MDR and IVDR Declaration of Conformity Preparation and UTS Document Registration
The EU declaration of conformity from your overseas manufacturer was rejected in ÜTS, or they never send the declaration at all

Who Issues the Declaration, and Who Cannot
The EU declaration of conformity is issued by the legal manufacturer. The manufacturer prepares it, signs it and takes full responsibility for it. An importer cannot write its own letterhead declaration and upload it to UTS. We can prepare the draft, but the manufacturer is the party that issues and signs the declaration.
For a domestic manufacturer, the rule in Article 13(1)(ç) of the UTS Guide applies. When the "Is the original document in Turkish?" option is set to "Yes", the original of the document signed electronically by the person registered as "Mesul Müdür" or "İmza Yetkilisi" of the domestic manufacturer must be uploaded in full under the "Original Document" field.
Two Routes: Self-Declaration and Certified Declaration
For a Class I device, the manufacturer declares conformity itself. Under Article 52(6) of the Medical Devices Regulation, Class I manufacturers draw up the EU declaration of conformity referred to in Article 19 after preparing the technical documentation set out in Annexes II and III. The same route applies to Class A in vitro diagnostic devices under Article 48(10) of the IVDR.
Class I sterile devices, Class I devices with a measuring function, Class I reusable surgical instruments, and all Class IIa, IIb and III devices, as well as Class B, C and D in vitro diagnostic devices, rely on a notified body certificate. The certificate number, the name of the notified body and its identification number are entered in field 8 of the declaration.
In both routes we prepare the declaration. In the self-declaration route we review the technical documentation against the headings of Annexes II and III. In the certified route we write the declaration according to the scope of the certificate. The conformity assessment and the notified body process take place between the manufacturer and the notified body, and we do not take part in them.
Fields Required in an EU Declaration of Conformity
Article 11(2) of the Guide lists the fields that must appear in the declaration and that are checked during document review in UTS. The sentence at the end of this paragraph is critical: the information entered into the system must match the information in the document.
- Manufacturer information (a): the name or trade name and the address of the manufacturer.
- Authorised representative (b): the name or trade name and the address of the EU Authorised Representative, where applicable.
- SRN (c): the Single Registration Number of the manufacturer, where already issued. In UTS this information is entered in the "Unique ID No" field.
- Responsibility statement (ç): a statement that the manufacturer has drawn up the EU declaration of conformity entirely under its own responsibility.
- Basic UDI-DI (d): the Basic UDI-DI.
- Identification and traceability (e): the product name and trade name, product code, catalogue number or other precise references such as photographs where applicable, together with the intended purpose, in a way that allows the device to be identified and traced.
- Risk class (g): the risk class of the device.
- Conformity statement (ğ): a statement that the device complies with the Medical Devices Regulation under which it is covered and, where applicable, with other relevant EU legislation.
- Common specifications (h): references to the common specifications used and against which conformity is declared.
- Notified body and certificate (ı): the name and identification number of the notified body, a description of the conformity assessment procedure carried out, and identification of the EU certificate or certificates issued. This item does not apply to declarations covering only Class I other or Class A devices.
- Dates (j, k, l): the start (issue) date of the document in (j), the end (validity) date in (k), and the place and date of issue of the declaration in (l).
- Signature (m): the name and surname of the authorised person, the position, and the signature or stamp. An electronic signature is also valid.
In declarations under the old regulations the list is shorter. Article 10(2) of the Guide lists the manufacturer's name and address, the EU Authorised Representative where applicable, device details (model, reference number, product name and, where relevant, a detailed description), the regulation the device falls under, the device class and/or any additional information, the start and end dates of the document, and the details of the authorised person.
One Declaration for Several Legislations
When a device is also subject to another piece of legislation requiring an EU declaration of conformity, two separate declarations are not written. Under Article 19(2) of the Medical Devices Regulation, a single EU declaration of conformity is prepared covering compliance with all the legislation applicable to the device. Electrical or radio-equipped devices are examples. You tell us which legislation applies, and we write all of it into the same declaration.
Can the Basic UDI-DI Be Used Instead of the Product Name?
Under Article 11(2)(e), first sub-item, the identification and traceability information can be provided through the Basic UDI-DI, except for the product name or trade name. The product name or trade name always appears in the declaration. Other precise references such as product code, catalogue number and photographs may be replaced by the Basic UDI-DI. This is the route we recommend to manufacturers who want to simplify declarations with long product lists.
Checks Before the Declaration Is Finalised
The final version of the declaration comes from the manufacturer, but four checks made beforehand shorten the process considerably. These are checks that your own company can carry out with its own records.
- Final form of the document: Article 10(2)(g) and Article 11(2)(m) of the Guide require the name, surname, position and signature or stamp of the authorised person. Article 11(2)(l) requires the place and date of issue. If the draft is sent unsigned or undated, it does not proceed to the UTS stage.
- Certificate number match: Article 11(2)(ı) requires the EU certificate to be identified in the declaration. The numbers stated in the declaration must match the certificates you hold and the certificates registered in UTS.
- Barcode and scope overlap: the identification and traceability information in Article 11(2)(e) must match the barcode list you intend to import. A product that is not in the declaration cannot be linked to a device registration.
- Duration: under Article 11(2)(k), first sub-item, if the document has no expiry date, the option "Unlimited Duration" is selected on the UTS screens. For a declaration to be unlimited in duration, this must be clear from the text of the declaration.
Our part, after these four checks, is to go through the declaration field by field in the order of the items in Article 11(2), list the missing items in an English correction list, and send it to the manufacturer. The manufacturer revises the declaration. We state which item is missing and why, citing the article number.
How an Unlimited-Duration Declaration Is Marked in UTS
Article 10(2)(f), first sub-item, and Article 11(2)(k), first sub-item, of the Guide set the same rule: if the document is of unlimited duration, "Unlimited Duration" is selected on the UTS screens. The second sub-items say that where a document end date exists, it is entered in the relevant field.
This distinction creates two different flows in an MDR transition extension. Under Article 47(3)(h) of the Guide, where the declaration linked to a device covered by an EC certificate with a transition extension is "Limited Duration" and the declaration is registered in UTS, the company creates a request without opening a document registration or update application, using the "Report Request" screen and selecting "MDR/IVDR Transition Extension" as the request field. The second sub-item of the same paragraph requires the UTS Document Number of the declaration to be entered in this request, and states that the applicant company is responsible for any incorrect processing or delay resulting from failing to enter it.
The fourth sub-item sets out the opposite case: "Unlimited Duration" declarations, meaning declarations with no end date, do not go through these steps. If such a declaration is not in "Registered" status in UTS, the normal document registration or update notification and application is carried out under the provisions of Articles 12 or 13 and Article 39 of the Guide. For IVDR, Article 50(2)(h) of the Guide sets out the same distinction.
Routes of Acceptance Are Narrower Than Certificates
This is where importers most often get stuck. Article 12(1) of the Guide lists four items of documents that may be uploaded under "Original Document" for the first registration or update notification of a foreign manufacturer's "Declaration of Conformity" or "EU Declaration of Conformity (MDR/IVDR)".
- Apostille (a): under the first sub-item, if the manufacturer is in an EU member state, an apostille issued by the competent authority of that country is accepted. If the manufacturer is established outside the EU, the apostille of the competent authority of the country where the manufacturer or its EU Authorised Representative is established is accepted. Under the second sub-item, the original apostilled document must be uploaded. A copy issued by a Turkish notary or any other country's notary does not need to be uploaded.
- Consulate or embassy approval (b): the document is accepted with the approval of the Turkish embassy or consulate in the manufacturer's country. If the manufacturer is established outside the EU, approval from the representation in the country of the EU Authorised Representative is also accepted.
- Original of an e-Apostille document (c): the documents must be verifiable through the query tools stated on the e-Apostille. Under the first sub-item the original, either digitally issued by the authority or downloaded from the authority's website, must be uploaded. Under the third sub-item, documents whose validity cannot be verified because the original was not uploaded, or invalid e-signed documents, are not accepted.
