When your buyer abroad or the health authority of the country you are going to asks you for a "free sale certificate", the Turkish counterpart is the free sale certificate issued by the Agency in Türkiye

Free Sale Certificate (FSC): Who Can Get One and How to Apply Through ÜTS?

When your buyer abroad or the health authority of the country you are going to asks you for a “free sale certificate”, the Turkish counterpart is the free sale certificate issued by the Agency in Türkiye. This document does not replace ÜTS registration; it is built on top of ÜTS registration and is issued only at the request of certain company capacities. Below we explain what the certificate is, who can request it, every step on the application screen, and the conditions that cause the application to be rejected, based on Articles 41 and 42 of the Guide on Registration Processes of Medical Devices to the Product Tracking System.

What Is a Free Sale Certificate?

According to the definitions article of the Guide, the free sale certificate is a certificate issued by the Agency, at the request of a domestic manufacturer or of the EU authorised representative of a foreign manufacturer established in Türkiye, so that devices under the medical device regulations can be exported; it declares that the relevant devices bearing the CE mark can be marketed in Türkiye and in EU member states.

Paragraph 1 of Article 41 of the Guide repeats the same definition functionally: for export purposes and on request, the Agency issues a free sale certificate declaring that the manufacturer or, as applicable, the authorised representative has a registered place of business in Türkiye, and that the device or devices bearing the CE mark under the medical device regulations can be marketed in Türkiye and in EU member states.

  • For which frameworks it is issued: According to paragraph 2 of Article 41 of the Guide, the Agency issues the free sale certificate both for devices under the new medical device regulations (MDR or IVDR) and for devices under the old medical device regulations (MDD, IVDD or AIMDD).

  • What it declares: The document declares that the device is placed on the market by a manufacturer or authorised representative with a registered place of business in Türkiye, and that because it bears the CE mark it can be marketed in Türkiye and in EU member states.

  • Its basis: The same document is also regulated in Article 60 of the Medical Device Regulation; that article states that the certificate will indicate the Basic UDI-DI of the device entered in the UDI database.

Who Can Request One? Can Importers Get One?

This is the point that causes the most confusion in practice. Paragraph 1 of Article 41 of the Guide ties the right to request to two capacities:

  • Domestic manufacturer: The legal manufacturer that produces the device in Türkiye can request a certificate for its own products.

  • EU authorised representative established in Türkiye: The EU authorised representative of a foreign manufacturer who is established in Türkiye can make a request for the devices of the manufacturer it represents. According to the definitions article of the Guide, an EU authorised representative is a person who, where the manufacturer is not established in Türkiye or in EU member states, is appointed in writing by the manufacturer to perform on its behalf certain obligations under the medical device regulations.

Paragraph 1 of Article 42 of the Guide repeats the same distinction: the certificate is issued, at the request of the relevant domestic manufacturer or of the EU authorised representative in question, for devices produced by a domestic manufacturer or devices belonging to manufacturers whose EU authorised representative is established in Türkiye.

The conclusion is clear: the Guide has not defined a route for requesting a certificate solely in the capacity of importer. An importer company that brings a device from abroad and places it on the market in Türkiye can apply for a free sale certificate for that device only if it holds the capacity of the manufacturer’s EU authorised representative established in Türkiye and is registered in ÜTS in that capacity. In practice, the first thing importers planning a re-export should check is their capacities in their own ÜTS records.

What Information Does the Certificate Contain?

The content of the certificate varies according to the regulation the device falls under. According to paragraph 3 of Article 41 of the Guide, a certificate issued for devices under the new medical device regulations contains two additional pieces of information:

  • Basic UDI-DI: The Basic UDI-DI of the relevant device or devices appears on the certificate. Article 60 of the Medical Device Regulation also says that the certificate will indicate the Basic UDI-DI of the device entered in the UDI database. According to the definitions article of the Guide, the Basic UDI-DI is the identifier code assigned at the level of the device’s unit of use, which is the main key in the database and in the relevant documentation for linking devices with the same intended purpose, risk class and essential design and manufacturing characteristics.

  • Unique number of the EU certificate: If a notified body has issued an EU certificate or certificates in relation to the relevant devices under Article 56 of MDR or Article 53 of IVDR, the unique number identifying that EU certificate appears on the certificate.

The Difference for Devices Under the Old Regulations

Paragraph 4 of Article 41 of the Guide is short and clear: a free sale certificate prepared for devices under the old medical device regulations does not contain the information mentioned in paragraph 3. That is, the certificate you receive for your devices under MDD, IVDD or AIMDD does not contain the Basic UDI-DI and the EU certificate number.

This difference matters, because some countries’ authorities may want to see the Basic UDI-DI on the certificate. If your device is still under the old regulation, the absence of this information on the certificate is not a deficiency but the situation the legislation provides for.

The Application Screen and Steps

For domestic manufacturers the application is carried out entirely within ÜTS. Paragraph 2 of Article 42 of the Guide gives the steps in order.

  • Section: The application is made from the “Free Sale Certificate Application Transactions” section under the “Medical Device” tab in ÜTS.

  • Creating the application: In this section the “Add Certificate Application” tab is clicked to enter the “Add Free Sale Certificate Application” screen.

  • Completing the selections: After the required selections are made, the certificate application is created by clicking the “Create” field on the screen.

  • Listing and details: The application that has been created appears in the list on the “List Free Sale Certificate Applications” screen with its application number, in “Awaiting Signature” status. Clicking the application number opens the “Free Sale Certificate Details” screen.

  • Adding products: From the “Add Products from Excel” field on the details screen you move to the “Add Products to Free Sale Certificate Application” screen.

  • Signing: After the Excel file is uploaded, the application is signed with an e-signature and sent to the Agency through ÜTS for evaluation.

At the end of these steps the process runs in this order: creation of the application, adding the product list by Excel, e-signature, Agency evaluation, creation of the accrual if found suitable, payment, and issue and delivery of the certificate. If anything is missing at any link of the chain, the next step does not open; for example, an application that has not been e-signed does not reach the Agency, and no certificate is issued for an application that has not been paid.

The Single-Country Rule

The free sale certificate is a document issued specifically for a country. According to paragraph 2 of Article 42 of the Guide, if the country to which the certificate will be sent is known, the name of the relevant country is selected from the “Country” field on the “Add Free Sale Certificate Application” screen, and only one country can be selected in that field. The certificate is issued specifically for the selected country.

  • More than one country: If certificates specific to more than one country are wanted, a separate free sale certificate application is made for each country concerned.

  • If the country is not known: If the country to which the certificate will be sent is not known, no selection is made in the “Country” field. In this case no particular country name appears on the certificate. The same approach applies to applications made through EBS; if the country is not known, no “Country” name is stated in the application petition.

If you plan to export to three countries, this means three separate applications and a separate fee for each. Clarifying from the start for which countries a document is really needed reduces the number of unnecessary applications. Not selecting a country at all is also an option; if the authority of the buyer country does not require its name to appear on the certificate, a certificate issued without a country can be used in more than one market. Before making this choice you should confirm the document format the buyer requests.

Number of English Copies and the Turkish Certificate Option

The number of certificates is also determined on the application screen and is directly reflected in the fee. Paragraph 2 of Article 42 of the Guide regulates these two fields separately.

  • Number of English copies: The requested number of English copies is entered in the “Number of English Copies” field on the “Add Free Sale Certificate Application” screen. A separate application fee is charged for each copy.

  • Turkish certificate: If a Turkish copy of the certificate is also wanted, the checkbox in the “Do you also want a Turkish certificate?” field must be ticked. If the application is suitable, a Turkish free sale certificate is also issued and a fee is also charged for the Turkish certificate.

The number of copies should be determined according to the real need. Asking in advance how many original copies the import file in the buyer country will require is more practical than opening a second application after the certificate has been issued.

Excel Product List: Barcode and English Product Name

The devices subject to the application are not entered one by one on the screen; they are added in bulk with an Excel file. According to paragraph 2 of Article 42 of the Guide, using the template downloaded from the “Download Sample Excel File” field on the “Add Products to Free Sale Certificate Application” screen, an Excel file is created containing the barcode numbers and English product names of the devices subject to the application.

  • English language: The product names for the device barcodes in the file must be in English.

  • Exact match: These English product names must match the Turkish product names in ÜTS exactly. This means the equivalent of the name registered in ÜTS, not a free translation.

  • Upload: The prepared file is uploaded to the “Upload the Excel file” field on the same screen.

In practice the error in this field arises from confusing the Turkish product name in ÜTS with the marketing name. The English name you write into the Excel file must be the equivalent of the product name in the ÜTS record, not the trade name in the product catalogue. The same care is needed on the barcode side: the barcode entered in the file is the device’s primary product number in ÜTS. Entering a barcode with a single wrong digit means that line finds no match in ÜTS and that the application is evaluated incompletely.

The Requirement That Products Be in “Registered” Status

The free sale certificate is issued for devices whose ÜTS registration is complete. According to paragraph 2 of Article 42 of the Guide, all devices whose barcode numbers appear in the Excel file must already be in “Registered” status in ÜTS.

  • Rejected records: If the application contains devices in “Rejected” status, the application is rejected by the Agency.

  • Unique Product Exists in the System: Likewise, if it contains devices in “Unique Product Exists in the System” status, the application is rejected.

  • Status at the time of issue: According to the same paragraph, after the application fee is paid, the certificate is issued only for devices that appear in the application and are in “Registered” status in ÜTS on the date the certificate is to be issued.

