Selecting the Device Class in ÜTS: The MDR and IVDR Class List

The class selected in ÜTS is more detailed than the class in the regulation. The MDR and IVDR class options, the classes that require a notified body, and the document chain.

You know your device's class. The problem is not what the class is, but what the system will ask of you once you select that class in ÜTS. The moment you select the class, the document list narrows, the document type you are looking for does not appear in the list, or the document set from the manufacturer does not fit the set ÜTS expects for that class. This is not a system fault; ÜTS uses the class as the key to the document business rules.

This page does not explain the classification rules; it explains the consequences of the class inside ÜTS. With the relevant articles of the Manual, we set out which class brings a notified body into the process, how the document chain changes as the class changes, where the class is entered on the registration screen, and at which stage a wrong class comes back. We rely on two texts: the Manual on Registration Procedures of Medical Devices in the Product Tracking System (Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz) and the Medical Device Regulation (Tıbbi Cihaz Yönetmeliği) published in the Official Gazette No. 31499 (bis) of 2/6/2021.

If you have not yet determined the class, the full text of the rules is on our medical device classification rules page; to see the likely class through questions, you can use the classification wizard in our MDR risk class article.

The class you select in ÜTS is more detailed than the class in the regulation

Article 51(1) of the Medical Device Regulation divides devices into four: class I, IIa, IIb or III. ÜTS does not use these four classes as they are. After saying that devices under the MDR are divided into four classes, Article 56(1) of the Manual adds expressly, in brackets, that additional classifications have been made in ÜTS according to characteristic product features such as whether the device is packaged sterile or has a measuring function, and that a device that is "Class-Ir" in ÜTS and a device that is "Class-I Other" in ÜTS are both classified as Class I under the MDR.

This additional distinction is reflected on the registration screen. Article 27(4)(a) of the Manual says that the "Class" field is mandatory in a device record under the MDR and that the appropriate one is selected from the list of options: "Class-I Other (not sterile and without a measuring function)", "Class-Ir (Reusable surgical instruments)", "Class-Im (With a measuring function)", "Class-Is (With a sterile function)", "Class-IIa", "Class-IIb" and "Class-III". So on the MDR side ÜTS sees seven options, not four.

The practical result is this: the manufacturer writing "Class I" in the EU declaration of conformity does not tell you which option to tick in ÜTS. If the device is supplied packaged sterile, Class-Is is selected; if it has a measuring function, Class-Im; if it is a reusable surgical instrument, Class-Ir. These three options do not behave the same as "Class-I Other" in terms of document business rules.

The class list is even longer on the IVDR side

For devices under the IVDR, Article 57(1) of the Manual defines four classes: Class A, Class B, Class C and Class D. The same paragraph states that additional classifications have been made in ÜTS according to characteristic product features and intended purposes, such as whether the device is packaged sterile or is a companion diagnostic device, and gives the example that a device that is "Near-Patient Test Device (Class C)" in ÜTS is counted as Class C under the IVDR.

Article 30(4)(a) of the Manual lists the options that can be selected in a record under the IVDR: "Class A Sterile", "Class A", "Class B", "Class C", "Class D", "Self-Test Device (Class B)", "Self-Test Device (Class C)", "Self-Test Device (Class D)", "Near-Patient Test Device (Class B)", "Near-Patient Test Device (Class C)", "Near-Patient Test Device (Class D)", "Companion Diagnostic Device (Class C)" and "Companion Diagnostic Device (Class D)". The same point states that this field is mandatory and that the device's risk class under the IVDR is selected here.

These thirteen options are not decorative. Article 67 of the Manual divides the IVDR document business rules exactly along these groups; separate sets of document business rules are provided for self-test devices, near-patient test devices and companion diagnostic devices.

Which class brings a notified body into the process

Whether a notified body is involved is the first distinction that determines which document types you will register in ÜTS. On the MDR side, this distinction is made in Article 61 of the Manual.

Article 61(4) of the Manual says that, except for custom-made or investigational devices, manufacturers of Class-I devices draw up technical documentation as set out in MDR Annexes II and III and then declare conformity by drawing up a declaration of conformity as referred to in MDR Annex IV and in Article 11 of the Manual. The same paragraph goes on to say that if the devices are placed on the market in sterile condition (Class-Is), have a measuring function (Class-Im) or are reusable surgical instruments (Class-Ir), the manufacturer will also carry out one of the procedures in MDR Annex IX (Chapters I and III) or MDR Annex XI (Part A). Points (a), (b) and (c) of the same paragraph draw the limit of the notified body's involvement: it is limited, respectively, to establishing and maintaining sterile conditions, to conformity with metrological requirements, and to matters relating to the reuse of the device.

This is a frequently missed point. For Class-Is, Class-Im and Class-Ir devices there is a notified body, but its certificate covers not the conformity of the whole device but only that limited subject. The general conformity of the device still rests on the manufacturer's declaration.

From Class IIa onwards the situation changes. Article 61(5)(a) of the Manual says that manufacturers of Class-IIa devices are subject to the MDR Annex IX (Chapters I and III) procedure, together with at least one representative device for each device category being subject to the MDR Annex IX (Chapter II - Article 4) procedure; point (b) of the same paragraph says that, alternatively, MDR Annex II and Annex III technical documentation may be prepared together with an MDR Annex XI (Article 10) or MDR Annex XI (Article 18) conformity assessment.

On the IVDR side the threshold begins after Class A. Article 62(4) of the Manual says that manufacturers of Class A devices prepare IVDR Annex II and Annex III technical documentation and draw up a declaration of conformity; if the devices are supplied sterile (Class A Sterile), one of the IVDR Annex IX (Chapters I and III) or IVDR Annex XI procedures is also carried out, but the notified body's involvement is limited to sterile conditions. Article 62(5) of the Manual regulates the notified body procedures for Class B, paragraph 6 for Class C, and paragraph 8 for Class D, separately.

The document chain is built by class, not by device

A device record in ÜTS does not stand on its own; the record is linked to documents registered in ÜTS beforehand. Article 66(1) of the Manual sets this up as follows: for a device registration notification under the MDR, according to the type and risk class of the device concerned, the appropriate one or ones of the documents that are in "Registered" ("Kayıtlı") status in ÜTS as a result of the application made under Article 39 of the Manual and that are mentioned in Articles 7, 11, 17, 20, 22, 23 and 24 of the Manual must be linked to the device in question.

The counterparts of these articles are as follows: Article 7 is the EU Certificate (MDR/IVDR), Article 11 the EU Declaration of Conformity (MDR/IVDR), Article 17 the Instructions for Use, Article 20 the Quality Management Certificate - ISO 13485, Article 22 the Domestic Goods Certificate ("Yerli Malı Belgesi"), Article 23 the Authorised Distributorship Certificate and Article 24 the Catalogue. What determines which of them are mandatory for which device is the class.

Article 66(2) of the Manual says that this linking is carried out in ÜTS according to the document business rules determined by the type and risk class of the device and on the basis of the conformity assessment procedures mentioned in Article 61 of the Manual. So the document business rule is the ÜTS counterpart of the conformity assessment procedure.

The continuation of the same article defines six separate document business rule sets on the MDR side:

  • Class I family: Article 66(3) of the Manual states that, for devices to be notified as "Class-I Other", "Class-Is", "Class-Im" or "Class-Ir", the appropriate one of the document business rules shown in Annex-88 of the Manual will be applied.

  • Class IIa: Paragraph 4 of the same article refers to the document business rules shown in Annex-89 for Class-IIa devices.

  • Class IIb, non-implantable: Paragraph 5 points to Annex-90 for Class-IIb devices excluding implantable devices and the devices under MDR Annex VIII Rule 12.

  • Class IIb, implantable: Paragraph 6 points to Annex-91 for implantable Class-IIb devices other than sutures, staples, dental fillings, dental braces, dental crowns, screws, wedges, plates, wires, pins, clips and connectors.

  • Class IIb, under Rule 12: Paragraph 7 points to Annex-92 for Class-IIb devices under MDR Annex VIII Rule 12.

  • Class III: Paragraph 8 states that, for Class-III devices, the appropriate one of the document business rules shown in Annex-93 will be applied.

The real information here is this: "Class-IIb" is not a single document behaviour. The same class is spread over three separate business rule sets, depending on whether the device is implantable and whether it falls under Rule 12.

On the IVDR side, Article 67 of the Manual builds a similar distribution. Paragraph 3 points to Annex-94 for Class A and Class A Sterile, paragraph 4 to Annex-95 for Class B and Class C, paragraph 5 to Annex-96 for Class D, paragraph 6 to Annex-97 for Self-Test Device (Class B) and Near-Patient Test Device (Class B), paragraph 7 to Annex-98 for Self-Test Device (Class C) and (Class D), paragraph 8 to Annex-99 for Near-Patient Test Device (Class C) and (Class D), and paragraph 9 to Annex-100 for Companion Diagnostic Device (Class C) and (Class D).

The instructions-for-use requirement depends directly on the class

The most visible effect of the class on the document chain appears in the instructions for use. Article 17(2)(a) of the Manual says that, for Class-III and Class-IIb devices certified under the MDR and the MDD, uploading to ÜTS the instructions for use issued by the manufacturer is mandatory. The same point says that, for Class-IIa devices, the instructions for use will be uploaded if the manufacturer has issued them; as an exception, only for Class-IIa devices under the MDD, if the manufacturer has not issued instructions for use, an intended purpose form/declaration signed by the responsible manager of the manufacturer or importer firm will be accepted.

Point (b) of the same paragraph makes uploading instructions for use mandatory for Class B, Class C and Class D devices certified under the IVDR, and for Self-Test Devices (Class B/C/D), Near-Patient Test Devices (Class B/C/D) and Companion Diagnostic Devices (Class C/D).

Article 17(3) of the Manual separates the other risk groups: uploading instructions for use is optional for Class-I Other, Class-Im, Class-Is, Class-Ir, Class A, Class A Sterile and IVD Other devices. The same paragraph makes a note: where the Agency deems it necessary, for example if a doubt arises about product safety, or if the original label of the device refers to the instructions for use, it may request the instructions for use to be uploaded for devices in this risk group too. So "optional" is not a permanent exemption.

In the case of accessories the rule changes once more. Article 17(4) of the Manual says that, for Class-IIa devices that are accessories of another device, if the field "Is it a component/accessory of a device?" is selected as "Yes" during device registration in applications under the MDR and the IVDR, the requirement to upload instructions for use will not be sought. By contrast, paragraph 5 of the same article says that, for Class-IIb and Class-III devices that are accessories, where the manufacturer of the main device is different from the manufacturer of the accessory, the accessory's own instructions for use will be uploaded, or if there are none, those of the main device.

The content of the declaration of conformity also changes by class

The EU declaration of conformity is drawn up in every class, but it does not say the same thing in every class. Article 11(2)(g) of the Manual says that the risk class of the device is among the fields that must be in the EU declaration of conformity and whose control is carried out during document review in ÜTS. The last sentence of the same paragraph says that the information entered into the system and the information in the document must be the same.

The critical distinction is in point (ı) of the same paragraph: for EU declarations of conformity other than those covering only Class-I Other or Class A devices, the name and identification number of the notified body, a description of the conformity assessment procedure performed and the identification of the EU certificate or certificates issued must be in the declaration. This single point summarises the effect of the class on the declaration of conformity: notified body information is not looked for only in Class-I Other and Class A declarations.

On the EU certificate side too, the risk class is a mandatory field. Article 7(2)(h)(1) of the Manual looks, for EU technical documentation assessment certificates, EU type-examination certificates and EU product verification certificates, for a clear description of the device, the intended purpose, the risk class and the Basic UDI-DI information. Sub-point 2 of the same point looks, for EU quality management system certificates and EU production quality assurance certificates, for the description of the devices or device groups and the risk class, and requires the intended purpose only for Class-IIb devices.

The class must be entered before document selection

In ÜTS the order is not free. Article 27(5)(a)(2) of the Manual says that the documents that must be added, according to the appropriate one of the document business rules mentioned in Article 66 of the Manual in line with the risk class of the device concerned, are selected from among the documents in registered status in ÜTS as a result of the application made under Article 39 of the Manual. Sub-point 3 of the same point ties the order expressly: before document selection in this field, the class of the relevant device must have been selected on the ÜTS registration screen, and the selection is made from among the documents that can be selected according to the class of the device.

This means: if you cannot see the document type you are looking for in the document list, the problem is often not in the document but in the class you selected. You cannot expect the document list to widen without correcting the class.

The same order exists on the document side. Article 66(1) of the Manual requires the documents to be linked to be in "Registered" status. A document that has not yet passed the Agency's review and reached registered status cannot be linked to the device; even if the class is right, the chain cannot be built.

A wrong class comes back at the latest in the document chain

A wrong class selection may not give an error at the time of registration. The problem appears later, at the correction stage. Article 43(1) of the Manual says that, where a device registered in ÜTS under the same barcode has records belonging to different firms and there are data differences between these records, the relevant product parameters must be made identical and consistent before the device application.

Paragraph 7 of the same article states that data change requests relating to fields requiring Agency assessment, such as "Class", "GMDN" and "EMDN", are in every case submitted to the Agency's assessment and approval. While other fields can be corrected by the unanimous agreement of the firms, a class correction does not go through this route.

The point where the chain breaks is in Article 43(11) of the Manual: if, as a result of updating the "Class" data field, the documents linked to the relevant device records become incompatible with the device class, a revision is issued to the devices with the relevant barcode. So correcting the class may require re-handling all the records tied to that barcode.

Article 44(7) of the Manual lists which fields are in the "Field to Request a Change" ("Değiştirme Talebinde Bulunulacak Alan") list; "Class" is in point (b), "Is it implantable?" ("Vücuda implante edilebilir mi?") in point (d), and "Was it packaged sterile?" ("Steril paketlendi mi?") in point (ı). That these three fields are in the same list is no coincidence; two of them directly determine the class and the document business rule.

Other fields that must be kept consistent with the class

Even if the class alone is right, an application does not proceed soundly if some fields on the registration screen contradict the class. Article 27(7)(h) of the Manual says that the "Is it implantable?" field must be selected in ÜTS. Sub-point 3 of the same point says that when this field is selected as "Yes", ÜTS automatically marks the "Is it single-use?" and "Is there a limited number of uses?" fields as "Yes" and the "Limited Number of Uses" field as "1", and that these fields cannot be changed. Points (r) and (ş) of the same paragraph say that the "Was it packaged sterile?" field and the "Is Sterilisation Required Before Use?" field are mandatory.

Consistency is also sought on the GMDN side. Article 27(4)(b)(5) of the Manual says that consistency is mandatory between the information in the description of the GMDN code to which the device is matched and product features such as whether the device is single-use and/or sterile.

Consistency on the barcode side, in turn, carries a direct sanction. Article 27(2)(b)(4) of the Manual says that a device application made under Article 40 of the Manual for a device record whose barcode has not been entered into ÜTS as it appears in its technical documents will end negatively.

The validity period of certificates also affects the document chain

A correct class is not enough to keep the chain going. Article 7(3) of the Manual says that EU certificates issued under the New Medical Device Regulations remain valid for the period stated in the certificate, not exceeding 5 years, provided that no change or update has been made to the certificate during this period, and that validity may be extended, on the manufacturer's application and on the basis of a re-assessment, for further periods each not exceeding 5 years. The same provision is also in Article 56(2) of the Medical Device Regulation.

On the quality management certificate side the period is shorter. Article 20(5) of the Manual says that, for documents of the type "Quality Management Certificate - ISO 13485", the validity period cannot be more than 3 years.

These periods should be read together with the class: in every class where a notified body comes into play, the chain's standing depends on the validity of the relevant certificate. Article 56(4) of the Medical Device Regulation says that the notified body will suspend, withdraw or restrict the certificate if it finds that conformity with the requirements is not maintained and corrective action is not taken.

What we do on the class decision, and what we cannot do

Within medical device ÜTS registration consultancy, our concrete output on the class side is this:

  • Class mapping table: We map the risk class written in the manufacturer's documents to the seven MDR options in Article 27(4)(a) of the Manual or to the IVDR options in Article 30(4)(a), and base the sterile, measuring-function and reusability distinctions on the document.

  • Document chain plan: We determine which of the business rule sets in Articles 66 and 67 of the Manual will apply and list which documents must be brought to "Registered" status before registration.

  • Intra-document consistency check: We compare the risk class, intended purpose and Basic UDI-DI information on the EU certificate and the EU declaration of conformity against the mandatory field lists in Articles 7 and 11 of the Manual.

  • Cross-field contradiction scan: We detect, before the application, inconsistencies between the class and the "Is it implantable?", "Was it packaged sterile?" and GMDN fields.

What we cannot do is clear too. We do not determine the class; Article 51(1) of the Medical Device Regulation ties classification to Annex VIII and the intended purpose of the device, and responsibility lies with the manufacturer. We cannot write into ÜTS a risk class that is not written in the manufacturer's document; Article 11(2) of the Manual requires the information entered into the system and the information in the document to be the same. Nor can we get a notified body to widen the scope of its certificate.

The way of working is this: the firm defines a user authorisation for us from its own ÜTS firm page, and we carry out the registration and application preparation with the "Medical Device Registration Operator" ("Tıbbi Cihaz Kayıt Elemanı") authorisation. Article 39(11) of the Manual leaves the "Approve" ("Onayla") action in a document application only to users with this authorisation; paragraph 13 of the same article states that an application in "Awaiting Signature" status will be electronically signed by the firm's signatory who is the applicant. The signature always stays with the firm.

Tıbbi Cihaz ÜTS Kayıt Danışmanlığı

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara