Class IIa and IIb Medical Devices: A Guide to ÜTS Document Registration

Certificate combinations for Class IIa and IIb medical devices, conformity assessment routes, certificates issued under the old regulations, and the ÜTS document business rules.


The file from your manufacturer holds a single EU certificate. It does not carry the device name, only a device group description and the words "Annex IX Chapters I and III". You are trying to register a Class IIa device in ÜTS, and the document business rule is not completed by this one certificate. The problem is not a fake document; in Class IIa and IIb, conformity assessment runs not on a single certificate but on a combination of certificates, and each certificate type names a different thing.

This page opens up the certificate logic of IIa and IIb: which annex defines which route, what information an EU certificate must contain and what you should know in advance it will not contain, and where the ÜTS document requirements of IIa and IIb part ways. Every provision is taken from two sources: the Medical Device Regulation (Tıbbi Cihaz Yönetmeliği) published in the Official Gazette No. 31499 (bis) of 2/6/2021 and the Manual on Registration Procedures of Medical Devices in the Product Tracking System (Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz).

In IIa and IIb, a "certificate" is not a single document but a combination

Article 52(1) of the Medical Device Regulation says that, before placing a device on the market, manufacturers carry out a conformity assessment in accordance with the relevant conformity assessment procedures set out in Annexes IX to XI. These three annexes are not alternatives to one another but parts that complement one another.

The title of Annex IX is "Conformity assessment based on a quality management system and on assessment of technical documentation"; its Chapter I governs the quality management system. Point 1 of Annex IX says that the manufacturer establishes, documents, implements and maintains a quality management system as described in Article 10(9) and maintains its effectiveness over the life cycle of the devices.

The title of Annex X is "Conformity assessment based on type-examination". Point 1 of Annex X defines EU type-examination as the procedure by which a notified body ascertains and certifies that a device, including its technical documentation and relevant life cycle processes and a representative sample of the device production envisaged, fulfils the relevant provisions of the Regulation.

The title of Annex XI is "Conformity assessment based on product conformity verification". Point 1 of Annex XI sets the aim of this assessment as ensuring that devices conform to the type for which an EU type-examination certificate has been issued and meet the provisions of the Regulation that apply to them. Point 2 of Annex XI says that, if an EU type-examination certificate has been issued under Annex X, the manufacturer applies either the procedure in Part A of this Annex (production quality assurance) or the procedure in Part B (product verification).

Put side by side, these three annexes give two basic routes in IIa and IIb: the route based on the quality management system (Annex IX) and the route based on type-examination (Annex X and Annex XI together). Without understanding which route the document set in your hands belongs to, you cannot decide which documents to register in ÜTS.

The two routes of Class IIa

Article 52(5) of the Medical Device Regulation defines two options for manufacturers of class IIa devices, other than custom-made or investigational devices. Point (a) of the same paragraph says that the manufacturer is subject to a conformity assessment as specified in Chapters I and III of Annex IX, including an assessment of the technical documentation as specified in point 4 of Annex IX for at least one representative device for each device category. Point (b) of the same paragraph says that, alternatively, the manufacturer may choose to draw up the technical documentation set out in Annexes II and III, accompanied by a conformity assessment as specified in point 10 or point 18 of Annex XI; the assessment of technical documentation is applied to at least one representative device for each device category.

Article 61(5) of the Manual repeats the same two routes in ÜTS language. Point (a) of the same paragraph says that the manufacturer is subject to the MDR Annex IX (Chapters I and III) procedure together with the MDR Annex IX (Chapter II - Article 4) procedure for at least one representative device for each device category; point (b) says that MDR Annex II and Annex III technical documentation may be prepared accompanied by an MDR Annex XI (Article 10) or MDR Annex XI (Article 18) conformity assessment.

Point 3 of Annex XI gives the basis for this second way expressly: by way of derogation from points 1 and 2 of this Annex, the procedures in this Annex, together with technical documentation drawn up as set out in Annexes II and III, may also be applied by manufacturers of class IIa devices. So in IIa, Annex XI can also be used as a route that does not require an EU type-examination certificate first.

The point to watch in Class IIa is the expression "representative device". On both routes, the assessment of technical documentation is carried out not for every device but for at least one representative device from each device category. For this reason the certificate in your hands may not carry the name of the device you are registering.

The three possibilities of Class IIb

Article 52(4)(a) of the Medical Device Regulation says that manufacturers of class IIb devices, other than custom-made or investigational devices, are subject to a conformity assessment as specified in Chapters I and III of Annex IX, including an assessment of the technical documentation as specified in point 4 of Annex IX for at least one representative device for each generic device group. The unit that is "device category" in IIa changes to "generic device group" in IIb.

Point (b) of the same paragraph sets the critical exception: for class IIb implantable devices other than sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors, the assessment of the technical documentation as specified in point 4 of Annex IX is applied for every device. So for implantable IIb devices the representative-device logic disappears; the certificate must cover your device by name.

Point (c) of the same paragraph gives the third possibility: the manufacturer may alternatively choose to apply a conformity assessment based on type-examination as specified in Annex X, accompanied by a conformity assessment based on product conformity verification as specified in Annex XI. Article 61(6)(c) of the Manual separates this choice into MDR Annex XI (Part A) or MDR Annex XI (Part B).

Points (a) and (b) of Article 61(6) of the Manual make the same distinction as the Regulation and keep the same twelve items (sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors) on the exception list under the name "WET devices".

Class IIb active devices under Rule 12 follow a separate path

Article 61(10) of the Manual says that, in addition to the procedures applicable under paragraphs 6 or 7, for Class III implantable devices or Class IIb active devices intended to administer and/or remove a medicinal product, as referred to in point 6.4 of MDR Annex VIII and in Article 56(5)(ç) of the Manual, the MDR Annex IX (Article 5.1) or MDR Annex X (Article 6) procedure, as applicable, will also be applied.

Article 54(1)(b) of the Medical Device Regulation ties the same device group to the clinical evaluation consultation procedure. Point 5.1(a) of Annex IX says that, for these devices, the notified body draws up a clinical evaluation assessment report and submits it, together with the manufacturer's clinical evaluation documentation, to the Commission for transmission to the expert panel.

The result is this: if a Class-IIb active device administers or removes a medicinal product, its certificate set differs from that of an ordinary IIb device, and ÜTS meets this with a separate document business rule set.

What an EU certificate names, and what it does not

Article 7(2) of the Manual lists the fields that must be in an EU certificate and whose control is carried out during document review in ÜTS; the last sentence of the same paragraph says that the information entered into ÜTS and the information in the certificate must be the same. For scope information, point (h) of the same paragraph splits in two, and this split is the most misunderstood point in IIa and IIb.

  • Certificates that look at the product: Sub-point 1 of point (h) looks, for EU technical documentation assessment certificates, EU type-examination certificates and EU product verification certificates, for a clear description of the device or devices, including name, model and type, and the intended purpose, risk class and Basic UDI-DI information relating to the conformity assessment procedure the device has undergone, as attached by the manufacturer to the instructions for use.

  • Certificates that look at the system: Sub-point 2 of point (h) looks, for EU quality management system certificates and EU production quality assurance certificates, for the description of the devices or device groups, the risk class, and, for Class-IIb devices, the intended purpose information.

The difference between these two sub-points concerns you directly. A quality management system certificate does not have to name your device by name, model and Basic UDI-DI; a device group description is enough. On the other hand, the same certificate must carry the risk class, and if the device is Class-IIb, the intended purpose as well. For Class-IIa devices, the intended purpose is not a field looked for in a quality management system certificate.

The practical consequence is this: not finding your device's name on the Annex IX Chapter I certificate in your hands is not by itself a non-conformity. But if there is a technical documentation assessment certificate or a type-examination certificate, that certificate is expected to identify the device at the level of name, model and type.

The other mandatory identity fields of the certificate are listed in the same paragraph: point (c) looks for the certificate number, point (d) for the name and identification number of the notified body, point (f) for the regulation it belongs to, point (g) for the annex applied, that is, a reference to the relevant annex under which the conformity assessment was carried out, point (ğ) for the issue and validity dates, point (m) for the conditions or limitations on the certificate's validity, and point (n) for the legally binding signature of the notified body. The "annex applied" field in point (g) is the only field that shows which route your device has gone through; you read the IIa and IIb distinction here.

Point (ı) of the same paragraph asks, if the certificate has been amended, reissued or added to, for a reference to every previous certificate and its issue date, together with a description of the changes. Point (j) asks, where applicable, for a reference to the relevant parts of the technical documentation or to other certificates required for placing the covered device on the market. You will most often find the second part of the combination through this point.

Certificates under the old regulations are a separate document type

Article 7(1) of the Manual says that the document type "EU Certificate (MDR/IVDR)" is used only for certificates prepared by a notified body under the New Medical Device Regulations; for certificates issued under the Old Medical Device Regulations, a document notification and application is made in the "EC Certificate" type.

The mandatory fields of an EC certificate are shorter. Article 6(2) of the Manual lists these fields: the name and address of the manufacturer, the EU Authorised Representative information if any, the certificate number, the name, identification number and address of the notified body, the regulation it belongs to, the annex applied, the document start and end dates, the product information or scope, and the binding signature of the notified body. The Basic UDI-DI and the risk class are not in this list. If the certificate of a Class-IIa or Class-IIb device from the MDD period does not state the risk class, this is not by itself a deficiency.

On the validity period, Article 6(3) of the Manual says that EC certificates, except for the EC Verification Certificate and the EC Type Verification Certificate, have a validity period and that this period cannot be more than 5 years. Article 7(3) of the Manual repeats the same upper limit of 5 years for EU certificates and says that validity may be extended for further periods each not exceeding 5 years. Article 56(2) of the Medical Device Regulation carries the same provision.

How a certificate is registered in ÜTS is a separate discipline

A correct certificate is not enough; it must be entered into ÜTS correctly. Article 7(4) of the Manual says that, if the document has been revised, the revision date is entered as the document start date; if it has not been revised and the start date is not clearly stated on the certificate, the most recent issue (signature) date on the certificate is entered.

Paragraph 5 of the same article says that, if a revision number is stated on the EU certificate and shown together with the certificate number, the certificate is registered with its revision number, and gives the example "XXXXX.Rev.1". Paragraph 6 requires the EU certificate to be uploaded and submitted to the Agency in full, with all its annexes and additional pages. In IIa and IIb certificates, the list of covered devices is mostly in the annexes; a missing annex means the scope cannot be proved.

Paragraph 7 resolves another sticking point: if the notified body that issued the EU certificate is not defined in ÜTS, before the document registration procedures a request is opened through the "Submit Request" ("Talep Bildir") screen in the area "Notified/Conformity Assessment Body Procedures"; if the NANDO query shows that the body is authorised, it is added to ÜTS. Article 6(7) of the Manual defines the same path for EC certificates.

IIa and IIb part ways in the ÜTS document business rules

Article 66(2) of the Manual says that document linking is carried out according to the document business rules determined by the type and risk class of the device and on the basis of the conformity assessment procedures mentioned in Article 61 of the Manual. This is the direct ÜTS counterpart of the conformity assessment route.

Article 66(4) of the Manual says that, for devices to be notified under the MDR as "Class-IIa", the appropriate one of the document business rules shown in Annex-89 of the Manual will be applied. Class IIa has a single business rule set.

In Class IIb, the situation splits into three. Paragraph 5 of Article 66 of the Manual points to Annex-90 for devices to be notified as "Class-IIb", other than implantable devices or devices under MDR Annex VIII Rule 12; paragraph 6 to Annex-91 for implantable Class-IIb devices other than sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors; and paragraph 7 to Annex-92 for Class-IIb devices under MDR Annex VIII Rule 12.

This split depends on a field on the registration screen. Article 27(7)(h) of the Manual says that the "Is it implantable?" ("Vücuda implante edilebilir mi?") field is a mandatory field in ÜTS. Selecting "Yes" or "No" in this field for a Class-IIb device determines which of the three different document business rule sets will run.

The difference between IIa and IIb in instructions for use is clear

Article 17(2)(a)(1) of the Manual says that, for Class-III and Class-IIb devices certified under the MDR and the MDD, uploading to ÜTS the instructions for use issued by the manufacturer is mandatory. In Class-IIb the instructions for use are mandatory without condition.

Sub-point 2 of the same point makes Class-IIa conditional: for Class-IIa devices, the instructions for use are uploaded if the manufacturer has issued them; as an exception, only for Class-IIa devices under the MDD, if the manufacturer has not issued instructions for use, an intended purpose form/declaration signed by the responsible manager of the manufacturer or importer firm and mentioned in Article 19 of the Manual is accepted.

Article 19(1) of the Manual narrows the scope of this form: for Class-IIa devices under the MDD, if the manufacturer has not issued instructions for use, an intended purpose form/declaration is prepared by the manufacturer or importer in the document format whose example is in Annex-15 of the Manual. Point (a) of paragraph 2 of the same article says that when this document is uploaded to ÜTS, "Instructions for Use" is selected as the document type; point (c) says that the question "Is it an Intended Purpose Form/Declaration?" is answered "Yes"; point (d) says that the document is uploaded unchanged and either wet-signed or e-signed. Paragraph 3 of the same article states that the form is for registration purposes only and is not physically on the device packaging. This route is not open for Class-IIa devices under the MDR.

In the case of accessories the distinction sharpens once more. Article 17(4) of the Manual says that, for Class-IIa devices that are accessories of another device, if the field "Is it a component/accessory of a device?" is selected as "Yes" in applications under the MDR and the IVDR, the requirement to upload instructions for use will not be sought. By contrast, paragraph 5 of the same article says that, for Class-IIb and Class-III devices that are accessories, if the manufacturer of the main device is different from the manufacturer of the accessory, the accessory's own instructions for use will be uploaded, or if there are none, the instructions for use created by the manufacturer of the main device, without regard to criteria such as differences in the notified body and address. Class IIa has an accessory exemption; Class IIb does not.

IIa and IIb also part ways in post-registration obligations

Once the certificate and the registration are complete, the two classes still do not carry the same obligation. Article 84(1)(a) of the Medical Device Regulation says that manufacturers of class IIa, class IIb and class III devices prepare a periodic safety update report. Point (b) of the same paragraph says that manufacturers of class IIb and class III devices update this report at least annually; point (c) says that manufacturers of class IIa devices update it when necessary and at least every 2 years. Both points say that the report is part of the technical documentation as set out in Annexes II and III.

Paragraph 2 of the same article says that, for class III devices or implantable devices, manufacturers submit these reports to the notified body through the electronic system; paragraph 3 says that, for devices other than these, the reports are submitted to the notified body involved in the conformity assessment and, upon request, to the Agency. An implantable Class-IIb device follows the same path as Class-III in this respect.

The standing of the certificate also depends on monitoring. Article 56(4) of the Medical Device Regulation says that, if the notified body finds that the requirements are no longer met by the manufacturer and corrective action is not taken within a set period, it will, taking account of the principle of proportionality, suspend, withdraw or restrict the certificate. Paragraph 3 of the same article says that notified bodies may impose restrictions on the intended purpose for certain groups of patients or require specific post-market clinical follow-up studies. These restrictions are written in the certificate and, under Article 7(2)(m) of the Manual, are among the fields checked in ÜTS review.

What we do on IIa and IIb files, and what we cannot do

Within medical device ÜTS registration consultancy, our concrete output on the Class IIa and IIb side is this:

  • Route identification: By reading the "annex applied" field in the certificates from the manufacturer, we determine whether the device has gone through the Annex IX-based route or the Annex X and Annex XI combination, and we request the missing part by name.

  • Scope matching: We compare the scope text of the certificate with the barcode, brand, model and Basic UDI-DI information to be registered in ÜTS; based on Article 7(2)(h) of the Manual, we distinguish that a device name is not looked for in quality management system certificates but is looked for in technical documentation assessment and type-examination certificates.

  • IIb sub-group decision: We base on the documents whether the device is implantable and whether it falls under MDR Annex VIII Rule 12, and we establish in writing which of paragraphs 5, 6 and 7 of Article 66 of the Manual applies.

  • Document chain preparation: We plan the registrations of the EU certificate, EU declaration of conformity, instructions for use and, where required, the quality management certificate in the order in which the applications under Article 39 of the Manual will bring them to "Registered" status.

  • Validity calendar: We check certificate end dates, revision numbers and the completeness of annexes before the application.

What we cannot do is clear too. We cannot widen the scope of a notified body's certificate or register a device outside the scope as if it were within it; Article 7(2) of the Manual requires the information entered into ÜTS and the information in the certificate to be the same. We cannot produce the manufacturer's technical documentation, clinical evaluation or quality management system in its place; paragraphs 4 and 5 of Article 52 of the Medical Device Regulation tie conformity assessment to the manufacturer, and point 1 of Annex IX ties the quality management system to the manufacturer. The choice of notified body and the certification process also run on the manufacturer's own contract.

The way of working is this: the firm defines a user authorisation for us from its own ÜTS firm page, and we create the document and device records with the "Medical Device Registration Operator" ("Tıbbi Cihaz Kayıt Elemanı") authorisation. Article 39(11) of the Manual leaves the "Approve" ("Onayla") action in a document application only to users with this authorisation; paragraph 13 of the same article states that an application in "Awaiting Signature" status will be electronically signed by the firm's signatory who is the applicant.

Tıbbi Cihaz ÜTS Kayıt Danışmanlığı

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara