Declaration of Conformity (DoC) and the ÜTS Document Registration Rules

There are two types of declaration of conformity in ÜTS. The manufacturer draws up the declaration and the importer registers it; the fields that are checked, the scope statement, the foreign manufacturer's declaration, translation.


Your manufacturer has sent you a PDF; it says "Declaration of Conformity" on it, and there is an image of a signature beneath. You upload it to ÜTS and your application comes back on the ground of "original document". The problem is not in the content of the declaration; most often it is in the way the declaration reached you: ÜTS accepts the declaration of conformity of a foreign manufacturer only through one of four defined routes, and none of these routes is "a scanned PDF the manufacturer sent by e-mail".

This page looks at the declaration of conformity from two sides. The first side is the document itself: there are two separate document types in ÜTS, which one is chosen depends on the regulation under which the device falls, and for both types there is a list of fields that the system checks one by one during review. The second side is the admissibility of the document: apostille, Turkish Republic consulate certification, e-Apostille and the manufacturer's own electronic signature. If you are a domestic manufacturer, you use none of these four routes; you have an entirely separate application type.

The sources are the Manual on Registration Procedures of Medical Devices in the Product Tracking System (Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz) and the Medical Device Regulation (Tıbbi Cihaz Yönetmeliği) published in the Official Gazette No. 31499 (bis) of 2/6/2021. We do not go into the certificate side here, that is, the documents issued by the notified body; this page deals only with the manufacturer's own declaration.

There are two separate types of declaration of conformity in ÜTS

Article 10(1) of the Manual says that the document of the "Declaration of Conformity" ("Uygunluk Beyanı") type is used only in notifications and applications concerning declarations of conformity prepared under the Old Medical Device Regulations, and that for declarations of conformity issued under the New Medical Device Regulations, a registration or update notification and application of the "EU Declaration of Conformity (MDR/IVDR)" ("AB Uygunluk Beyanı (MDR/IVDR)") type is made. Article 11(1) of the Manual repeats the same distinction from the opposite direction and states that for declarations under the Old Medical Device Regulations an application of the "Declaration of Conformity" type is made.

Which regulations these two definitions correspond to is written in Article 4 of the Manual. Point (m) of paragraph 1 of the same article defines the Old Medical Device Regulations as the repealed Medical Device Regulation (MDD), the repealed Active Implantable Medical Devices Regulation (AIMDD) and the repealed Regulation on In Vitro Medical Diagnostic Devices (IVDD). Point (şşş) of the same paragraph defines the New Medical Device Regulations as the Medical Device Regulation (MDR) published in the Official Gazette No. 31499 (bis) of 2/6/2021 and the In Vitro Diagnostic Medical Device Regulation (IVDR).

The practical result is this: you choose the document type not by looking at the class of the device or at the certificate you hold, but by looking at which regulation the declaration refers to. If "93/42/EEC" or "Directive 98/79/EC" appears in the text of the declaration, the type is "Declaration of Conformity". If "Regulation (EU) 2017/745" or "2017/746" appears, the type is "EU Declaration of Conformity (MDR/IVDR)". In a notification where the wrong type is selected, the field list also works wrongly, because the mandatory fields of the two types are not the same.

The declaration is drawn up by the manufacturer; the importer only registers it

The declaration of conformity is not a document an importer can write. Article 10(3) of the Manual says that the declaration of conformity is drawn up by the legal manufacturer firm; Article 11(3) of the Manual says that the EU declaration of conformity is drawn up by the legal manufacturer firm. These two sentences close the way for a firm registering in ÜTS as an importer to prepare a declaration on its own letterhead.

The source of the obligation is in the Regulation. Article 10(6) of the Medical Device Regulation says that manufacturers of devices other than custom-made or investigational devices, where they have demonstrated the conformity of their devices as a result of the applicable conformity assessment procedures, draw up an EU declaration of conformity under Article 19 and affix the CE marking under Article 20. Article 19(3) of the same Regulation says that, by drawing up the EU declaration of conformity, the manufacturer assumes responsibility for compliance with the requirements of the Regulation and with all other legislation applicable to the device.

The importer's role is verification. Article 13(2)(a)(1) of the Medical Device Regulation says that, in order to place a device on the market, the importer verifies that the device is CE marked and that the EU declaration of conformity has been drawn up. Paragraph 9 of the same article says that importers keep a copy of the EU declaration of conformity and, where applicable, a copy of every certificate issued under Article 56, for the period referred to in Article 10(8). Article 19(1) of the same Regulation places the keeping of the declaration up to date on the manufacturer.

When these three provisions are put together, the division of roles in ÜTS becomes clear: the manufacturer is responsible for the content of the declaration, and the firm making the record is responsible for entering the declaration into the system correctly and completely.

The fields ÜTS checks in a declaration under the old regulations

Article 10(2) of the Manual lists the fields that must be in a declaration of conformity prepared under the provisions of the Old Medical Device Regulations and whose checking is ensured during document review in ÜTS. Point (a) of the same paragraph looks for the name or trade name and address of the manufacturer, point (b) the name or trade name and address of the EU authorised representative, if any, point (c) the information on the medical devices (model, reference number, product name and, if any, detailed description), point (ç) the regulation it belongs to, point (d) the class of the medical devices and/or the annex followed, point (e) the document start date, that is, the date of issue, point (f) the document end date, that is, the validity date, and point (g) the name and surname of the authorised person together with their title and signature or stamp. The same point expressly states that the electronic signature is also valid.

There are two options in the duration field. Sub-point (1) of point (f) of the same paragraph says that, if the document is unlimited-term, "Unlimited-Term" ("Sınırsız Süreli") is selected on the ÜTS screens; sub-point (2) says that, if the document has an end date, it is entered in the relevant field. The last sentence of the paragraph sets a single rule for all of these fields: the information entered into the system must be the same as the information in the document. Even a one-letter difference between the manufacturer address written in the declaration and the address of the manufacturer record you select in ÜTS will come up at the review stage.

In the EU declaration of conformity the field list expands markedly

For MDR and IVDR declarations, the second paragraph of Article 11 of the Manual gives a much longer list. Point (a) of the same paragraph looks for the name or trade name and address of the manufacturer, point (b) the name or trade name and address of the EU authorised representative where applicable, and point (c) the manufacturer's Single Registration Number (SRN), if one has already been issued, and states that this number is entered in the "Münferit Kimlik No" (Single Identification No) field in ÜTS.

The sentences that give the document its character as a declaration are also mandatory fields. Point (ç) of the same paragraph looks for a statement that the EU declaration of conformity is drawn up under the sole responsibility of the manufacturer; point (ğ) looks for a statement that the device is in conformity with the Medical Device Regulation it falls under and, where applicable, with other relevant EU legislation that requires an EU declaration of conformity to be drawn up. Most templates pass over these two sentences in a single line; ÜTS counts them as separately checked fields.

On the product identification side, point (d) of the same paragraph asks for the Basic UDI-DI, and point (e) asks for the product name and trade name, product code, catalogue number or, where applicable, other unambiguous references such as a photograph, which allow the device covered by the declaration to be identified and traced, together with its intended purpose. The first sub-point of the same point allows some flexibility: except for the product name or trade name, the information enabling identification and traceability may be provided by the Basic UDI-DI. So if catalogue numbers are not written out one by one, the Basic UDI-DI can fill that gap, but the product name never drops out.

The fields on conformity assessment are in the same paragraph. Point (g) asks for the risk class of the device and point (h) for references to the common specifications used and in relation to which conformity is declared. Point (ı) ties the declaration to the notified body: in all EU declarations of conformity except those covering only Class I Other or Class A devices, the name and identification number of the notified body, a description of the conformity assessment procedure carried out and the identification of the EU certificate or certificates issued, that is, the certificate number or numbers, must be present. For every device that goes through a notified body process, the declaration and the certificate are linked to each other through this point.

The remaining fields are on the date and signature side: point (i) asks for additional information where applicable, point (j) the document start date, point (k) the document end date, point (l) the place and date of issue of the declaration, and point (m) the name and surname of the authorised person with their title and signature or stamp; here too an electronic signature is valid. The last sentence of the same paragraph again requires that the information entered into the system be the same as the information in the document.

For domestic manufacturers, the seventh paragraph of Article 11 of the Manual states that the EU Declaration of Conformity templates prepared as examples, particularly for domestic manufacturing firms, are in Annex 7 and Annex 8 of the Manual.

The scope statement must identify the product; a general wording is not accepted

The part of the declaration where problems come up most often is the product scope. The fifth paragraph of Article 10 of the Manual says that the scope of the medical devices in a declaration of conformity must be able to identify the product, that general wording is not enough to identify a product, and it gives a direct example: instead of "surgical hand instruments", detail such as forceps, retractor and dilator is expected. The same paragraph also shows the solution: if a document issued for imported products conflicts with this requirement, a document listing the branch types or product descriptions that detail the products described in general terms in the document is uploaded separately to the "Technical Document Related to the Product(s) in the Document" ("Belgedeki Ürün(ler)e Ait Teknik Doküman") field.

The Agency may ask for more in the same field if it sees fit. The sixth paragraph of Article 10 of the Manual lists what is uploaded to this field when it is considered necessary for any of the devices in the declaration: a technical document containing the effect and mechanism of action of the medical device, its intended purpose, place of use, duration of use and target population, the class of the device, the classification rule and its justification, the product content and, depending on the form of the product, percentage component information, together with label information and the instructions for use. The fifth paragraph of Article 11 of the Manual gives the same list in more detail for MDR and IVDR declarations: the name and general description of the device, the name and address of the manufacturer, the UDI-DI number and the EU certificate number if any, intended purpose, place of use, type of contact, duration of use, population to be used on, mechanism of action, class and classification rule, device content and percentage component information, and label information and instructions for use are in this list. The same paragraph also says that these documents must be in the technical file of the device and that, where considered necessary, the whole technical file will be kept ready to be delivered to the Agency.

Contract manufacturing is a separate heading. The fourth paragraph of Article 10 and the fourth paragraph of Article 11 of the Manual say that where the manufacturer has the devices covered by the declaration made by another manufacturer, the title and address details of that manufacturer are also expected to be in the declaration or the certificate; if this information is not on the declaration, the certificate, the device label or the sales packaging, the Agency will, if it considers it necessary, ask for the pages of the technical file containing this information to be uploaded.

Finally, the seventh paragraph of Article 10 and the sixth paragraph of Article 11 of the Manual require the declaration to be uploaded completely, with all its annexes and additional pages. In most declarations the scope list sits in the annex; an annex that is uploaded incomplete means the scope cannot be proven.

A foreign manufacturer's declaration enters ÜTS by one of four routes

This is the most important section of the page. The first paragraph of Article 12 of the Manual lists, as four alternatives, the document to be uploaded to the "Original Document" ("Orijinal Doküman") field in the first registration or update notifications that importing firms make for the "Declaration of Conformity" or "EU Declaration of Conformity (MDR/IVDR)" belonging to a foreign manufacturer, and the last sentence of the paragraph states that these are uploaded completely, together with their annexes.

Original of the apostille-certified document

Point (a) of the same paragraph accepts an apostille-certified document. The first sub-point of the same point determines from which authority the apostille is obtained: if the manufacturer is in an EU member state, an apostille from the competent authority of the country where the manufacturer is established is accepted; if the manufacturer is established in a non-EU country, an apostille from the competent authority of the country where the manufacturer or the EU authorised representative appointed by the manufacturer is established is accepted. The second sub-point of the same point says that the original of the apostille-certified document must be uploaded to ÜTS and that a copy issued by a Turkish notary or by a notary of another country need not be uploaded.

Document certified by the Turkish Republic embassy or consulate

Point (b) of the same paragraph defines, as the second route, a document certified by the Turkish Republic embassy or consulate in the manufacturer's country. The first sub-point of the same point states that, if the manufacturer is established in a non-EU country, certification by the Turkish Republic embassy or consulate in the country where the manufacturer's EU authorised representative is established is also accepted. The second sub-point says that the certified document itself must be uploaded and that a notarised copy is not needed.

e-Apostille: the file to upload is the original file downloaded from the authority

The third route is the electronic apostille, and its conditions are written in the most detail. Point (c) of the first paragraph of Article 12 of the Manual says that where the apostille-certified document mentioned in point (a) is created electronically by the competent authorities of states that are members of or parties to the Hague Convention, the original of the e-Apostille-certified document is accepted. For this, one of two alternative verification conditions must be met: the documents to which an apostille has been applied can be reached and verified through the query tools indicated on the e-Apostille; or the document can be verified by uploading the apostilled document and/or entering the apostille details through the same query tools.

The critical point is in the sub-points of the same point. So that the e-signature of an e-apostilled document can be verified, the first sub-point makes it mandatory to upload to ÜTS the original of the document that the authority issuing the e-apostilled document gave to the party concerned digitally or that was downloaded from the authority's website. In other words, printing and scanning the file downloaded from the authority's site, taking a screenshot of it or re-saving it in another program breaks this condition. The second sub-point says that, in the e-signature query of the document, the e-signature of the authority that carried out the apostille must be validly verifiable. The third sub-point states the result: e-apostilled documents whose e-signature cannot be verified, because the original was not uploaded or for other reasons, and/or which carry an invalid e-signature, are not accepted.

The manufacturer's own electronic signature

The fourth route does not require an apostille chain at all. Point (ç) of the first paragraph of Article 12 of the Manual says that, for documents signed electronically by the manufacturer, the original of the e-signed document is accepted provided that three things can be verified through e-signature verification tools or data: that the document was signed electronically by its manufacturer, that the e-signature is valid, and that no change was made to the document afterwards. These three elements are required together; the existence of a signature is not enough on its own, it must be possible to show that the signature belongs to the manufacturer and that the document was not changed after signing.

Update 2026/KKB-2: With TİTCK Announcement No. 2026/KKB-2 of 02.11.2026, a new route was introduced in place of the option of the manufacturer uploading an e-signed original document, for the declaration of conformity and the SİP (system/procedure pack) declaration of a foreign manufacturer:

"it has been decided to apply the option under which the said document is uploaded to the relevant field in ÜTS together with an undertaking letter signed electronically by the authorised person of the importing firm authorised by the foreign manufacturer, with an e-signature (secure e-signature) created on the basis of a qualified electronic certificate obtained from an electronic certificate service provider recognised by Türkiye."

Source: TİTCK Announcement No. 2026/KKB-2. Required documents and steps: ÜTS document registration guide.

The first sub-point of the same point makes uploading the original of the e-signed document to ÜTS mandatory so that the e-signature can be verified. The second sub-point gives the same result as the e-Apostille point: documents whose e-signature cannot be verified, because the original was not uploaded or for other reasons, and/or which carry an invalid e-signature, are not accepted.

The shared logic of these two points is this: when the electronic route is chosen, the file that goes to ÜTS must be the file that carries the signature itself. A signature that cannot be verified or is invalid does not make a document stronger than a wet-signed one; it makes a document that cannot be accepted at all.

The Turkish translation is a separate field and has its own rules

Uploading the original document correctly does not finish the job. The second paragraph of Article 12 of the Manual says that, where the condition in the first paragraph applies, a complete Turkish translation of the document uploaded to the "Original Document" field, made by a sworn translator, is uploaded to the "Turkish Document" ("Türkçe Doküman") field. Point (a) of the same paragraph says that only the Turkish part of the document certified by the sworn translator is uploaded to this field; point (b) says that the document will carry the name and surname, stamp, signature or seal of the sworn translator who made the translation, and a translator's declaration that the translation was made in conformity with the original document.

The third paragraph of the same article says that it is preferred that the documents uploaded to both fields be in a searchable format, so that the review expert can assess them more easily and correctly. The translation obligation has a counterpart on the Regulation side too: the first paragraph of Article 19 of the Medical Device Regulation says that, for devices placed on the market in Türkiye, if the declaration is drawn up in other languages, a Turkish translation of the declaration is added.

On physical submission, the fourth paragraph of Article 12 of the Manual is clear: the documents uploaded to the "Original Document" and "Turkish Document" sections need not be sent physically to the Agency. The same paragraph says that, after the provision mentioned in paragraph (a) of Article 69 of the Manual enters into force, the petition whose example is in Annex 2 need not be sent physically either, and that applying through ÜTS is sufficient.

The EU authorised representative is proven together with the declaration

If you are registering the EU declaration of conformity of a foreign manufacturer, one more document is requested. The seventh paragraph of Article 12 of the Manual says that, where an EU declaration of conformity belonging to a manufacturer not established in an EU member state or in Türkiye is registered, the agreement made with the EU authorised representative appointed by the manufacturer and/or the authorisation document drawn up between them must be declared to the Agency through ÜTS.

Point (a) of the same paragraph makes it mandatory to make an entry in the "EU Authorised Representative" ("AB Yetkili Temsilcisi") field on the ÜTS registration screen in such registration notifications. Point (b) says that the agreement or authorisation document between the manufacturer and the EU authorised representative is uploaded to the "Authorisation Document" ("Yetki Belgesi") field that appears below this field; alternatively, a printout of the actor screen from the EUDAMED website, belonging to the manufacturer concerned and clearly showing the details of the EU authorised representative in question, may also be uploaded.

Point (c) brings a convenience: the original of the document and its Turkish translation by a sworn translator are uploaded together to the "Authorisation Document" field, but no apostille or embassy or consulate certification is required on the original of the authorisation document. The first sub-point of the same point says that, if a EUDAMED printout has been uploaded to this field, the sworn Turkish translation of the printout need not be uploaded separately. Point (ç) says that this document too is not sent physically to the Agency. Point (d) draws the limit: manufacturers established in an EU member state or in Türkiye do not appoint an EU authorised representative; this requirement exists only for manufacturers not established in an EU member state or in Türkiye.

The Regulation basis of this field is summarised in the first sub-point of point (b) of the second paragraph of Article 11 of the Manual: under the first paragraph of Article 11 of the MDR and the first paragraph of Article 12 of the IVDR, manufacturers established outside the EU member states and Türkiye must appoint an authorised representative from Türkiye or from one of the EU member states in order to supply devices to the EU market.

A domestic manufacturer registers its own declaration with an entirely separate application type

If you are a domestic manufacturer, you do not use any of the four routes above. Point (a) of the first paragraph of Article 13 of the Manual requires that, in the first registration or update notifications and applications concerning the "Declaration of Conformity" or "EU Declaration of Conformity (MDR/IVDR)" of their own products, domestic manufacturing firms apply by selecting "e-Signed Document Application for Manufacturer Firms" ("İmalatçı Firmalar için e-imzalı Belge Başvurusu") as the Document Application Type. Point (b) of the same paragraph says that the option "Are you the manufacturer of the products covered by the document?" ("Belge kapsamındaki ürünlerin üreticisi firması siz misiniz?") is selected as "Yes".

The language condition is different too. Point (c) of the same paragraph says that documents whose original language is Turkish are accepted; point (ç) says that, by selecting "Yes" for the option "Is the original document Turkish?" ("Orijinal Doküman Türkçe mi?"), the original of the document, signed electronically by the person registered as "Responsible Manager" or "Signatory" on behalf of the domestic manufacturing firm concerned, is uploaded completely to the "Original Document" field.

The signature check is done by the system. The first sub-point of the same point says that the e-signed document will be checked automatically by the system and must have been signed electronically by the firm's "Responsible Manager" or "Signatory", and that ÜTS allows a document application only in that case. The second sub-point adds a name-matching condition: if the name and surname of the signing official are written in the body of the document, the name and surname in the document must match those of the person registered as "Responsible Manager" or "Signatory" who signed the document electronically. Point (e) of the same paragraph says that this document need not be sent physically to the Agency and that applying through ÜTS as an "e-Signed Document Application for Manufacturer Firms" is sufficient.

The choice of application type is also defined in Article 39 of the Manual. Point (c) of the fourth paragraph of the same article lists the "e-Signed Document Application for Manufacturer Firms" type; the seventh paragraph of the same article says this type is selected for the document types mentioned in Articles 9, 13 and 16 of the Manual. On the importer side, the "Document Application for Importer Firms" ("İthalatçı Firmalar için Belge Başvurusu") type in point (a) of the fourth paragraph of the same article is used; the fifth paragraph of the same article ties the document types mentioned in Articles 8, 12 and 15 of the Manual to this type.

Date threshold: what decides is when the old-regulation declaration was issued

Whether a declaration under the old regulations can be accepted in ÜTS depends on the date it was issued. The eighth paragraph of Article 10 of the Manual says that, under Article 110 and point (d) of the first paragraph of Article 111 of the MDR, the MDD and AIMDD were repealed as of 26/5/2021, and that devices under the MDD or AIMDD that were lawfully placed on the market before that date may continue to be made available on the market or put into service only with a declaration of conformity issued before 26/5/2021. The same paragraph repeats this expressly: devices under the MDD or AIMDD cannot be placed on the market with a declaration of conformity whose first date of issue is 26/5/2021 or later.

The same paragraph makes one exception. A declaration of conformity issued for devices under the MDD or AIMDD may be updated, provided that the devices benefit from the MDR transitional provisions mentioned in Article 45 of the Manual, for changes other than "significant changes" in the product scope of the declaration and in the design and intended purpose of the devices, such as a change of manufacturer address, a change of manufacturing site, a change of EU authorised representative and a change of GMDN code. The same paragraph says that a declaration updated in this way is expected to be linked to its pre-26/5/2021 version; if a link cannot be made, the version or versions of the declaration prepared before 26/5/2021 must be uploaded to the "Technical Document Related to the Product(s) in the Document" ("Belgedeki Ürün(ler)e Ait Teknik Doküman") field.

The ninth paragraph of Article 10 of the Manual sets up the same structure for in vitro diagnostic devices. Under Article 107 and point (d) of the first paragraph of Article 108 of the IVDR, the IVDD was repealed as of 26/5/2022; the first date of issue of a declaration of conformity issued for devices under the IVDD must be before 26/5/2022. The update exception is tied here to the IVDR transitional provisions mentioned in Article 48 of the Manual.

In practice, these two paragraphs mean this: if the MDD declaration you hold looks recently dated, you first need to determine whether it is an updated version of a pre-26/5/2021 declaration or a newly issued one. The outcome in ÜTS is not the same in the two cases.

Once the declaration is registered, it is linked to the device

The declaration of conformity is not an end in itself; it is a precondition for registering the device. The first paragraph of Article 66 of the Manual says that, for a device registration notification under the MDR, depending on the type and risk class of the device, the appropriate one or ones of the documents mentioned in Articles 7, 11, 17, 20, 22, 23 and 24 of the Manual, which are in "Registered" ("Kayıtlı") status in ÜTS as a result of an application made under the provisions of Article 39 of the Manual, must be linked to the device. Article 11 in this list is the EU declaration of conformity. The first paragraph of Article 67 of the Manual sets up the same structure under the IVDR.

The order is therefore fixed: the declaration is registered first, reaches "Registered" status, and is then linked to the device. The eleventh paragraph of Article 39 of the Manual says that, once the documents to be attached to the application are complete, the "Approve" ("Onayla") action is taken on the "Document Application Summary" ("Belge Başvurusu Özeti") screen, and that only users of the "Medical Device Registration Operator" type of the firm concerned can take this action. The twelfth paragraph of the same article says that, if there is nothing contrary to procedure and business rules in the notification, the application is created in "Awaiting Signature" ("İmza Bekliyor") status; the thirteenth paragraph says that this application is signed electronically by the firm's applicant signatory. Point (a) of the same paragraph says that applications containing document types mentioned in Articles 6, 7, 10, 11 and 14 of the Manual move to "Awaiting Payment" ("Ödeme bekliyor") status after signing, and to "Under Review" ("İncelemede") status after payment and approval. The declaration of conformity and the EU declaration of conformity are in this list.

There is a volume limit too. The fifth paragraph of Article 12 of the Manual says that a document application whose procedures and principles are mentioned in Article 39 allows at most 5 document registration notifications; the fifteenth paragraph of Article 39 of the Manual states that a document application contains at most 5 document notifications.

What we do on a declaration file, and what we cannot do

Within Medical Device ÜTS Registration Consultancy, our concrete output on the declaration side is this:

  • Document type determination: Looking at the regulation the declaration refers to, we determine whether "Declaration of Conformity" or "EU Declaration of Conformity (MDR/IVDR)" is to be selected, based on the first paragraphs of Articles 10 and 11 of the Manual.

  • Field review: We compare the declaration you hold with the list of points in the second paragraph of Article 10 or the second paragraph of Article 11 of the Manual, and report any missing field to the manufacturer by point number.

  • Acceptance route planning: According to the manufacturer's country and what it has available, we work out which of the apostille, consulate certification, e-Apostille or manufacturer e-signature routes is applicable under the first paragraph of Article 12 of the Manual, and we check the originality of an e-signed file before upload.

  • Scope matching: We assess whether the product description in the declaration meets the "able to identify the product" criterion in the fifth paragraph of Article 10 of the Manual, and if it does not, we prepare the supplementary document set.

  • Domestic manufacturer application: We set up the e-signed application flow in Article 13 of the Manual and check in advance that the name of the signatory matches the name in the document.

  • Date check: For old-regulation declarations, we review the 26/5/2021 and 26/5/2022 thresholds and the version link before the application.

What we cannot do is just as clear. We cannot draw up a declaration of conformity on the manufacturer's behalf; paragraph 3 of Article 10 and paragraph 3 of Article 11 of the Manual assign the declaration to the legal manufacturer, and paragraph 3 of Article 19 of the Medical Device Regulation ties responsibility for conformity to the manufacturer. We cannot complete a field in the declaration for you, and we cannot sign in place of the manufacturer. We also cannot issue the apostille or the consulate certification; these are acts of the competent authorities and missions named in the first paragraph of Article 12 of the Manual. Nor is it possible to make an unverifiable e-signature verifiable; points (c) and (ç) of the same paragraph treat such documents as not accepted.

The way of working is this: the firm assigns us a user authorisation from its own ÜTS firm page, and we create the declaration records and the document application with the "Medical Device Registration Operator" ("Tıbbi Cihaz Kayıt Elemanı") authorisation. The eleventh paragraph of Article 39 of the Manual ties the "Approve" ("Onayla") action to this authorisation; under the thirteenth paragraph of the same article, the person who signs the application electronically remains the firm's signatory.

Tıbbi Cihaz ÜTS Kayıt Danışmanlığı

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application