- Original of an e-signed document (ç): for documents electronically signed by the manufacturer, the original of the e-signed document is uploaded, provided that the e-signature verification tools confirm that the document was signed by its manufacturer, that the signature is valid and that the document has not been altered afterwards.
2026/KKB-2 update: under the TİTCK Announcement No. 2026/KKB-2 dated 02.11.2026, a new route was introduced in place of the option for the manufacturer to upload an e-signed original for the foreign manufacturer's declaration of conformity and SIP declaration. The announcement states that the document may be uploaded to the relevant field in UTS together with an undertaking signed electronically, using a qualified electronic signature based on a qualified electronic certificate issued by a provider recognised in Türkiye, by the authorised person of the importer company authorised by the foreign manufacturer.
Certificates have two acceptance routes that are not part of the declaration. These are the publication of the notified body's certificate on its official query site under Article 8(1)(d) of the Guide, and its upload to EUDAMED under Article 8(1)(e). These routes are specific to certificates and are not in the four-item list in Article 12(1).
For translation, Article 12(2) of the Guide applies. A complete sworn Turkish translation of the original document made by a sworn translator must be uploaded to the "Turkish Document" field. Under Article 12(2)(a), only the Turkish part of the sworn translation approved by the sworn translator is uploaded to this field. Under Article 12(2)(b), the document must contain the translator's name and surname, stamp, signature or seal, and the translator's declaration that the translation was made in accordance with the original.
Date Errors in Old-Regulation Declarations
For devices under the MDD, AIMDD and IVDD, the date on which the declaration was issued determines its fate.
Article 10(2)(8) of the Guide, together with Article 110 of the MDR and Article 111(1)(d), states that the MDD and AIMDD were repealed from 26/5/2021. Under the same paragraph, devices lawfully placed on the market before that date may remain on the market only with a declaration of conformity issued before 26/5/2021. In other words, the first issue date of the declaration of conformity for MDD or AIMDD devices must be before 26/05/2021.
Article 10(2)(9) sets the same structure for the IVDD: under Article 107 of the IVDR and Article 108(1)(d), the IVDD was repealed from 26/5/2022, and the first issue date of the declaration of conformity for IVDD devices must be before 26/05/2022.
The linking rule is in the same paragraphs. An updated declaration is expected to be linked to the version or versions issued before 26/5/2021 (26/5/2022 for IVDD). If the content of the declaration cannot be linked to older versions, the version or versions of the declaration prepared before that date must be uploaded to the "Technical Documentation of the Product(s) in the Document" field.
Permitted Changes to Old-Regulation Declarations
An old-regulation declaration is not frozen, but the room for updating it is narrow. Article 10(2)(8) and (9) of the Guide draw the same frame: for devices that benefit from the transitional provisions of the MDR or IVDR, the declaration may be updated for administrative changes, such as a change of the manufacturer's address, a change of production site, a change of EU Authorised Representative, or a change of GMDN code, but not for significant changes to the product scope or to the design and intended purpose of the devices.
In practice, administrative changes may be updated, while adding new products to the scope or changing the intended purpose cannot be done this way. This is the first point we discuss with companies that want to add new products.
The Declaration Must Be Kept Up to Date
A declaration is not a document written once and left aside. The first paragraph of Article 19 is clear: the manufacturer keeps the EU declaration of conformity up to date. When the name, variants, class, applied standard or certificate of the device changes, the declaration also changes. We keep the revision history of the declaration, and the UTS document registration is updated to the new revision. We also update declarations from the old directive period. Our guide on the manufacturer's declaration of conformity on the UTS side is available on our site.
How the Product Scope Is Written Affects Acceptance
Article 10(2)(5) of the Guide requires the scope of medical devices in the declaration of conformity to be descriptive enough to identify the product, and states that general wording is not sufficient to identify a product. The example given in the Guide is clear: "surgical hand instruments" must be detailed as forceps, retractors, dilators, and so on.
The same paragraph provides a way out. For products imported, where the document does not meet this requirement, a document listing the branch types or product descriptions in which products defined by general wording are detailed is uploaded separately to the "Technical Documentation of the Product(s) in the Document" field. This route is used when the manufacturer does not re-issue the declaration.
The Link Between the Declaration and the Device Application
Even when the declaration is correct, a device application does not progress if its structure is wrong. Article 40(5)(a) of the Guide states that the declaration linked to the relevant device or devices is selected from the "Declaration" field on the "Medical Device Application Summary" screen.
One declaration rule: under (c), medical device applications can only be created from devices linked to a single declaration, so only one declaration can be selected from the list. Products linked to two different declarations cannot be combined in a single application.
Status filter: under (b), only declarations linked to devices in "Ready for Application" status are listed on the "Select EU Declaration of Conformity / EU Declaration of Conformity" screen. If your declaration does not appear in the list, the problem lies in the status of the devices, not the declaration.
Wrong selection: under (ç), an incorrectly selected declaration can be removed from the "Selected Declaration" field using the "Remove" button.
Is There a Declaration Template?
There is, but for a limited purpose. Article 11(7) of the Guide states that templates of the EU declaration of conformity created to serve as examples, especially for domestic manufacturers, are included in Annex 7 and Annex 8 of the Guide.
These templates are not a form to impose on a foreign manufacturer; the manufacturer uses its own format. The function of a template is to show how the fields in Article 11(2) appear in a document. When working with a foreign manufacturer, we use these annexes as a reference to explain the missing fields.
The SIP Declaration Is a Separate Document
For companies that bring together systems or procedure packs, UTS has a separate document type. Article 14(1) of the Guide states that the "SIP Declaration (MDR) (Article 22)" document is the declaration prepared by the natural or legal persons defined as SIP manufacturers when they bring together devices under the MDR bearing the CE mark with other CE-marked devices, in vitro diagnostic devices bearing the CE mark under the IVDR, and, where justified, other products compliant with the legislation applicable to them, in order to place them on the market as a SIP.
Article 14(2) lists three undertakings the declaration must contain:
- Mutual compatibility (a): the SIP manufacturer confirms the mutual compatibility of the devices and, where applicable, other products in accordance with the instructions of the manufacturers, and carries out its activity in accordance with those instructions.
- Packaging and information (b): the SIP manufacturer packages the system or procedure pack and provides users with the relevant information, including information supplied by the manufacturers of the devices or other products brought together.
- Internal monitoring, verification and validation (c): the bringing-together activity is subject to appropriate internal monitoring, verification and validation methods.
Article 14(5) addresses a frequently asked question: a SIP does not carry an additional CE mark. However, it does carry the name of the person or company, the registered trade name or the registered trademark, together with an address at which the SIP manufacturer can be reached. Article 14(3) states that where the SIP manufacturer sterilises the pack before placing it on the market, one of the conformity assessment procedures under Article 61(9) of the Guide applies, and the involvement of the notified body is limited to matters relating to the assurance of sterility.
Translation of the Declaration of Conformity
When the manufacturer's declaration of conformity is uploaded to UTS, a Turkish translation is also required alongside the original document. Article 12(2) of the Guide states that the complete Turkish translation of the original document made by a sworn translator is uploaded to the "Turkish Document" field.
What a Sworn Translation Must Contain
Under Article 12(2)(a) and (b):
- Only the Turkish section of the sworn-translator-approved document is uploaded to the "Turkish Document" field.
- The document contains the translator's name and surname, stamp, signature or seal.
- The document contains the translator's declaration that the translation was made in accordance with the original.
Common Errors in Declaration Translations
The Guide requires the translation to be "complete". The annexes and the product list of the declaration are therefore translated as well, and model and reference numbers are carried over exactly as in the original. The original-language text is not added to the Turkish Document field; only the Turkish section is uploaded.
EU Authorised Representative and Power of Attorney
When an EU declaration of conformity for a manufacturer not established in an EU member state or in Türkiye is registered, the EU Authorised Representative information is also declared. Article 12(7)(a) of the Guide states:
For EU declaration of conformity (MDR/IVDR) registration notifications of manufacturers not established in an EU member state or in Türkiye, entry in the "EU Authorised Representative" field on the UTS registration screen is mandatory.
Under Article 12(7)(c), the "Power of Attorney" field is uploaded with the original of the document and its Turkish translation by a sworn translator together; the original of the power of attorney does not require an apostille or consulate approval. Under Article 12(7)(d), manufacturers established in an EU member state or in Türkiye do not appoint an EU authorised representative.
Who This Service Is For
Domestic Manufacturers
A domestic manufacturer placing its own product on the market issues the EU declaration of conformity under its own responsibility. We prepare the declaration according to the device class, the conformity assessment route and the notified body certificate, if any, submit it to the manufacturer for signature, and complete the UTS document registration.
Importers Bringing in a Foreign Manufacturer's Product
For importers, the declaration belongs to the foreign manufacturer. In this case we check the declaration's suitability for UTS, determine the approval route (apostille, consulate, e-Apostille or e-signature), prepare the sworn Turkish translation, and complete the document registration.
For the document types and registration rules for the UTS declaration of conformity, see our EU declaration of conformity and CE mark guide.
What We Deliver
- A draft declaration prepared according to Annex IV, with ten sections.
- A Turkish translation, if the declaration is in another language (Art. 19(1)).
- A revision history.
- A comparison of the product name, brand name, Basic UDI-DI and class in the declaration with the label, instructions for use and UTS record.
- The UTS document registration.
For your UTS product registrations, see our UTS product registration consultancy page.
What We Need From You
- Device list: product name, brand name, model or catalogue number, risk class, Basic UDI-DI.
- Technical documentation or table of contents.
- Notified body certificate, if any.
- Label and instructions for use.
- Manufacturer company details and the name and position of the person who will sign the declaration.
Scope and Limits of the Service
The output we provide is concrete: preparation of a draft declaration according to Annex IV or a field-by-field review of an existing declaration against the items in Article 11(2), listing the missing items in a written correction list to be sent to the manufacturer, comparison of certificate numbers and barcode scope with the declaration, checking the date of issue and link to earlier versions for old-regulation declarations, selecting the acceptance route (apostille, consulate, e-Apostille, e-signature) according to the document's nature, preparing the sworn translation file in accordance with Article 12(2), and creating the document registration in UTS.
What we cannot do is also clear. Under Article 11(3) and Article 10(3) of the Guide, the authority to issue the declaration rests with the legal manufacturer. A consultant cannot issue the declaration on behalf of the manufacturer, cannot change the text of the declaration under its own signature, and cannot commit to the outcome of the authority's document review. We prepare the draft; the signature and responsibility belong to the manufacturer. If the manufacturer does not respond, what we can do is document which item is missing and repeat the request.
Out of Scope
- Notified body conformity assessment and CE certification.
- Testing and laboratory processes, including laboratory selection.
- Interpretation or modification of your technical specifications.
- Finding importers or distributors.
Frequently Asked Questions
Does Medex issue the declaration of conformity?
No. The manufacturer issues and signs the declaration. We prepare the draft according to Annex IV, submit it to the manufacturer for signature, and complete the UTS document registration.
Do you prepare a declaration for a Class IIb device?
Yes, if you hold a notified body certificate. We write the declaration according to the scope of the certificate; we do not take part in the conformity assessment.
Does a Class I device need a notified body?
As a general rule, no: the manufacturer declares conformity itself under Article 52(6). If the device is supplied sterile, has a measuring function, or is a reusable surgical instrument, the notified body is involved only with respect to those aspects.
Does the declaration have to be in Turkish?
For devices placed on the market in Türkiye, if the declaration is issued in another language, a Turkish translation is added (Art. 19(1)).
If a device is subject to both the MDR and another EU legislation, how many declarations are needed?
One declaration. Article 19(2) requires a single EU declaration of conformity covering all legislation applicable to the device.
When does the declaration need to be updated?
When any of the information in the declaration changes. Under Article 19(1), the manufacturer keeps the declaration up to date.