The last item is overlooked by most companies. If the status of a device that was “Registered” at the time of application changes before the certificate is issued, that device does not enter the certificate. If you are updating your product records during the application process, you need to take into account the effect of this on the certificate.

At Most 1000 Devices

Paragraph 2 of Article 42 of the Guide states that at most 1000 devices can be added to one free sale certificate application. This limit directly determines the number of rows in the Excel file.

For companies working with a broad catalogue this means splitting the list. When splitting, the two rules under the following headings must also be taken into account: devices under different regulations cannot appear in the same application, and in MDR/IVDR applications the Basic UDI-DI information of the devices must be the same. These three rules together largely determine how you split the list. The same limit also applies to the PDF product list uploaded in applications made through EBS by an EU authorised representative.

MDR, IVDR and Old-Regulation Products Cannot Be Together

According to paragraph 2 of Article 42 of the Guide, devices under MDR, IVDR and MDD/AIMDD/IVDD cannot appear together in an application. This means three separate groupings: devices under MDR, devices under IVDR and devices under the old regulations.

A company with a mixed portfolio may have to open more than one application even for a document going to the same country. Before separating the applications, you need to be sure that the regulation information of each device in ÜTS is correct; a device that has completed its MDR transition but whose ÜTS record still appears under the old framework would lead you to put the list on the wrong side. If part of the same product family has moved to MDR and part remains under MDD by using the transition provisions, these two groups cannot be combined in a single certificate.

The Same Basic UDI-DI Requirement

For applications under the new regulations an additional grouping rule applies. According to paragraph 2 of Article 42 of the Guide, when a free sale certificate application is made for devices under MDR or IVDR, the Basic UDI-DI information of the devices in the application must be the same. A different free sale certificate application is made for devices with different Basic UDI-DIs.

This rule is consistent with the content of the certificate: because the Basic UDI-DI of the devices appears on the certificate, a single certificate contains a single Basic UDI-DI group. In practice, for a broad product family under MDR you may need to open as many applications as there are Basic UDI-DIs. No such requirement is sought in applications under the old regulations.

Payment: Halkbank Corporate Collection

The free sale certificate is subject to a fee, and the payment method is specifically set out in the Guide. According to paragraph 2 of Article 42 of the Guide, the application is evaluated by the Agency; if it is suitable, an accrual is created and a payment reference number is sent to the e-mail address registered in ÜTS of the applicant company.

  • Corporate Collection Application: A Corporate Collection Application is used for the fees to be paid for the issue of the certificate.

  • Payment channel: The corporate payment must be made at a Halkbank branch or through Halkbank internet banking, using the reference number sent to the e-mail address registered in ÜTS.

  • Bank transfer and EFT: A bank transfer or EFT must absolutely not be made. If one is made, the payment information cannot be matched.

Keeping the e-mail address registered in ÜTS up to date becomes decisive at this stage. The reference number goes there; if it does not arrive, payment cannot be made, and the period in the next heading continues to run. Corporate collection is the method that makes the payment match the application automatically; the reference number must therefore be used in full at the time of payment.

If a Bank Transfer/EFT Was Made by Mistake

The way to resolve an erroneous payment is defined in the Guide. If an erroneous bank transfer or EFT has been made to the Agency, and an application is made to the Agency with a petition containing the application number, the receipt and the account/IBAN number, the erroneous payment is refunded to the person concerned.

This refund does not mean the application has been paid. Payment must still be made using the reference number and through the corporate collection channel. The refund process and the payment period of the application are independent of each other.

10-Day Period for Payment

According to paragraph 2 of Article 42 of the Guide, payment for the free sale certificate must be made within at most 10 days. This is a period the legislation grants to the applicant; it is not the time in which the work will be completed.

The same paragraph also writes the consequence: applications for which payment is not made within 10 days are automatically cancelled by ÜTS. For a cancelled application you must create the application from scratch, re-upload the Excel file and e-sign again. It is therefore sensible to assign payment follow-up to one person once the accrual notification arrives.

Issue of the Certificate and Delivery by Post

After payment is completed, the certificate is issued. According to paragraph 2 of Article 42 of the Guide, after the application fee is paid, the certificate is issued only for devices that appear in the relevant application and are in “Registered” status in ÜTS on the date the certificate is to be issued, and is sent by post to the company’s address registered in ÜTS.

This provision turns the currency of the company address in ÜTS directly into a delivery matter. A company that has changed its address but has not updated its ÜTS record will find that the certificate has been sent to its old address.

Report Request for Hand Delivery and 3 Business Days

Post is not the only option. According to paragraph 2 of Article 42 of the Guide, if, after the application fee is paid, a request is submitted on the “Report Request” screen in ÜTS by selecting the “Medical Device Document Return Transactions” request area, stating that the certificate will be collected by hand, the certificate is held at the Agency for 3 business days to be delivered by hand to the company’s representative.

The same provision also says what happens when the period ends: if the company’s representative does not collect it by hand within 3 business days, the certificate is sent by post to the company’s address registered in ÜTS. If hand delivery is to be preferred, the request must be submitted after payment and in time.

The EBS Route for an EU Authorised Representative Established in Türkiye

Applications made in the capacity of EU authorised representative run not through the ÜTS screen but through a different channel. According to paragraph 3 of Article 42 of the Guide, for an application to be made for devices of manufacturers whose EU authorised representative is established in Türkiye, the relevant EU authorised representative, who is registered in ÜTS under the “manufacturer/importer/dealer/exporter” activity area type, must apply to the Agency by petition through EBS.

  • Petition content: The petition and/or its annexes state the number of English copies, the name of the country to which the certificate will be sent, and whether a Turkish copy is requested.

  • Product list: The file containing the barcode numbers and English product names of the devices subject to the application is uploaded to EBS in PDF format as an attachment to the application file. The English product names must match the Turkish product names in ÜTS exactly.

  • The same limits: At most 1000 devices can be added to the file; devices under MDR, IVDR and MDD/AIMDD/IVDD cannot appear together; and in MDR or IVDR applications the Basic UDI-DI information of the devices must be the same.

  • Product status: The device barcodes in the file must be in “Registered” status in ÜTS; if there are devices in “Rejected” status in the application, the application is rejected.

  • Payment and delivery: If the application is found suitable, the fee is paid in line with the instructions the Agency sends; the certificate is issued for devices in “Registered” status on the date of issue and is sent by post to the company’s address registered in ÜTS.

On this route it is a precondition that the company record in ÜTS is in the right activity area. An EU authorised representative that has no record of the “Manufacturer/Importer/Dealer/Exporter” type cannot apply through EBS.

A Free Sale Certificate and ÜTS Registration Are Not the Same Thing

The two transactions are often confused, although their purposes are opposite. As stated in Article 1 of the Guide, ÜTS registration relates to carrying out registration processes in line with the medical device regulations and concerns the device being placed on the market in Türkiye. The free sale certificate, as Article 41 of the Guide states explicitly, is issued for export purposes.

  • Direction: ÜTS registration looks inward, the free sale certificate looks outward.

  • Order: The precondition of the certificate is that the device is “Registered” in ÜTS. A certificate cannot be obtained without registration; obtaining a certificate also does not replace registration.

  • Requester: ÜTS registration is made by the company placing the device on the market; the certificate is requested by the domestic manufacturer or the EU authorised representative established in Türkiye.

  • Channel: ÜTS registration and the device application run through ÜTS. The free sale certificate also runs through ÜTS for domestic manufacturers, whereas applications made in the capacity of EU authorised representative are made by petition through EBS.

The certificate is also not a document that re-evaluates the conformity of the device. The Agency does not redo the conformity assessment; for devices in “Registered” status in ÜTS, it issues a declaration for export based on the information in the record. The content of the certificate is therefore only as accurate as the content of your ÜTS record.

Pre-Application Checklist

When the conditions in Article 42 of the Guide are turned into a checklist, most of the reasons for rejection of the application are eliminated in advance.

  • Product statuses: Are all devices in the list “Registered” in ÜTS? Is any device left in “Rejected” or “Unique Product Exists in the System” status?

  • Regulation separation: Are devices under MDR, IVDR and the old regulations mixed in the list?

  • Basic UDI-DI equality: If it is an MDR or IVDR application, is the Basic UDI-DI information the same for all devices in the list?

  • 1000-device limit: Does the number of devices in the Excel file exceed 1000?

  • English name match: Do the English product names in the Excel file match the Turkish product names in ÜTS exactly?

  • Country selection: Was a single country selected? If more than one country is needed, was a separate application opened for each?

  • Number of copies: Was the number of English copies entered correctly? Was the Turkish certificate checkbox ticked according to need?

  • Contact and address currency: Is the e-mail address registered in ÜTS able to receive the payment reference number? Is the registered company address the current address to which the certificate will be posted?

  • Payment channel: Is payment planned as a corporate payment with the reference number at a Halkbank branch or through internet banking?

All of these checks are made through your company record in ÜTS. Companies that do not want to carry out the process with their own team can open the “Medical Device Registration Officer” authority for a consultant from the company page in ÜTS and have the product and document records prepared with this authority. The e-signature remains with the company’s signatory, and the party who signs the application and is responsible for the accuracy of the data is the company.

If you also offer your device to the EU market, you can look at our English guide on the EU free sale certificate.

ÜTS Registration

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara