Doctor Information in an ITC Registration: E-Prescription

This is the heading where an ITC registration differs most from a mass-produced device registration. Under the thirteenth paragraph of Article 29 of the Guide, the first field under the heading "Patient and Doctor/Dentist Information" is the question "Is there an e-prescription number?", for which a selection is mandatory.

  • If there is an e-prescription: on the screen opened from the "Fetch E-Prescription" field, the "Patient ID Number" and "E-Prescription Number" are entered, and the patient, doctor/dentist and healthcare facility information on the prescription is brought to the registration screen automatically. The information brought must, where applicable, be consistent with the information in the ITC Statement.

  • If there is no e-prescription (patient): on the "Select Patient" screen, "ID No" is selected and the information is brought automatically, or "Other" is selected and the patient's name, surname and ID number are entered.

  • If there is no e-prescription (doctor and facility): on the "Select Doctor/Dentist" screen, the personnel type is selected and the "ÇKYS Diploma Registration No" is entered. For the healthcare facility, if the question "Will the transaction be made with the SGK Facility Code?" is "Yes", the "SGK Facility Code" field is filled in; if "No", the "Company Identifier Number" field is filled in.

  • Prescription or medical report: the prescription or medical report given to the patient, or its printout from the SGK query screen, is uploaded to ÜTS.

The common rule in all these fields is that the information entered must match entirely what is written in the prescription or medical report.

Barcode and Specific Characteristics of the Product in an ITC

An ITC often has no number allocated by a barcode organization. The Guide resolves this case separately.

  • If none has been allocated: under the tenth paragraph of Article 29 of the Guide, if no primary product number has been allocated by one of the authorized barcode organizations, the "UTS" option is selected. In this case the number is generated automatically by ÜTS with the prefix "UTS-" and cannot be changed by the company.

  • If one has been allocated: the barcode must be entered into the system exactly as it is, without any change. If the barcode in the technical documents is not entered as written in those documents, the application will be concluded negatively.

  • Specific Characteristics of the Product: this is a mandatory field in ÜTS. The specific characteristics of the product, as stated in the e-prescription or in the prescription/medical report, are entered into the field, and this information must match the prescription exactly.

Class III Implantable Custom-Made Devices

In ITCs in general the notified body is not involved; it is involved in Class III implantable custom-made devices. Under the twelfth paragraph of Article 29 of the Guide, the field "Is it a Class III implantable custom-made medical device?" appears under the heading "Declaration Information"; if "Yes" is selected, a selection in the field "Was an application made under Announcement No. 2023/KK-1?" becomes mandatory.

  • Additional conformity assessment: under the eighth paragraph of Article 61 of the Guide, manufacturers of Class III custom-made implantable devices are subject, in addition to the procedure in MDR Annex XIII, to a conformity assessment as set out either in MDR Annex IX (Chapter I) or in MDR Annex XI (Part A).

  • Certificate selection: if no application was made under 2023/KK-1, in the "EU Certificate" field that opens, the "EU Quality Management Certificate" or "EU Production Quality Assurance Certificate" issued as a result of one of these procedures is selected. The certificate must have been registered in ÜTS beforehand and its document application must have concluded positively.

  • 2023/KK-1 route: if an application was made under this announcement, the field "Evidence Documents for MDR Transition Period Extension" opens and the documents mentioned in Article 47 of the Guide are uploaded.

  • Date: under the third paragraph of Article 45 of the Guide, provided the conditions in the relevant Regulation are met, Class III implantable ITCs under MDD can be placed on the market or put into service until 26.05.2026.

Patient Privacy in ITC Images

Applications in these two registration routes tend to get stuck at the same points. The items below are based directly on Guide provisions and can be used as a pre-application checklist.

  • A device in the contents is not "Registered": if the devices that make up the SİP include records in "Rejected" or "Deleted" status, ÜTS automatically blocks the application.

  • No dealership record: if you are not the importer or manufacturer of a device in the contents, an ÜTS dealership record for that device is mandatory.

  • Sterility marked wrongly: the devices in the contents being sterile does not mean the SİP is sterile. The field is filled in according to whether the pack is additionally sterilized before placing on the market; on the MDR side, choosing "Yes" creates an additional EU Certificate requirement.

  • Shelf life exceeded: the shelf life entered for the SİP cannot exceed that of the shortest-lived device in its contents.

  • 5-product limit exceeded: SİP and ITC applications contain at most 5 medical device notifications.

  • Mismatch between declaration and registration: if the identifying information in the SİP Declaration does not match what is entered in ÜTS, the application may be concluded negatively. On the ITC side, the same consistency is required for the prescription and medical report information.

  • Document order skipped: for documents such as instructions for use, catalogue and EU Certificate to be selected in the product registration, they must first be registered as documents and the document application must have concluded.

All of these processes are carried out inside ÜTS with the company's own user authorizations. Companies that do not want to proceed with their own team can open a "Medical Device Registration Officer" authorization to a consultant from the company page in ÜTS, so that document and product records are created with that authorization. In this model the e-signature stays with the company's signatory; the party that signs the application and is responsible for the accuracy of the data is the company.

You can find the place of the same subject in EU legislation in our English article on custom-made devices (MDR Annex XIII).

ÜTS Registration

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara