The ÜTS Registration Chain: Company, Document and Product Registration
ÜTS registration consists of three tracks: company, document and product registration. The fields verified before each application, content checks and barcode uniqueness. Medex ÜTS.
When you start the ÜTS work, three things come to the table at the same time: your firm registration is not yet complete, the documents from the manufacturer are sitting in a folder, and the sales side asks when the products will show as registered. It is not possible to run these three in parallel, because there is a one-way link between them: a product cannot be registered without its document, and a document cannot be registered without its firm.
This page sets out the concrete work content of ÜTS registration consultancy as three lines of work. For each line of work it states where the information comes from, what the firm itself has to provide, what we verify before an application is sent, and in which area the three lines of work lock into one another. The conceptual framework of the sharing of responsibility and of the consultancy relationship is a separate subject; here the work itself is described.
The source is the Manual on Registration Procedures of Medical Devices in the Product Tracking System (Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz). In places reference is also made to the Medical Device Regulation (Tıbbi Cihaz Yönetmeliği) published in the Official Gazette No. 31499 (bis) of 2/6/2021 and to the Medical Device Sales, Advertising and Promotion Regulation (Tıbbi Cihaz Satış, Reklam ve Tanıtım Yönetmeliği) published in the Official Gazette No. 29001 of 15/5/2014.
Three lines of work and the one-way link between them
Before making a work plan, the direction of this link needs to be seen. Article 5(1) of the Manual says that the registration to ÜTS of firms carrying out the sale and/or application of medical devices is carried out if they are authorised under the relevant legislation. Article 39(1) of the Manual says that an application must be made to the Agency for the assessment of document registration or update notifications, and Article 40(1) of the Manual says that, after the information required for the device registration notification has been added to ÜTS, a separate device application must be made for the medical devices that are ready for application to reach "Registered" status.
The provision that establishes the link is Article 66(1) of the Manual: for a device registration notification under the MDR, according to the type and risk class of the device, the appropriate one of the documents in "Registered" status in ÜTS as a result of the application made under the provisions of Article 39 must be linked to the device. Article 67(1) of the Manual sets up the same link for the IVDR.
What this means for planning is this: the calendar is built on the number of documents, not on the number of products. The real bottleneck in front of a portfolio of a hundred items is the four or five documents behind that portfolio.
Firm registration line of work: which source the identity will come from
Firm registration is not a form filled in by hand; it is fed from another system. Article 5(2) of the Manual says that firm registrations are carried out from MERSİS or VEDOP sources, that if a firm has both a MERSİS and a VEDOP number the operation is carried out over the MERSİS number, and that the firm registration operations of pharmacies are carried out from the EBS source.
The first thing we ask of the firm in this line of work is not a document but a verification: whether the title, address and representation information in the trade registry is current. Because if the source record is wrong, it arrives in ÜTS wrong, and the correction is not made in ÜTS.
The firm type also determines whether physical documents are needed. Article 5(3) of the Manual says that, when the firm type "Firm with a MERSİS record" or "Sole proprietorship without a MERSİS record" is selected, no physical documents are needed for the application; that if the firm type is "General partnership firm" ("Adi ortaklık firması"), a firm registration application must also be made through EBS in order to complete the firm registration process; and, by points (a) and (b) of the same paragraph, it requires that the original or notarised copy of the general partnership agreement and the originals or notarised copies of the signature circulars or signature declarations of the persons named in the agreement be sent to the Agency with a standard petition. Paragraph 4 of the same article says that, if this application ends negatively, the physical documents will be returned to the firm with a cover letter, and if it ends positively, the originals will not be returned but kept in the Agency's archive.
What we verify in firm registration before the application
The condition most often skipped in this line of work is the ÇKYS link. Article 5(5) of the Manual says that firms that will register in ÜTS under the type "Manufacturer/Importer/Dealer/Exporter" must hold a medical device sales centre licence and that the ÇKYS number must be entered as mandatory information during firm registration. The same paragraph says expressly that, for places where the devices listed in Annex-3 of the Sales, Advertising and Promotion Regulation are sold, a sales centre authorisation certificate is not required, except for their manufacturer or importer, but that this exception does not remove the obligations set by the Medical Device Regulations.
There is a lock on the commercial side as well. Article 26(1) of the Sales, Advertising and Promotion Regulation says that devices cannot be sold by natural or legal persons not authorised under the provisions of the Regulation; paragraph 2 of the same article says that sales centres cannot sell devices that are not registered in the Agency's registration and information management system. These two paragraphs tie firm registration and product registration to the same commercial outcome.
On the staffing side, we check two channels separately. Article 5(6) of the Manual says that, in transfer or merger operations, the signatories and users defined in ÜTS must be updated; paragraph 7 says that, if changes of responsible manager, title and address are updated in ÇKYS, they are updated automatically in ÜTS. One runs by hand and the other automatically; treating the two as substitutes for each other quietly leaves the firm registration out of date.
Document registration line of work: the manufacturer's folder and ÜTS's document are not the same thing
The first thing we do in this line of work is not to produce a document but to map the incoming file to the types that ÜTS recognises. Article 7(1) of the Manual says that the document type "EU Certificate (MDR/IVDR)" ("AB Sertifikası (MDR/IVDR)") is used only for certificates issued under the New Medical Device Regulations; Article 6(1) of the Manual says that the "EC Certificate" ("EC Sertifikası") type is used only in notifications relating to certificates under the Old Medical Device Regulations. The same distinction exists on the declaration side: Article 11(1) of the Manual ties the "EU Declaration of Conformity (MDR/IVDR)" ("AB Uygunluk Beyanı (MDR/IVDR)") type to the new regulations, and Article 10(1) of the Manual ties the "Declaration of Conformity" ("Uygunluk Beyanı") type to the old regulations.
Once the mapping is done, the content of the document is checked. Article 11(2) of the Manual lists, point by point, the information that must be in the EU declaration of conformity: point (a) asks for the manufacturer's name or trade name and address, point (b) for the name or trade name and address of the EU authorised representative where applicable, point (c) for the manufacturer's single registration number if already issued, point (ç) for the statement that the declaration is issued under the sole responsibility of the manufacturer, point (d) for the Basic UDI-DI, point (e) for the information allowing identification and traceability of the device and its intended purpose, point (g) for the risk class of the device, and points (j) and (k) for the start and end date of the document.
What we ask of the firm at this point is clear: we cannot complete an incomplete declaration. The declaration is issued by the manufacturer; our job is to detect the gap before the application and notify it in writing.
In which fields the content check is done on the document side
The check runs separately for each document type and looks at the provisions specific to the type.
Link between declaration and certificate: Article 11(2)(ı) of the Manual asks that, in all EU declarations of conformity other than those whose scope covers only Class I Other or Class A devices, the name and identification number of the notified body, a description of the conformity assessment procedure carried out and the identification of the EU certificate or certificates issued be present.
Quality management document: Article 20(2) of the Manual says that, if the document has been revised, the revision date will be entered as the "Document Start Date" ("Belge Başlangıç Tarihi"); paragraph 3 says that, where the revision number is stated next to the certificate number, the document number will be recorded together with the revision number; and paragraph 5 says that the validity period of such documents cannot be more than 3 years.
Instructions for use: Article 17(4) of the Manual says that, for Class IIa devices that are accessories of another device, if the "Is it a component/accessory of a device?" field is selected "Yes" during registration, no obligation to upload instructions for use is sought; paragraph 5 says that, for higher-class devices that are accessories, the accessory's own instructions for use will be uploaded, or, if there are none, the instructions for use of the main device.
Authorised distributorship document: Article 23(1) of the Manual defines this document as a document issued by the manufacturer of the device on behalf of importer firms operating in our country.
Catalogue: Article 24(2) of the Manual says that uploaded catalogues are recorded in ÜTS without being subject to checking by the review expert.
These checks are done before the application, not after. If there is a violation of a business rule in a document notification, the application is not formed at all; Article 39(12) of the Manual writes that the application will be created in "Awaiting Signature" ("İmza Bekliyor") status only if there is no situation contrary to the procedure and business rules.
Product registration line of work: the device record consists of ten headings
The device record is not a single screen. Article 27(1) of the Manual says that the information that must be entered into the system during device registration under the MDR is under the headings "Descriptive Information" ("Tanımlayıcı Bilgiler"), "Import/Manufacture Information" ("İthal/İmal Bilgileri"), "Classification Information" ("Sınıflandırma Bilgileri"), "Product Documents" ("Ürün Belgeleri"), "Product Images" ("Ürün Görselleri"), "Features" ("Özellikler"), "Measurement Information" ("Ölçü Bilgileri"), "Storage/Use Condition Information" ("Saklama/kullanım Koşulu Bilgileri"), "CMR and/or Endocrine Disrupting Substance Information" ("CMR ve/veya Endokrin Bozucu Madde Bilgileri") and "Application Status" ("Başvuru Durumu"). Article 30(1) of the Manual sets up a similar structure for the IVDR, and Article 25(1) of the Manual for the old regulations.
The sources of the fields under these headings differ, and the data we ask of the firm varies accordingly.
The source of the barcode is the manufacturer: Article 27(2)(a) of the Manual says that the standard of the UDI issuing entity will be selected and that the issuing entities authorised under the MDR are GS1, HIBCC, ICCBBA and IFA. Sub-point (3) of point (b) of the same paragraph says that the barcode must be entered into ÜTS exactly as it is, without any change, and sub-point (4) says that the application made for a device record whose barcode is not entered as it appears in its technical documents will end negatively. The answer to the question of imported or manufactured is in the declaration: sub-point (2) of Article 27(3)(a) says that "Manufactured" ("İmal") is selected if the manufacturer that issued the declaration of conformity to which the device is linked is established in Turkey, and "Imported" ("İthal") if not; sub-point (4) makes it mandatory to fill in the "Country of Import" and "GTİP Number" fields when "Imported" is selected. Point (b) of the same paragraph says that the country where the device is physically produced is entered in the "Country of Origin" field, and, in contract manufacturing, the country where the producing manufacturer is established is selected.
Class and terminology are read from the document: Sub-point (3) of Article 27(4)(b) of the Manual says that, if the GMDN code identifying the device is in at least one of the documents belonging to the device, that code will be selected; sub-point (5) makes it mandatory that the description of the GMDN code is consistent with features of the device such as whether it is single-use and sterile. Sub-point (5) of point (ç) of the same paragraph requires that, if the EMDN code is in the documents, the most detailed code and terminal term be selected.
Substance content is a separate heading: Point (a) of Article 27(10) of the Manual makes it mandatory to select whether the device contains CMR and/or endocrine disrupting substances as stated in point 10.4.1 of Annex I of the MDR; sub-point (4) of point (b) of the same paragraph makes it mandatory that information on the selected substances and the precautions to be taken be present in the instructions for use. This field is data that must be asked of the manufacturer separately; it is not written in most documents.
A record with the same barcode links you to another firm's record
This is the least anticipated item in the product registration line of work. Article 27(2)(d) of the Manual says that in the "Products with the Same Barcode" ("Aynı Barkodlu Ürünler") field, the list of records of devices previously registered in ÜTS by other firms with the same UDI-DI number can be viewed.
If there is a record in the list and the data do not match, the matter turns into a request. Article 43(1) of the Manual says that, if there are data differences between the records of different firms of a device registered under the same barcode, the relevant product parameters must be made identical and consistent in ÜTS before the first device application. Paragraph 2 of the same article says that a request application will be created for this according to the work steps in Article 44 of the Manual; paragraph 3 says that the request will be submitted to the other firms for their opinion, that the period for giving an opinion is 5 days, and that firms that do not give an opinion within this period are deemed to have accepted the request.
The result runs two ways. Article 43(4)(d) of the Manual says that, if the request relates to fields other than those requiring Agency assessment, such as "Class", "GMDN" and "EMDN", the change will be made automatically by ÜTS if the decision is unanimous; point (e) of the same paragraph says that, for fields requiring Agency assessment, the request will be submitted to the Agency in "Under Review" ("İncelemede") status. Paragraphs 5 and 6 of the same article say that, if one of the firms rejects or none of them gives an opinion, the request will still go to the Agency.
For this reason we run the portfolio scan before starting the device record. If a barcode clash is noticed later, the whole application calendar shifts.
Bulk notification changes the scale, not the rule
When the number of items goes up, entering them one by one does not work. Article 32(1) of the Manual says that the information that must be entered into the system for device registration under the MDR can be entered one by one as in Article 27, and can also be entered in bulk through the "Bulk MDR Add/Update Medical Device" ("Toplu MDR Tıbbi Cihaz Ekle/Güncelle") screen; Article 33(1) of the Manual sets up the same for the IVDR.
Bulk notification has its own format rules. Article 32(4) of the Manual says that in the template Excel file the red columns indicate mandatory fields and the yellow columns indicate fields that are mandatory depending on another field being filled in. Paragraph 5 of the same article says that the Notification Information File is created by compressing the template file together with the label information files and the different packaging image files in zip format; paragraph 6 says that these files must be directly inside the zip and not in separate folders; paragraph 7 says that the size of the zip file can be at most 200 MB; and paragraph 8 says that Turkish characters cannot be used in file names.
Bulk entry does not remove the application obligation. Paragraph 10 of the same article says that, for the assessment of a device registration or update notification made under this article, a device application must be made to the Agency as mentioned in Article 40 of the Manual. Paragraph 1 of Articles 37 and 38 of the Manual says that storage/use condition and measurement information values that are not present as parametric data fields in the template files can be added from separate bulk screens.
Where the three lines of work lock together: document selection and the Basic UDI-DI
The point where the lines of work join is the "Product Documents" heading on the device registration screen. Sub-point (1) of Article 27(5)(a) of the Manual says that selecting the "Documents" field is a mandatory field; sub-point (2) says that the documents to be added will be selected from among the documents already in registered status in ÜTS as a result of the application made under Article 39 of the Manual; and sub-point (3) says that it is mandatory for the class of the device to have been selected before the document selection.
The last link of the chain is the Basic UDI-DI. Sub-point (3) of Article 27(5)(c) of the Manual says that, before making a selection on the "Select Basic UDI-DI" ("Temel UDI-DI Seç") screen, selecting from the "Documents" field the EU Declaration of Conformity and the EU Certificate associated with the device and with the device's Basic UDI-DI is mandatory; sub-point (4) says that through this screen a selection can be made only from among the Basic UDI-DI numbers associated with the selected EU Declaration of Conformity, and that more than one selection cannot be made in this field. Article 30(5)(a) and (c) of the Manual set up the same order for the IVDR.
The order becomes this: firm registration, class decision, document application, the document reaching "Registered" status, document selection in the device record, Basic UDI-DI selection, device application. No step in this order can be brought forward.
The final check made before the application is sent
When the record is completed, the notification has not yet gone to the Agency. Sub-point (2) of Article 27(11)(a) of the Manual says that the "Ready for application?" ("Başvuruya hazır mı?") field is selected "No" if the notification will be updated later or is not yet ready to be sent for application, and "Yes" if it has reached its final form; sub-point (3) says that if "No" is selected the notification is saved in ÜTS as "Draft" ("Taslak"); and sub-point (4) says that if "Yes" is selected it is saved as "Ready for Application" ("Başvuruya Hazır") and listed on the "List My Medical Devices" ("Tıbbi Cihazlarımı Listele") screen.
This field is the gate of the device application. Article 40(5)(a) of the Manual says that the declaration linked to the relevant device is selected from the "Declaration" ("Beyan") field on the "Medical Device Application Summary" ("Tıbbi Cihaz Başvurusu Özeti") screen; point (b) says that only declarations linked to devices in "Ready for Application" status are listed on this screen; and point (c) says that, since medical device applications can be formed only from devices linked to a single declaration, only one declaration can be selected from the list.
Our pre-application checklist comes from these provisions: the barcode being identical to the technical document, consistency between the class and the selected document set, grouping on a per-declaration basis, leaving no item that has not moved to "Ready for Application" status, and data consistency in records with the same barcode. We do not create an application until these five headings are completed.
Work that continues after registration is finished
"Registered" status is not the end of the process. Article 36(1) of the Manual says that the operations of associating box/packaging and container barcodes containing multiple products with the unique product barcode of a device that has had a device record created and has reached "Registered" status with a positive application result, that is, the operation of adding placing-on-the-market information, can be carried out through the "Add Placing on the Market Information" ("Piyasaya Arz Bilgisi Ekle") screen; paragraph 2 of the same article says that this operation can also be done from the bulk screen.
Some updates do not require a new application. Article 35(1) of the Manual says that device information for which updating without application is permitted, such as associating a new catalogue with the relevant device or uploading a new product image, can also be entered in bulk through the "Bulk Update Product Without Application" ("Toplu Başvurusuz Ürün Güncelle") screen. Paragraph 4 of the same article points to a trap here: if the "Product Catalogue List" ("Ürün Katalog Listesi") field in the template is left empty and a product catalogue linked to the device exists before the operation, the relationship between the device and the catalogue is cut.
There is a separate application on the export side. Article 41(1) of the Manual says that the Agency issues a free sale certificate for export purposes and upon the request of the domestic manufacturer or of the EU authorised representative established in Turkey of a foreign manufacturer; paragraph 3 of the same article says that the certificate issued for devices under the new regulations will contain the Basic UDI-DI of the device and, if any, the unique number identifying the EU certificate. This certificate becomes meaningful only after the device registration is completed.
What we do in the three lines of work, and what we cannot do
Within medical device ÜTS registration consultancy, our concrete outputs in the three lines of work are these:
Firm registration file: We check the registration source, the firm type and the ÇKYS condition under paragraphs 2, 3 and 5 of Article 5 of the Manual, and for general partnership firms we prepare the document set in points (a) and (b) of paragraph 3 of the same article.
Document mapping and content report: We map each document from the manufacturer to the types in Articles 6, 7, 10, 11, 17, 20, 23 and 24 of the Manual, and we notify content gaps in writing before the application, according to the list of points in Article 11(2) of the Manual.
Device registration set: We fill in the fields in paragraphs 2, 3, 4 and 10 of Article 27 of the Manual, verifying them from the source document, and, when the number of items requires it, we prepare the bulk notification templates in Articles 32 and 33 of the Manual.
Barcode clash scan: We list the records with the same barcode through Article 27(2)(d) of the Manual and, if necessary, set up the data correction request under Articles 43 and 44 of the Manual.
Pre-application verification: We check item by item the conditions in paragraphs 5 and 11 of Article 27 and in points (a), (b) and (c) of Article 40(5) of the Manual, and we do not include an incomplete item in the application.
What we cannot do is clear as well. We cannot produce documents on your behalf: the EU declaration of conformity is issued by the manufacturer, the EU certificate by the notified body, and the instructions for use are prepared by the manufacturer. We cannot assign the barcode; Article 27(2)(a) of the Manual ties the UDI-DI to the issuing entity. We cannot electronically sign the application on your behalf; Article 40(5)(g) of the Manual requires that the device application be signed by the firm's signatory. We do not commit to the Agency's review result or its duration; Article 40(9) of the Manual describes the route to take in case of a negative decision, not the outcome.
The way of working is as follows: the firm defines a user authorisation for us from its own ÜTS firm page, we prepare and approve the notifications and application packages in the three lines of work with the "Medical Device Registration Operator" ("Tıbbi Cihaz Kayıt Elemanı") authorisation, and the person who electronically signs the application remains the firm's signatory.